Estimation of Radiation Exposure for Brain Perfusion CT: Standard Protocol Compared With Deviations in Protocol

Abstract

OBJECTIVE: The purpose of this study was to measure the organ doses and estimate the effective dose for the standard brain perfusion CT protocol and erroneous protocols. MATERIALS AND METHODS: An anthropomorphic phantom with metal oxide semiconductor field effect transistor (MOSFET) detectors was scanned on a 64-MDCT scanner. Protocol 1 used a standard brain perfusion protocol with 80 kVp and fixed tube current of 200 mA. Protocol 2 used 120 kVp and fixed tube current of 200 mA. Protocol 3 used 120 kVp with automatic tube current modulation (noise index, 2.4; minimum, 100 mA; maximum, 520 mA). RESULTS: Compared with protocol 1, the effective dose was 2.8 times higher with protocol 2 and 7.8 times higher with protocol 3. For all protocols, the peak dose was highest in the skin, followed by the brain and calvarial marrow. Compared with protocol 1, the peak skin dose was 2.6 times higher with protocol 2 and 6.7 times higher with protocol 3. The peak skin dose for protocol 3 exceeded 3 Gy. The ocular lens received significant scatter radiation: 177 mGy for protocol 2 and 435 mGy for protocol 3, which were 4.6 and 11.3 times the dose for protocol 1, respectively. CONCLUSION: Compared with the standard protocol, erroneous protocols of increasing the tube potential from 80 kVp to 120 kVp will lead to a three- to fivefold increase in organ doses, and concurrent use of high peak kilovoltage with incorrectly programmed tube current modulation can increase dose to organs by 7- to 11-fold. Tube current modulation with a low noise index can lead to doses to the skin and ocular lens that are close to thresholds for tissue reactions.

Overview

Integrity signals

Registry

Crossref · 10.2214/ajr.12.10031

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Flagged first, then by citations

Research works printed but not linked: 2

  1. Hirata M, 2005, Radiat Med, 23, 97
  2. Wang C, 2012, Poster presentation at 2012 annual meeting of the American Association of Physicists in Medicine

Cited by

Sorted by citations

Metadata

Authors
Hoang, J. K., Wang, C., Frush, D. P., Enterline, D. S., Samei, E., Toncheva, G., Lowry, C., & Yoshizumi, T. T.
Affiliations
  • Duke Medical Center (US)
  • Duke University (US)
Venue
American Journal of Roentgenology
Published
2013-11
Volume
201
Issue
5
Pages
W730-W734
ISSN
  • 0002-9580 (ISSN-L)
  • 0361-803X
  • 1546-3141
OpenAlex
W2112260118
Cited by
37
References
12
Citations by year
2026: 1 · 2025: 1 · 2024: 3 · 2023: 1 · 2022: 4 · 2021: 5 · 2020: 3 · 2019: 1 · 2018: 6 · 2017: 2 · 2016: 1 · 2015: 6 · 2014: 2
FWCI
2.32
Type
journal article
Language
en
Open access
closed
Locations
  • American Journal of Roentgenology (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
  • Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Digital Radiography and Breast Imaging
Keywords
radiation dose assessment, effective dose, organ dose, automatic tube current modulation, noise index, tube potential, tube voltage, skin dose, scattered radiation, anthropomorphic phantom, MOSFET dosimetry, protocol deviations
MeSH terms
Brain, Humans, Radiation Dosage, Radiometry, Tomography, X-Ray Computed, Phantoms, Imaging

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review