Accuracy of Fitbit Devices: Systematic Review and Narrative Syntheses of Quantitative Data

Open access

Published version of a preprint

The DOI matches its Crossref record; 86 of the 86 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Although designed as a consumer product to help motivate individuals to be physically active, Fitbit activity trackers are becoming increasingly popular as measurement tools in physical activity and health promotion research and are also commonly used to inform health care decisions. OBJECTIVE: The objective of this review was to systematically evaluate and report measurement accuracy for Fitbit activity trackers in controlled and free-living settings. METHODS: We conducted electronic searches using PubMed, EMBASE, CINAHL, and SPORTDiscus databases with a supplementary Google Scholar search. We considered original research published in English comparing Fitbit versus a reference- or research-standard criterion in healthy adults and those living with any health condition or disability. We assessed risk of bias using a modification of the Consensus-Based Standards for the Selection of Health Status Measurement Instruments. We explored measurement accuracy for steps, energy expenditure, sleep, time in activity, and distance using group percentage differences as the common rubric for error comparisons. We conducted descriptive analyses for frequency of accuracy comparisons within a ±3% error in controlled and ±10% error in free-living settings and assessed for potential bias of over- or underestimation. We secondarily explored how variations in body placement, ambulation speed, or type of activity influenced accuracy. RESULTS: We included 67 studies. Consistent evidence indicated that Fitbit devices were likely to meet acceptable accuracy for step count approximately half the time, with a tendency to underestimate steps in controlled testing and overestimate steps in free-living settings. Findings also suggested a greater tendency to provide accurate measures for steps during normal or self-paced walking with torso placement, during jogging with wrist placement, and during slow or very slow walking with ankle placement in adults with no mobility limitations. Consistent evidence indicated that Fitbit devices were unlikely to provide accurate measures for energy expenditure in any testing condition. Evidence from a few studies also suggested that, compared with research-grade accelerometers, Fitbit devices may provide similar measures for time in bed and time sleeping, while likely markedly overestimating time spent in higher-intensity activities and underestimating distance during faster-paced ambulation. However, further accuracy studies are warranted. Our point estimations for mean or median percentage error gave equal weighting to all accuracy comparisons, possibly misrepresenting the true point estimate for measurement bias for some of the testing conditions we examined. CONCLUSIONS: Other than for measures of steps in adults with no limitations in mobility, discretion should be used when considering the use of Fitbit devices as an outcome measurement tool in research or to inform health care decisions, as there are seemingly a limited number of situations where the device is likely to provide accurate measurement.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2196/10527

