Adaptive iterative dose reduction using 3D processing for reduced-and low-dose pulmonary CT: comparison with standard-dose CT for image noise reduction and radiological findings
Abstract
Objective The purpose of this study was to determine the utility of adaptive iterative dose reduction using 3D processing (AIDR 3D) for image noise reduction and assessment of radiologic findings obtained with reduced- and low-dose chest CT in patients with various pulmonary diseases. Subjects and methods Chest CT examinations at three different tube current settings and using 16- and 64-MDCT scanners were performed for 37 patients. Standard-dose (150 mAs) data were reconstructed as thin-section CT without AIDR 3D, and low-dose (25 mAs) and reduced-dose (50 mAs) data were reconstructed as thin-section CT without and with AIDR 3D. To compare image quality, image noises at all CT doses were quantitatively assessed by region of interest measurements. For comparison of radiologic finding assessments, likelihoods of occurrence of emphysema, ground-glass opacity, reticular opacity, bronchiectasis, honeycomb pattern, and nodules were evaluated on a 5-point scale. Then, image noise and agreements of radiologic findings between standard-dose CT and others were statistically evaluated. Results The image quality scores of reduced- and low-dose CT without AIDR 3D were significantly lower than those of both protocols with AIDR 3D and standard-dose CT (p 0.81). Conclusion AIDR 3D is useful for image noise reduction and assessment of radiologic findings obtained with reduced- and low-dose CT for patients with various pulmonary diseases.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Based on the fact that the amount of noise on an image is inversely proportional to the square root of the radiation dose, reducing tube current and tube voltage can inevitably lead to the degradation of image quality, thereby generating a larger amount of noise on the image [30,31].”
From A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation (2020) · cites this paper as [31]
In the citing paper
“Based on the fact that the amount of noise on an image is inversely proportional to the square root of the radiation dose, reducing tube current and tube voltage can inevitably lead to the degradation of image quality, thereby generating a larger amount of noise on the image [30,31].”
From A Novel Singular Value Decomposition-Based Denoising Method in 4-Dimensional Computed Tomography of the Brain in Stroke Patients with Statistical Evaluation (2020) · cites this paper as [31]
Citations
Cites
Silva, A. C., Lawder, H. J., Hara, A., Kujak, J., & Pavlicek, W. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.09.2953 · Cited by 560 · cited 2×
“by using other techniques, such as 3D adaptive raw data filter [11], adaptive statistical iterative reconstruction [2–5], and iterative reconstruction in image space [6, 7].”
MethodReproducibility of histomorphologic diagnoses with special reference to the kappa statistic
SVANHOLM, H., STARKLINT, H., GUNDERSEN, H. J. G., FABRICIUS, J., BARLEBO, H., & OLSEN, S. (1989) · APMIS · doi:10.1111/j.1699-0463.1989.tb00464.x · Cited by 453 · cited 2×
“00) [12].”
ResultRadiation Dose Reduction in Chest CT: A Review
Kubo, T., Lin, P.-J. P., Stiller, W., Takahashi, M., Kauczor, H.-U., Ohno, Y., & Hatabu, H. (2008) · American Journal of Roentgenology · doi:10.2214/ajr.07.2556 · Cited by 256 · cited 2×
“construction [1].”
MethodSelf-citationGervaise, A., Osemont, B., Lecocq, S., Noel, A., Micard, E., Felblinger, J., & Blum, A. (2012) · European Radiology · doi:10.1007/s00330-011-2271-7 · Cited by 251 · cited 3×
“On the other hand, AIDR 3D processing uses a scanner model and a statistical noise model that takes both the number of photons and electrical noise into consideration to eliminate noise due to photon starvation for the projection data [8, 10].”
BackgroundMethodInfluentialSöderberg, M., & Gunnarsson, M. (2010) · Acta Radiologica · doi:10.3109/02841851003698206 · Cited by 239 · cited 1×
“Second, it has been suggested in previous publications that AEC is useful for radiation dose reduction [24–31].”
MethodPontana, F., Duhamel, A., Pagniez, J., Flohr, T., Faivre, J.-B., Hachulla, A.-L., Remy, J., & Remy-Jardin, M. (2011) · European Radiology · doi:10.1007/s00330-010-1991-4 · Cited by 230 · cited 2×
“by using other techniques, such as 3D adaptive raw data filter [11], adaptive statistical iterative reconstruction [2–5], and iterative reconstruction in image space [6, 7].”
MethodPrakash, P., Kalra, M. K., Digumarthy, S. R., Hsieh, J., Pien, H., Singh, S., Gilman, M. D., & Shepard, J.-A. O. (2010) · Journal of Computer Assisted Tomography · doi:10.1097/rct.0b013e3181b26c67 · Cited by 188 · cited 2×
“by using other techniques, such as 3D adaptive raw data filter [11], adaptive statistical iterative reconstruction [2–5], and iterative reconstruction in image space [6, 7].”
