A Non-Contact Vital Signs Monitor
Open accessThe DOI matches its Crossref record.
Abstract
The expansion and contraction of the lungs and heart result in movement of the chest wall that can be detected and monitored to determine respiration and heart rate. A prototype non-contact Vital Signs Monitor (VSM) has been developed which uses very low power, high frequency Doppler radar to detect these motions. Digital signal processing (DSP) techniques, imbedded in the VSM, are used to extract heart and respiration rate information from the resultant waveform. A 10-GHz prototype VSM was developed for the Air Force in the mid-1980s using analog technology. The objective was to assess a fallen soldier's clinical condition at distances up to 100 meters before committing resources to assist that individual. An updated and improved version of the original VSM was developed in 1997. This device was designed to operate at shorter distances, use a higher frequency carrier, and provide more specific heart and respiration rate information using digital signal processing techniques. The VSM radar system is a straightforward homodyne receiver. It operates using frequency modulated continuous wave (FM-CW) transmission, which allows for very low power levels. The safe human power density exposure level at its operating frequency of 35 GHz is 10 mW/cm2. A simple approximation using uniform distribution and an antenna aperture of 2 cm by 3 cm gives a power density at the antenna face of 0.017 mW/cm2, nearly a factor of 1000 below the safe level. When the VSM's antenna is trained on the chest wall of a subject, the VSM is capable of measuring and distinguishing minute movements resulting from the mechanical activity of the heart and lungs. As the subject's chest wall moves, the exact phase of the return signal changes. To avoid the possibility of phase-related dead spots, two signals differing in phase by 90 degrees are used to demodulate the signal to baseband (DC). The two resulting "time-varying DC" signals represent the sine and cosine of a phase angle corresponding to the changing position of the target, in this case the motion of the chest wall. The current VSM operates at a frequency of 35 GHz with a corresponding wavelength of only 8.6 mm. This provides a response sensitive enough to detect the small motions caused by cardiac function. The Vital Signs Monitor has several possible application areas. The fact that it is noncontacting would make it especially attractive for monitoring patients in burn units, NICUs, or trauma centers, where attaching electrodes is either inconvenient or not feasible. Results to date indicate a strong correlation between the cardiac component of the motion signal and an electrocardiogram (ECG). With careful signal processing and analysis, it may be possible to extract clinically useful information about cardiac condition, function, or performance from the surface-motion waveform. This could provide a safe, inexpensive, and painless addition to the diagnostic and monitoring tools currently available to cardiologists. Although there are technical obstacles to overcome in filtering gross motions of the subject, the VSM offers significant advances over conventional methods of measuring heart and respiration rate.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Remote Measurements of Heart and Respiration Rates for Telemedicine
Zhao, F., Li, M., Qian, Y., & Tsien, J. Z. (2013) · PLoS ONE · doi:10.1371/journal.pone.0071384 · Cited by 189
“These kinds of contact-based methods may cause undesirable skin irritation, discomfort and soreness to the subject to be measured, especially for neonate [7,8].”
BackgroundReal-Time Monitoring of Respiration Rhythm and Pulse Rate During Sleep
Zhu, X., Chen, W., Nemoto, T., Kanemitsu, Y., Kitamura, K., Yamakoshi, K., & Wei, D. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.884641 · Cited by 171
“There are numerous conventional methods for respiration measurement, such as spirometry, nasal thermocouples, body volume changes, inductance pneumography, impedance plethysmography, strain gauge measurements of thoracic circumference, pneumatic respiration transducers, whole-body plethysmography [6], the fiber-optic…”
MethodNoninvasive Monitoring of Vital Signs Based on Highly Sensitive Fiber Optic Mattress
Wang, S., Ni, X., Li, L., Wang, J., Liu, Q., Yan, Z., Zhang, L., & Sun, Q. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2974313 · Cited by 71
Fekr, A. R., Radecka, K., & Zilic, Z. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2015.2445783 · Cited by 61
Seismocardiogram and Ballistocardiogram Sensing
Postolache, O., Girão, P., & Postolache, G. (2013) · Advanced Instrument Engineering · doi:10.4018/978-1-4666-4165-5.ch017 · Cited by 5
Metadata
- Authors
- Matthews, G., Sudduth, B., & Burrow, M.
- Venue
- Critical Reviews in Biomedical Engineering
- Publisher
- Begell House
- Published
- 2000
- Volume
- 28
- Issue
- 1-2
- Pages
- 173-178
- ISSN
- 0278-940X (ISSN-L)
- OpenAlex
- W3102014066
- Cited by
- 59
- References
- 0
- FWCI
- 0.00
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Non-Contact Vital Signs Monitor. https://app.cochat.ai/papers/10.1615/critrevbiomedeng.v28.i12.290
- The paper
- Matthews, G., Sudduth, B., & Burrow, M. (2000). A Non-Contact Vital Signs Monitor. Critical Reviews in Biomedical Engineering. https://doi.org/10.1615/critrevbiomedeng.v28.i12.290
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.