Benefit of Time-of-Flight in PET: Experimental and Clinical Results
Open accessAbstract
UNLABELLED: Significant improvements have made it possible to add the technology of time-of-flight (TOF) to improve PET, particularly for oncology applications. The goals of this work were to investigate the benefits of TOF in experimental phantoms and to determine how these benefits translate into improved performance for patient imaging. METHODS: In this study we used a fully 3-dimensional scanner with the scintillator lutetium-yttrium oxyorthosilicate and a system timing resolution of approximately 600 ps. The data are acquired in list-mode and reconstructed with a maximum-likelihood expectation maximization algorithm; the system model includes the TOF kernel and corrections for attenuation, detector normalization, randoms, and scatter. The scatter correction is an extension of the model-based single-scatter simulation to include the time domain. Phantom measurements to study the benefit of TOF include 27-cm- and 35-cm-diameter distributions with spheres ranging in size from 10 to 37 mm. To assess the benefit of TOF PET for clinical imaging, patient studies are quantitatively analyzed. RESULTS: The lesion phantom studies demonstrate the improved contrast of the smallest spheres with TOF compared with non-TOF and also confirm the faster convergence of contrast with TOF. These gains are evident from visual inspection of the images as well as a quantitative evaluation of contrast recovery of the spheres and noise in the background. The gains with TOF are higher for larger objects. These results correlate with patient studies in which lesions are seen more clearly and with higher uptake at comparable noise for TOF than with non-TOF. CONCLUSION: TOF leads to a better contrast-versus-noise trade-off than non-TOF but one that is difficult to quantify in terms of a simple sensitivity gain improvement: A single gain factor for TOF improvement does not include the increased rate of convergence with TOF nor does it consider that TOF may converge to a different contrast than non-TOF. The experimental phantom results agree with those of prior simulations and help explain the improved image quality with TOF for patient oncology studies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Surti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Self-citationList-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET
Parra, L., & Barrett, H. H. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.700734 · Cited by 357
Time of flight in PET revisited
Moses, W. W. (2003) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2003.817319 · Cited by 336
Time-of-flight positron emission tomography: status relative to conventional PET.
Budinger, T. F. (1983) · PubMed · Cited by 283
Current trends in scintillator detectors and materials
Moses, W. W. (2002) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/s0168-9002(02)00955-5 · Cited by 275
PET performance measurements using the NEMA NU 2-2001 standard.
Daube-Witherspoon, M. E., Karp, J. S., Casey, M., DiFilippo, F. P., Hines, H., Muehllehner, G., Simcic, V. J., Stearns, C. W., Adam, L.-E., Kohlmyer, S. G., et al. (2002) · PubMed · Cited by 222
Self-citationA Matheematical Model for Positron-Emission Tomography Systems Having Time-of-Flight Measurements
Snyder, D. L., Thomas, L. J., & Ter-Pogossian, M. M. (1981) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1981.4332168 · Cited by 217
Medical radioisotope scintigraphy
McCready, V. R. (1970) · The International Journal of Applied Radiation and Isotopes · doi:10.1016/0020-708x(70)90109-2 · Cited by 216
Fast accurate iterative reconstruction for low-statistics positron volume imaging
Reader, A. J., Erlandsson, K., Flower, M. A., & Ott, R. J. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/4/012 · Cited by 190
Tomitani, T. (1981) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1981.4335769 · Cited by 181
Huesman, R. H., Klein, G. J., Moses, W. W., Qi, J., Reutter, B. W., & Virador, P. R. G. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.870263 · Cited by 174
Investigation of time-of-flight benefit for fully 3-DPET
Surti, S., Karp, S., Popescu, L. M., Daube-Witherspoon, E., & Werner, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.871419 · Cited by 166
Self-citationExtension of Single Scatter Simulation to Scatter Correction of Time-of-Flight PET
Watson, C. (2007) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2007.901227 · Cited by 164
First experimental results of time-of-flight reconstruction on an LSO PET scanner
Conti, M., Bendriem, B., Casey, M., Chen, M., Kehren, F., Michel, C., & Panin, V. (2005) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/19/006 · Cited by 164
Lewellen, T. K. (1998) · Seminars in Nuclear Medicine · doi:10.1016/s0001-2998(98)80031-7 · Cited by 160
PETT VI: a positron emission tomograph utilizing cesium fluoride scintillation detectors.
