State of the art in total body PET
Open accessAbstract
The idea of a very sensitive positron emission tomography (PET) system covering a large portion of the body of a patient already dates back to the early 1990s. In the period 2000–2010, only some prototypes with long axial field of view (FOV) have been built, which never resulted in systems used for clinical research. One of the reasons was the limitations in the available detector technology, which did not yet have sufficient energy resolution, timing resolution or countrate capabilities for fully exploiting the benefits of a long axial FOV design. PET was also not yet as widespread as it is today: the growth in oncology, which has become the major application of PET, appeared only after the introduction of PET-CT (early 2000).The detector technology used in most clinical PET systems today has a combination of good energy and timing resolution with higher countrate capabilities and has now been used since more than a decade to build time-of-flight (TOF) PET systems with fully 3D acquisitions. Based on this technology, one can construct total body PET systems and the remaining challenges (data handling, fast image reconstruction, detector cooling) are mostly related to engineering. The direct benefits of long axial FOV systems are mostly related to the higher sensitivity. For single organ imaging, the gain is close to the point source sensitivity which increases linearly with the axial length until it is limited by solid angle and attenuation of the body. The gains for single organ (compared to a fully 3D PET 20-cm axial FOV) are limited to a factor 3–4. But for long objects (like body scans), it increases quadratically with scanner length and factors of 10–40 × higher sensitivity are predicted for the long axial FOV scanner. This application of PET has seen a major increase (mostly in oncology) during the last 2 decades and is now the main type of study in a PET centre. As the technology is available and the full body concept also seems to match with existing applications, the old concept of a total body PET scanner is seeing a clear revival. Several research groups are working on this concept and after showing the potential via extensive simulations; construction of these systems has started about 2 years ago. In the first phase, two PET systems with long axial FOV suitable for large animal imaging were constructed to explore the potential in more experimental settings. Recently, the first completed total body PET systems for human use, a 70-cm-long system, called PennPET Explorer, and a 2-m-long system, called uExplorer, have become reality and first clinical studies have been shown. These results illustrate the large potential of this concept with regard to low-dose imaging, faster scanning, whole-body dynamic imaging and follow-up of tracers over longer periods. This large range of possible technical improvements seems to have the potential to change the current clinical routine and to expand the number of clinical applications of molecular imaging. The J-PET prototype is a prototype system with a long axial FOV built from axially arranged plastic scintillator strips.This paper gives an overview of the recent technical developments with regard to PET scanners with a long axial FOV covering at least the majority of the body (so called total body PET systems). After explaining the benefits and challenges of total body PET systems, the different total body PET system designs proposed for large animal and clinical imaging are described in detail. The axial length is one of the major factors determining the total cost of the system, but there are also other options in detector technology, design and processing for reducing the cost these systems. The limitations and advantages of different designs for research and clinical use are discussed taking into account potential applications and the increased cost of these systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Attenuation correction for a combined 3D PET/CT scanner
Kinahan, P. E., Townsend, D. W., Beyer, T., & Sashin, D. (1998) · Medical Physics · doi:10.1118/1.598392 · Cited by 946
National electrical manufacturers association
(1926) · Journal of the A.I.E.E. · doi:10.1109/jaiee.1926.6537344 · Cited by 860
Analytic 3D image reconstruction using all detected events
Kinahan, P. E., & Rogers, J. G. (1989) · IEEE Transactions on Nuclear Science · doi:10.1109/23.34585 · Cited by 767
Cherry, S. R., Jones, T., Karp, J. S., Qi, J., Moses, W. W., & Badawi, R. D. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.184028 · Cited by 763
Self-citationFirst Human Imaging Studies with the EXPLORER Total-Body PET Scanner*
Badawi, R. D., Shi, H., Hu, P., Chen, S., Xu, T., Price, P. M., Ding, Y., Spencer, B. A., Nardo, L., Liu, W., et al. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.119.226498 · Cited by 722
Surti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Self-citationDevelopment of new scintillators for medical applications
Lecoq, P. (2016) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2015.08.041 · Cited by 610
Benefit of Time-of-Flight in PET: Experimental and Clinical Results
Karp, J. S., Surti, S., Daube-Witherspoon, M. E., & Muehllehner, G. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.044834 · Cited by 584
Self-citationLevin, C. S., & Hoffman, E. J. (1999) · Physics in Medicine and Biology · doi:10.1088/0031-9155/44/3/019 · Cited by 551
Image Reconstruction is a New Frontier of Machine Learning
Wang, G., Ye, J. C., Mueller, K., & Fessler, J. A. (2018) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2018.2833635 · Cited by 525
Cited by
New PET technologies – embracing progress and pushing the limits
Aide, N., Lasnon, C., Kesner, A., Levin, C. S., Buvat, I., Iagaru, A., Hermann, K., Badawi, R. D., Cherry, S. R., Bradley, K. M., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05390-4 · Cited by 73
Role of Artificial Intelligence in Theranostics
Brosch-Lenz, J., Yousefirizi, F., Zukotynski, K., Beauregard, J.-M., Gaudet, V., Saboury, B., Rahmim, A., & Uribe, C. (2021) · PET Clinics · doi:10.1016/j.cpet.2021.06.002 · Cited by 42
Alhorani, Q., Alkhybari, E., Rawashdeh, M., Sabarudin, A., Latiff, R. A., Al-Ibraheem, A., & Mohamad, M. (2024) · Journal of Radiological Protection · doi:10.1088/1361-6498/ad2c62 · Cited by 6
Metadata
- Authors
- Vandenberghe, S., Moskal, P., & Karp, J. S.
- Venue
- EJNMMI Physics
- Published
- 2020-05-25
- Volume
- 7
- Issue
- 1
- Pages
- 35-35
- ISSN
- 2197-7364 (ISSN-L)
- OpenAlex
- W3030779614
- Cited by
- 390
- References
- 55
- FWCI
- 24.65
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: State of the art in total body PET. https://app.cochat.ai/papers/10.1186/s40658-020-00290-2
- The paper
- Vandenberghe, S., Moskal, P., & Karp, J. S. (2020). State of the art in total body PET. EJNMMI Physics. https://doi.org/10.1186/s40658-020-00290-2
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.