Glucose Monitoring at the Arm
Open accessAbstract
OBJECTIVE—We have examined whether rapid changes in blood glucose (BG) result in clinically relevant differences between capillary BG values measured at the forearm and the fingertip and whether local rubbing of the skin before blood sampling can diminish such differences. RESEARCH DESIGN AND METHODS—Capillary BG samples were collected every 15 min for 3–5 h from the fingertip and the forearm of 17 insulin-treated diabetic patients and analyzed with different glucose monitors (FreeStyle, One Touch Ultra, and Soft-Sense). In a subgroup of patients (n = 8), local rubbing of the forearm skin was performed before blood sampling. A rapid increase in BG was induced by oral administration of glucose, and subsequently, a rapid decrease in glucose was induced by intravenous administration of insulin. RESULTS—In the fasting state, the BG values at the fingertip and at the forearm were similar (7.8 ± 2.4 vs. 7.2 ± 2.3 mmol/l, P = 0.06). However, during rapid increase in glucose, BG values at the fingertip were consistently higher than at the forearm (maximal difference 4.6 ± 1.2 mmol/l, P < 0.001). During rapid decrease in glucose, lower BG values were recorded at the fingertip (maximal difference to forearm 5.0 ± 1.0 mmol/l, P < 0.001). At the forearm, BG was delayed by a median of 35 min (P < 0.01) in relation to the fingertip. Rubbing of forearm skin decreased the observed differences but with a large intraindividual and interindividual variability. There were no obvious device-specific differences. CONCLUSIONS—To avoid risky delays of hyperglycemia and hypoglycemia detection, BG monitoring at the arm should be limited to situations in which ongoing rapid changes in BG can be excluded.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“When glucose concentrations are rising or have just risen sharply, fingertip capillary glucose concentrations are greater than those in capillary blood from other sites including forearm, thigh and abdomen (Ellison et al. 2002;Jungheim & Koschinsky, 2002;van der Valk et al. 2002), with variable differences in repeated tests on the same subject (Ellison et al. 2002);”
From Glycaemic index methodology (2005) · cites this paper as Jungheim & Koschinsky, 2002
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“conversely, when blood glucose concentrations fall, fingertip capillary glucose concentrations are lower than those on the forearm (Jungheim & Koschinsky, 2002).”
From Glycaemic index methodology (2005) · cites this paper as Jungheim & Koschinsky, 2002
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Gray's Anatomy: The Anatomical Basis of Medicine and Surgery
Gray, H., Williams, P. L., & Bannister, L. H. (1996) · Medical Entomology and Zoology · Cited by 943
Sensors for glucose monitoring: technical and clinical aspects
Koschinsky, T., & Heinemann, L. (2001) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.188 · Cited by 288
Self-citationGlucose measurement in patients with diabetes mellitus with dermal interstitial fluid
Bantle, J. P., & Thomas, W. (1997) · Journal of Laboratory and Clinical Medicine · doi:10.1016/s0022-2143(97)90044-5 · Cited by 128
Microvascular Blood Flow, Volume, and Velocity Measured by Laser Doppler Techniques in IDDM
Rendell, M., Bergman, T., O'Donnell, G., Drobny, E., Borgos, J., & Bonner, R. F. (1989) · Diabetes · doi:10.2337/diab.38.7.819 · Cited by 125
Stansberry, K. B., Peppard, H. R., Babyak, L. M., Popp, G. M., McNitt, P. M., & Vinik, A. (1999) · Diabetes Care · doi:10.2337/diacare.22.9.1549 · Cited by 101
Comparison of Glucose Levels in Dermal Interstitial Fluid and Finger Capillary Blood
Stout, P. J., Peled, N., Erickson, B., Hilgers, M. E., Racchini, J. R., & Hoegh, T. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/152091501750220046 · Cited by 76
FreeStyle ™ : A Small-Volume Electrochemical Glucose Sensor for Home Blood Glucose Testing
Feldman, B., McGarraugh, G., Heller, A., Bohannon, N. J. V., Skyler, J. S., DeLeeuw, E. R., & Clarke, D. H. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050025177 · Cited by 70
