Accuracy of Fitbit Devices: Systematic Review and Narrative Syntheses of Quantitative Data
Open accessPublished version of a preprint
The DOI matches its Crossref record; 86 of the 86 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Although designed as a consumer product to help motivate individuals to be physically active, Fitbit activity trackers are becoming increasingly popular as measurement tools in physical activity and health promotion research and are also commonly used to inform health care decisions. OBJECTIVE: The objective of this review was to systematically evaluate and report measurement accuracy for Fitbit activity trackers in controlled and free-living settings. METHODS: We conducted electronic searches using PubMed, EMBASE, CINAHL, and SPORTDiscus databases with a supplementary Google Scholar search. We considered original research published in English comparing Fitbit versus a reference- or research-standard criterion in healthy adults and those living with any health condition or disability. We assessed risk of bias using a modification of the Consensus-Based Standards for the Selection of Health Status Measurement Instruments. We explored measurement accuracy for steps, energy expenditure, sleep, time in activity, and distance using group percentage differences as the common rubric for error comparisons. We conducted descriptive analyses for frequency of accuracy comparisons within a ±3% error in controlled and ±10% error in free-living settings and assessed for potential bias of over- or underestimation. We secondarily explored how variations in body placement, ambulation speed, or type of activity influenced accuracy. RESULTS: We included 67 studies. Consistent evidence indicated that Fitbit devices were likely to meet acceptable accuracy for step count approximately half the time, with a tendency to underestimate steps in controlled testing and overestimate steps in free-living settings. Findings also suggested a greater tendency to provide accurate measures for steps during normal or self-paced walking with torso placement, during jogging with wrist placement, and during slow or very slow walking with ankle placement in adults with no mobility limitations. Consistent evidence indicated that Fitbit devices were unlikely to provide accurate measures for energy expenditure in any testing condition. Evidence from a few studies also suggested that, compared with research-grade accelerometers, Fitbit devices may provide similar measures for time in bed and time sleeping, while likely markedly overestimating time spent in higher-intensity activities and underestimating distance during faster-paced ambulation. However, further accuracy studies are warranted. Our point estimations for mean or median percentage error gave equal weighting to all accuracy comparisons, possibly misrepresenting the true point estimate for measurement bias for some of the testing conditions we examined. CONCLUSIONS: Other than for measures of steps in adults with no limitations in mobility, discretion should be used when considering the use of Fitbit devices as an outcome measurement tool in research or to inform health care decisions, as there are seemingly a limited number of situations where the device is likely to provide accurate measurement.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures










Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The Fitbit Charge HR was chosen as a lower-cost alternative to actigraphy with comparable performance on sleep/wake classification measures (43).”
From Nightly sleep duration predicts grade point average in the first year of college (2023) · cites this paper as [43]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The Fitbit Flex 2 was chosen as it provided an objective measure of sleep with acceptable accuracy at a reasonable cost (43).”
From Nightly sleep duration predicts grade point average in the first year of college (2023) · cites this paper as [43]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“however, a recent review into the use of Fitbit activity trackers suggests that steps are overestimated in free-living conditions [81].”
From Consumer-based wearable activity trackers increase physical activity participation: systematic review and meta-analysis (2019) · cites this paper as [81]
In the citing paper
“using a modification of the modified Consensus-Based Standards for the Selection of Health Status Measurement Instruments (COSMIN) validation subscale used by Feehan et al [13] and an a priori modified COSMIN reliability subscale (Multimedia Appendix 2) to assess the quality and risk of bias of each study (EC and DF)”
From Reliability and Validity of Commercially Available Wearable Devices for Measuring Steps, Energy Expenditure, and Heart Rate: Systematic Review (2020) · cites this paper as [13]
In the citing paper
“Our interpretation of measurement accuracy was focused on acceptable limits of percentage difference of ±3% in controlled settings and percentage difference of ±10% in free-living settings, as outlined in previous work [13].”
From Reliability and Validity of Commercially Available Wearable Devices for Measuring Steps, Energy Expenditure, and Heart Rate: Systematic Review (2020) · cites this paper as [13]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Preferred Reporting Items for Systematic Reviews and Meta-Analyses: the PRISMA statement
Moher, D., Liberati, A., Tetzlaff, J., Altman, D. G., & for the PRISMA Group (2009) · BMJ · doi:10.1136/bmj.b2535 · Cited by 83,823
Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement
Moher, D., Liberati, A., Tetzlaff, J., & Altman, D. G. (2010) · International Journal of Surgery · doi:10.1016/j.ijsu.2010.02.007 · Cited by 12,156
Mokkink, L. B., Terwee, C. B., Patrick, D. L., Alonso, J., Stratford, P. W., Knol, D. L., Bouter, L. M., & de Vet, H. C. W. (2010) · Quality of Life Research · doi:10.1007/s11136-010-9606-8 · Cited by 4,260 · cited 1×
“All articles were independently assessed for risk of bias by 2 independent reviewers (JG, CP) using a modification of the validation subscale from the checklist for assessing the methodological quality of studies on measurement properties of health status measurement instruments (Consensus-Based Standards for the Sele…”
MethodSteffen, T. M., Hacker, T. A., & Mollinger, L. (2002) · Physical Therapy · doi:10.1093/ptj/82.2.128 · Cited by 2,070
Terwee, C. B., Mokkink, L. B., Knol, D. L., Ostelo, R. W. J. G., Bouter, L. M., & de Vet, H. C. W. (2012) · Quality of Life Research · doi:10.1007/s11136-011-9960-1 · Cited by 1,978 · cited 1×
“We rated each dimension as excellent, good, fair, or poor quality based on a priori modifications to the COSMIN validation subscale scoring criteria appropriate for accuracy studies (Multimedia Appendix 4) [16].”