Ter‐Pogossian, M. M., Ficke, D. C., Hood, J. T., Yamamoto, M., & Mullani, N. A. (1982) · PubMed · Cited by 147
Snyder, D. L., & Politte, D. G. (1983) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1983.4332660 · Cited by 139
Ter‐Pogossian, M. M., Ficke, D. C., Hood, J. T., Yamamoto, M., & Mullani, N. A. (1982) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198202000-00021 · Cited by 132
Optimization of a fully 3D single scatter simulation algorithm for 3D PET
Accorsi, R., Adam, L.-E., Werner, M. E., & Karp, J. S. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/12/008 · Cited by 123
Self-citationPOTENTIAL ADVANTAGES OF A CESIUM FLUORIDE SCINTILLATOR FOR A TIME-OF-FLIGHT POSITRON CAMERA
Allemand, R., Gressel, O., Vacher, J., & Sank, V. J. (1980) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198008000-00054 · Cited by 116
Iterative image reconstruction using geometrically ordered subsets with list-mode data
Popescu, L. M., Matej, S., & Lewitt, R. M. (2005) · IEEE Symposium Conference Record Nuclear Science 2004. · doi:10.1109/nssmic.2004.1466649 · Cited by 108
Physical characteristics of TTV03, a new high spatial resolution time-of-flight positron tomograph
Mazoyer, B. M., Trébossen, R., Schoukroun, C., Verrey, B., Syrota, A., Vacher, J., Lemasson, P., Monnet, O., Bouvier, A., & Lecomte, J. L. (1990) · IEEE Transactions on Nuclear Science · doi:10.1109/23.106714 · Cited by 102
Systematic and Distributed Time-of-Flight List Mode PET Reconstruction
Wang, W., Hu, Z., Gualtieri, E., Parma, M. J., Walsh, E. S., Sebok, D. A., Hsieh, Y., Tung, C., Song, X., Griesmer, J. J., et al. (2006) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2006.354229 · Cited by 96
Self-citationImplementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling
Werner, M. E., Surti, S., & Karp, J. S. (2006) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2006.354238 · Cited by 88
Self-citationPerformance measurements of the SP3000/UW time-of-flight positron emission tomograph
Lewellen, T. K., Bice, A. N., Harrison, R. L., Pencke, M. D., & Link, J. M. (1988) · IEEE Transactions on Nuclear Science · doi:10.1109/23.12808 · Cited by 80
Brownell, G. L., Burnham, C. A., Wilensky, S., Aronow, S., Kazemi, H., & Strieder, D. J. (1969) · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · Cited by 45
SURVEY OF RADIOISOTOPE CAMERAS.
Anger, H. O. (1966) · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · Cited by 43
Instrumentation trends in nuclear medicine
Budinger, T. F. (1977) · Seminars in Nuclear Medicine · doi:10.1016/s0001-2998(77)80037-8 · Cited by 35
Snyder, D. L. (1982) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1982.4336009 · Cited by 32
Politte, D. G., & Snyder, D. L. (1984) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.1984.4333330 · Cited by 30
Performance characteristics of the University of Texas TOFPET-I PET camera
Wong, W. H. M., Mullani, N. A., Philippe, E. A., Hartz, R. K., Bristow, D., Yerian, K., Gaeta, J. M., & Ketharnavaz, N. (1986) · Journal of Nuclear Medicine · Cited by 19
Effect of random and scatter fractions in variance reduction using time-of-flight information
Kimdon, J. A., Qi, J., & Moses, W. W. (2004) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2003.1352415 · Cited by 17
Cited by
Physical and clinical performance of the mCT time-of-flight PET/CT scanner
Jakoby, B. W., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/8/004 · Cited by 492
Performance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484
PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging Applications
Pichler, B. J., Kolb, A., Nägele, T., & Schlemmer, H.-P. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.061853 · Cited by 418
Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system
Zaidi, H., Ojha, N., Morich, M., Griesmer, J., Hu, Z., Maniawski, P., Ratib, O., Izquierdo-Garcia, D., Fayad, Z. A., & Shao, L. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/10/013 · Cited by 402
State of the art in total body PET
Vandenberghe, S., Moskal, P., & Karp, J. S. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00290-2 · Cited by 390
Self-citation
Metadata
- Authors
- Karp, J. S., Surti, S., Daube-Witherspoon, M. E., & Muehllehner, G.
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2008-02-20
- Volume
- 49
- Issue
- 3
- Pages
- 462-470
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- OpenAlex
- W2148488773
- Cited by
- 584
- References
- 32
- FWCI
- 23.89
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.