MethodPontana, F., Pagniez, J., Flohr, T., Faivre, J.-B., Duhamel, A., Remy, J., & Remy-Jardin, M. (2011) · European Radiology · doi:10.1007/s00330-010-1990-5 · Cited by 180 · cited 2×
“by using other techniques, such as 3D adaptive raw data filter [11], adaptive statistical iterative reconstruction [2–5], and iterative reconstruction in image space [6, 7].”
MethodDiffuse Lung Disease: CT of the Chest with Adaptive Statistical Iterative Reconstruction Technique
Prakash, P., Kalra, M. K., Ackman, J. B., Digumarthy, S. R., Hsieh, J., Do, S., Shepard, J.-A. O., & Gilman, M. D. (2010) · Radiology · doi:10.1148/radiol.10091487 · Cited by 162 · cited 2×
“by using other techniques, such as 3D adaptive raw data filter [11], adaptive statistical iterative reconstruction [2–5], and iterative reconstruction in image space [6, 7].”
MethodSchindera, S. T., Odedra, D., Raza, S. A., Kim, T. K., Jang, H.-J., Szucs-Farkas, Z., & Rogalla, P. (2013) · Radiology · doi:10.1148/radiol.13122349 · Cited by 159
CT of the chest: minimal tube current required for good image quality with the least radiation dose.
Mayo, J. R., Hartman, T. E., Lee, K. S., Primack, S. L., Vedal, S., & Müller, N. L. (1995) · American Journal of Roentgenology · doi:10.2214/ajr.164.3.7863879 · Cited by 154 · cited 1×
“The results of previous studies [32, 33] of image quality and diagnostic quality showed that degradation of image quality can occur at tube current settings higher than those compromising the capability of depicting abnormalities.”
BackgroundRadiation Exposure and Image Quality in Chest CT Examinations
Ravenel, J. G., Scalzetti, E. M., Huda, W., & Garrisi, W. (2001) · American Journal of Roentgenology · doi:10.2214/ajr.177.2.1770279 · Cited by 142
Standard-Dose and 50%—Reduced-Dose Chest CT: Comparing the Effect on Image Quality
Prasad, S. R., Wittram, C., Shepard, J.-A. O., McLoud, T. C., & Rhea, J. T. (2002) · American Journal of Roentgenology · doi:10.2214/ajr.179.2.1790461 · Cited by 130
Israel, G. M., Cicchiello, L. A., Brink, J. A., & Huda, W. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.09.3331 · Cited by 114 · cited 1×
“Second, it has been suggested in previous publications that AEC is useful for radiation dose reduction [24–31].”
MethodPapadakis, A. E., Perisinakis, K., & Damilakis, J. E. (2008) · Medical Physics · doi:10.1118/1.2977535 · Cited by 95 · cited 1×
“Second, it has been suggested in previous publications that AEC is useful for radiation dose reduction [24–31].”
MethodLow-Dose Spiral Computed Tomography of the Thorax: Comparison with the Standard-Dose Technique
Takahashi, M., Maguire, W., Ashtari, M., Khan, A., PAPP, Z., Alberico, R. A., CAMPBELL, W., Eacobacci, T. M., & Herman, P. G. (1998) · Investigative Radiology · doi:10.1097/00004424-199802000-00002 · Cited by 89 · cited 1×
“The results of previous studies [32, 33] of image quality and diagnostic quality showed that degradation of image quality can occur at tube current settings higher than those compromising the capability of depicting abnormalities.”
BackgroundLow-Dose Chest CT: Optimizing Radiation Protection for Patients
Zhu, X., Yu, J., & Huang, Z. (2004) · American Journal of Roentgenology · doi:10.2214/ajr.183.3.1830809 · Cited by 82
Comparison of low-dose and standard-dose helical CT in the evaluation of pulmonary nodules
Karabulut, N., Törü, M., Gelebek, V., Gülsün, M., & Ariyurek, M. O. (2002) · European Radiology · doi:10.1007/s00330-002-1368-4 · Cited by 67
Asbestos-related Pleuropulmonary Diseases: Evaluation with Low-Dose Four–Detector Row Spiral CT
Rémy‐Jardin, M., Sobaszek, A., Duhamel, A., Mastora, I., Zanetti, C., & Rémy, J. (2004) · Radiology · doi:10.1148/radiol.2331031133 · Cited by 67
Adaptive Statistical Iterative Reconstruction Technique for Pulmonary CT
Yanagawa, M., Honda, O., Yoshida, S., Kikuyama, A., Inoue, A., Sumikawa, H., Koyama, M., & Tomiyama, N. (2010) · Academic Radiology · doi:10.1016/j.acra.2010.05.014 · Cited by 63 · cited 3×
“Although doses lower than those used in the present study have been suggested as acceptable without and with adaptive statistical iterative reconstruction, those studies could not assess artifacts from chest wall, fat, vertebrae, ribs, and mediastinal structures such as trachea, great vessels, and cardiac motions [5].”