MethodWhite Paper: “Walking Speed: the Sixth Vital Sign”
Fritz, S., & Lusardi, M. (2009) · Journal of Geriatric Physical Therapy · doi:10.1519/00139143-200932020-00002 · Cited by 1,390 · cited 1×
“When we further explored factors potentially influencing step count accuracy, we observed that accuracy of step count in controlled settings seemed to vary with speed of ambulation (jog, normal, self-paced, slow, or very slow) [50,51], with body placement (torso, wrist, or ankle), and with variations in how the body m…”
Systematic review of the validity and reliability of consumer-wearable activity trackers
Evenson, K. R., Goto, M. M., & Furberg, R. D. (2015) · International Journal of Behavioral Nutrition and Physical Activity · doi:10.1186/s12966-015-0314-1 · Cited by 1,232 · cited 2×
“[Fitbit devices] and their ability to estimate steps, distance, physical activity, energy expenditure, and sleep” [11].”
BackgroundPedometer Measures of Free-Living Physical Activity: Comparison of 13 Models
SCHNEIDER, P. L., CROUTER, S. E., & BASSETT, D. R. (2004) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000113486.60548.e9 · Cited by 752 · cited 2×
“We focused our interpretation of measurement accuracy based on predefined acceptable limits for measurement accuracy in controlled settings as a percentage difference of ±3% and acceptable limits for relative accuracy in free-living settings as a percentage difference of ±10% [19-22].”
MethodAccuracy and Reliability of 10 Pedometers for Measuring Steps over a 400-m Walk
SCHNEIDER, P. L., CROUTER, S. E., LUKAJIC, O., & BASSETT, D. R. (2003) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000089342.96098.c4 · Cited by 633 · cited 2×
“We focused our interpretation of measurement accuracy based on predefined acceptable limits for measurement accuracy in controlled settings as a percentage difference of ±3% and acceptable limits for relative accuracy in free-living settings as a percentage difference of ±10% [19-22].”
MethodHenriksen, A., Haugen Mikalsen, M., Woldaregay, A. Z., Muzny, M., Hartvigsen, G., Hopstock, L. A., & Grimsgaard, S. (2018) · Journal of Medical Internet Research · doi:10.2196/jmir.9157 · Cited by 542
Research works printed but not linked: 5
- Fitbit Reports $571M Q4’17 and $1.616B FY’17 Revenue201802262018-08-01fitbit.comhttps://investor.fitbit.com/press/press-releases/press-release-details/2018/Fitbit-Reports-571M-Q417-and-1616B-FY17-Revenue/default.aspx
- Rosenberg, D, 2016, AMIA Annu Symp Proc, 2016, 1050
- Popay, J, 2006, Guidance on the Conduct of Narrative Synthesis in Systematic Reviews. A Product From the ESRC Methods Programme. Version 1: April 2006
- Montes, J, 2017, J Exerc Physiol Online, 20, 134
- Stackpool, CM, 2014, J Fit Res, 3, 32
Cited by
Fuller, D., Colwell, E., Low, J., Orychock, K., Tobin, M. A., Simango, B., Buote, R., Van Heerden, D., Luan, H., Cullen, K., et al. (2020) · JMIR mHealth and uHealth · doi:10.2196/18694 · Cited by 715
“Studies with fewer than 10 participants were excluded, as has been done in previous work [13].”
BackgroundMethodResultInfluentialBrickwood, K.-J., Watson, G., O'Brien, J., & Williams, A. D. (2019) · JMIR mHealth and uHealth · doi:10.2196/11819 · Cited by 689
“however, a recent review into the use of Fitbit activity trackers suggests that steps are overestimated in free-living conditions [81].”
Tazreean, R., Nelson, G., & Twomey, R. (2022) · Journal of Comparative Effectiveness Research · doi:10.2217/cer-2021-0258 · Cited by 334
Wearable Devices in Cardiovascular Medicine
Hughes, A., Shandhi, M. M. H., Master, H., Dunn, J., & Brittain, E. (2023) · Circulation Research · doi:10.1161/circresaha.122.322389 · Cited by 291
Fitbit-based interventions for healthy lifestyle outcomes: systematic review and meta-analysis
Ringeval, M., Wagner, G., Denford, J., Paré, G., & Kitsiou, S. (2020) · Journal of Medical Internet Research · doi:10.2196/23954 · Cited by 225
“We focused on Fitbit devices because they are among the most accurate commercially available wearables [39-43] and are, in some cases, comparable to research-grade monitors [44].”
ResultGermini, F., Noronha, N., Borg Debono, V., Abraham Philip, B., Pete, D., Navarro, T., Keepanasseril, A., Parpia, S., de Wit, K., & Iorio, A. (2022) · Journal of Medical Internet Research · doi:10.2196/30791 · Cited by 220
“Feehan et al [77] conducted a systematic review on the accuracy of Fitbit devices for measuring physical activity.”
Diving into Sweat: Advances, Challenges, and Future Directions in Wearable Sweat Sensing
Childs, A., Mayol, B., Lasalde-Ramírez, J. A., Song, Y., Sempionatto, J. R., & Gao, W. (2024) · ACS Nano · doi:10.1021/acsnano.4c10344 · Cited by 217
The Influence of Wearables on Health Care Outcomes in Chronic Disease: Systematic Review
Mattison, G., Canfell, O., Forrester, D., Dobbins, C., Smith, D., Töyräs, J., & Sullivan, C. (2022) · Journal of Medical Internet Research · doi:10.2196/36690 · Cited by 171
“Initially designed for the health and fitness industry to track wellness [14], most commercially available wearables can be used to monitor key health-related metrics, including heart rate, sleep quality, energy expenditure, and step count.”
