Evaluating the Clinical Accuracy of Two Continuous Glucose Sensors Using Continuous Glucose–Error Grid Analysis
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
OBJECTIVE—To compare the clinical accuracy of two different continuous glucose sensors (CGS) during euglycemia and hypoglycemia using continuous glucose–error grid analysis (CG-EGA). RESEARCH DESIGN AND METHODS—FreeStyle Navigator (Abbott Laboratories, Alameda, CA) and MiniMed CGMS (Medtronic, Northridge, CA) CGSs were applied to the abdomens of 16 type 1 diabetic subjects (age 42 ± 3 years) 12 h before the initiation of the study. Each system was calibrated according to the manufacturer’s recommendations. Each subject underwent a hyperinsulinemic-euglycemic clamp (blood glucose goal 110 mg/dl) for 70–210 min followed by a 1–mg · dl−1 · min−1 controlled reduction in blood glucose toward a nadir of 40 mg/dl. Arterialized blood glucose was determined every 5 min using a Beckman Glucose Analyzer (Fullerton, CA). CGS glucose recordings were matched to the reference blood glucose with 30-s precision, and rates of glucose change were calculated for 5-min intervals. CG-EGA was used to quantify the clinical accuracy of both systems by estimating combined point and rate accuracy of each system in the euglycemic (70−180 mg/dl) and hypoglycemic (<70 mg/dl) ranges. RESULTS—A total of 1,104 data pairs were recorded in the euglycemic range and 250 data pairs in the hypoglycemic range. Overall correlation between CGS and reference glucose was similar for both systems (Navigator, r = 0.84; CGMS, r = 0.79, NS). During euglycemia, both CGS systems had similar clinical accuracy (Navigator zones A + B, 88.8%; CGMS zones A + B, 89.3%, NS). However, during hypoglycemia, the Navigator was significantly more clinically accurate than the CGMS (zones A + B = 82.4 vs. 61.6%, Navigator and CGMS, respectively, P < 0.0005). CONCLUSIONS—CG-EGA is a helpful tool for evaluating and comparing the clinical accuracy of CGS systems in different blood glucose ranges. CG-EGA provides accuracy details beyond other methods of evaluation, including correlational analysis and the original EGA.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312 · cited 2×
“Continuous glucose–error grid analysis (CG-EGA) has been developed as a method of evaluating the clinical accuracy of CGS utilizing both blood glucose point accuracy (CGS readings versus reference blood glucose determinations) and rate and direction of change accuracy (3).”
BackgroundMethodSelf-citationNewer portable glucose meters—analytical improvement compared with previous generation devices?
Weitgasser, R., Gappmayer, B., & Pichler, M. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.10.1821 · Cited by 85
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760
Self-citationGlucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584
Non-invasive glucose monitoring technology in diabetes management: A review
Vashist, S. K. (2012) · Analytica Chimica Acta · doi:10.1016/j.aca.2012.03.043 · Cited by 575
Relationship between glycated haemoglobin levels and mean glucose levels over time
Nathan, D. M., Turgeon, H., & Regan, S. (2007) · Diabetologia · doi:10.1007/s00125-007-0803-0 · Cited by 517
“Although some studies have suggested less than acceptable correlations between CGMS interstitial glucose results and venous plasma or capillary glucose levels, most recent studies, including the present study, have noted generally strong correlations [19, 20].”
Result
Metadata
- Authors
- Clarke, W. L., Anderson, S., Farhy, L., Breton, M., Gonder-Frederick, L., Cox, D., & Kovatchev, B.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2005-10-01
- Volume
- 28
- Issue
- 10
- Pages
- 2412-2417
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2022278688
- Cited by
- 155
- References
- 3
- FWCI
- 7.12
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluating the Clinical Accuracy of Two Continuous Glucose Sensors Using Continuous Glucose–Error Grid Analysis. https://app.cochat.ai/papers/10.2337/diacare.28.10.2412
- The paper
- Clarke, W. L., Anderson, S., Farhy, L., Breton, M., Gonder-Frederick, L., Cox, D., & Kovatchev, B. (2005). Evaluating the Clinical Accuracy of Two Continuous Glucose Sensors Using Continuous Glucose–Error Grid Analysis. Diabetes Care. https://doi.org/10.2337/diacare.28.10.2412
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.