Cites

86 of 86 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) flow chart.
Figure 1 Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) flow chart.CC BY · PubMed Central
Figure 2 Percentage error distribution by sample size. Top: Step count in controlled settings. The blue oval indicates extreme outliers (n=4 comparisons). Bottom: Energy expenditure in controlled settings. Solid blue lines indicate mean error estimation. Dotted blue lines indicate 95% CI.
Figure 2 Percentage error distribution by sample size. Top: Step count in controlled settings. The blue oval indicates extreme outliers (n=4 comparisons). Bottom: Energy expenditure in controlled settings. Solid blue lines indicate mean error estimation. Dotted blue lines indicate 95% CI.CC BY · PubMed Central
Figure 3 Step count percentage error in controlled settings. Speed (jog, normal, self-paced, slow, very slow) by body placement (torso, wrist, ankle) of the Fitbit device. Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Gray shading indicates ±3% measurement error.
Figure 3 Step count percentage error in controlled settings. Speed (jog, normal, self-paced, slow, very slow) by body placement (torso, wrist, ankle) of the Fitbit device. Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Gray shading indicates ±3% measurement error.CC BY · PubMed Central
Figure 4 Step count percentage error in controlled settings. Body placement (torso, wrist, ankle) of the Fitbit device by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Gray shading indicates ±3% measurement error.
Figure 4 Step count percentage error in controlled settings. Body placement (torso, wrist, ankle) of the Fitbit device by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Gray shading indicates ±3% measurement error.CC BY · PubMed Central
Figure 5 Step count percentage error in controlled settings. Body motion (normal, constrained, exaggerated, variable) by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal).
Figure 5 Step count percentage error in controlled settings. Body motion (normal, constrained, exaggerated, variable) by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal).CC BY · PubMed Central
Figure 6 Energy expenditure percentage error in controlled settings. Activity versus rest by body placement (torso, wrist). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.
Figure 6 Energy expenditure percentage error in controlled settings. Activity versus rest by body placement (torso, wrist). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.CC BY · PubMed Central
Figure 7 Energy expenditure percentage error in controlled settings. Body placement (torso, wrist) by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.
Figure 7 Energy expenditure percentage error in controlled settings. Body placement (torso, wrist) by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.CC BY · PubMed Central
Figure 8 Energy expenditure percentage error in controlled settings. Speed (jog, normal, slow, self-paced, very slow) by body placement (torso, wrist). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.
Figure 8 Energy expenditure percentage error in controlled settings. Speed (jog, normal, slow, self-paced, very slow) by body placement (torso, wrist). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.CC BY · PubMed Central
Figure 9 Energy expenditure percentage error in controlled settings. Motion limitations (normal, constrained, exaggerated, variable) by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.
Figure 9 Energy expenditure percentage error in controlled settings. Motion limitations (normal, constrained, exaggerated, variable) by speed (jog, normal, self-paced, slow, very slow). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.CC BY · PubMed Central
Figure 10 Energy expenditure percentage error in controlled settings. Type of ambulation (continuous no incline, continuous incline, intermittent) by body placement (torso, wrist). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.
Figure 10 Energy expenditure percentage error in controlled settings. Type of ambulation (continuous no incline, continuous incline, intermittent) by body placement (torso, wrist). Dark lines indicate mean (horizontal). Dashed lines indicate median (horizontal). Triangles indicate measurement by direct calorimetry. Gray shading indicates ±3% measurement error.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

2 more claims about this paper

Create a free account to see every claim other papers make about this one.

See all claims — free account

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 86 references

Research works printed but not linked: 5

  1. Fitbit Reports $571M Q4’17 and $1.616B FY’17 Revenue201802262018-08-01fitbit.comhttps://investor.fitbit.com/press/press-releases/press-release-details/2018/Fitbit-Reports-571M-Q417-and-1616B-FY17-Revenue/default.aspx
  2. Rosenberg, D, 2016, AMIA Annu Symp Proc, 2016, 1050
  3. Popay, J, 2006, Guidance on the Conduct of Narrative Synthesis in Systematic Reviews. A Product From the ESRC Methods Programme. Version 1: April 2006
  4. Montes, J, 2017, J Exerc Physiol Online, 20, 134
  5. Stackpool, CM, 2014, J Fit Res, 3, 32

Cited by

Items 301–400 of 615

Metadata

Authors
Feehan, L. M., Geldman, J., Sayre, E. C., Park, C., Ezzat, A. M., Yoo, J. Y., Hamilton, C. B., & Li, L. C.
Affiliations
University of British Columbia (CA)
Corresponding author
Feehan, Lynne M
Venue
JMIR mHealth and uHealth
Published
2018-08-09
Volume
6
Issue
8
Pages
e10527
ISSN
2291-5222 (ISSN-L)
OpenAlex
W2883411829
Cited by
616
References
86
Citations by year
2026: 69 · 2025: 72 · 2024: 86 · 2023: 62 · 2022: 84 · 2021: 80 · 2020: 130 · 2019: 30 · 2018: 2
FWCI
7.92
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • JMIR mhealth and uhealth (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • Simon Fraser University
  • Canadian Institutes of Health Research (316595)
Topics
  • Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Physical Activity and Health
  • Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
  • Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
Keywords
Fitbit, wearable activity trackers, wearable devices, measurement accuracy, step count, energy expenditure, sleep monitoring, distance measurement, accelerometer, free-living conditions, physical activity measurement, health promotion research, consumer health technology, measurement bias, adults with disabilities, healthy adults, healthcare decision-making

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.