BackgroundMethodJUNG, K.-J., Lee, K. S., Kim, S. Y., Kim, T. S., Pyeun, Y. S., & LEE, J. Y. (2000) · Investigative Radiology · doi:10.1097/00004424-200009000-00007 · Cited by 63 · cited 1×
“Second, it has been suggested in previous publications that AEC is useful for radiation dose reduction [24–31].”
Methodd’Agostino, A. G., Rémy-Jardin, M., Khalil, C., Delannoy-Deken, V., Flohr, T. G., Duhamel, A., & Rémy, J. (2006) · European Radiology · doi:10.1007/s00330-006-0218-1 · Cited by 62 · cited 1×
“Body mass index has been shown to influence artifacts on chest CT [34–36].”
BackgroundZaporozhan, J., Ley, S., Weinheimer, O., Eberhardt, R., Tsakiris, I., Noshi, Y., Herth, F. J. F., & Kauczor, H. (2006) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-200605000-00018 · Cited by 60
Chiu, C., Chern, M., Wu, M., Hsu, W., Wu, Y., Huang, M.-H., & Chang, S.-C. (2003) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/s0022-5223(03)00033-3 · Cited by 49
Joemai, R. M. S., Veldkamp, W. J. H., Kroft, L. J. M., Giron, I. H., & Geleijns, J. (2013) · American Journal of Roentgenology · doi:10.2214/ajr.12.9780 · Cited by 42
Multidetector CT of Bronchiectasis: Effect of Radiation Dose on Image Quality
Yi, C. A., Lee, K. S., Kim, T. S., Han, D., Sung, Y. M., & Kim, S. (2003) · American Journal of Roentgenology · doi:10.2214/ajr.181.2.1810501 · Cited by 40 · cited 1×
“Second, it has been suggested in previous publications that AEC is useful for radiation dose reduction [24–31].”
MethodDinkel, H., Sonnenschein, M., Hoppe, H., & Vock, P. (2003) · European Radiology · doi:10.1007/s00330-002-1647-0 · Cited by 29
Use of 3D Adaptive Raw-Data Filter in CT of the Lung: Effect on Radiation Dose Reduction
Kubo, T., Ohno, Y., Gautam, S., Lin, P. P., Kauczor, H., Hatabu, H., & iLEAD Study Group (2008) · American Journal of Roentgenology · doi:10.2214/ajr/07.2630 · Cited by 29
Self-citationComputer‐aided detection and quantification of cavitary tuberculosis from CT scans
Xu, Z., Bağcı, U., Kübler, A., Luna, B. M., Jain, S. K., Bishai, W. R., & Mollura, D. J. (2013) · Medical Physics · doi:10.1118/1.4824979 · Cited by 28
Botelho, M. P. F., Agrawal, R., Gonzalez-Guindalini, F. D., Hart, E. M., Patel, S. K., Töre, H. F., & Yaghmai, V. (2013) · European Journal of Radiology · doi:10.1016/j.ejrad.2013.07.011 · Cited by 24
Matsumoto, K., Ohno, Y., Koyama, H., Kono, A., Inokawa, H., Onishi, Y., Nogami, M., Takenaka, D., Araki, T., & Sugimura, K. (2009) · European Journal of Radiology · doi:10.1016/j.ejrad.2009.09.012 · Cited by 23
Self-citationWeng, M.-J., Wu, M., Pan, H.-B., Kan, Y.-Y., & Yang, C.-F. (2004) · Clinical Imaging · doi:10.1016/s0899-7071(03)00246-8 · Cited by 22
Low-dose CT of the thorax in cancer follow-up
Yamada, T., Ono, S., Tsuboi, M., Saito, H., Sato, A., Matsuhashi, T., Ishibashi, T., & Takahashi, S. (2003) · European Journal of Radiology · doi:10.1016/j.ejrad.2003.09.017 · Cited by 22
Ley‐Zaporozhan, J., Ley, S., Krummenauer, F., Ohno, Y., Hatabu, H., & Kauczor, H. (2008) · European Journal of Radiology · doi:10.1016/j.ejrad.2008.10.006 · Cited by 19 · cited 1×
“Second, it has been suggested in previous publications that AEC is useful for radiation dose reduction [24–31].”
