Open-Flow Microperfusion of Subcutaneous Adipose Tissue for On-Line Continuous Ex Vivo Measurement of Glucose Concentration
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: To evaluate a novel technique for on-line continuous glucose measurement in subcutaneous adipose tissue, and to investigate its accuracy for detection of hypoglycemia. RESEARCH DESIGN AND METHODS: The method combined an open-flow microperfusion of subcutaneous adipose tissue using a double lumen catheter and an extracorporeal sensor cell. An isotonic ion-free solution was perfused through the inner lumen of the catheter, equilibrated with the subcutaneous tissue fluid, and sampled through the outer lumen. The recovery was continuously monitored as the ratio between the measured sampled fluid conductivity and the subcutaneous tissue fluid conductivity (assumed to have a constant value of 1.28 S/m at 25°C). Glucose concentration was calculated on-line from the measured glucose in the sampled fluid and the measured recovery in healthy volunteers during hyperglycemic glucose loads (n = 8), hypoglycemic hyperinsulinemic clamp (n = 6), and a 24-h monitoring period (n = 7). RESULTS: Subcutaneous glucose concentrations in the fasting state were 94% of the plasma glucose concentrations in arterialized venous samples. According to the error grid analysis, 96.9% of the on-line measured subcutaneous glucose concentrations during hyperglycemia and 96.3% during hypoglycemia were in accurate or acceptable zones. The mean differences between the measured subcutaneous glucose and the actual plasma glucose concentration were −0.06–3.3 mmol/l (hyperglycemia), and −0.6–1.1 mmol/l (hypoglycemia). CONCLUSIONS: By combining open-flow microperfusion, glucose sensor, and conductivity measurement, glucose concentration in the subcutaneous adipose tissue can be monitored on-line, extracorporeally, and continuously without any in vivo calibration, and gives accurate measurements during hyper- and hypoglycemia.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Schwartz, N. S., Clutter, W. E., Shah, S. D., & Cryer, P. E. (1987) · Journal of Clinical Investigation · doi:10.1172/jci112884 · Cited by 394
Ertefai, S., & Gough, D. A. (1989) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(89)90097-6 · Cited by 52 · cited 1×
“However, since various approaches (in vivo [2-8] and ex vivo measurements [10-19], wick technique [11], microdialysis [12-18], filtrate collection [12], reverse iontophoresis [19]) and calibration techniques (one point in vitro/one point in vivo [17], two points in vivo [6], no net flux protocol [18, 28], or zero flow…”
Result
Cited by
Cline, G. W., Petersen, K. F., Krssak, M., Shen, J., Hundal, R. S., Trajanoski, Z., Inzucchi, S., Dresner, A., Rothman, D. L., & Shulman, G. I. (1999) · New England Journal of Medicine · doi:10.1056/nejm199907223410404 · Cited by 675
Self-citationParkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Methods for reducing biosensor membrane biofouling
Wisniewski, N., & Reichert, M. (2000) · Colloids and Surfaces B: Biointerfaces · doi:10.1016/s0927-7765(99)00148-4 · Cited by 543
Sensors for glucose monitoring: technical and clinical aspects
Koschinsky, T., & Heinemann, L. (2001) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.188 · Cited by 288
Dudde, R., Vering, T., Piechotta, G., & Hintsche, R. (2006) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2006.864477 · Cited by 47
“There are implantable sensors, including needle-type sensors and extracorporal sensors, coupled to iontophoresis [8], microdialysis [10], or microperfusion [11] systems.”
Method
Metadata
- Authors
- Trajanoski, Z., Brunner, G. A., Schaupp, L., Ellmerer, M., Wach, P., Pieber, T. R., Kotanko, P., & Skrabal, F.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 1997-07-01
- Volume
- 20
- Issue
- 7
- Pages
- 1114-1121
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2002225039
- Cited by
- 67
- References
- 4
- FWCI
- 2.87
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Open-Flow Microperfusion of Subcutaneous Adipose Tissue for On-Line Continuous Ex Vivo Measurement of Glucose Concentration. https://app.cochat.ai/papers/10.2337/diacare.20.7.1114
- The paper
- Trajanoski, Z., Brunner, G. A., Schaupp, L., Ellmerer, M., Wach, P., Pieber, T. R., Kotanko, P., & Skrabal, F. (1997). Open-Flow Microperfusion of Subcutaneous Adipose Tissue for On-Line Continuous Ex Vivo Measurement of Glucose Concentration. Diabetes Care. https://doi.org/10.2337/diacare.20.7.1114
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.