State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography–Tandem Mass Spectrometry Methods
Abstract
BACKGROUND: Vitamin D testing is increasing worldwide. Recently several diagnostic manufacturers including Abbott and Siemens have launched automated 25-hydroxy vitamin D (25OH-D) immunoassays. Furthermore, preexisting assays from DiaSorin and Roche have recently been modified. We compared the performance of 5 automated immunoassays, an RIA and 2 liquid chromatography–tandem mass spectrometry (LC-MS/MS) methods. METHODS: Aliquots of 170 randomly selected patient samples were prepared and 25OH-D was measured by 2 LC-MS/MS methods, an RIA (DiaSorin), and automated immunoassays from Abbott (Architect), DiaSorin (LIAISON), IDS (ISYS), Roche (E170, monoclonal 25OH-D3 assay), and Siemens (Centaur). Within-run and between-run imprecision were evaluated by measurement of 5 replicates of 2 serum pools on 5 consecutive days. RESULTS: The LC-MS/MS methods agreed, with a concordance correlation coefficient (CCC) of 0.99 and bias of 0.56 μg/L (1.4 nmol/L). The RIA assay showed a performance comparable to LC-MS/MS, with a CCC of 0.97 and a mean bias of 1.1 μg/L (2.7 nmo/L). All immunoassays measured total 25OH-D (including D3 and D2), with the exception of the Roche assay (D3 only). Among the immunoassays detecting total 25OH-D, the CCCs varied between 0.85 (Abbott) to 0.95 (LIAISON). The mean bias ranged between 0.2 μg/L (0.5 nmol/L) (LIAISON) and 4.56 μg/L (11.4 nmol/L) (Abbott). The Roche 25OH-D3 assay demonstrated small mean bias [−2.7 μg/L (−6.7 nmol/L)] [−2.7 μg/L (−6.7 nmol/L)] but a low CCC of just 0.66. Most assays demonstrated good intra- and interassay precision, with CV <10%. CONCLUSIONS: Automated immunoassays demonstrated variable performance and not all tests met our minimum performance goals. It is important that laboratories be aware of the limitations of their assay.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Holick, M. F. (2007) · New England Journal of Medicine · doi:10.1056/nejmra070553 · Cited by 13,720
Holick, M. F., Binkley, N. C., Bischoff-Ferrari, H. A., Gordon, C. M., Hanley, D. A., Heaney, R. P., Murad, M. H., & Weaver, C. M. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2011-0385 · Cited by 10,538
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
Overview of general physiologic features and functions of vitamin D
DeLuca, H. F. (2004) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/80.6.1689s · Cited by 2,327
Passing, H., & Bablok, W. (1983) · cclm · doi:10.1515/cclm.1983.21.11.709 · Cited by 1,513
Rosen, C. J. (2011) · New England Journal of Medicine · doi:10.1056/nejmcp1009570 · Cited by 816
Holick, M. F., Biancuzzo, R. M., Chen, T. C., Klein, E. K., Young, A., Bibuld, D., Reitz, R., Salameh, W., Ameri, A., & Tannenbaum, A. D. (2008) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2007-2308 · Cited by 771
Biological Variation: From Principles to Practice
Fraser, C. G. (2001) · Cited by 653
Singh, R. J., Taylor, R. L., Reddy, G. S., & Grebe, S. K. G. (2006) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2006-0710 · Cited by 478
Christakos, S., Ajibade, D. V., Dhawan, P., Fechner, A. J., & Mady, L. J. (2010) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2010.02.002 · Cited by 430
Cited by
Global summary of maternal and newborn vitamin D status – a systematic review
Saraf, R., Morton, S. M. B., Camargo, C. A., & Grant, C. C. (2016) · Maternal & Child Nutrition · doi:10.1111/mcn.12210 · Cited by 365
Mass spectrometry and immunoassay: how to measure steroid hormones today and tomorrow
Taylor, A. E., Keevil, B., & Huhtaniemi, I. T. (2015) · European Journal of Endocrinology · doi:10.1530/eje-15-0338 · Cited by 320
Smoking is a significant determinant of low serum vitamin D in young and middle-aged healthy males
Kassi, E. N., Stavropoulos, S., Kokkoris, P., Galanos, A., Moutsatsou, P., Dimas, C., Papatheodorou, A., Zafeiris, C., & Lyritis, G. (2015) · Hormones · doi:10.14310/horm.2002.1521 · Cited by 79
Dirks, N. F., Vesper, H. W., van Herwaarden, A. E., van den Ouweland, J. M. W., Kema, I. P., Krabbe, J. G., & Heijboer, A. C. (2016) · Clinica Chimica Acta · doi:10.1016/j.cca.2016.08.016 · Cited by 40
Metadata
- Authors
- Farrell, C.-J. L., Martin, S., McWhinney, B., Straub, I., Williams, P., & Herrmann, M.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2012-03-01
- Volume
- 58
- Issue
- 3
- Pages
- 531-542
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2166970322
- Cited by
- 373
- References
- 26
- FWCI
- 32.42
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography–Tandem Mass Spectrometry Methods. https://app.cochat.ai/papers/10.1373/clinchem.2011.172155
- The paper
- Farrell, C.-J. L., Martin, S., McWhinney, B., Straub, I., Williams, P., & Herrmann, M. (2012). State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography–Tandem Mass Spectrometry Methods. Clinical Chemistry. https://doi.org/10.1373/clinchem.2011.172155
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.