Accuracy and Longevity of an Implantable Continuous Glucose Sensor in the PRECISE Study: A 180-Day, Prospective, Multicenter, Pivotal Trial
Open accessAbstract
OBJECTIVE: It is known that continuous glucose monitoring (CGM) systems can lower mean glucose compared with episodic self-monitoring of blood glucose. Implantable CGM systems may provide additional benefits. RESEARCH DESIGN AND METHODS: We studied the Eversense (Senseonics Inc.) implantable CGM sensor in 71 participants aged 18 years and older with type 1 and type 2 diabetes in a 180-day multinational, multicenter pivotal trial. Participants used the CGM system at home and in the clinic. CGM accuracy was assessed during eight in-clinic visits with the mean absolute relative difference (MARD) for venous reference glucose values >4.2 mmol/L as the primary end point. Secondary end points included Clarke Error Grid Analysis and alarm performance. The primary safety outcome was device-related serious adverse events. This trial is registered with ClinicalTrials.gov, number NCT02154126. RESULTS: The MARD value against reference glucose values >4.2 mmol/L was 11.1% (95% CI 10.5, 11.7). Clarke Error Grid Analysis showed 99.2% of samples in the clinically acceptable error zones A and B. Eighty-one percent of hypoglycemic events were detected by the CGM system within 30 min. No device-related serious adverse events occurred during the study. CONCLUSIONS: Our results indicate the safety and accuracy of this new type of implantable CGM system and support it as an alternative for transcutaneous CGM.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Pickup, J. C., Freeman, S. C., & Sutton, A. J. (2011) · BMJ · doi:10.1136/bmj.d3805 · Cited by 541 · cited 2×
“When used regularly, CGM can effectively lower mean glucose compared with fingerstick glucose measurements only (3).”
BackgroundMethodMiller, K. M., Beck, R. W., Bergenstal, R. M., Goland, R. S., Haller, M. J., McGill, J. B., Rodriguez, H., Simmons, J. H., Hirsch, I. B., & for the T1D Exchange Clinic Network (2013) · Diabetes Care · doi:10.2337/dc12-1770 · Cited by 505
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168 · cited 1×
“performance was less in the hypoglycemic range, as is also seen with other CGM products (7,8,11,12).”
Background
Cited by
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069
“In Europe, a new type of implantable rtCGM system is available as an alternative for transcutaneous CGM (26) [C].”
MethodSelf-citationContinuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831
Self-citationContinuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458
Mayer, M., & Baeumner, A. J. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00719 · Cited by 300
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
Heinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Improving the Clinical Value and Utility of CGM Systems: Issues and Recommendations
Petrie, J. R., Peters, A. L., Bergenstal, R. M., Holl, R. W., Fleming, G. A., & Heinemann, L. (2017) · Diabetes Care · doi:10.2337/dci17-0043 · Cited by 135
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
“108,156 Another aspect closely related to accuracy is the reliability of hypo-and hyperglycemic threshold alerts which has been reported in 24.8% of studies, respectively.”
BackgroundInfluentialLorenz, C., Sandoval, W., & Mortellaro, M. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0028 · Cited by 80
“The Eversense CGM sensor measures glucose in aqueous media (ISF) when inserted in subcutaneous tissue.(22,26,33,34) In vitro tests were performed in model solutions composed of phosphate-buffered saline (10 mM phosphate buffer solution, NaH2PO4-H2O = 1.”
A Century of Diabetes Technology: Signals, Models, and Artificial Pancreas Control
Kovatchev, B. (2019) · Trends in Endocrinology & Metabolism · doi:10.1016/j.tem.2019.04.008 · Cited by 80
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
A Review of Research Progress on Glycemic Variability and Gestational Diabetes
Yu, W., Wu, N., Li, L., OuYang, H., Qian, M., & Shen, H. (2020) · Diabetes, Metabolic Syndrome and Obesity: Targets and Therapy · doi:10.2147/dmso.s261486 · Cited by 44
Deiss, D., Irace, C., Carlson, G., Tweden, K. S., & Kaufman, F. R. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0159 · Cited by 44
Real-World Data from the First U.S. Commercial Users of an Implantable Continuous Glucose Sensor
Sanchez, P., Ghosh-Dastidar, S., Tweden, K. S., & Kaufman, F. R. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0234 · Cited by 40
Diabetes and technology in 2030: a utopian or dystopian future?
Kerr, D., Axelrod, C., Hoppe, C., & Klonoff, D. C. (2018) · Diabetic Medicine · doi:10.1111/dme.13586 · Cited by 35
Metadata
- Authors
- Kropff, J., Choudhary, P., Neupane, S., Barnard, K., Bain, S. C., Kapitza, C., Forst, T., Link, M., Dehennis, A., & DeVries, J. H.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2016-11-04
- Volume
- 40
- Issue
- 1
- Pages
- 63-68
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2547651330
- Cited by
- 171
- References
- 3
- FWCI
- 20.49
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.