Tanizawa, Y., Matsuda, K., Matsuo, M., Ohta, Y., Ochi, N., Adachi, M., Koga, M., Mizuno, S., Kajita, M., Tanaka, Y., et al. (2000) · Diabetes · doi:10.2337/diabetes.49.1.114 · Cited by 66
Risky Delay of Hypoglycemia Detection by Glucose Monitoring at the Arm
Jungheim, K., & Koschinsky, T. (2001) · Diabetes Care · doi:10.2337/diacare.24.7.1303-a · Cited by 62
Self-citationUse of an Automated Device for Alternative Site Blood Glucose Monitoring
Fineberg, S. E., Bergenstal, R. M., Bernstein, R. M., Laffel, L. M., & Schwartz, S. L. (2001) · Diabetes Care · doi:10.2337/diacare.24.7.1217 · Cited by 59
Cited by
Sorted by citations
Brouns, F., Bjorck, I., Frayn, K. N., Gibbs, A. L., Lang, V., Slama, G., & Wolever, T. M. S. (2005) · Nutrition Research Reviews · doi:10.1079/nrr2005100 · Cited by 1,093
“…glucose concentrations are greater than those in capillary blood from other sites including forearm, thigh and abdomen (Ellison et al. 2002; Jungheim & Koschinsky, 2002; van der Valk et al. 2002), with variable differences in repeated tests on the same subject (Ellison et al. 2002);…”
BackgroundGoldstein, D. E., Little, R. R., Lorenz, R. A., Malone, J. I., Nathan, D., Peterson, C. M., & Sacks, D. B. (2004) · Diabetes Care · doi:10.2337/diacare.27.7.1761 · Cited by 625
A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
Cengiz, E., & Tamborlane, W. V. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0002 · Cited by 390
“, forearm) as compared with finger tip samples.(12) However, at times of rapid change, mainly due to blood flow variability, levels from alternate sites differ considerably.”
Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy
Yadav, J., Rani, A., Singh, V., & Murari, B. M. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.01.005 · Cited by 394
Assessment and monitoring of glycemic control in children and adolescents with diabetes
Rewers, M., Pihoker, C., Donaghue, K. C., Hanås, R., Swift, P. G. F., & KLINGENSMITH, G. J. (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2009.00582.x · Cited by 349
“The use of SMBG during exercise may also allow improved insulin management and a decreased risk for hypoglycemia during and following exercise (47).”
BackgroundAssessment and monitoring of glycemic control in children and adolescents with diabetes
Rewers, M. J., Pillay, K., de Beaufort, C., Craig, M. E., Hanas, R., Acerini, C. L., & Maahs, D. M. (2014) · Pediatric Diabetes · doi:10.1111/pedi.12190 · Cited by 316
Assessment and monitoring of glycemic control in children and adolescents with diabetes
Rewers, M., Pihoker, C., Donaghue, K. C., Hanås, R., Swift, P. G. F., & KLINGENSMITH, G. J. (2007) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2007.00352.x · Cited by 107
“In the fasting state, glucose readings from the forearm are similar to the fingertip (47) (B).”
MethodBerard, L., Blumer, I. R., Houlden, R. L., Miller, D. B., & Woo, V. C. (2013) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2013.01.017 · Cited by 98
Hydrogel‐Forming Microneedles and Applications in Interstitial Fluid Diagnostic Devices
Aroche, A. F., Nissan, H. E., & Daniele, M. A.-A. (2024) · Advanced Healthcare Materials · doi:10.1002/adhm.202401782 · Cited by 93
Continuous Glucose Monitoring: Evidence and Consensus Statement for Clinical Use
Liebl, A., Henrichs, H. R., Heinemann, L., Freckmann, G., Biermann, E., & Thomas, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700227 · Cited by 77
Comparison of Glucose Levels in Dermal Interstitial Fluid and Finger Capillary Blood
Stout, P. J., Peled, N., Erickson, B., Hilgers, M. E., Racchini, J. R., & Hoegh, T. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/152091501750220046 · Cited by 76
Togashi, Y., Shirakawa, J., Okuyama, T., Yamazaki, S., Kyohara, M., Miyazawa, A., Suzuki, T., Hamada, M., & Terauchi, Y. (2016) · Scientific Reports · doi:10.1038/srep25465 · Cited by 78
“In humans, blood glucose measured in the forearm is known to lag more than 30 minutes behind the blood glucose measured at the fingertip during the rapid changes in the blood glucose levels induced by oral glucose loading or intravenous insulin administration 13 .”