BackgroundDoryab, A., Villalba, D. K., Chikersal, P., Dutcher, J. M., Tumminia, M., Liu, X., Cohen, S., Creswell, K., Mankoff, J., Creswell, J. D., et al. (2019) · JMIR mHealth and uHealth · doi:10.2196/13209 · Cited by 164
“Fitbit Flex has shown to have moderate validity to track activities compared with ActiGraph [16].”
Bauer, M., Glenn, T., Geddes, J., Gitlin, M., Grof, P., Kessing, L. V., Monteith, S., Faurholt-Jepsen, M., Severus, E., & Whybrow, P. C. (2020) · International Journal of Bipolar Disorders · doi:10.1186/s40345-019-0164-x · Cited by 163
Rykov, Y., Thach, T.-Q., Bojic, I., Christopoulos, G., & Car, J. (2021) · JMIR mHealth and uHealth · doi:10.2196/24872 · Cited by 151
“According to a systematic review [42], studies have consistently indicated that Fitbit devices were likely to provide accurate measures of daily step counts.”
Wearable inertial sensors for human movement analysis: a five-year update
Picerno, P., Iosa, M., D’Souza, C., Benedetti, M. G., Paolucci, S., & Morone, G. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.1988849 · Cited by 174
Tedesco, S., Sica, M., Ancillao, A., Timmons, S., Barton, J., & O'Flynn, B. (2019) · JMIR mHealth and uHealth · doi:10.2196/13084 · Cited by 139
“12%), which are largely consistent with findings reported in other studies adopting Fitbit devices to investigate sleep measurement accuracy in healthy young adults in free-living settings (MAPE approximately 10%) [37].”
InfluentialMicrodosing with psilocybin mushrooms: a double-blind placebo-controlled study
Cavanna, F., Muller, S. L., de la Fuente, L. A., Zamberlán, F., Palmucci, M., Janeckova, L., Kuchař, M., Pallavicini, C., & Tagliazucchi, E. (2022) · Translational Psychiatry · doi:10.1038/s41398-022-02039-0 · Cited by 171
“A systematic review showed that these metrics are among the most accurate provided by the wristband [53].”
BackgroundCOVID-19-related mobility reduction: heterogenous effects on sleep and physical activity rhythms
Ong, J. L., Lau, T., Massar, S. A. A., Chong, Z. T., Ng, B. K. L., Koek, D., Zhao, W., Yeo, B. T. T., Cheong, K., & Chee, M. W. L. (2020) · SLEEP · doi:10.1093/sleep/zsaa179 · Cited by 145
“However, validation studies on Fitbit devices conducted by several research groups (see [52] for a review of validation studies for sleep, and [53] for PA) have been encouraging, demonstrating overall high correlation values (r’s > 0.”
BackgroundDoherty, C., Baldwin, M., Keogh, A., Caulfield, B., & Argent, R. (2024) · Sports Medicine · doi:10.1007/s40279-024-02077-2 · Cited by 133
“None of the reviews deemed to be of moderate quality performed a meta-analysis [15, 38, 45].”
InfluentialTang, M. S. S., Moore, K., McGavigan, A., Clark, R. A., & Ganesan, A. N. (2020) · JMIR mhealth and uhealth · doi:10.2196/15576 · Cited by 134
Tedesco, S., Sica, M., Ancillao, A., Timmons, S., Barton, J., & O’Flynn, B. (2019) · PLOS ONE · doi:10.1371/journal.pone.0216891 · Cited by 116
Lang, C. E., Barth, J., Holleran, C. L., Konrad, J. D., & Bland, M. D. (2020) · Sensors · doi:10.3390/s20205744 · Cited by 117
“There was wide variability in the accuracy of these devices in individuals with normal gait speed [42,43].”
BackgroundThe role of wearables in heart failure
Singhal, A., & Cowie, M. R. (2020) · Current Heart Failure Reports · doi:10.1007/s11897-020-00467-x · Cited by 85
Validity of the ActivPAL monitor to distinguish postures: A systematic review
O’Brien, M. W., Wu, Y., Petterson, J. L., Bray, N. W., & Kimmerly, D. S. (2022) · Gait & Posture · doi:10.1016/j.gaitpost.2022.03.002 · Cited by 81
Bonten, T. N., Rauwerdink, A., Wyatt, J. C., Kasteleyn, M. J., Witkamp, L., Riper, H. M., van Gemert‐Pijnen, J. E. W. C., Cresswell, K. M., Sheikh, A., Schijven, M. P., et al. (2020) · Journal of Medical Internet Research · doi:10.2196/17774 · Cited by 101
“In addition, due to rising demands such as time and cost restrictions from policymakers and commercial interests, the quality of eHealth evaluation studies is under pressure [7-9].”
BackgroundHomewood, S. (2023) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3544548.3581505 · Cited by 99
“This is in spite of the fact that ftness tracking devices are often known to be inaccurate, for example Fitbit was found to meet acceptable accuracy for step count approximately half the time and is especially inaccurate in contexts where the walking speed is reduced [32, 103].”
BackgroundChevance, G., Golaszewski, N. M., Tipton, E., Hekler, E. B., Buman, M. P., Welk, G. J., Patrick, K., & Godino, J. (2022) · JMIR mhealth and uhealth · doi:10.2196/35626 · Cited by 92
“The underestimation of activity intensity appears consistent with previous systematic reviews, including different brands of activity monitors, older Fitbits, and/or other criterion measures than those considered in this study (see the study by O’Driscoll et al [17] for energy expenditure, the study by Evenson et al […”
BackgroundMethodInfluentialWearable Devices to Improve Physical Activity and Reduce Sedentary Behaviour: An Umbrella Review
Longhini, J., Marzaro, C., Bargeri, S., Palese, A., Dell’Isola, A., Turolla, A., Pillastrini, P., Battista, S., Castellini, G., Cook, C. E., et al. (2024) · Sports Medicine - Open · doi:10.1186/s40798-024-00678-9 · Cited by 108
“WDs for physical activity tracking, are electronic non-invasive monitoring devices, mainly in the form of wrist devices that enable the tracking of PA metrics (e.g., the number of steps taken, energy expenditure, time spent sleeping and time spent in different activities levels) [7, 8].”
