Microdialysis of Glucose in Subcutaneous Adipose Tissue up to 3 Weeks in Healthy Volunteers

Wientjes, Klaas J, Vonk, Pieter, Vonk-van Klei, Yvonne, Schoonen, Adelbert JM, Kossen, Nico W · 1998 · Diabetes Care
Cited by 58

The DOI matches its Crossref record.

Abstract

OBJECTIVE: To measure possible changes in dialysate glucose concentrations over time, to validate the diffusional model for glucose transport from tissue to the probe, and to evaluate the actual glucose concentration in adipose tissue. RESEARCH DESIGN AND METHODS: Glucose concentrations in the subcutaneous adipose tissue of five healthy subjects (age 25 ± 2.7 years, BMI 23.2 ± 2.3 kg/m2 [mean ± SD]) were measured by the microdialysis technique and compared with blood glucose. We applied microdialysis probes with hollow fibers of various membrane length (10–35 mm), used eight perfusion flow rates (0.5–20 μl/min), and perfused four glucose solutions (0.0, 2.8, 8.3, 11.1 mmol/l). RESULTS: After implantation, a substantial decrease in glucose recovery to the lowest value of 26 ± 10% of the final plateau value was noted during the first few hours (n = 4). Recovery increased and stabilized after 5–9 days at 84.0 ± 7.4% of capillary blood glucose when a flow rate of 0.5 μ/min was applied. According to the zero net-flux method, the glucose concentration in equilibrium, Cequi, with the surrounding tissue can be obtained. This concentration also decreases; however, 1 h after recovery, Cequi increases again over 1 or 2 days to a stable value that is not significantly different from the measured capillary blood glucose (P < 0.05). Using various perfusion flow rates and probes (membrane length 10–35 mm), it is shown that diffusion is the rate-limiting process for glucose transport through tissue. CONCLUSIONS: Insertion of the microdialysis probes causes damage to the adipose cells and the vascular bed around the probe. Glucose recovery decreases because of a lower blood supply. In 5–9 days, glucose recovery increases; apparently, this time is needed to repair the microstructure of tissue around the probe. After stabilization of the recovery, no loss of probe permeability, which is due to biocompatibility problems, was seen. The change during the 2 days in equilibrium concentration is probably caused by an inflammation reaction that consumes glucose around the probe. The individual increase in recovery during the 1st days after probe insertion until a stable plateau value is reached (flow rate >0 μ/min) is complicated for short-term clinical glucose measurements in adipose tissue. After stabilization, the mean equilibrium concentration of all subjects was equal to the mean capillary blood glucose concentration. Therefore, we conclude that capillary blood glucose concentration probably is the driving force for diffusion through the capillary wall into the probe and is not some interstitial concentration.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2337/diacare.21.9.1481

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 58

Metadata

Authors
Wientjes, K. J., Vonk, P., Vonk-van Klei, Y., Schoonen, A. J., & Kossen, N. W.
Venue
Diabetes Care
Published
1998-09-01
Volume
21
Issue
9
Pages
1481-1488
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W2068127732
Cited by
58
References
0
FWCI
2.33
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Microdialysis of Glucose in Subcutaneous Adipose Tissue up to 3 Weeks in Healthy Volunteers. https://app.cochat.ai/papers/10.2337/diacare.21.9.1481
The paper
Wientjes, K. J., Vonk, P., Vonk-van Klei, Y., Schoonen, A. J., & Kossen, N. W. (1998). Microdialysis of Glucose in Subcutaneous Adipose Tissue up to 3 Weeks in Healthy Volunteers. Diabetes Care. https://doi.org/10.2337/diacare.21.9.1481

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.