Gebhart, S. S. P., Faupel, M. L., Fowler, R. T., Kapsner, C., Lincoln, D., McGee, V., Pasqua, J. J., Steed, L., Wangsness, M., Xu, F., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503321827812 · Cited by 69
Berard, L. D., Siemens, R., & Woo, V. (2018) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2017.10.007 · Cited by 69
“after meals, after exercise and during hypoglycemia), fingertip testing has been shown to more accurately reflect glycemic status than forearm or thigh testing (50,51).”
MethodMinimally invasive capillary blood sampling methods
Hoffman, M. S. F., McKeage, J. W., Xu, J., Ruddy, B. P., Nielsen, P. M. F., & Taberner, A. J. (2023) · Expert Review of Medical Devices · doi:10.1080/17434440.2023.2170783 · Cited by 67
Bina, D. M., Anderson, R., Johnson, M. L., Bergenstal, R. M., & Kendall, D. M. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.981 · Cited by 66
“These current results support the previous findings of Jungheim and Koschinsky (3), who demonstrated that local rubbing of the forearm, while reducing variation in results for some patients, was not a predictably reliable method to decrease observed fingertip-forearm BG differences.”
ResultKoschinsky, T., Jungheim, K., & Heinemann, L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322527030 · Cited by 64
“9,10 Similar findings could be confirmed in daily-life settings.”
Self-citationGlycemic response and glycemic index of semolina spaghetti enriched with barley β-glucan
Chillo, S., Ranawana, D. V., Pratt, M., & Henry, C. J. (2010) · Nutrition · doi:10.1016/j.nut.2010.07.003 · Cited by 61
Stout, P. J., Racchini, J. R., & Hilgers, M. E. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.635 · Cited by 59
Monitoring glycemic control: The cornerstone ofdiabetes care
LeRoith, D., & Smith, D. O. (2005) · Clinical Therapeutics · doi:10.1016/j.clinthera.2005.10.010 · Cited by 58
Strauss, S. M., Wheeler, A. J., Russell, S. L., Brodsky, A., Davidson, R. M., Gluzman, R., Lin, L., Malo, R. G., Salis, B., Schoor, R., et al. (2009) · Journal of Periodontology · doi:10.1902/jop.2009.080542 · Cited by 51
Dunn, T. C., Eastman, R. C., & Tamada, J. A. (2004) · Diabetes Care · doi:10.2337/diacare.27.9.2161 · Cited by 50
Assessing Glucose Monitor Performance—A Standardized Approach
Mahoney, J. J., & Ellison, J. M. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0245 · Cited by 50
Forst, T., Caduff, A., Talary, M., Weder, M., Brändle, M., Kann, P., Flacke, F., Friedrich, C., & Pfützner, A. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.94 · Cited by 51
The Diabetes Research in Children Network (Direcnet) Study Group (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.675 · Cited by 48
The Diabetes Research In Children, N. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322640971 · Cited by 44
Austin, M. M., Haas, L. B., Johnson, T., Parkin, C. G., Parkin, C. L., Spollett, G. R., & Volpone, M. T. (2006) · The Science of Diabetes Self-Management and Care · doi:10.1177/0145721706295873 · Cited by 41
“It has been shown that significant variations in blood glucose levels exist between samples obtained from finger sticks and alternate sites (ie, forearm) when blood glucose levels are changing rapidly.(24,25) It is therefore recommended that patients obtain blood samples from either the fingertips or base of the thumb.”
Physiological Influences on Off-Finger Glucose Testing
McGarraugh, G., Price, D. A., Schwartz, S. L., & Weinstein, R. L. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/15209150152607141 · Cited by 40
Self-monitoring of blood glucose in diabetes: is it worth it?
