High-Density μLED-Based Optical Cochlear Implant With Improved Thermomechanical Behavior
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
This study reports the realization of an optical cochlear implant (oCI) with optimized thermomechanical properties for optogenetic experiments. The oCI probe comprises 144 miniaturized light-emitting diodes (µLEDs) distributed along a bendable, 1.5-cm-long, 350-µm-wide and 26-µm-thick probe shaft, individually controlled via a n×p matrix interconnection. In contrast to our earlier approach based on polyimide (PI) and epoxy resin with different thermal expansion coefficients, the µLEDs and interconnecting wires are now embedded into a triple-layer stack of a single, biocompatible and highly transparent epoxy material. The new material combination results in a pronounced reduction of thermomechanical bending in comparison with the material pair of the earlier approach. We developed a spin-coating process enabling epoxy resin layers down to 5 µm at thickness variations of less than 7 % across the entire carrier wafer. We observed that the cross-linking of epoxy resin layers strongly depends on the spin-coating parameters which were found to be correlated to a potential separation of epoxy resin components of different densities. Furthermore, various metallization layers and corresponding adhesion promoting layers were investigated. We identified the combination of silicon carbide with a titanium-based metallization to provide the highest peeling strength, achieving an adhesion to epoxy improved by a factor of two. In order to obtain a high process yield, we established a stress-free implant release using the electrochemical dissolution of a sacrificial aluminum layer. The direct comparison of oCI probe variants using a single epoxy material and the combination of PI and epoxy resin revealed that the epoxy-resin-only probe shows minimal thermomechanical probe bending with a negligible hysteresis. The thermal probe characterization demonstrated that the temperature increase is limited to 1 K at µLED DC currents of up to 10 mA depending on the stimulation duration and the medium surrounding the probe. The optical output power and peak wavelengths of the new oCI variant were extracted to be 0.82 mW and 462 nm when operating the µLEDs at 10 mA, 10 kHz and a duty cycle of 10 %. The optical power corresponds to a radiant emittance of 407 mW/mm2, sufficient for optogenetic experiments using channelrhodopsin-2.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Optogenetics in neural systems
Yizhar, O., Fenno, L. E., Davidson, T. J., Mogri, M., & Deisseroth, K. (2011) · Neuron · doi:10.1016/j.neuron.2011.06.004 · Cited by 1,881
Cochlear Implants: System Design, Integration, and Evaluation
Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., & Feng, H. (2008) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2008.2008250 · Cited by 878
A MEMS-based flexible multichannel ECoG-electrode array
Rubehn, B., Bosman, C., Oostenveld, R., Fries, P., & Stieglitz, T. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/3/036003 · Cited by 387
Kwon, K. Y., Sirowatka, B., Weber, A., & Li, W. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2013.2282318 · Cited by 166 · cited 1×
“, as optogenetic cardiac pacemaker or defibrillator (Kwon et al., 2013b; Bruegmann et al., 2016; Crocini et al., 2016; Diaz-Maue et al., 2018).”
BackgroundCochlear-implant spatial selectivity with monopolar, bipolar and tripolar stimulation
Zhu, Z., Tang, Q., Zeng, F.-G., Guan, T., & Ye, D. (2012) · Hearing Research · doi:10.1016/j.heares.2011.11.005 · Cited by 115
Cited by
Novel electrode technologies for neural recordings
Hong, G., & Lieber, C. M. (2019) · Nature Reviews Neuroscience · doi:10.1038/s41583-019-0140-6 · Cited by 695
MicroLED technologies and applications: characteristics, fabrication, progress, and challenges
Chen, Z., Yan, S., & Danesh, C. (2021) · Journal of Physics D: Applied Physics · doi:10.1088/1361-6463/abcfe4 · Cited by 336
Wasisto, H. S., Prades, J. D., Gülink, J., & Waag, A. (2019) · Applied Physics Reviews · doi:10.1063/1.5096322 · Cited by 309
Conformable hybrid systems for implantable bioelectronic interfaces
Fallegger, F., Schiavone, G., & Lacour, S. P. (2020) · Advanced Materials · doi:10.1002/adma.201903904 · Cited by 113
Compliant peripheral nerve interfaces.
Paggi, V., Akouissi, O., Micera, S., & Lacour, S. P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abcdbe · Cited by 60
Metadata
- Authors
- Klein, E., Gossler, C., Paul, O., & Ruther, P.
- Venue
- Frontiers in Neuroscience
- Publisher
- Frontiers Media SA
- Published
- 2018-10-01
- Volume
- 12
- Pages
- 659-659
- ISSN
- 1662-453X (ISSN-L)
- OpenAlex
- W2894736135
- Cited by
- 81
- References
- 5
- FWCI
- 4.50
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: High-Density μLED-Based Optical Cochlear Implant With Improved Thermomechanical Behavior. https://app.cochat.ai/papers/10.3389/fnins.2018.00659
- The paper
- Klein, E., Gossler, C., Paul, O., & Ruther, P. (2018). High-Density μLED-Based Optical Cochlear Implant With Improved Thermomechanical Behavior. Frontiers in Neuroscience. https://doi.org/10.3389/fnins.2018.00659
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.