BackgroundMair, J. L., Hayes, L. D., Campbell, A. K., Buchan, D. S., Easton, C., & Sculthorpe, N. F. (2022) · JMIR Formative Research · doi:10.2196/34662 · Cited by 79
St Fleur, R. G., St George, S. M., Leite, R., Kobayashi, M., Agosto, Y., & Jake-Schoffman, D. E. (2021) · JMIR mHealth and uHealth · doi:10.2196/23411 · Cited by 74
“Overall, there have been mixed findings about the accuracy of Fitbit measurements, with some studies indicating step count accuracy 50% of the time compared with research-grade accelerometers [15] and others reporting high validity in step count measurements [16,17].”
BackgroundResultPerformance assessment of new-generation Fitbit technology in deriving sleep parameters and stages
Haghayegh, S., Khoshnevis, S., Smolensky, M. H., Diller, K. R., & Castriotta, R. J. (2019) · Chronobiology International · doi:10.1080/07420528.2019.1682006 · Cited by 78
“Nonetheless, persisting major concern of the public and medical professionals is accuracy (Baron et al. 2018; Evenson et al. 2015; Feehan et al. 2018).”
BackgroundWang, Y., & Wang, X. (2024) · Frontiers in Public Health · doi:10.3389/fpubh.2024.1484848 · Cited by 151
“In terms of health and fitness monitoring tools, Fitbit (46–48) can track users’ heart rate, steps, sleep quality and exercise data in real time.”
Tailored physical activity behavior change interventions: challenges and opportunities
Ma, J. K., Floegel, T. A., LI, L. C., Leese, J., Vera, M. A. D., Beauchamp, M. R., Taunton, J., Liu-Ambrose, T. Y. L., & Allen, K. D. (2021) · Translational Behavioral Medicine · doi:10.1093/tbm/ibab106 · Cited by 80
Self-citationKim, W.-P., Kim, H.-J., Pack, S. P., Lim, J., Cho, C., & Lee, H. (2023) · JAMA Network Open · doi:10.1001/jamanetworkopen.2023.3502 · Cited by 84
A Multimodal Data Fusion Technique for Heartbeat Detection in Wearable IoT Sensors
John, A., Redmond, S. J., Cardiff, B., & John, D. (2021) · IEEE Internet of Things Journal · doi:10.1109/jiot.2021.3093112 · Cited by 76
“Therefore, it is highly beneficial to fuse ECG and PPG signals for improving detection accuracy in ambulatory conditions rather than independently interpreting features from them [25].”
BackgroundComprehensive comparison of Apple Watch and Fitbit monitors in a free-living setting
Bai, Y., Tompkins, C. L., Gell, N., Dione, D. R., Zhang, T., & Byun, W. (2021) · PLoS ONE · doi:10.1371/journal.pone.0251975 · Cited by 78
“Extensive validation studies have been conducted in controlled settings with Fitbit products and literature to date has shown relatively high accuracy in estimating steps and heart rate [19].”
BackgroundMethodSmart Shirts for Monitoring Physiological Parameters: Scoping Review
Khundaqji, H., Hing, W. A., Furness, J. W., & Climstein, M. (2020) · JMIR mhealth and uhealth · doi:10.2196/18092 · Cited by 80
“However, despite demonstrating validity, reliability, and acceptability for their estimates of physiological metrics such as heart rate (HR), the use of these devices is largely targeted toward fitness enthusiasts rather than researchers or clinicians [11-13].”
BackgroundPhysical activity measurement in older adults: Wearables versus self-report
VandeBunte, A. M., Gontrum, E. Q., Goldberger, L., Fonseca, C. S., Djukic, N. A., You, M. Y., Kramer, J. H., & Casaletto, K. B. (2022) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.869790 · Cited by 74
“However, there are still mixed findings in current literature with regard to its validity and reliability (42).”
BackgroundThe Opportunity and Obstacles for Smartwatches and Wearable Sensors
Foster, K. R., & Torous, J. (2019) · IEEE Pulse · doi:10.1109/mpuls.2018.2885832 · Cited by 62
“(University of British Columbia), the situation is mixed [3].”
BackgroundMosler, A., Kemp, J. L., King, M., Lawrenson, P. R., Semciw, A. I., Freke, M., Jones, D., Casartelli, N. C., Wörner, T., Ishøi, L., et al. (2019) · British Journal of Sports Medicine · doi:10.1136/bjsports-2019-101457 · Cited by 67
“27 32 Therefore, the IHiPRN participants recommended that clinicians quantify physical activity using objective measures in the management of hip-related pain in young and middle-aged active adults and also include self-reported measures as appropriate.”