O’Kane, M., & Pickup, J. C. (2009) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1258/acb.2009.009011 · Cited by 42
Howard, R., Guo, J., & Hall, K. D. (2020) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqaa198 · Cited by 37
Self-monitoring of blood glucose in children and teens with diabetes
Bui, H., Perlman, K., & Daneman, D. (2005) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2005.00095.x · Cited by 37
“Meters that offer AST are US Food and Drug Administration (FDA) approved for use on the forearm, upper arm, palm, abdomen, thigh, and calf (14).”
BackgroundMethodInfluentialHaupt, A., Berg, B., Paschen, P., Dreyer, M., Häring, H.-U., Smedegaard, J., & Matthaei, S. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.597 · Cited by 36
Zhang, Y., Wu, G., Wei, H., Guo, Z., Yang, H., He, Y., Xie, S., & Liu, Y. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.000990 · Cited by 32
Schmid, C., Haug, C., Heinemann, L., & Freckmann, G. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0047 · Cited by 30
Evaluation of CGMS ® During Rapid Blood Glucose Changes in Patients with Type 1 Diabetes
Wilhelm, B., Forst, S., Weber, M. M., Larbig, M., Pfützner, A., & Forst, T. M. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.146 · Cited by 28
Vahlsing, T., Delbeck, S., Leonhardt, S., & Heise, H. M. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818798347 · Cited by 28
Clinical Performance of a Continuous Viscometric Affinity Sensor for Glucose
Diem, P., Kalt, L., Haueter, U., Krinelke, L., Fajfr, R., Reihl, B., & Beyer, U. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.790 · Cited by 27
Laboratory Tests in Diagnosis and Management of Diabetes Mellitus. Practical Considerations
Miedema, K. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.193 · Cited by 26
“Unfortunately, the results from this alternate-site testing do not always equal fingertip values (18).”
BackgroundHe, R., Wei, H., Gu, H., Zhu, Z., Zhang, Y., Guo, X., & Cai, T. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.10.101513 · Cited by 24
Cohen, O., Fine, I., Monashkin, E., & Karasik, A. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503763816418 · Cited by 25
Gingival crevicular blood: As a non-invasive screening tool for diabetes mellitus in dental clinics
Shetty, N., Shankarapillai, R., Mathur, L. K., Manohar, B., Mathur, A., & Jain, M. (2013) · Journal of Indian Society of Periodontology · doi:10.4103/0972-124x.118319 · Cited by 25
“Notably, although some research [25,26] found that the concurrence of glucose measures from CFB and from alternate testing sites is diminished in the postprandial period, other studies [27,28] failed to find that the length of time since last food intake affected this concurrence.”
Heinemann, L., & Boecker, D. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500420 · Cited by 25
Alternate site blood glucose testing: do patients prefer it?
Tieszen, K. L., & New, J. P. (2003) · Diabetic Medicine · doi:10.1046/j.1464-5491.2003.00927.x · Cited by 24
Self-Monitoring of Blood Glucose, Cutaneous Finger Injury, and Sensory Loss in Diabetic Patients
Floch, J. P. L., Bauduceau, B., Lévy, M., Mosnier-Pudar, H., Sachon, C., & Kakou, B. (2008) · Diabetes Care · doi:10.2337/dc08-1174 · Cited by 22
Recommendations for self-monitoring in pediatric diabetes: a consensus statement by the ISPED
Diabetes Study Group of the Italian Society for Pediatric Endocrinology and Diabetology, Scaramuzza, A. E., Cherubini, V., Tumini, S., Bonfanti, R., Buono, P., Cardella, F., d’Annunzio, G., Frongia, A. P., Lombardo, F., et al. (2013) · Acta Diabetologica · doi:10.1007/s00592-013-0521-7 · Cited by 22
“their effectiveness persists [15, 16].”
BackgroundAlternative Site Blood Glucose Testing: A Multicenter Study
Fedele, D., Corsi, A. M., Noacco, C., Prisco, F., Squatrito, S., Torre, E. L. M., Iafusco, D., Errico, M. K., Toniato, R., Nicolucci, A., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322641033 · Cited by 22
“7,8,13 When this study was planned, the initial reports of differences between arm and finger values associated with rapidly changing glucose levels had not yet appeared.”
