Investigation of the Accuracy of 18 Marketed Blood Glucose Monitors
The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: Cleared blood glucose monitors (BGMs) for personal use may not always deliver levels of accuracy currently specified by international and U.S. regulatory bodies. This study’s objective was to assess the accuracy of 18 such systems cleared by the U.S. Food and Drug Administration representing approximately 90% of commercially available systems used from 2013 to 2015. RESEARCH DESIGN AND METHODS: A total of 1,035 subjects were recruited to have a capillary blood glucose (BG) level measured on six different systems and a reference capillary sample prepared for plasma testing at a reference laboratory. Products were obtained from consumer outlets and tested in three triple-blinded studies. Each of the three participating clinical sites tested a different set of six systems for each of the three studies in a round-robin. In each study, on average, a BGM was tested on 115 subjects. A compliant BG result was defined as within 15% of a reference plasma value (for BG ≥100 mg/dL [5.55 mmol/L]) or within 15 mg/dL (0.83 mmol/L) (for BG <100 mg/dL [5.55 mmol/L]). The proportion of compliant readings in each study was compared against a predetermined accuracy standard similar to, but more lenient than, current regulatory standards. Other metrics of accuracy included the overall compliance proportion; the proportion of extreme outlier readings differing from the reference value by >20%; modified Bland-Altman analysis including average bias, coefficient of variation, and 95% limits of agreement; and proportion of readings with no clinical risk as determined by the Surveillance Error Grid. RESULTS: The different accuracy metrics produced almost identical BGM rankings. Six of the 18 systems met the predetermined accuracy standard in all three studies, 5 systems met it in two studies, and 3 met it in one study. Four BGMs did not meet the accuracy standard in any of the three studies. CONCLUSIONS: Cleared BGMs do not always meet the level of analytical accuracy currently required for regulatory clearance. This information could assist patients, professionals, and payers in choosing products and regulators in evaluating postclearance performance.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, comparability of the study by Klonoff et al and our study is limited because Klonoff included BGMS which were cleared by the Food and Drug Administration (FDA) when accuracy criteria were milder than current criteria and applied a non-official accuracy standard which is more lenient than ISO and FDA specifications.29”
From System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015 (2020) · cites this paper as [29]
In the citing paper
“In another accuracy study with the 18 best-selling BGMS in the USA by Klonoff et al,29 12 out of 18 tested BGMS were not meeting accuracy requirements similar to ISO 15197.”
From System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015 (2020) · cites this paper as [29]
In the citing paper
“Most meters meet the accuracy standards established by the International Organization for Standardization”
In the citing paper
“In one analysis, 6 of the top 18 best-selling glucose meters met the accuracy standard (19).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [19]
In the citing paper
“Although measurement quality can come close to or equal that of the glucose monitoring systems used by healthcare professionals, several available SMBG systems still do not meet internationally accepted accuracy standards, such as the International Organization for Standardization 15197 standard.”
From Self-Monitoring of Blood Glucose as an Integral Part in the Management of People with Type 2 Diabetes Mellitus (2022) · cites this paper as [24]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
International Organization for Standardization
(2015) · Encyclopedia of Astrobiology · doi:10.1007/978-3-662-44185-5_100574 · Cited by 6,716
Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406
Self-citationKlonoff, D. C., Lias, C., Vigersky, R., Clarke, W., Parkes, J. L., Sacks, D. B., Kirkman, M. S., Kovatchev, B., Panel, T. E. G., Aurand, L., et al. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814539589 · Cited by 174
Self-citationComparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127 · cited 1×
“A BGM was considered to have met the standard unless the proportion compliant was in the “clear rejection zone” (7).”