Is That Lawful? Data Privacy and Fitness Trackers in the Workplace
Collins, P., & Marassi, S. (2021) · International Journal of Comparative Labour Law and Industrial Relations · doi:10.54648/ijcl2021003 · Cited by 63
Artificial Intelligence: The Milestone in Modern Biomedical Research
Athanasopoulou, K., Daneva, G. N., Adamopoulos, P. G., & Scorilas, A. (2022) · BioMedInformatics · doi:10.3390/biomedinformatics2040049 · Cited by 91
Digital Health: Implications for Heart Failure Management
Singhal, A., & Cowie, M. (2021) · Cardiac failure review · doi:10.15420/cfr.2020.28 · Cited by 64
Viceconti, M., Penna, S. H., Dartee, W., Mazzà, C., Caulfield, B., Becker, C., Maetzler, W., García‐Aymerich, J., Davico, G., & Rochester, L. (2020) · Sensors · doi:10.3390/s20205920 · Cited by 66
Fiedler, J., Eckert, T., Burchartz, A., Wöll, A., & Wunsch, K. (2021) · Sensors · doi:10.3390/s21082672 · Cited by 67
“While wearables such as Fitbit can easily be attached to the body for 24 h, they are mainly designed for commercial purposes, show limited validity and reliability, and can only provide accurate step counts in adults under certain conditions (no mobility limitation and worn at the torso) but not for energy estimation…”
BackgroundCollins, J. E., Yang, H. Y., Trentadue, T. P., Gong, Y., & Losina, E. (2019) · PLOS ONE · doi:10.1371/journal.pone.0211231 · Cited by 55
“Our findings differ from literature reporting that Fitbit underestimates sedentary time[37]; this may be attributed, in part, to our specific sample of participants with musculoskeletal conditions that limit mobility and impact gait and, in part, to our consideration of HR, which allows us to clearly distinguish no we…”
ResultDigital Technologies to Support Better Outcome and Experience of Care in Patients with Heart Failure
McBeath, K., Angermann, C. E., & Cowie, M. (2022) · Current Heart Failure Reports · doi:10.1007/s11897-022-00548-z · Cited by 58
Lozano‐Lozano, M., Cantarero‐Villanueva, I., Martín‐Martín, L., Galiano‐Castillo, N., Sánchez, M., Fernández‐Lao, C., Postigo-Martín, P., & Arroyo‐Morales, M. (2019) · JMIR mhealth and uhealth · doi:10.2196/14136 · Cited by 58
“Furthermore, monitoring using globally extended systems, such as Fitbit wristbands, is being questioned [30].”
Nightly sleep duration predicts grade point average in the first year of college
Creswell, J. D., Tumminia, M. J., Price, S., Sefidgar, Y., Cohen, S., Ren, Y., Brown, J., Dey, A. K., Dutcher, J. M., Villalba, D., et al. (2023) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2209123120 · Cited by 53
“The Fitbit Charge HR was chosen as a lower-cost alternative to actigraphy with comparable performance on sleep/wake classification measures (43).”
Polar Vantage and Oura Physical Activity and Sleep Trackers: Validation and Comparison Study
Henriksen, A., Svartdal, F., Grimsgaard, S., Hartvigsen, G., & Hopstock, L. A. (2022) · JMIR Formative Research · doi:10.2196/27248 · Cited by 63
“Several systematic reviews on consumer-based activity tracker validity have been published in the last few years, and they reported similar conclusions [27,31,33-38].”
BackgroundSemanik, P. A., Lee, J. J., Pellegrini, C. A., Song, J., Dunlop, D. D., & Chang, R. W. (2019) · ACR Open Rheumatology · doi:10.1002/acr2.11099 · Cited by 54
“In their review, Feehan et al (12) focused on the accuracy of measures derived from Fitbit devices and noted that there was a tendency for the Fitbit to overestimate MVPA in free-living settings compared with an ActiGraph accelerometer, similar to our study.”
ResultMikkelsen, M.-L. K., Berg‐Beckhoff, G., Frederiksen, P., Horgan, G. W., O’Driscoll, R., Palmeira, A. L., Scott, S., Stubbs, R. J., Heitmann, B. L., & Larsen, S. C. (2020) · PLoS ONE · doi:10.1371/journal.pone.0234426 · Cited by 59
Biosensors for Personal Mobile Health: A System Architecture Perspective
Arumugam, S., Colburn, D. A. M., & Sia, S. K. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201900720 · Cited by 51
Pino‐Ortega, J., Gómez‐Carmona, C. D., & Rico-González, M. (2021) · Gait & Posture · doi:10.1016/j.gaitpost.2021.04.015 · Cited by 54
Keogh, A., Alcock, L., Brown, P. M., Buckley, E. E., Brozgol, M., Gazit, E., Hansen, C., Scott, K., Schwickert, L., Becker, C., et al. (2023) · Digital Health · doi:10.1177/20552076221150745 · Cited by 54
“While wrist worn devices are typically well accepted by participants but may be less accurate in terms of gait monitoring.(15,33,43,44) With regards to the second component of the interviews, results need to be considered in line with the limitations of the study.”
ResultGiurgiu, M., Timm, I., Becker, M. L., Schmidt, S., Wunsch, K., Nissen, R., Davidovski, D., Bussmann, J. B. J., Nigg, C. R., Reichert, M., et al. (2022) · JMIR mhealth and uhealth · doi:10.2196/36377 · Cited by 54
“Following practical consideration when applying wearables [255,256], a large number of studies defined a valid day if ≥10 hours of wear time during waking hours were captured.”
LifeSnaps, a 4-month multi-modal dataset capturing unobtrusive snapshots of our lives in the wild
Yfantidou, S., Karagianni, C., Efstathiou, S., Vakali, A. I., Palotti, J., Giakatos, D. P., Marchioro, T., Kazlouski, A., Ferrari, E., & Girdzijauskas, Š. (2022) · Scientific Data · doi:10.1038/s41597-022-01764-x · Cited by 61
Saito, T., Suzuki, H., & Kishi, A. (2022) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.861808 · Cited by 48
Extending factory digital Twins through human characterisation in Asset Administration Shell
Cutrona, V., Bonomi, N., Montini, E., Ruppert, T., Delinavelli, G., & Pedrazzoli, P. (2023) · International journal of computer integrated manufacturing (Print) · doi:10.1080/0951192x.2023.2278108 · Cited by 31
“The classification of the data collected via wearable devices (Feehan et al. 2018) as health data, as the ones modelled in Table 2, is crucial, as any processing of data concerning health is, in principle, prohibited under the GDPR, unless one of the enumerated exceptions applies.”
