Influence of Contact Forces on Wrist Photo plethysmography – Prestudy for a Wearable Patient Monitor / Einfluß von Kontaktkräften auf die photoplethysmografische Pulsmessung am Handgelenk – Vorstudie zu einem tragbaren Patientenmonitor
Abstract
The practical setting shows that gentle manual pressure on a photosensor suffices to improve the detectability of arterial pulses, but changes in the pressure applied may also produce signal artefacts. To study these effects, stepwise increasing contact forces (0.5 to 4 N) were applied to a photosensor placed over the radial artery. Additionally, the influence of optical coupling between sensor and skin surface was examined by introducing an elastic distance ring. The AC and DC components from the recorded photoplethysmogram were analysed. The AC component (absorption due to arterial pulsation) increased with the pressure applied; at lower forces (0.5 to 3 N) an introduction of the ring enhanced this effect. The characteristic of the DC component (backscattering from non-pulsating tissues) depends on optical coupling: without the ring the DC component increased stepwise with force (slope 0.035 V/N), but with the ring in place this component decreased (slope −0.075 V/N). Since the sensitivity of the pulse signal to artifacts is related to the slope of the DC component, such artefacts can be minimized by making the slope small. The utilization of these results to improve pulse detection and reduce motion artefacts in a wearable wrist device is discussed. Oft genügt schon ein geringer manueller Andruck auf einen Photosensor, um eine arterielle Pulswelle besser zu detektieren, bei Änderung des Andrucks kommt es dabei aber auch zu Artefakten im Pulssignal. Um hier den Einfluß auf photoplethysmografische Signale zu quantifizieren, wurden mit einem Galvanometer schrittweise zunehmende Andruckkräfte (0.5 bis 4 N) auf das Sensorgehäuse appliziert. Um dabei auch den Einfluß des optischen Übergangs zwischen Haut und Sensor zu studieren, wurde dieser Andrucktest zuerst ohne und dann mit einem elastischen Distanzring durchgeführt. Zur Analyse wurden die aufgezeichneten photoplethysmografischen Signale in ihre AC (0.1 bis 20 Hz) und DC-Komponenten zerlegt. Bei der AC-Komponente (Absorption infolge arterieller Pulsation) zeigte sich eine Zunahme mit der Andruckkraft, wobei diese Charakteristik durch den Ring noch verstärkt wurde. Die Kraftcharakeristik der DC-Komponente (nichtpulsierendes reflektiertes Licht) hingegen war unterschiedlich: Ohne Ring kam es, ausgehend von niedrigen Werten, zu einer stufenweisen Zunahme (Steigung 0.035 N/V) und mit Ring zu einer Abnahme (Steigung −0.075 V/N) der Reflektanz. Da die Artefaktempfindlichkeit des Pulssignals von der Steigung der DC-Charakteristik abhängt, ergibt sich durch die Gestaltung des optischen Übergangs hier eine Möglichkeit zur Artefaktunterdrükkung. Der Transfer dieser Ergebnisse zur Verbesserung der Pulsdetektion und der Artefaktunterdrückung bei einem Handgelenkmonitor wurde diskutiert.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
PPG sensors used to measure HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles.
“Prior research has suggested that PPG sensors used to measure the HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles [36-38].”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [36]
During activities involving sustained muscle contractions or higher intensity exercises, the contact between the device's PPG sensor and skin is decreased, leading to a disruption in the signal quality and causing poor quality data.
“It has been suggested that during activities involving sustained muscle contractions or higher intensity exercises, the contact between the device’s PPG sensor and skin is decreased, leading to a disruption in the signal quality and causing poor quality data [36,37].”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [36]
Cites
A review of the optical properties of biological tissues
Cheong, W. F., Prahl, S. A., & Welch, A. J. (1990) · IEEE Journal of Quantum Electronics · doi:10.1109/3.64354 · Cited by 2,530 · see citing works
Photoplethysmograph fingernail sensors for measuring finger forces without haptic obstruction
Mascaro, S. A., & Asada, H. H. (2001) · IEEE Transactions on Robotics and Automation · doi:10.1109/70.964669 · Cited by 192 · see citing works
Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry
Schmitt, J. M. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.137285 · Cited by 118 · see citing works
Standing Up with Denervated Muscles in Humans Using Functional Electrical Stimulation
Kern, H., Hofer, Strohhofer, Mayr, Richter, & Stöhr (1999) · Artificial Organs · doi:10.1046/j.1525-1594.1999.06376.x · Cited by 63 · see citing works
Pessenhofer, H. (1986) · Basic Research in Cardiology · doi:10.1007/bf01907432 · Cited by 32 · see citing works
Gallasch, E., Rafolt, D., Moser, M., Hindinger, J., Eder, H. C., Wießpeiner, G., & Kenner, T. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.486291 · Cited by 24 · see citing works
Self-citationPhotoplethysmographic evaluation of the relationship between skin reflectance and skin blood volume
Higgins, J. L., & Fronek, A. S. (1986) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(86)90047-6 · Cited by 19 · see citing works
Transducer system for the noninvasive recording of arterial pressure contours
Karr, S. G., Karwoski, T., Jacobs, J. E., & Mockros, L. F. (1985) · Annals of Biomedical Engineering · doi:10.1007/bf02407770 · Cited by 17 · see citing works
Surface myomechanical responses recorded on a scanner galvanometer
Rafolt, D., & Gallasch, E. (2002) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345460 · Cited by 6 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Validity of Wearable Activity Monitors during Cycling and Resistance Exercise
BOUDREAUX, B. D., HEBERT, E. P., HOLLANDER, D. B., WILLIAMS, B. M., CORMIER, C. L., NAQUIN, M. R., GILLAN, W. W., GUSEW, E. E., & KRAEMER, R. R. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001471 · Cited by 205 · see citing works
Reddy, R. K., Pooni, R., Zaharieva, D. P., Senf, B., El Youssef, J., Dassau, E., Doyle III, F. J., Clements, M. A., Rickels, M. R., Patton, S. R., et al. (2018) · JMIR mHealth and uHealth · doi:10.2196/10338 · Cited by 157 · see citing works
Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking.
