The effect of varying LED intensity on pulse oximeter accuracy
Abstract
Most pulse oximeters automatically alter the intensity of their light-emitting diodes (LEDs) according to the absorption of the finger, toe or earlobe to which they are attached. This paper investigates the effect of changing LED intensity on pulse oximeter accuracy. Our results show that the peak wavelength of a red LED typically increases by 8 nm as its intensity is increased ten-fold. To determine whether this shift introduces a significant error, a simple theoretical model based on the Beer-Lambert law is used. The model predicts that a 10:1 change in LED intensity results in a 2.5% error at 50% arterial oxygen saturation (SpO2). At high saturations (SpO2 greater than or equal to 85%) the model predicts little loss of accuracy and thus any effect due to changes in LED intensity will be apparent only at low saturations.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In low saturation conditions, a 10:1 variation in LED intensity can lead to an error of 2.5%.
“Another study to determine the effect of varying LED intensity on pulse oximeter accuracy showed that in low saturation conditions, a 10:1 variation in LED intensity can lead to an error of 2.5% [103].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [103]
Cites
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617328 · Cited by 305 · see citing works
Spectrophotometry of Haemoglobin Derivatives.
(1971) · Annals of Internal Medicine · doi:10.7326/0003-4819-75-1-150 · Cited by 262 · see citing works
Kelleher, J. F. (1989) · The Journal of Clinical Monitoring · doi:10.1007/bf01618369 · Cited by 205 · see citing works
Effect of anemia on pulse oximeter accuracy at low saturation
Severinghaus, J. W., & Koh, S. O. (1990) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02828282 · Cited by 129 · see citing works
Pulse Oximetry: Technical Aspects of Machine Design
Pologe, J. A. (1987) · International Anesthesiology Clinics · doi:10.1097/00004311-198702530-00009 · Cited by 110 · see citing works
Role of Light Scatterng in Whole Blood Oximetry
Steinke, J. M., & Shepherd, A. P. (1986) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1986.325713 · Cited by 78 · see citing works
Postoperative monitoring of conjunctival oxygen tension and temperature
Fink, S., Abraham, E., & Ehrlich, H. (1988) · International journal of clinical monitoring and computing · doi:10.1007/bf01739231 · Cited by 77 · see citing works
Mendelson, Y., & Kent, J. C. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.30810 · Cited by 60 · see citing works
Payne, J. P., & Severinghaus, J. W. (1986) · Springer eBooks · doi:10.1007/978-1-4471-1423-9 · Cited by 42 · see citing works
Optical fibre telecommunications
Midwinter, J. E. (1980) · Optics & Laser Technology · doi:10.1016/0030-3992(80)90074-2 · Cited by 37 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Humphreys, K., Ward, T. E., & Markham, C. (2007) · Review of Scientific Instruments · doi:10.1063/1.2724789 · Cited by 214 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
An assessment of the accuracy of pulse oximeters
Milner, Q. J. W., & Mathews, G. R. (2012) · Anaesthesia · doi:10.1111/j.1365-2044.2011.07021.x · Cited by 79 · see citing works
Reynolds, K. J., Palayiwa, E., Moyle, J., Sykes, M. K., & Hahn, C. E. W. (1993) · The Journal of Clinical Monitoring · doi:10.1007/bf01616919 · Cited by 41 · see citing works
Self-citationVahlsing, T., Delbeck, S., Leonhardt, S., & Heise, H. M. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818798347 · Cited by 28 · see citing works
TEMPERATURE DEPENDENCE OF LED AND ITS THEORETICAL EFFECT ON PULSE OXIMETRY
Reynolds, K. J., de Kock, J. P., Tarassenko, L., & Moyle, J. (1991) · British Journal of Anaesthesia · doi:10.1093/bja/67.5.638 · Cited by 26 · see citing works
Self-citationOpsins as main regulators of skin biology
Castrucci, A. M. D. L., Baptista, M. S., & de Assis, L. V. M. (2023) · Journal of Photochemistry and Photobiology · doi:10.1016/j.jpap.2023.100186 · Cited by 9 · see citing works
Evaluation of a Pulse Oximeter Sensor Tester
Dugani, S., Hodzovic, I., Sindhakar, S., Nadra, A., Dunstan, C., Wilkes, A. R., & Mecklenburgh, J. S. (2011) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-011-9283-3 · Cited by 7 · see citing works
Evaluation of a new pulse oximeter testing device
Fisher, J. A., Martire, T., & VOLGYESI, G. A. (1996) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03011260 · Cited by 3 · see citing works
Characterizing the influence of skin pigmentation on pulse oximetry
Putcha, A., Schichlein, K. D., Nguyen, T., Davis, B., Choffin, R., Malik, S., Sharma, A., Sosa, G., Saffari, S., Ribaudo, J. G., et al. (2025) · Biophotonics discovery. · doi:10.1117/1.bios.2.3.032506 · Cited by 2 · see citing works
Martín-Escudero, P., & Cabanas, A. M. (2025) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2025.117213 · Cited by 2 · see citing works
DOVE: Shoulder-based Opioid Overdose Detection and Reversal Device
Lingamoorthy, A., Watson, A., Gordon, D., Mandal, A. S., Henderson, K., Ma, X., Weimer, J. E., Brenner, J. S., & Kandasamy, N. (2023) · IEEE/ACM Conference on Connected Health: Applications, Systems and Engineering Technologies · doi:10.1145/3580252.3586973 · Cited by 1 · see citing works
Finger blood content, light transmission, and pulse oximetry errors
Craft, T. M., LAWSON, R. A., & Young, D. (1992) · Anaesthesia · doi:10.1111/j.1365-2044.1992.tb01955.x · Cited by 1 · see citing works
Design of portable sleep apnea hypopnea monitor instrumentation
Hu, H., Chen, H., & Chen, Z. (2011) · JOURNAL OF ELECTRONIC MEASUREMENT AND INSTRUMENT · doi:10.3724/sp.j.1187.2011.00812
Iron Deficiency in Cyanotic Heart Disease
Broberg, C. S., & Lunde, P. (2010) · Elsevier eBooks · doi:10.1016/b978-0-443-06712-9.00072-6
Improving Pulse Oximetry Accuracy by Mitigating Photoplethysmographic Non-Linearities
Dervieux, E., & Uhring, W. (2026) · IEEE Interregional NEWCAS Conference (NEWCAS) · doi:10.1109/newcas64543.2026.11674001
- Authors
- De Kock, J. P., Reynolds, K. J., Tarassenko, L., & Moyles, J. T. B.
- Affiliations
- John Radcliffe Hospital (GB)
- University of Oxford (GB)
- Milton Keynes Hospital (GB)
- Venue
- Journal of Medical Engineering & Technology
- Publisher
- Informa UK Limited
- Published
- 1991-01
- Volume
- 15
- Issue
- 3
- Pages
- 111-115
- ISSN
- 0309-1902 (ISSN-L)
- 1464-522X
- OpenAlex
- W2093531986
- Cited by
- 16
- References
- 10
- Citations by year
- 2026: 1 · 2025: 2 · 2023: 2 · 2022: 1 · 2018: 1 · 2012: 1
- FWCI
- 1.28
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Medical Engineering & Technology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Keywords
- pulse oximetry, oxygen saturation measurement, red LED, peak wavelength shift, Beer-Lambert law, measurement error, arterial oxygen saturation
- MeSH terms
- Equipment Design, Hemoglobins, Humans, Hydrogen-Ion Concentration, Models, Theoretical, Oximetry, Oxygen, Predictive Value of Tests, Reference Values
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review