MethodBreton, M. D., & Kovatchev, B. P. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400309 · Cited by 103
Self-citationHeinemann, L. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0076 · Cited by 62
Performance of Cleared Blood Glucose Monitors
Klonoff, D. C., & Prahalad, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815584797 · Cited by 59
Self-citationComputing the Surveillance Error Grid Analysis: Procedure and Examples
Kovatchev, B. P., Wakeman, C. A., Breton, M. D., Kost, G. J., Louie, R. F., Tran, N. K., & Klonoff, D. C. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814539590 · Cited by 47
Self-citationSchmid, C., Haug, C., Heinemann, L., & Freckmann, G. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0047 · Cited by 34
Blood Glucose Pattern Management in Diabetes: Creating Order from Disorder
Choudhary, P., Genovese, S., & Reach, G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700618 · Cited by 28
Research works printed but not linked: 1
- Fein, 2015, 2014-15 Economic Report on Retail, Mail, and Specialty Pharmacies
Cited by
Items 101–155 of 155
Ober, R. A., & Geist, G. E. (2020) · Journal of the American Association for Laboratory Animal Science · doi:10.30802/aalas-jaalas-19-000140 · Cited by 4
Preoperative optimization of diabetes
Stewart, E., & Selzer, A. R. (2021) · International Anesthesiology Clinics · doi:10.1097/aia.0000000000000351 · Cited by 3
Pleus, S., Heinemann, L., Freckmann, G., Nauck, M. A., Tytko, A., Kaiser, P., & Petersmann, A. (2022) · Diabetes aktuell · doi:10.1055/a-1851-3751 · Cited by 3
Hospital Diabetes Meeting 2022
Huang, J. K., Yeung, A. M., Nguyen, K. T., Xu, N. Y., Preiser, J., Rushakoff, R. J., Seley, J. J., Umpierrez, G. E., Wallia, A., Drincic, A., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221110878 · Cited by 4
Self-citationRimskaya, A. M., Krasnovidova, A. E., & Vitebskaya, A. V. (2023) · Diabetes Mellitus · doi:10.14341/dm12889 · Cited by 3
New technical approach to the diabetes therapy
Dovč, K., & Battelino, T. (2020) · Minerva Pediatrics · doi:10.23736/s0026-4946.20.05865-x · Cited by 3
Schweiger, D. Š., & Battelino, T. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00001-6 · Cited by 3
Comparative study of i-SENS glucometers in neonates using capillary blood samples
Kim, H. N., & Yoon, S. (2021) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2020-1367 · Cited by 3
Setford, S. J., Liu, Z., McColl, D., Phillips, S., Cameron, H., & Grady, M. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211039465 · Cited by 5
Евгеньевна, Р. Г. (2021) · Meditsinskiy sovet = Medical Council · doi:10.21518/2079-701x-2021-12-286-292 · Cited by 3
Factors determining the prognosis of people with type 1 diabetes – current perspective
Araszkiewicz, A., Trzaska, K., Soukharev, A., Spławski, J., Wolbach, J., Marek, P., Dubel, K., Wojak, W., & Zozulińska‐Ziółkiewicz, D. (2025) · Journal of Medical Science · doi:10.20883/medical.e1122 · Cited by 2
Choo, M., Hoy, G. E., Dugan, S. P., McEwen, L. N., Gunaratnam, N. T., Wyckoff, J., Jeevaraaj, T., Saththiyaseelan, A., Ganeikabahu, B., Katulanda, P., et al. (2020) · BMJ Open · doi:10.1136/bmjopen-2020-038148 · Cited by 2
Moriuchi, T., Otaki, Y., Satou, H., Chai, F., Hayashida, Y., Aikawa, R., Sugiyama, T., & Sode, K. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211060865 · Cited by 2
Updated Post-Market Surveillance of the Clinical Accuracy of a Blood Glucose Test Strip Platform
Setford, S. J., Phillips, S., Cameron, H., & Grady, M. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231158394 · Cited by 2
Yoo, S. J., Choi, R., Park, H., Chun, S., & Um, T. H. (2024) · Laboratory Medicine Online · doi:10.47429/lmo.2024.14.4.382 · Cited by 2
Comparative Evaluation of the Precision and Accuracy of Five Available Glucometers in India
Kesavadev, J., Shankar, A., Chandran, G. B., Sreelakshmi, R., & Basanth, A. (2024) · International Journal of Diabetes and Technology · doi:10.4103/ijdt.ijdt_44_24 · Cited by 2
Utility of point‐of‐care vs reference laboratory testing for the evaluation of glucose levels
