An improved approach to modelling patient reclassification: HbA 1c as an example
The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
Background: External quality assurance (EQA) programs must assess whether the analytical performance of commonly used in vitro diagnostic (IVD) devices is fit for their intended clinical use. We developed a method for this assessment based on reclassification of results at clinical decision limits and applied it to results from a haemoglobin A 1c (HbA 1c ) EQA program. Content: Using a commutable EQA program to provide analytical performance characteristics, we simulated 10,000 measurements of the HbA 1c value of each of 6,326 subjects in the National Health and Nutrition Survey (NHANES) by common IVD devices and the NHANES device. A parametric empirical Bayes approach was used to account for regression to the mean. We estimated the rates at which the IVD devices would reclassify results and compared them with the estimated reclassification rate for the NHANES device, calculating the difference between them as the ‘incremental reclassification rate’ (IRR). IRRs were ‘optimal’ if <1.5 %, ‘desirable’ if <3.0 %, and ‘minimal’ if <4.7 %. Summary and outlooks: EQA performance data were available for five HbA 1c devices: Abbott Afinion, Bio-Rad D series, Roche Cobas, Sebia Capillarys, and Siemens DCA. The median IRRs were optimal for all methods for diabetes diagnosis (0.11–0.22 %) and for diabetes monitoring (−0.51–1.01 %), with negative IRRs representing lower reclassification rates than the NHANES device. For prediabetes diagnosis, the Afinion failed the minimal standard (8.62 %), but median IRRs were desirable for the other methods (1.43–2.69 %). EQA providers should consider reporting IRR for suitable analytes to help identify poorly performing IVD devices.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
6. Glycemic Targets: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s006 · Cited by 1,403 · cited 1×
“While patient-speci fi c targets are used, the target for many non-pregnant patients is ≤ 7.0 % (53 mmol/mol; [9, 10]).”
InfluentialSandberg, S., Fraser, C. G., Horvath, A. R., Jansen, R., Jones, G., Oosterhuis, W., Petersen, P. H., Schimmel, H., Sikaris, K., & Panteghini, M. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0067 · Cited by 563
Self-citationSacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351 · cited 3×
“Diabetes is diagnosed when HbA 1c is ≥ 6.5 % (48 mmol/mol), with con fi rmation testing on a di ff erent sample recommended [6 – 8].”
InfluentialSelf-citationCeriotti, F., Fernandez-Calle, P., Klee, G. G., Nordin, G., Sandberg, S., Streichert, T., Vives-Corrons, J.-L., & Panteghini, M. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0091 · Cited by 181
Little, R. R., Rohlfing, C., & Sacks, D. B. (2019) · Clinical Chemistry · doi:10.1373/clinchem.2018.296962 · Cited by 161
Chai, J. H., Ma, S., Heng, D., Yoong, J., Lim, W.-Y., Toh, S.-A., & Loh, T. P. (2017) · Scientific Reports · doi:10.1038/s41598-017-14172-8 · Cited by 50 · cited 1×
“Firstly, the results originally measured in studies are often used as the ‘ true ’ reference values [19 – 21].”
Self-citation
Research works printed but not linked: 18
- English, E, Lenters-Westra, E. HbA1c method performance: the great success story of global standardization. Crit Rev Clin Lab Sci 2018;55:408–19. https://doi.org/10.1080/10408363.2018.1480591. · DOI
- Chaila, MZ, Viniegra, M, Gagliardino, JJ, Martínez, A, Simesen de Bielke, MG, Frusti, M, et al.. Glycated hemoglobin measurement: comparison of three methods versus high performance liquid chromatography. J Diabetes Sci Technol 2022;16:724–31. https://doi.org/10.1177/1932296821997179. · DOI
- Lenters-Westra, E, English, E. Are hemoglobin A1c point-of-care analyzers fit for purpose? The story continues. Clin Chem Lab Med 2020;59:765–74. https://doi.org/10.1515/cclm-2020-1308. · DOI
- Lenters-Westra, E, Singh, P, Vetter, B, English, E. Challenges in HbA1c point-of-care testing: only 5 of 19 HbA1c point-of-care devices meet IFCC and NGSP certification criteria on independent evaluation. Clin Chem 2025;71:775–88. https://doi.org/10.1093/clinchem/hvaf059. · DOI
- International Expert Committee. International expert committee report on the role of the A1c assay in the diagnosis of diabetes. Diabetes Care 2009;32:1327–34. https://doi.org/10.2337/dc09-9033. · DOI · Paper page