Jourdan, T., Debs, N., & Frindel, C. (2021) · Sensors · doi:10.3390/s21144808 · Cited by 47
“Previous literature reviews on the validation of wearable sensors were interested in monitoring activity on healthy subjects [15,20–22] or have often placed a focus on a very specific medical application [18,23,24].”
Quantified Self-Using Consumer Wearable Device: Predicting Physical and Mental Health
Pardamean, B., Soeparno, H., Budiarto, A., Mahesworo, B., & Baurley, J. W. (2020) · Healthcare Informatics Research · doi:10.4258/hir.2020.26.2.83 · Cited by 44
“Feehan et al. [12] 2018 The accuracy of Fitbit devices in clinical and research settings.”
BackgroundNissar, G., Khan, R. A., Mushtaq, S., Lone, S. A., & Moon, A. H. (2024) · Multimedia Tools and Applications · doi:10.1007/s11042-024-18580-7 · Cited by 42
RetractedAzza, Y., Grueschow, M., Karlen, W., Seifritz, E., & Kleim, B. (2019) · SLEEP · doi:10.1093/sleep/zsz310 · Cited by 44
Deviations from normal bedtimes are associated with short-term increases in resting heart rate
Faust, L. J., Feldman, K., Mattingly, S. M., Hachen, D. S., & Chawla, N. V. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0250-6 · Cited by 38
The Importance of Data Quality Control in Using Fitbit Device Data From the Research Program
Lederer, L. E., Breton, A., Jeong, H., Master, H., Roghanizad, A. R., & Dunn, J. (2023) · JMIR mhealth and uhealth · doi:10.2196/45103 · Cited by 46
“For example, the reliability of Fitbit Flex’s step count and moderate to vigorous physical activity data was found to be dependent on the activity type (walking, stair stepping, jogging, and incline walking) [24], and step count error was shown to be higher during activity than during rest [20].”
InfluentialCharacterizing and predicting person-specific, day-to-day, fluctuations in walking behavior
Chevance, G., Baretta, D., Heino, M. T. J., Perski, O., Olthof, M., Klasnja, P., Hekler, E. B., & Godino, J. (2021) · PLoS ONE · doi:10.1371/journal.pone.0251659 · Cited by 47
“, physical activity intensity) because the estimation of steps from Fitbit sensors range from good to excellent in comparison with direct observation [40], while the validity of other metrics remains questionable [41].”
BackgroundGiurgiu, M., Kolb, S., Nigg, C., Burchartz, A., Timm, I., Becker, M. L., Rulf, E., Doster, A.-K., Koch, E. D., Bussmann, J. B. J., et al. (2022) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2021-001267 · Cited by 42
Seton, K. A., Espejo, J., Giménez-Orenga, K., Haagmans, R., Ramadan, D. J., & Mehlsen, J. (2024) · Journal of Clinical Medicine · doi:10.3390/jcm13020325 · Cited by 46
“Validation studies concluded that step count has the best accuracy in measuring overall activity levels [277].”
BackgroundJamieson, A., Chico, T., Jones, S. H., Chaturvedi, N., Hughes, A. D., & Orini, M. (2025) · npj Cardiovascular Health · doi:10.1038/s44325-025-00082-6 · Cited by 51
Hamilton, A. C., Lee, N., Stilphen, M., Hu, B., Schramm, S., Frost, F. S., Fox, J., & Rothberg, M. B. (2019) · Journal of Hospital Medicine · doi:10.12788/jhm.3153 · Cited by 38
Wearable activity sensors and early pain after total joint arthroplasty
Patterson, J. T., Wu, H., Chung, C. C., Bendich, I., Barry, J. J., & Bini, S. A. (2020) · Arthroplasty Today · doi:10.1016/j.artd.2019.12.006 · Cited by 33
“The accuracy of this sensor for medical research has been previously validated [6].”
MethodDeep Sleep and Beeps: Sleep Quality Improvement Project in General Surgery Patients
Allen, R. W., Burney, C. P., Davis, A., Henkin, J. R., Kelly, J. L., Judd, B. G., & Ivatury, S. J. (2021) · Journal of the American College of Surgeons · doi:10.1016/j.jamcollsurg.2021.02.010 · Cited by 36
Nakagata, T., Murakami, H., Kawakami, R., Julien Tripette, A., Nakae, S., Yamada, Y., Ishikawa‐Takata, K., Tanaka, S., & Miyachi, M. (2022) · Gait & Posture · doi:10.1016/j.gaitpost.2022.08.004 · Cited by 32
Hartung, V., Sarshar, M., Karle, V. L., Shammas, L., Rashid, A., Roullier, P., Eilers-Petri, C., Mäurer, M., Flachenecker, P., Pfeifer, K., et al. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17249314 · Cited by 35
Femiano, R., Werner, C., Wilhelm, M., & Eser, P. (2022) · Gait & Posture · doi:10.1016/j.gaitpost.2021.11.035 · Cited by 31
Promoting child and adolescent health through wearable technology: A systematic review
Zhang, W., Xiong, K., Zhu, C., Evans, R. D., Zhou, L. M., & Podrini, C. (2024) · Digital Health · doi:10.1177/20552076241260507 · Cited by 43
“Existing reviews provide important insights into the reliability, accuracy, and acceptability of wristband wearables 11,74,75 ; however, there is a need for studies that speci fi cally address the accuracy of activities among adolescents and children.”
