State of the art for measurement of urine albumin: comparison of routine measurement procedures to isotope dilution tandem mass spectrometry
Open accessAbstract
BACKGROUND: Urine albumin is the primary biomarker for detection and monitoring of kidney damage. Because fixed decision criteria are used to identify patients with increased values, we investigated if commonly used routine measurement procedures gave comparable results. METHODS: Results from 17 commercially available urine albumin measurement procedures were investigated vs an isotope dilution mass spectrometry (IDMS) procedure. Nonfrozen aliquots of freshly collected urine from 332 patients with chronic kidney disease, diabetes, cardiovascular disease, and hypertension were distributed to manufacturers to perform urine albumin measurements according to the respective instructions for use for each procedure. Frozen aliquots were used for measurements by the IDMS procedure. An error model was used to determine imprecision and bias components. RESULTS: Median differences between the largest positive and negative biases vs IDMS were 45%, 37%, and 42% in the concentration intervals of 12–30 mg/L, 31–200 mg/L, and 201–1064 mg/L, respectively. Biases varied with concentration for most procedures and exceeded ±10% over the concentration interval for 14 of 16 quantitative procedures. Mean biases ranged from −35% to 34% at 15 mg/L. Dilution of samples with high concentrations introduced bias for 4 procedures. The combined CV was >10% for 5 procedures. It was not possible to estimate total error due to dependence of bias on concentration. CVs for sample-specific influences were 0% to 15.2%. CONCLUSIONS: Bias was the dominant source of disagreement among routine measurement procedures. Consequently, standardization efforts will improve agreement among results. Variation of bias with concentration needs to be addressed by manufacturers.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Mean biases of urine albumin measurement procedures were large and ranged from -35% to 34% at 15 mg/L.
“Mean biases were large and ranged from À35% to 34% at 15 mg/L.”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [498]
Calibration bias was the main source of error for differences among urine albumin measurement methods, and precision was adequate for most assays.
“The authors concluded that calibration bias was the main source of error for differences among methods and precision was adequate for most assays.”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [498]
Dipstick methods were compared with local laboratory methods, which exhibit large biases.
“An important limitation of all these data is that the dipstick methods were compared with local laboratory methods, which, as indicated above, exhibit large biases (498).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [498]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A New Equation to Estimate Glomerular Filtration Rate
Levey, A. S., Stevens, L. A., Schmid, C. H., Zhang, Y. (., Castro, A. F., Feldman, H. I., Kusek, J. W., Eggers, P., Van Lente, F., Greene, T., et al. (2009) · Annals of Internal Medicine · doi:10.7326/0003-4819-150-9-200905050-00006 · Cited by 25,603 · see citing works
Standards of Medical Care in Diabetes
Power, D. V. (2006) · Diabetes Care · doi:10.2337/diacare.29.02.06.dc05-1593 · Cited by 8,672 · see citing works
KDIGO 2024 clinical practice guideline for the evaluation and management of chronic kidney disease
Stevens, P. E., Ahmed, S. B., Carrero, J. J., Foster, B., Francis, A., Hall, R. K., Herrington, W. G., Hill, G., Inker, L. A., Kazancıoğlu, R., et al. (2024) · Kidney International · doi:10.1016/j.kint.2023.10.018 · Cited by 8,530 · see citing works
Standards of medical care in diabetes—2013
American Diabetes Association (2013) · Diabetes Care · doi:10.2337/dc13-s011 · Cited by 4,369 · see citing works
(2012) · Kidney International Supplements · doi:10.1038/kisup.2012.73 · Cited by 3,247 · see citing works
Basevi, V., Mario, S. D., Morciano, C., Nonino, F., & Magrini, N. (2011) · Diabetes Care · doi:10.2337/dc11-0174 · Cited by 3,058 · see citing works
Plasma aldosterone regulation in anephric man