- Cheung, NW, Conn, JJ, D’Emden, MC, Gunton, JE, Jenkins, AL, Ross, GP, et al.. Position statement of the Australian diabetes society: individualisation of glycated haemoglobin targets for adults with diabetes mellitus. Med J Aust 2009;191:339–44. https://doi.org/10.5694/j.1326-5377.2009.tb02819.x. · DOI · Paper page
- Horvath, AR, Bossuyt, PM, Sandberg, S, John, AS, Monaghan, PJ, Verhagen-Kamerbeek, WD, et al.. Setting analytical performance specifications based on outcome studies – is it possible? Clin Chem Lab Med 2015;53:841–8. https://doi.org/10.1515/cclm-2015-0214. · DOI · Paper page
- Loh, TP, Smith, AF, Bell, KJL, Lord, SJ, Ceriotti, F, Jones, G, et al.. Setting analytical performance specifications using HbA1c as a model measurand. Clin Chim Acta 2021;523:407–14. https://doi.org/10.1016/j.cca.2021.10.016. · DOI
- Petersen, PH, Jørgensen, LGM, Brandslund, I, De Fine Olivarius, N, Stahl, M. Consequences of bias and imprecision in measurements of glucose and HbA1c for the diagnosis and prognosis of diabetes mellitus. Scand J Clin Laboratory Suppl 2005;65:51–60. https://doi.org/10.1080/00365510500236135. · DOI · Paper page
- Åsberg, A, Odsæter, IH, Carlsen, SM, Mikkelsen, G. Using the likelihood ratio to evaluate allowable total error – an example with glycated hemoglobin (HbA1c). Clin Chem Lab Med 2014;53:1459–64. https://doi.org/10.1515/cclm-2014-1125. · DOI
- Røys, EÅ, Viste, K, Farrell, CJ, Kellmann, R, Alaour, B, Sylte, MS, et al.. A parametric empirical bayes approach to personalized reference intervals and reference change values. Clin Chem 2025;71:1147–57. https://doi.org/10.1093/clinchem/hvaf092. · DOI
- Gough, A, Sitch, A, Ferris, E, Marshall, T. Within-subject variation of HbA1c: a systematic review and meta-analysis. PLoS One 2023;18:e0289085. https://doi.org/10.1371/journal.pone.0289085. · DOI
- Fraser, CG, Petersen, PH, Libeer, JC, Ricos, C. Proposals for setting generally applicable quality goals solely based on biology. Ann Clin Biochem 1997;34:8–12. https://doi.org/10.1177/000456329703400103. · DOI · Paper page
- Diabetes Control and Complications Trial Research Group, Nathan, DM, Genuth, S, Lachin, J, Cleary, P, Crofford, O. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977–86. https://doi.org/10.1056/NEJM199309303291401. · DOI · Paper page
- UK Prospective Diabetes Study Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). Lancet 1998;352:837–53. · DOI · Paper page
- Weykamp, C, Siebelder, C, Lenters, E, Slingerland, R, English, E. The risk of clinical misinterpretation of HbA1c: modelling the impact of biological variation and analytical performance on HbA1c used for diagnosis and monitoring of diabetes. Clin Chim Acta 2023;548:117495. https://doi.org/10.1016/j.cca.2023.117495. · DOI
- Çubukçu, HC, Vanstapel, F, Thelen, M, van Schrojenstein Lantman, M, Bernabeu-Andreu, FA, Meško Brguljan, P, et al.. APS calculator: a data-driven tool for setting outcome-based analytical performance specifications for measurement uncertainty using specific clinical requirements and population data. Clin Chem Lab Med 2023;62:597–607. https://doi.org/10.1515/cclm-2023-0740. · DOI
- Çubukçu, HC. Computer simulation approaches to evaluate the interaction between analytical performance characteristics and clinical (mis)classification: a complementary tool for setting indirect outcome-based analytical performance specifications. Clin Chem Lab Med 2025;63:1292–300. https://doi.org/10.1515/cclm-2024-1195. · DOI
Metadata
- Authors
- Farrell, C.-J. L., Loh, T. P., Badrick, T., Jones, G. R. D., & Horvath, A. R.
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2026-06-24
- ISSN
- 1434-6621
- 1437-4331
- Cited by
- 1
- References
- 6
- Language
- en
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: An improved approach to modelling patient reclassification: HbA 1c as an example. https://app.cochat.ai/papers/10.1515/cclm-2026-0259
- The paper
- Farrell, C.-J. L., Loh, T. P., Badrick, T., Jones, G. R. D., & Horvath, A. R. (2026). An improved approach to modelling patient reclassification: HbA 1c as an example. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2026-0259
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.