BackgroundPostoperative recovery of accelerometer-based physical activity in older cancer patients
Jonker, L. T., Hendriks, S., Lahr, M. M. H., Munster, B. C. V., de Bock, G. H., & van Leeuwen, B. L. (2020) · European Journal of Surgical Oncology · doi:10.1016/j.ejso.2020.06.012 · Cited by 35
Greenberg, J., Mace, R. A., Popok, P., Kulich, R. J., Patel, K. V., Burns, J. W., Somers, T. J., Keefe, F. J., Schatman, M. E., & Vrancenanu, A. M. (2020) · Journal of Pain Research · doi:10.2147/jpr.s266455 · Cited by 32
Kim, B. R., Kwon, H., Chun, M. Y., Park, K. D., Lim, S. M., Jeong, J. H., & Kim, G. H. (2020) · Frontiers in Aging Neuroscience · doi:10.3389/fnagi.2020.549983 · Cited by 36
Anchouche, K., Elharram, M., Oulousian, E., Razaghizad, A., Avram, R., Marquis‐Gravel, G., Randhawa, V. K., Nkulikiyinka, R., Ni, W., Fiuzat, M., et al. (2021) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2021.07.001 · Cited by 31
Strain, T., Wijndaele, K., Pearce, M., & Brage, S. (2022) · Journal for the Measurement of Physical Behaviour · doi:10.1123/jmpb.2021-0046 · Cited by 39
“However, time spent in intensity bands or energy expenditure is generally poorly quantified (Evenson et al., 2015; Feehan et al., 2018; Fuller et al., 2020).”
BackgroundInfluentialHorwitz, A. G., Czyz, E. K., Al‐Dajani, N., Dempsey, W., Zhao, Z., Nahum-Shani, I. B., & Sen, S. (2022) · Journal of Affective Disorders · doi:10.1016/j.jad.2022.06.064 · Cited by 38
Darvall, J. N., Wang, A., Nazeem, M. N., Harrison, C. L., Clarke, L., Mendoza, C., Parker, A. V., Harrap, B., Teale, G. R., Story, D., et al. (2020) · JMIR mhealth and uhealth · doi:10.2196/15112 · Cited by 35
Biondi, A., Laiou, P., Bruno, E., Viana, P. F., Schreuder, M., Hart, W. L., Nurse, E. S., Pal, D. K., & Richardson, M. P. (2020) · JMIR Research Protocols · doi:10.2196/25309 · Cited by 33
“This device was selected from a list of wrist-worn devices based on dependability, durability (battery life of up to 7 days), and user acceptability [31].”
MethodMair, J. L., Hayes, L. D., Campbell, A. K., & Sculthorpe, N. F. (2021) · Journal for the Measurement of Physical Behaviour · doi:10.1123/jmpb.2021-0012 · Cited by 30
“While the accuracy of wearable activity trackers in measuring step count in free-living settings is considered acceptable for normal walking pace (Evenson & Spade, 2020; Feehan et al., 2018; Fokkema et al., 2017), they do not yet provide a valid measure of moderate to vigorous PA (Redenius et al.”
BackgroundResultInfluentialValente, A., Lopes, D. S., Nunes, N., & Esteves, A. (2022) · IEEE Conference on Virtual Reality and 3D User Interfaces (VR) · doi:10.1109/vr51125.2022.00034 · Cited by 34
“Commercial wearables such as Fitbit 1 already provide reliable information on sleep quality by measuring body acceleration, and on stress recognition by measuring electrodermal activity [4,11,12,29].”
BackgroundActivity Tracker–Based Metrics as Digital Markers of Cardiometabolic Health: Cross-Sectional Study
Rykov, Y., Thach, T.-Q., Dunleavy, G., Roberts, A. C., Christopoulos, G., Soh, C. K., & Car, J. (2020) · JMIR mHealth and uHealth · doi:10.2196/16409 · Cited by 29
“We used steps data for cluster analysis because step counts are probably the most accurate Fitbit metric [49].”
BackgroundMethodInfluentialSong, S., Seo, Y., Hwang, S., Kim, H., & Kim, J. (2024) · JMIR mhealth and uhealth · doi:10.2196/55842 · Cited by 35
“Previous studies have confirmed the accuracy of Fitbit activity data [32-34].”
ResultThe Physical Activity Wearables in the Police Force (PAW-Force) study: acceptability and impact
Buckingham, S. A., Morrissey, K., Williams, A. J., Price, L., & Harrison, J. (2020) · BMC Public Health · doi:10.1186/s12889-020-09776-1 · Cited by 34
Wearables for Neurologic Conditions
Minen, M. T., & Stieglitz, E. J. (2020) · Neurology Clinical Practice · doi:10.1212/cpj.0000000000000971 · Cited by 28
“Researchers use commercially available wearables and may wish to specifically contract with the manufacturer to ensure that the research subject’s information has stricter protections than those available to the general public under the provider’s own privacy policy or ensure that data are only shared with the researc…”
BackgroundViciana, J., Casado-Robles, C., Guijarro-Romero, S., & Mayorga-Vega, D. (2022) · Journal of Sports Science and Medicine · doi:10.52082/jssm.2022.356 · Cited by 37
“…that the majority of the previous studies have been performed in adults or special samples such as people with different health problems (e.g., Feehan et al., 2018 or Fokkema et al., 2017), this study expands and complements this area of research, especially among the high school student…”
ResultMontgomery, S. M., Nair, N., Chen, P., & Dikker, S. (2024) · Measurement Sensors · doi:10.1016/j.measen.2024.101075 · Cited by 37
Zhang, Y., Folarin, A., Dineley, J., Conde, P., de Angel, V., Sun, S., Ranjan, Y., Rashid, Z., Stewart, C., Laiou, P., et al. (2024) · Journal of Affective Disorders · doi:10.1016/j.jad.2024.03.106 · Cited by 39
O’Driscoll, R., Turicchi, J., Hopkins, M., Duarte, C., Horgan, G. W., Finlayson, G., & Stubbs, R. J. (2021) · JMIR mhealth and uhealth · doi:10.2196/23938 · Cited by 32
“Feehan et al [45] compared estimates of time in intensities with devices such as ActiGraph and Actical, and concluded that more than 80% of studies reported errors >10% with mean differences ranging between 44% and 632% for estimations of activity above light intensity.”