(2016) · PubMed Central · Cited by 150 · see citing works
Are Activity Wrist-Worn Devices Accurate for Determining Heart Rate during Intense Exercise?
Martín-Escudero, P., Cabanas, A. M., Dotor-Castilla, M. L., Galindo-Canales, M., Miguel‐Tobal, F., Fernández, C., Fuentes, M. E., & Giannetti, R. (2023) · Bioengineering · doi:10.3390/bioengineering10020254 · Cited by 40 · see citing works
Determinants of photoplethysmography signal quality at the wrist
Charlton, P. H., Marozas, V., Mejía‐Mejía, E., Kyriacou, P. A., & Mant, J. (2025) · PLOS Digital Health · doi:10.1371/journal.pdig.0000585 · Cited by 38 · see citing works
Carek, A. M., & Inan, O. T. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2683801 · Cited by 20 · see citing works
Validity of the Polar Vantage M watch when measuring heart rate at different exercise intensities
Shumate, T., Link, M., Furness, J. W., Kemp‐Smith, K., Simas, V., & Climstein, M. (2021) · PeerJ · doi:10.7717/peerj.10893 · Cited by 18 · see citing works
Validation of a Self‐Monitoring Tool for Use in Exercise Therapy
Powierza, C. S., Clark, M., Hughes, J. M., Carneiro, K. A., & Mihalik, J. P. (2017) · PM&R · doi:10.1016/j.pmrj.2017.03.012 · Cited by 17 · see citing works
Blood pressure assessment with in-ear photoplethysmography
Xing, X., Ma, Z., Xu, S., Zhang, M., Zhao, W., Song, M., & Dong, W. (2021) · Physiological Measurement · doi:10.1088/1361-6579/ac2a71 · Cited by 16 · see citing works
The Impact of Contact Force on Signal Quality Indices in Photoplethysmography Measurements
Cause, J. L., Morillo, Á. S., Naranjo, J. C. G., Stiens, J. H., & da Silva, B. (2024) · Applied Sciences · doi:10.3390/app14135704 · Cited by 14 · see citing works
Ceugniez, M., Devanne, H., & Hermand, É. (2024) · JMIR mhealth and uhealth · doi:10.2196/54871 · Cited by 11 · see citing works
Cause, J. L., Morillo, Á. S., da Silva, B., Naranjo, J. C. G., & Stiens, J. H. (2023) · Sensors · doi:10.3390/s23146628 · Cited by 9 · see citing works
Evaluating Vascular Depth-Dependent Changes in Multi-Wavelength PPG Signals Due to Contact Force
Cause, J. L., Morillo, Á. S., da Silva, B., Naranjo, J. C. G., & Stiens, J. H. (2024) · Sensors · doi:10.3390/s24092692 · Cited by 8 · see citing works
Machine-Learning Classification of Pulse Waveform Quality
OuYoung, T., Weng, W.-L., Hu, T.-Y., Lee, C.-C., Wu, L., & Hsiu, H. (2022) · Sensors · doi:10.3390/s22228607 · Cited by 8 · see citing works
Liu, Z.-J., Xiang, T., Zhou, R.-S., Ji, N., & Zhang, Y. (2024) · Connected Health And Telemedicine · doi:10.20517/chatmed.2023.11 · Cited by 4 · see citing works
Kayabaş, O., Cuğ, M., & BUDAK, C. (2022) · Ankara Üniversitesi Beden Eğitimi ve Spor Yüksekokulu SPORMETRE Beden Eğitimi ve Spor Bilimleri Dergisi · doi:10.33689/spormetre.1019059 · Cited by 1 · see citing works
Comparison of Single Foot IPG and PPG as a Distal PAT in Wearable Devices
García-Limón, J. A., Casas, Ó., Alvarado-Serrano, C., & Casanella, R. (2025) · IEEE International Workshop on Metrology for Industry 4.0 & IoT (MetroInd4.0&IoT) · doi:10.1109/metroind4.0iot66048.2025.11122060
- Authors
- Rafolt, D., & Gallasch, E.
- Affiliations
- Vienna General Hospital (AT)
- Corresponding author
- Rafolt, D
- ORCID
- Rafolt, D
- Venue
- Biomedizinische Technik/Biomedical Engineering
- Publisher
- Walter de Gruyter GmbH
- Published
- 2004-01-01
- Volume
- 49
- Issue
- 1-2
- Pages
- 22-26
- ISSN
- 0013-5585 (ISSN-L)
- 1862-278X
- OpenAlex
- W2017198338
- Cited by
- 19
- References
- 9
- Citations by year
- 2025: 2 · 2024: 4 · 2023: 3 · 2022: 2 · 2021: 2 · 2018: 2 · 2017: 3 · 2016: 1
- FWCI
- 0.00
- Type
- journal article
- Language
- de
- Open access
- closed
- Locations
- Biomedizinische Technik/Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Keywords
- photoplethysmography, contact force, radial artery, optical coupling, DC component, pulse detection, motion artifacts, pulse wave amplitude, pulse wave reflection
- MeSH terms
- Equipment Design, Heart Rate, Humans, Monitoring, Physiologic, Motor Activity, Paraplegia, Pulse, Signal Processing, Computer-Assisted, Wrist, Photoplethysmography, Radial Artery
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review