InfluentialPrecision and accuracy of two blood glucose meters: FreeStyle Flash versus One Touch Ultra
Rivers, S. M., Kane, M. P., Bakst, G., Busch, R. S., & Hamilton, R. A. (2006) · American Journal of Health-System Pharmacy · doi:10.2146/ajhp050473 · Cited by 22
Laione, F., & Marques, J. L. B. (2005) · Medical & Biological Engineering & Computing · doi:10.1007/bf02344732 · Cited by 20
“There is also the possibility of a delay in the detection time during rapid changes in glucose (JUNGHEIM and KOSCHINSKY, 2002).”
BackgroundLucidarme, N., Alberti, C., Zaccaria, I., Claude, E., & Tubiana‐Rufi, N. (2005) · Diabetes Care · doi:10.2337/diacare.28.3.710 · Cited by 20
“It should be noted that published studies usually evaluated these differences in individuals experiencing rapid blood glucose variations during oral glucose tolerance testing (4,10).”
BackgroundMethodResultInfluentialGingival Crevicular Blood as a Potential Screening Tool: A Cross Sectional Comparative Study
Rapone, B., Ferrara, E., Santacroce, L., Topi, S., Converti, I., Gnoni, A., Scarano, A., & Scacco, S. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17207356 · Cited by 19
Kempe, K. C., Budd, D., Stern, M., Ellison, J. M., Saari, L. A., Adiletto, C. A., Olin, B., Price, D. A., & Horwitz, D. L. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.916 · Cited by 19
Kempe, K. C., Price, D. A., Ellison, J., Marhoul, J. C., Morrow, L., Win, K., Kuschma, K., & Hompesch, M. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0023 · Cited by 17
“5,6 For patients who use alternative testing sites, these physiological differences can complicate data interpretation, pattern analysis, and therapy adjustments.”
Ito, T., Kamoi, K., Minagawa, S., Kimura, K., & Kobayashi, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400420 · Cited by 16
“(2) Did you think it may be frightening to insert?”
BackgroundAlternative-Site Blood Glucose Measurement at the Abdomen
van der Valk, P. R., van der Schatte Olivier-Steding, I., Wientjes, K.-J. C., Schoonen, A. J., & Hoogenberg, K. (2002) · Diabetes Care · doi:10.2337/diacare.25.11.2114 · Cited by 15
Evaluation of Portable Blood Glucose Meters. Problems and Recommendations
Slingerland, R. J., & Miedema, K. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.186 · Cited by 13
Knapp, P. E., Showers, K. M., Phipps, J. C., Speckman, J. L., Sternthal, E., Freund, K. M., Ash, A. S., & Apovian, C. M. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0060 · Cited by 12
“These findings have been attributed to a delayed equilibration of arm and thigh skin with arterial blood glucose resulting from lesser blood flow.(10) Manufacturers of SMBG devices have recommended preparatory rubbing or heating of AST sites to increase blood flow and mitigate the difference due to equilibration time,…”
Wu, M., Fang, M.-Y., Jen, L.-N., Hsiao, H.-C., Müller, A., & Hsu, C. (2008) · Clinical Chemistry · doi:10.1373/clinchem.2008.106328 · Cited by 11
“Although commercially available glucose meters have been developed for alternative-site testing, the stabilityofmeasurementshasbeenanongoingtopicof studyinarmtests(18, 19 ).Theaimofthisstudywasto characterize and evaluate the clinical performance of the Bionime Rightest GM310 meter for alternative-site tests (e.g., fi…”
Method[History, accuracy and precision of SMBG devices].