BackgroundUsing self-monitoring technology for nutritional counseling and weight management
Carpenter, C. A., Ugwoaba, U. A., Cardel, M. I., & Ross, K. M. (2022) · Digital Health · doi:10.1177/20552076221102774 · Cited by 33
“kilocalories “burned” through physical activity), it is important to discourage overreliance on these devicecalculated estimates, as they have been shown to have low validity.(42,45)”
BackgroundValidity of the Polar M430 Activity Monitor in Free-Living Conditions: Validation Study
Henriksen, A., Grimsgaard, S., Horsch, A., Hartvigsen, G., & Hopstock, L. A. (2019) · JMIR Formative Research · doi:10.2196/14438 · Cited by 35
“Similarly, Feehan et al [49] systematically reviewed Fitbit devices and found that validity for EE was low, but validity for measuring steps was higher.”
White, J. W., Finnegan, O., Tindall, N., Nelakuditi, S., Brown, D. E., Pate, R. R., Welk, G. J., de Zambotti, M., Ghosal, R., Wang, Y., et al. (2024) · PLoS ONE · doi:10.1371/journal.pone.0286898 · Cited by 25
“The drawbacks of estimating physical activity based on proprietary algorithms are that it is unclear whether best practice recommendations were used to develop these algorithms, and the algorithms could be updated by these companies at any time unbeknownst to the user [13, 14].”
BackgroundMethodManners, J., Kemps, E., Lechat, B., CATCHESIDE, P. G., Eckert, D. J., & Scott, H. (2025) · Journal of Sleep Research · doi:10.1111/jsr.14480 · Cited by 39
“…highly accurate at bed occupancy timing (i.e. in/out of bed times) and duration compared with polysomnography with video (Ravindran et al., 2021), which wearable devices are typically less accurate at detecting in home environments (Feehan et al., 2018; Haghayegh et al., 2019; Hamill et al., 2020).”
Walsh, J. C., Richmond, J. P., McSharry, J., Groarke, A., Glynn, L. G., Kelly, M. G., Harney, O., & Groarke, J. M. (2021) · JMIR mhealth and uhealth · doi:10.2196/24915 · Cited by 27
Modelling the impacts of COVID-19 on nurse workload and quality of care using process simulation
Qureshi, S. M., Bookey‐Bassett, S., Purdy, N., Greig, M. A., Kelly, H., & Neumann, P. (2022) · PLoS ONE · doi:10.1371/journal.pone.0275890 · Cited by 33
“The Fitbit Alta Tracker provides an accurate measurement of steps in adults in comparison to the distance walked [45].”
BackgroundMethodPerson-specific dose-finding for a digital messaging intervention to promote physical activity.
Hojjatinia, S., Hojjatinia, S., Lagoa, C. M., Brunke-Reese, D., & Conroy, D. E. (2021) · Health Psychology · doi:10.1037/hea0001117 · Cited by 29
IoT Health Data in Electronic Health Records (EHR): Security and Privacy Issues in Era of 6G
Koren, A., & Prasad, R. (2022) · Journal of ICT Standardization · doi:10.13052/jicts2245-800x.1014 · Cited by 28
“Systematic review [14] and summary of 67 studies [15] concluded wearable devices meet acceptable accuracy and offer high reliability.”
BackgroundThe Effects of a Mobile Wellness Intervention with Fitbit Use and Goal Setting for Workers
Lee, S.-H., Ha, Y., Jung, M.-R., Yang, S.-K., & Kang, W. (2019) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2018.0185 · Cited by 30
“27 Multiple systematic reviews to examine the accuracy of Fitbit devices have found that Fitbit devices are most likely to provide an accurate measure of steps in adults.”
Metadata
- Authors
- Feehan, L. M., Geldman, J., Sayre, E. C., Park, C., Ezzat, A. M., Yoo, J. Y., Hamilton, C. B., & Li, L. C.
- Affiliations
- University of British Columbia (CA)
- Corresponding author
- Feehan, Lynne M
- ORCID
- Venue
- JMIR mHealth and uHealth
- Publisher
- JMIR Publications Inc.
- Published
- 2018-08-09
- Volume
- 6
- Issue
- 8
- Pages
- e10527
- ISSN
- 2291-5222 (ISSN-L)
- PMC
- PMC6107736
- OpenAlex
- W2883411829
- Semantic Scholar
- 2a26625e52767429985e16a354821d08a66c5799
- Cited by
- 616
- References
- 86
- Citations by year
- 2026: 69 · 2025: 72 · 2024: 86 · 2023: 62 · 2022: 84 · 2021: 80 · 2020: 130 · 2019: 30 · 2018: 2
- FWCI
- 7.92
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- JMIR mhealth and uhealth (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Simon Fraser University
- Canadian Institutes of Health Research (316595)
- Topics
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Physical Activity and Health
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
- Keywords
- Fitbit, wearable activity trackers, wearable devices, measurement accuracy, step count, energy expenditure, sleep monitoring, distance measurement, accelerometer, free-living conditions, physical activity measurement, health promotion research, consumer health technology, measurement bias, adults with disabilities, healthy adults, healthcare decision-making
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.