Spurious Reporting of Nocturnal Hypoglycemia by CGMS in Patients With Tightly Controlled Type 1 Diabetes
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
OBJECTIVE—The Medtronic MiniMed Continuous Glucose Monitoring System (CGMS) is designed to continuously monitor interstitial fluid glucose levels within a range of 40–400 mg/dl. It is considered an important tool for overnight glucose monitoring. The goal of this study was to determine the accuracy of this system in individuals with tightly controlled diabetes. RESEARCH DESIGN AND METHODS—Seven adolescents and young adults with HbA1c levels 6.6 ± 0.6% (range 5.7–7.1) were admitted to the Clinical Research Center. Simultaneous glucose measurements obtained by glucose analyzer, Accu-Check Advantage meter, and CGMS were compared. The analyzer levels were considered the standard. RESULTS—The CGMS results were lower than analyzer readings in 74% of simultaneous pairs of tests performed during the 24-h period; the average correlation was 0.76. There was a trend for the poorest correlation to occur in patients with the narrowest range in daily glucose levels. When the lowest CGMS reading of the night was compared with the simultaneous analyzer reading, the CGMS level was lower in all cases by an average of 38 ± 15%. In six of seven patients, the discrepancy was believed to be clinically significant; in at least four patients, overnight glucose levels reported by CGMS were falsely low, in a range that might have resulted in inappropriate reduction of overnight insulin dose. CONCLUSIONS—CGMS reports of asymptomatic nighttime hypoglycemia may be spurious and should be interpreted with caution in patients with tightly controlled diabetes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Bode, B. W., Gross, T. M., Thornton, K. R., & Mastrototaro, J. J. (1999) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(99)00113-8 · Cited by 232 · cited 1×
“Although the CGMS has proved to be useful for improving HbA1c levels in persons with poorly controlled type 1 diabetes (3,9,10), a high degree of unexpected asymptomatic nighttime hypoglycemia has been reported in these studies (3– 6,8).”
Background
Cited by
Cryer, P. E., Davis, S. N., & Shamoon, H. (2003) · Diabetes Care · doi:10.2337/diacare.26.6.1902 · Cited by 1,160
Risk of hypoglycaemia in types 1 and 2 diabetes: Effects of treatment modalities and their duration
UK Hypoglycaemia Study Group (2007) · Diabetologia · doi:10.1007/s00125-007-0599-y · Cited by 825
“It should identify hypoglycaemia irrespective of subjective awareness, but its ability to detect clinically relevant hypoglycaemia has been questioned [22].”
BackgroundKlonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Hypoglycaemia in diabetes mellitus: epidemiology and clinical implications
Frier, B. M. (2014) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2014.170 · Cited by 481
Continuous glucose monitoring and closed‐loop systems
Hovorka, R. (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01672.x · Cited by 443
Association of Hypoglycemia and Cardiac Ischemia
Desouza, C., Salazar, H., Cheong, B., Murgo, J., & Fonseca, V. (2003) · Diabetes Care · doi:10.2337/diacare.26.5.1485 · Cited by 406
“However, this accuracy diminishes toward the 40-mg/dl glucose level and may give false readings below this level (35).”
BackgroundBoyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
Chow, E., Bernjak, A., Williams, S., Fawdry, R. A., Hibbert, S., Freeman, J., Sheridan, P. J., & Heller, S. R. (2014) · Diabetes · doi:10.2337/db13-0468 · Cited by 403
“Although there are reports that CGM glucose may be falsely low during hypoglycemia in type 1 diabetes (41), in type 2 diabetes CGM glucose has been reported to be falsely high owing to differences in glucose distribution in the interstitial space (42).”
BackgroundA Novel Approach to Continuous Glucose Analysis Utilizing Glycemic Variation
McDonnell, C. M., Donath, S. M., Vidmar, S. I., Werther, G. A., & Cameron, F. J. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.253 · Cited by 358
Impact of Exercise on Overnight Glycemic Control in Children with Type 1 Diabetes Mellitus
Children, T. D. R. I. (2005) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2005.04.065 · Cited by 264
Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R. L., Bugler, J. R., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
Poolsup, N., Suksomboon, N., & Kyaw, A. M. (2013) · Diabetology & Metabolic Syndrome · doi:10.1186/1758-5996-5-39 · Cited by 179
“In addition, SMBG gives a single instant reading without any information on glucose trends and thus may miss important and significant glucose fluctuations [10-12].”
BackgroundSusceptibility of Interstitial Continuous Glucose Monitor Performance to Sleeping Position
Mensh, B. D., Wisniewski, N. A., Neil, B. M., & Burnett, D. R. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700408 · Cited by 97
“J Diabetes Sci Technol 2013;7(4):863–870 ORIGINAL ARTICLE”
Chetty, V. T., Almulla, A., Odueyungbo, A., & Thabane, L. (2008) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2008.02.014 · Cited by 85
“8 mmol/l [34].”
BackgroundThe role of continuous glucose monitoring in clinical decision-making in diabetes in pregnancy
McLACHLAN, K., JENKINS, A., & O’NEAL, D. (2007) · Australian and New Zealand Journal of Obstetrics and Gynaecology · doi:10.1111/j.1479-828x.2007.00716.x · Cited by 82
Burnett, D. R., Huyett, L. M., Zisser, H. C., Doyle, F. J., & Mensh, B. D. (2014) · Diabetes · doi:10.2337/db13-1649 · Cited by 52
“This is especially vexing during sleep, since sleeping on SQ sensors can cause large inaccuracies (6), and sleeping patients are at high risk for hypoglycemic complications (6-16).”
Background
Metadata
- Authors
- McGowan, K., Thomas, W., & Moran, A.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2002-09-01
- Volume
- 25
- Issue
- 9
- Pages
- 1499-1503
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2031438274
- Cited by
- 148
- References
- 1
- FWCI
- 10.51
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.