Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners

Open access
Borges, Nattai R., Driller, Matthew W. · 2016 · Journal of Strength and Conditioning Research

The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index; none of the 10 sources its claims cite is checked for support yet.

Abstract

Borges, NR and Driller, MW. Wearable lactate threshold predicting device is valid and reliable in runners. J Strength Cond Res 30(8): 2212–2218, 2016—A commercially available device claiming to be the world's first wearable lactate threshold predicting device (WLT), using near-infrared LED technology, has entered the market. The aim of this study was to determine the levels of agreement between the WLT-derived lactate threshold workload and traditional methods of lactate threshold (LT) calculation and the interdevice and intradevice reliability of the WLT. Fourteen (7 male, 7 female; mean ± SD ; age: 18–45 years, height: 169 ± 9 cm, mass: 67 ± 13 kg, V̇ o 2 max: 53 ± 9 ml·kg −1 ·min −1 ) subjects ranging from recreationally active to highly trained athletes completed an incremental exercise test to exhaustion on a treadmill. Blood lactate samples were taken at the end of each 3-minute stage during the test to determine lactate threshold using 5 traditional methods from blood lactate analysis which were then compared against the WLT predicted value. In a subset of the population ( n = 12), repeat trials were performed to determine both inter-reliability and intrareliability of the WLT device. Intraclass correlation coefficient (ICC) found high to very high agreement between the WLT and traditional methods (ICC > 0.80), with TEMs and mean differences ranging between 3.9–10.2% and 1.3–9.4%. Both interdevice and intradevice reliability resulted in highly reproducible and comparable results (CV < 1.2%, TEM <0.2 km·h −1 , ICC > 0.97). This study suggests that the WLT is a practical, reliable, and noninvasive tool for use in predicting LT in runners.

Overview

CoChat Summary

TL;DR

The wearable lactate threshold predicting device (WLT; BSXinsight multisport edition) showed no significant differences from five traditional blood lactate methods for treadmill workload at lactate threshold and had high test-retest and interdevice reliability in runners. It was most closely correlated with the 4 mmol/L method, though it tended to slightly overestimate LT workload compared with traditional methods.

Key findings

  • One-way ANOVA found no statistically significant differences between LT workloads determined by the WLT and traditional methods (p = 0.60); ICCs ranged from 0.80 to 0.98.
  • Highest agreement was with the 4 mmol/L method: ICC = 0.98, TEM = 0.48 km/h (3.9%), mean difference = 0.18 km/h (1.3%), 95% limits of agreement −0.82 to 1.18 km/h.
  • Interdevice reliability (2 devices during one test, n = 12): CV = 1.2% (0.9–2.1), TEM = 0.16 km/h (0.12–0.28), ICC = 0.97 (0.90–0.99); intradevice reliability (same device, two tests ~24 h apart): CV = 1.2% (0.9–2.1), TEM = 0.16 km/h (0.11–0.27), ICC = 0.97 (0.90–0.99).
  • The WLT tended to overestimate LT workload versus traditional methods; Bland-Altman plots showed outliers versus Dmax and modified Dmax and a downward trend versus the 1 mmol/L method with increasing workload.

Methods

  • Validity/reliability study in 14 runners (7 male, 7 female; VO2max range 39.7–67.0 ml/kg/min; LT speeds 10.9–16.2 km/h); reliability subset n = 12.
  • Treadmill graded exercise test to exhaustion with 3-minute stages set by WLT software, warm-up at 4.8 km/h, second stage 9.3–11.7 km/h, increments 0.3–1.1 km/h until the subject could not maintain the workload; minimum 4 stages required. Blood lactate was collected by finger-prick capillary samples in the final 10 seconds of each stage and 1 minute postexercise; the WLT NIRS device was fitted over the gastrocnemius.
  • Traditional LT methods: linear spline fitting, Dmax, modified Dmax, fixed 4 mmol/L, and first rise >1 mmol/L. Analysis used 1-way ANOVA, ICC with 95% CI, TEM, Bland-Altman plots; reliability data were log-transformed and reported as CV%, TEM, and ICC.

Limitations

  • Stated: the patented WLT design and algorithm make the exact LT prediction methodology difficult to evaluate; this is the first study comparing the WLT with traditional methods, so direct comparisons with existing research are difficult; heterogeneous NIRS methodology (device, measurement site, parameter) may affect compatibility of results.
  • Stated: the range of LT values produced by traditional methods (12.3–13.3 km/h) makes it difficult to conclude whether the WLT provides a valid prediction of LT; the WLT showed small overestimation and method-dependent agreement (outliers vs Dmax/modified Dmax, trend vs 1 mmol/L).
  • Evident: small sample (14 total; 12 reliability) and testing limited to treadmill running, gastrocnemius placement, and threshold range 10.9–16.2 km/h, which may limit generalizability.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1519/jsc.0000000000001307

Cites

23 of 23 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Verification

Claims the paper makes

0 Valid · 0 Not valid · 8 Not checked · 23 Own results and interpretations

7 more claims

Create a free account to see all 31 claims and every source verdict.

See all claims — free account
  • It may state two separate claims as one. 13
  • Probably not a substantial claim. 10

Create a free account to see these claims.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

1 more claim about this paper

Create a free account to see every claim other papers make about this one.

See all claims — free account

Citations

Cites

Show all 23 references

Research works printed but not linked: 1

  1. What is an acceptable amount of measurement error? The application of meaningful "analytical goals" to the reliability of sports science measurements made on a ratio scale. G Atkinson, A Nevill, B Edwards, J Sports Sci. 1718Atkinson, G, Nevill, A, and Edwards, B. What is an acceptable amount of measurement error? The application of meaningful "analytical goals" to the reliability of sports science measurements made on a ratio scale. J Sports Sci 17: 18, 1999.

Cited by

See all 59

Metadata

Authors
Borges, N. R., & Driller, M. W.
Affiliations
  • Central Queensland University (AU)
  • University of Waikato (NZ)
Corresponding author
Borges, Nattai R.
Venue
Journal of Strength and Conditioning Research
Published
2016-08
Volume
30
Issue
8
Pages
2212-2218
ISSN
1064-8011 (ISSN-L)
OpenAlex
W2319419678
Cited by
59
References
23
Citations by year
2026: 5 · 2025: 5 · 2024: 8 · 2023: 3 · 2022: 7 · 2021: 6 · 2020: 4 · 2019: 6 · 2018: 7 · 2017: 4 · 2016: 4
FWCI
4.93
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • The Journal of Strength and Conditioning Research (publishedVersion)
  • PubMed (publishedVersion)
  • Acquire (CQUniversity) (submittedVersion)
  • Figshare (submittedVersion, cc-by)
Topics
  • Health Sciences › Medicine › Orthopedics and Sports Medicine › Sports Performance and Training
  • Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
  • Physical Sciences › Engineering › Biomedical Engineering › Lower Extremity Biomechanics and Pathologies
Keywords
lactate threshold, near-infrared LEDs, runners, incremental exercise test, blood lactate, device reliability, inter-device agreement, treadmill running, endurance training
MeSH terms
Adolescent, Adult, Exercise Test, Female, Humans, Male, Middle Aged, Running, Reproducibility of Results, Anaerobic Threshold, Biomarkers, Spectroscopy, Near-Infrared, Lactic Acid, Young Adult

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

A language model drew the claims from the paper on 2026-11-23, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-11-23 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners. https://app.cochat.ai/papers/10.1519/jsc.0000000000001307
The paper
Borges, N. R., & Driller, M. W. (2016). Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners. Journal of Strength and Conditioning Research. https://doi.org/10.1519/jsc.0000000000001307

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.