MethodSelf-citationPatient Preparation and Scanning Techniques
Taylor, C. M., Blum, A. R., & Abbara, S. (2010) · Radiologic Clinics of North America · doi:10.1016/j.rcl.2010.04.011 · Cited by 14 · cited 1×
“Body mass index has been shown to influence artifacts on chest CT [34–36].”
BackgroundNitta, N., Ikeda, M., Sonoda, A., Nagatani, Y., Ohta, S., Takahashi, M., & Murata, K. (2013) · American Journal of Roentgenology · doi:10.2214/ajr.12.10364 · Cited by 12
Sun, Z. (2006) · European Journal of Radiology · doi:10.1016/j.ejrad.2006.10.007 · Cited by 11 · cited 1×
“Body mass index has been shown to influence artifacts on chest CT [34–36].”
BackgroundKoyama, H., Ohno, Y., Yamazaki, Y., Matsumoto, K., Onishi, Y., Takenaka, D., Yoshikawa, T., Nishio, M., Matsumoto, S., Murase, K., et al. (2011) · European Journal of Radiology · doi:10.1016/j.ejrad.2011.03.037 · Cited by 4
Self-citation
Research works printed but not linked: 2
- 10.2214/ajr.07.2630 · doi:10.2214/ajr.07.2630
- Markarian B, 1993, Heitzman’s the lung: radiologic-pathologic correlation, 4
Cited by
Schindera, S. T., Odedra, D., Raza, S. A., Kim, T. K., Jang, H.-J., Szucs-Farkas, Z., & Rogalla, P. (2013) · Radiology · doi:10.1148/radiol.13122349 · Cited by 159
Kim, Y., Kim, Y. K., Lee, B. E., Lee, S. J., Ryu, Y. J., Lee, J. H., & Chang, J. H. (2015) · American Journal of Roentgenology · doi:10.2214/ajr.14.13629 · Cited by 129
“Thus, the former are more affected by any increase in noise [7–9, 17, 18].”
Backgroundden Harder, A. M., Willemink, M. J., de Ruiter, Q. M. B., Schilham, A. M. R., Krestin, G. P., Leiner, T., de Jong, P. A., & Budde, R. P. J. (2015) · European Journal of Radiology · doi:10.1016/j.ejrad.2015.07.011 · Cited by 75
Zhang, W.-L., Li, M., Zhang, B., Geng, H.-Y., Liang, Y.-Q., Xu, K., & Li, S. (2013) · PLoS ONE · doi:10.1371/journal.pone.0081486 · Cited by 63
“The results are in accordance with those reported by Yamada et al [24] and Ohno et al [25].”
BackgroundResultComparing five different iterative reconstruction algorithms for computed tomography in an ROC study
Jensen, K., Martinsen, A. C. T., Tingberg, A., Aaløkken, T. M., & Fosse, E. T. (2014) · European Radiology · doi:10.1007/s00330-014-3333-4 · Cited by 60
Metadata
- Authors
- Ohno, Y., Takenaka, D., Kanda, T., Yoshikawa, T., Matsumoto, S., Sugihara, N., & Sugimura, K.
- Affiliations
- Kobe University (JP)
- Hyogo Prefectural Cancer Center (JP)
- Toshiba (Japan) (JP)
- Venue
- American Journal of Roentgenology
- Publisher
- American Roentgen Ray Society
- Published
- 2012-10
- Volume
- 199
- Issue
- 4
- Pages
- W477-W485
- ISSN
- 0002-9580 (ISSN-L)
- 0361-803X
- 1546-3141
- OpenAlex
- W1964060591
- Cited by
- 79
- References
- 39
- Citations by year
- 2026: 3 · 2025: 2 · 2024: 3 · 2023: 4 · 2022: 4 · 2021: 3 · 2020: 4 · 2019: 5 · 2018: 8 · 2017: 3 · 2016: 8 · 2015: 13 · 2014: 11 · 2013: 7 · 2012: 1
- FWCI
- 7.73
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- American Journal of Roentgenology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Lung Cancer Diagnosis and Treatment
- Keywords
- adaptive iterative dose reduction, low-dose CT, chest CT, image denoising, image quality, lung disease, emphysema, ground-glass opacities, bronchiectasis, honeycomb pattern, pulmonary nodules, interobserver agreement, tube current reduction
- MeSH terms
- Adult, Aged, Aged, 80 and over, Female, Humans, Image Processing, Computer-Assisted, Lung, Lung Diseases, Male, Middle Aged, Radiation Dosage, Radiographic Image Enhancement, Tomography, X-Ray Computed, Phantoms, Imaging, Imaging, Three-Dimensional, Signal-To-Noise Ratio, Multidetector Computed Tomography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.