Andriankaja, O. M., Muñoz‐Torres, F. J., Vergara, J. L., Pérez, C. M., & JOSHIPURA, K. J. (2019) · Diabetic Medicine · doi:10.1111/dme.13922 · Cited by 2
Accuracy of Two CE-Marked Blood Glucose Monitoring System Based on EN ISO 15197:2015
Kim, J.-H., Garo, M. L., Guerra, A., Paparo, M. T., & Russo, A. (2021) · Diabetology · doi:10.3390/diabetology2040021 · Cited by 2
AL-KHALIFA, I. I., Mohammed, S. M., M.Hameed, T., Alawami, M. F., Mohammed, O. A., Alaa, H., Khalid, M., Faraj, A., & Theeyzen, S. (2024) · Research Journal of Pharmacy and Technology · doi:10.52711/0974-360x.2024.00701 · Cited by 1
Accuracy of Glucose Meter Technologies in Low vs High Blood Sugar Ranges: A Bayesian Meta-Analysis
Sujimongkol, C., Daochai, S., Wichakhrueang, S., & Daochai, C. (2025) · Biomedical Engineering International Conference · doi:10.1109/bmeicon66226.2025.11113812 · Cited by 1
Jatmiko, S. W. (2025) · PengabdianMu Jurnal Ilmiah Pengabdian kepada Masyarakat · doi:10.33084/pengabdianmu.v10i6.8761 · Cited by 1
The Effects of Aspartame, Sucrose, and Water on Short-Term Cognitive Performance
Jacobson, M. M., Passey, J. K., Allen, S., Allen, J. G., Stevens, A., Golden-Appleton, A., Smith, M. W., Larson, M. J., Jefferies, L. K., Hancock, C. R., et al. (2026) · Perceptual and Motor Skills · doi:10.1177/00315125251410466 · Cited by 1
Baumstark, A., Jendrike, N. R. M., Pleus, S., Waldenmaier, D., Zschornack, E., Link, M., Kamecke, U., Mende, J., & Freckmann, G. (2019) · Diabetologie und Stoffwechsel · doi:10.1055/a-0987-0015 · Cited by 1
Modeling the SMBG measurement error
Vettoretti, M. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00005-3 · Cited by 1
In-Patient Management of Hyperglycemia during COVID-19 Pandemic
Usman, W. Y. K. M., Jindal, R., Sahu, D., Siddiqui, M. A., & WANGNOO, S. K. (2020) · Apollo Medicine · doi:10.4103/am.am_72_20 · Cited by 1
Braunstein, M. L. (2022) · Health informatics · doi:10.1007/978-3-030-91563-6_11 · Cited by 1
Safety and Performance of Continuous Glucose Monitoring: An Overview
Venkatesan, V., & Singh, H. (2022) · Current Diabetes Reviews · doi:10.2174/1573399818666220525111746 · Cited by 2
Tebaibia, A., Malek, R., Griene, L., Haddam, A. E. M., Nouri, N., Cherrak, A., Belkacem, K., Abrouk, S., Semrouni, M., Gasmi, K., et al. (2023) · Médecine des Maladies Métaboliques · doi:10.1016/j.mmm.2023.04.002 · Cited by 1
Reliability and Measurement Error
Newman, T. B., & Kohn, M. A. (2020) · Cambridge University Press eBooks · doi:10.1017/9781108500111.006
Self-citationPoint-of-care testing in diabetes management
Cozma, A., Vonica, C. L., Sitar-Tăut, A., & Fodor, A. (2019) · Revista română de medicină de laborator · doi:10.2478/rrlm-2019-0014
Konservative und medikamentöse Therapie des Gestationsdiabetes mellitus
Kleinwechter, H. J., & Laubner, K. (2019) · Der Diabetologe · doi:10.1007/s11428-019-00543-4
Herbots, N., Suresh, N. C., Khanna, S., Narayan, S. R., Chow, A. A., Sahal, M., Ram, S., Day, J. M., Pershad, Y., Thinakaran, H. L., et al. (2019) · MRS Advances · doi:10.1557/adv.2019.398
Clinical evaluation of SMBG systems
Clarke, W. L. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00003-x
Analytical performance of SMBG systems
Ahn, D. T., & Klonoff, D. C. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00002-8
Self-citationConsequences of SMBG systems inaccuracy
Campos-Náñez, E., & Fortwaengler, K. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00004-1
Medina, A. M. G., Jojoa, R. I. J., Henao, D. C., Parra, D., Múñoz, Ó., Rubio, C., & Rincón, M. P. (2022) · Revista Colombiana de Endocrinología Diabetes & Metabolismo · doi:10.53853/encr.8.3.625
Sahebkar, M., Ariafar, A., Attarzadeh, F., Rahimi, N., Malkemes, S. J., Rakhshani, M. H., & Assarroudi, A. (2024) · Journal of Research in Medical Sciences · doi:10.4103/jrms.jrms_20_23
Hepatic neuromodulation using temporal interference
Güemes, A., Missey, F., Sahalianov, I. Y., Głowacki, E. D., Malliaras, G. G., & Williamson, A. (2024) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc53108.2024.10782726
A Confront on Monitoring Pediatric Type 1 Diabetes
Sivakumar, T., Kataveni, S., Peter, V., & Yashaswini, T. (2024) · Journal of Pediatrics and Congenital Disorders · doi:10.47275/2379-6707-120