Dufaitre-Patouraux, L., Vague, P., & Lassmann-Vague, V. (2003) · PubMed · Cited by 11
Owiredu, W. K. B. A., Amegatcher, G., & Amidu, N. (2009) · Journal of Medical Sciences(Faisalabad) · doi:10.3923/jms.2009.185.193 · Cited by 10
In Vivo Evaluation of Fine Needle Amperometric Glucose Sensors Implanted in Rabbit's Blood Vessel
Edagawa, K., Fuchiwaki, Y., & Yasuzawa, M. (2013) · Journal of The Electrochemical Society · doi:10.1149/2.016402jes · Cited by 9
Park, K., Park, M.-S., Cha, Y.-J., Kim, W., Choi, S.-S., Kim, K. O., Cha, E., & Kim, K.-A. (2010) · Annals of Laboratory Medicine · doi:10.3343/kjlm.2010.30.3.264 · Cited by 9
“The ~35-min time delay between GA and GF under rapidly changing glucose concentrations observed by Jungheim and Koschinsky [29] may have been caused by the higher capillary circulation on the fingertip, which is 5-20 times higher than that at the forearm [32-34], thereby leading to a slower glucose dynamics.”
InfluentialKoschinsky, F. D. A. P. T., Junker, R., Luppa, P. B., & Schlebusch, H. (2009) · Journal of Laboratory Medicine · doi:10.1515/jlm.2009.058 · Cited by 9
Continuous glucose monitoring: key challenges to replacing episodic SMBG
Stout, P. J., Racchini, J. R., Hilgers, M. E., & Noujaim, S. E. (2006) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(06)70007-9 · Cited by 9
Stavrianos, C., & Anastasiou, E. (2004) · Diabetes Care · doi:10.2337/diacare.27.2.627 · Cited by 9
Zhang, Y., Wei, H., Yang, H., He, Y., Wu, G., Xie, S., Zhu, Z., & He, R. (2012) · Journal of Biophotonics · doi:10.1002/jbio.201200128 · Cited by 8
Pavithran, A. A., Ramamoorthy, L., Suryanarayana, B. S., Murugesan, R., & Kumari, M. J. (2020) · Europe PMC (PubMed Central) · doi:10.34172/jcs.2020.028 · Cited by 8
Estimating prevalence of diabetes in a Congolese town was feasible
Muyer, M. T., Muls, E., Mapatano, M. A., Makulo, R., Mvitu, M., Kimenyembo, W., Mandja, M., Kimbondo, P., Bieleli, E., Kaimbo, D. K. W., et al. (2010) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2010.03.014 · Cited by 8
“Additionally, during a fast, change of glycemia appears 30 minutes later in a venous sample taken from the forearm compared with a finger capillary sample [48].”
BackgroundTechnologie et fiabilité de l’autosurveillance glycémique : historique et état actuel
Dufaitre-Patouraux, L., Vague, P., & Lassmann-Vague, V. (2003) · Diabetes & Metabolism · doi:10.1016/s1262-3636(03)72768-2 · Cited by 8
Stephens, E., & Riddle, M. C. (2003) · Reviews in Endocrine and Metabolic Disorders · doi:10.1023/a:1027345927676 · Cited by 8
“However with rapid changes in glucose levels, changes at the forearm were delayed on average about 35 minutes in relation to the fingertip [ 38 ].”
MethodAlternative Site Self Blood Glucose Testing Is Preferred by Women with Gestational Diabetes
Clarke, P., Coleman, M. A. G., & Holt, R. I. G. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.604 · Cited by 7
Tomita, T., Goto, H., Sumiya, K., Yoshida, T., Kudo, K., & Kohda, Y. (2018) · Diabetology International · doi:10.1007/s13340-018-0372-0 · Cited by 7
Capillary blood in core laboratories: current and future challenges
González, Á., Deza, S., Maroto‐García, J., & Varo, N. (2025) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2025-0888 · Cited by 6
“The arm is another site used for blood collection, although it has less irrigation than the fi nger, since there are no anastomoses, making extraction slower and more complicated for patients or untrained personnel, and producing more hemolysis [77, 78].”
The accuracy and acceptability of performing capillary blood glucose measurements at the earlobe
Chan, H. Y. L., Lau, T. S. L., Ho, S. Y., Leung, D. Y. P., & Lee, D. (2016) · Journal of Advanced Nursing · doi:10.1111/jan.12944 · Cited by 6
“Rather, evidence showed that the palm is a more suitable alternative site in both steady and dynamic glycaemic conditions (Jungheim & Koschinsky 2002a, Bina et al. 2003, Kempe et al. 2005).”
