Automated 22-kD Growth Hormone–Specific Assay without Interference from Pegvisomant

Manolopoulou, Jenny, Alami, Younes, Petersenn, Stephan, Schopohl, Jochen, Wu, Zida, Strasburger, Christian J, Bidlingmaier, Martin · 2012 · Clinical Chemistry
Cited by 75

The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Large variability exists among different growth hormone (GH) assays owing to differences in calibration, antibody specificity, isoform recognition, and interference from GH binding protein (GHBP). The GH receptor antagonist Pegvisomant presents a new challenge because Pegvisomant interferes with many GH assays. A recent consensus conference established criteria for standardization and evaluation of GH assays. Following consensus recommendations, we developed a new GH assay on an automated analyzer (IDS-iSYS, Immunodiagnostic Systems). METHODS: A monoclonal antibody not cross-reacting with Pegvisomant was combined with a monoclonal antibody specific for 22-kD GH. Isoform specificity and interference from GHBP was tested and compared to that seen in 2 existing automated GH assays (Siemens Immulite, Diasorin Liaison). We also compared GH concentrations measured by the 3 assays for healthy volunteers and patients with acromegaly receiving different treatments. Using the iSYS assay, we also established nadir GH values during oral glucose load and analyzed changes in endogenous GH during Pegvisomant treatment. RESULTS: Analytical and functional sensitivities were 0.01 μg/L and 0.04 μg/L, with a dynamic range from 0.04 to 100 μg/L. Intraassay CVs were 2%–4%, whereas interassay CVs were 5%–7% at GH concentrations between 1.7 and 27.5 μg/L. The assay was specific for 22-kD GH and not affected by GHBP. The presence of Pegvisomant, which leads to a negative bias on the Immulite and dramatic overestimation of GH on the Liaison, had no impact on the iSYS GH assay. CONCLUSIONS: The new assay fulfils recent consensus recommendations and presents a useful new tool for reliable measurement of GH.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1373/clinchem.2012.188128

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 4 references

Cited by

Metadata

Authors
Manolopoulou, J., Alami, Y., Petersenn, S., Schopohl, J., Wu, Z., Strasburger, C. J., & Bidlingmaier, M.
Venue
Clinical Chemistry
Published
2012-10-01
Volume
58
Issue
10
Pages
1446-1456
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2166548501
Cited by
75
References
4
FWCI
3.91
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.