Rivadeneira, P. S., & Mira, A. (2025) · HardwareX · doi:10.1016/j.ohx.2025.e00684
Pavese, F. (2025) · Medical Research Archives · doi:10.18103/mra.v13i8.6697 · Cited by 2
Medical Devices Information Systems in Primary Care
Cardoso, D. F. B. (2022) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT) · Cited by 2
Warade, J. P. (2025) · Journal of Applied Biochemistry & Laboratory Medicine · doi:10.5005/jablm-11031-0015
Medication Deprescribing Practices Among Lifestyle Medicine Providers: A Cross-Sectional Survey
MacField, J. F., Staffier, K. L., Kelly, J. H., Moghe, R. A., & Karlsen, M. C. (2026) · American Journal of Lifestyle Medicine · doi:10.1177/15598276261430281
Ohanian, A., Ghazarian, V., Mkhitaryan, K., Aghajanova, E., Libaridian, L., & Selimyan, R. (2026) · BMJ Open · doi:10.1136/bmjopen-2026-116832
Baghous, H., Aouiche, S., & AZZOUG, S. (2026) · IOSR Journal of Dental and Medical Sciences · doi:10.9790/0853-2503024960
Ryu, O., & Lee, Y. M. (2026) · Health Insurance Review & Assessment Service Research · doi:10.52937/hira.26.6.1.e6
Hassan, F., Haq, A. U., Nadeem, H., Atta, A., Murtaza, Q., & Hashim, M. (2026) · Khyber Journal Of Medical Sciences · doi:10.70520/kjms.v19i2.914
Letelier, P. J., Fierro, G., Cuminao, K. A. H., Huenchuleo, A., Boguen, R. E., & Guzmán, N. H. (2026) · Journal of Medical Biochemistry · doi:10.5937/jomb0-64436
Prediction of Blood Glucose in Human Tissue Across Days by Non-Invasive Near-Infrared Spectroscopy
Gessell, A. J., Ye, M., Arnold, M. A., & Small, G. W. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968261475681
The Diabetes Technology Timeline
Sriravindrarajah, A., Forner, P. M., & SNAITH, J. R. (2026) · Springer eBooks · doi:10.1007/978-3-032-24854-1_1
Aconselhamento e Dispensa de Produtos de Saúde e Produtos de Puericultura em Farmácia Comunitária
Moreira, F. (2026) · Directory of Open access Books (OAPEN Foundation) · doi:10.26537/ed.p.porto.123
Sketch, A. E., Watson, S. E., Montgomery, E. L., Olajuyigbe, T., Wintergerst, K. A., & Thrasher, B. J. (2026) · Journal of Diabetology · doi:10.4103/jod.jod_69_26
A Comprehensive Diabetes Monitoring Device Design for Enhanced Health Management
Ruwaidy Mat Rasul, S. N. S. H. (2025) · South Eastern European Journal of Public Health
Rossi, A., Rossi, M. O., Jiménez, A. C., & Boolani, A. (2026) · Frontiers in Clinical Diabetes and Healthcare · doi:10.3389/fcdhc.2026.1978637
Metadata
- Authors
- Klonoff, D. C., Parkes, J. L., Kovatchev, B. P., Kerr, D., Bevier, W. C., Brazg, R. L., Christiansen, M., Bailey, T. S., Nichols, J. H., & Kohn, M. A.
- Affiliations
- Mills-Peninsula Medical Center, Diabetes Research Institute, San Mateo, CA
- Joan Lee Parkes Consulting, Inc., Bristol, IN
- University of Virginia (US)
- Sansum Diabetes Research Institute (US)
- Rainier Clinical Research Center (US)
- Diablo Clinical Research (US)
- AMCR Institute (US)
- Vanderbilt University (US)
- University of California, San Francisco (US)
- Corresponding author
- Klonoff, David C.
- ORCID
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2018-06-13
- Volume
- 41
- Issue
- 8
- Pages
- 1681-1688
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2808448482
- Cited by
- 144
- References
- 12
- Citations by year
- 2026: 9 · 2025: 13 · 2024: 14 · 2023: 12 · 2022: 15 · 2021: 24 · 2020: 26 · 2019: 25 · 2018: 6
- FWCI
- 16.23
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Diabetes Care (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Abbott Diabetes Care
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Keywords
- blood glucose meters, point-of-care glucose testing, accuracy, regulatory approval, capillary blood glucose, Bland-Altman analysis, consumer product evaluation, diabetes self-monitoring
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Blood Glucose, Commerce, Diabetes Mellitus, Double-Blind Method, Equipment and Supplies, Female, Hematocrit, Humans, Male, Middle Aged, Patient Compliance, Prediabetic State, Reference Values, United States, United States Food and Drug Administration, Blood Glucose Self-Monitoring, Reproducibility of Results, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.