BackgroundMethodNew Developments in Home Glucose Monitoring: Minimizing the Pain
Gabbay, R. A., Phd, M., & Gabbay, R. A. (2015) · Cited by 6
Hypoglycemia Detection Rate Differs Among Blood Glucose Monitoring Sites
Meguro, S., Funae, O., Hosokawa, K., & Atsumi, Y. (2005) · Diabetes Care · doi:10.2337/diacare.28.3.708 · Cited by 6
Blood glucose testing at alternate sites in newborn infants
Norman, M., Schiött, J., Åkerström, S., Lattunen, N., & Berggren-Broström, E. (2005) · Acta Paediatrica · doi:10.1080/08035250510039865 · Cited by 6
Diabetes diagnostics including analytical methods for glucose monitoring
Wahl, H. G., & Koschinsky, T. (2018) · Springer eBooks · doi:10.1007/978-3-662-54497-6_12 · Cited by 5
Self-citationHow and when to use an alternative site in self-monitoring of blood glucose
Dufaitre-Patouraux, L., Djemli, K., & Vague, P. (2004) · Diabetes & Metabolism · doi:10.1016/s1262-3636(07)70145-3 · Cited by 5
“In a study on 17 type 1 diabetic patients [16, 17] they show an approximate 30 minutes delay in displaying the glycaemia variations on AS both at the occasion of an hypoglycaemia (insulin hypoglycaemia) or an hyperglycaemia (after absorption of 75 g of carbohydrates).”
BackgroundDooley, G., Godbole, S. V., Wrobel, J., Henthorn, T., Brooks‐Russell, A., Limbacher, S. A., & Kosnett, M. J. (2025) · Journal of Analytical Toxicology · doi:10.1093/jat/bkaf043 · Cited by 4
“…in capillary blood glucose, it was observed that capillary glucose obtained at the nonglaborous skin of the forearm, abdomen, and thigh increased more slowly and to a lesser extent than that at the fingertip in the first two hours following an acute oral glucose challenge or meal [34–36].”
Pfützner, A., Hermanns, N., Schröder, S., Wegenstein, M., Larbig, M., Mondok, Á., Forst, T. M., & Haak, T. (2002) · Swiss Medical Weekly · doi:10.4414/smw.2002.09991 · Cited by 4
Frequent lancing for monitoring blood glucose may cause skin changes
Fruhstorfer, H. (2006) · Practical Diabetes · doi:10.1002/pdi.949 · Cited by 3
L’observance à la pratique de l’autosurveillance glycémique
Reach, G. (2010) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(10)70144-9 · Cited by 2
Arimoto, K., Sasaki, H., Ogawa, K., Yamasaki, H., Okamoto, K., Furuta, H., Hanabusa, T., Nishi, M., Nakao, T., & Nanjo, K. (2004) · Diabetes Care · doi:10.2337/diacare.27.6.1449 · Cited by 2
“CONCLUSIONS — Since Jungheim and Koschinsky (1,2) reported their findings about a risky delay of hypoglycemia detection by glucose monitoring at the arm site, it has been stated that during rapid blood glucose change, SMBG values obtained at alternate sites of diabetic patients lag behind the fingertip values (3,4).”
BackgroundAnalysis: Glucose Testing of Blood Samples from Alternate Sites—State of Affairs
Peled, N. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.609 · Cited by 2
Deza, S., Fernández-Montero, A., Calleja, S., Zabalza, N., Arias, M. Á., Gómez, C., Orbegozo, N., Fischer, T., Engelmann-Wadin, S., Blumhardt, P., et al. (2026) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2026-0671 · Cited by 1
“Warming the forearm to increase blood fl ow reduces these site-dependent di ff erences, bringing values closer to those from fi ngerstick sampling [38].”