Walker, W. G., & Cooke, C. R. (1973) · Kidney International · doi:10.1038/ki.1973.1 · Cited by 1,848 · see citing works
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360 · see citing works
Self-citationCurrent issues in measurement and reporting of urinary albumin excretion
Miller, W. G., Bruns, D. E., Hortin, G. L., Sandberg, S., Aakre, K. M., McQueen, M. J., Itoh, Y., Lieske, J. C., Seccombe, D. W., Jones, G., et al. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.106567 · Cited by 381 · see citing works
Self-citationLevey, A. S., Schoolwerth, A. C., Burrows, N. R., Williams, D. E., Stith, K. R., & McClellan, W. (2009) · American Journal of Kidney Diseases · doi:10.1053/j.ajkd.2008.11.019 · Cited by 227 · see citing works
Quantification of Urinary Albumin by Using Protein Cleavage and LC-MS/MS
Seegmiller, J. C., Barnidge, D. R., Burns, B. E., et al. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.115543 · Cited by 81 · see citing works
A reference system for urinary albumin: current status
Lieske, J. C., Bondar, O. P., Miller, W. G., Bachmann, L. M., Narva, A. S., Itoh, Y., Zegers, I., Schimmel, H. G., Phinney, K. W., & Bunk, D. M. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0768 · Cited by 47 · see citing works
Self-citationRecent Trends in Performance and Current State of Creatinine Assays
Killeen, A. A., Ashwood, E. R., Ventura, C. B., et al. (2013) · Archives of Pathology & Laboratory Medicine · doi:10.5858/arpa.2012-0134-cp · Cited by 44 · see citing works
Laboratory issues in measuring and reporting urine albumin
Miller, W. G., & Bruns, D. E. (2008) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfp022 · Cited by 30 · see citing works
Self-citationAnalysis of molecular forms of albumin in urine
Hortin, G. L., & Sviridov, D. O. (2008) · PROTEOMICS - CLINICAL APPLICATIONS · doi:10.1002/prca.200780145 · Cited by 15 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Diabetic kidney disease: a report from an ADA Consensus Conference
Tuttle, K. R., Bakris, G. L., Bilous, R. W., Chiang, J. L., de Boer, I. H., Goldstein-Fuchs, J., Hirsch, I. B., Kalantar-Zadeh, K., Narva, A. S., Navaneethan, S. D., et al. (2014) · Diabetes Care · doi:10.2337/dc14-1296 · Cited by 1,153 · see citing works
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351 · see citing works
Self-citationMcTaggart, M. P., Newall, R. G., Hirst, J. A., Bankhead, C. R., Lamb, E. J., Roberts, N. W., & Price, C. P. (2014) · Annals of Internal Medicine · doi:10.7326/m13-2331 · Cited by 39 · see citing works
- Authors
- Bachmann, L. M., Nilsson, G., Bruns, D. E., McQueen, M. J., Lieske, J. C., Zakowski, J. J., & Miller, W. G.
- Affiliations
- Virginia Commonwealth University (US)
- Nilsson Measurement Quality, Uppsala, Sweden
- University of Virginia (US)
- McMaster University (CA)
- Mayo Clinic (US)
- Beckman Coulter (United States) (US)
- Corresponding author
- Bachmann, Lorin M
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2014-03-01
- Volume
- 60
- Issue
- 3
- Pages
- 471-480
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2161581547
- Semantic Scholar
- 48f5d52a9321744beb1a9ca01615043e26a379e7
- Cited by
- 68
- References
- 16
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 4 · 2023: 7 · 2022: 4 · 2021: 3 · 2020: 5 · 2019: 6 · 2018: 5 · 2017: 10 · 2016: 5 · 2015: 7 · 2014: 6 · 2013: 1
- FWCI
- 4.36
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model (vor)
- Locations
- Clinical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Funders
- University of Pennsylvania (5-U01-DK-060990-10, 554511)
- National Institute of Diabetes and Digestive and Kidney Diseases
- NIH
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Nephrology › Chronic Kidney Disease and Diabetes
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
- Keywords
- urinary albumin, albuminuria, isotope dilution mass spectrometry, measurement bias, imprecision, method comparison, chronic kidney disease, kidney injury biomarkers, measurement standardization, cardiovascular disease, hypertension, diabetes mellitus
- MeSH terms
- Adult, Aged, Aged, 80 and over, Humans, Indicator Dilution Techniques, Middle Aged, Oligopeptides, Serum Albumin, Urinalysis, Tandem Mass Spectrometry, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review