Measurement Accuracy of Blood Glucose Test on the Forearm
Park, K.-S., Song, Y.-Y., Park, M., Kim, K.-O., Choi, S.-S., Kim, K., Ryu, J.-D., & Cha, E. (2009) · Journal of the Korea Academia-Industrial cooperation Society · doi:10.5762/kais.2009.10.9.2451 · Cited by 1
Determination of Glucose and Continuous Glucose Monitoring
Wang, Y. F., & Jia, W. (2018) · Springer eBooks · doi:10.1007/978-981-10-7074-7_1 · Cited by 1
Equivalence of Fasting and Non-Fasting Blood Glucose Readings at the Palm and Fingertip Sites
Owida, H. A., Al-Nabulsi, J., & Khatib, A. O. A. (2022) · International Conference on Information and Communication Systems (ICICS) · doi:10.1109/icics55353.2022.9811230 · Cited by 1
Automonitoraggio glicemico nei pazienti diabetici
Montagnana, M., & Lippi, G. (2013) · La Rivista Italiana della Medicina di Laboratorio - Italian Journal of Laboratory Medicine · doi:10.1007/s13631-013-0017-z · Cited by 1
Klinische Anwendungen von mikrobiologischen Schnelltests
Stürenburg, E. (2008) · Springer eBooks · doi:10.1007/978-3-540-79152-2_32 · Cited by 1
Cribado de diabetes mellitus tipo 2: algo más que la búsqueda de la diabetes ignorada
González-Clemente, J. M. (2004) · Medicina Clínica · doi:10.1016/s0025-7753(04)74451-6 · Cited by 1
Analysis: Glucose Monitoring and Physical Activity: The Present and Future Challenges
Colberg, S. R. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.681 · Cited by 1
Clinical Usefulness of Glucose Testing from the Forearm in Diabetic Patients
Lee, S.-W., Chon, S., Oh, S., Woo, J., Kim, S. W., Kim, J. W., Kim, Y. S., Paeng, J. R., Koh, G., & Nam, H. H. (2006) · Journal of Korean Endocrine Society · doi:10.3803/jkes.2006.21.4.281 · Cited by 1
“이러한 현상에 대해 Jungheim 등[10]은 식후나 인슐린 투여후 등의 혈당의 급속한 변화가 있을 때 전완부의 혈당변화의 반응이 지연 될 가능성을 제시하였다.”
Insulin, glucagon, and hypoglycemic drugs
Krans, H. M. J. (2003) · Side effects of drugs annual · doi:10.1016/s0378-6080(03)80049-1 · Cited by 1
Management of Point‐of‐Care Testing
Smy, L., Apfeld, K., Ampomah, S., & Thieme, M. (2024) · Wiley eBooks · doi:10.1002/9781683673941.ch22 · Cited by 1
Pharmakotherapie der sensomotorischen Neuropathie
Luft, D. (2008) · Der Diabetologe · doi:10.1007/s11428-008-0261-4
POCT in der Diabetesdiagnostik
Koschinsky, T., & Wahl, H. G. (2011) · Springer eBooks · doi:10.1007/978-3-642-20172-1_15
Self-citationGarcía-García, F. (2015) · UPM Digital Archive (Technical University of Madrid) · doi:10.20868/upm.thesis.37805
Surveillance du contrôle de la glycémie
Berard, L., Blumer, I. R., Houlden, R. L., Miller, D. B., & Woo, V. C. (2013) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2013.07.037
For the POCT Working Group: Theodor Koschinsky, Junker, R., Luppa, P. B., & Schlebusch, H. (2009) · Journal of Laboratory Medicine · doi:10.1515/jlm.2009.058et
Metadata
- Authors
- Jungheim, K., & Koschinsky, T.
- Affiliations
- Deutsches Diabetes-Zentrum e.V. (DE)
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2002-06-01
- Volume
- 25
- Issue
- 6
- Pages
- 956-960
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2082028634
- Cited by
- 114
- References
- 22
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 2 · 2023: 1 · 2022: 1 · 2021: 1 · 2020: 3 · 2018: 5 · 2017: 2 · 2016: 3 · 2015: 4 · 2014: 2 · 2013: 9 · 2012: 3
- FWCI
- 6.93
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Diabetes Care (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- capillary blood glucose, hyperglycemia detection, hypoglycemia detection, insulin-treated diabetes
- MeSH terms
- Awareness, Capillaries, Fingers, Forearm, Humans, Hyperglycemia, Hypoglycemia, Insulin, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review