Optimizing NV magnetometry for Magnetoneurography and Magnetomyography applications

Open access

The DOI matches its Crossref record; 66 of the 66 references it cites were found in a registry or an index.

Abstract

Magnetometers based on color centers in diamond are setting new frontiers for sensing capabilities due to their combined extraordinary performances in sensitivity, bandwidth, dynamic range, and spatial resolution, with stable operability in a wide range of conditions ranging from room to low temperatures. This has allowed for its wide range of applications, from biology and chemical studies to industrial applications. Among the many, sensing of bio-magnetic fields from muscular and neurophysiology has been one of the most attractive applications for NV magnetometry due to its compact and proximal sensing capability. Although SQUID magnetometers and optically pumped magnetometers (OPM) have made huge progress in Magnetomyography (MMG) and Magnetoneurography (MNG), exploring the same with NV magnetometry is scant at best. Given the room temperature operability and gradiometric applications of the NV magnetometer, it could be highly sensitive in the pT/Hz-range even without magnetic shielding, bringing it close to industrial applications. The presented work here elaborates on the performance metrics of these magnetometers to the state-of-the-art techniques by analyzing the sensitivity, dynamic range, and bandwidth, and discusses the potential benefits of using NV magnetometers for MMG and MNG applications.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3389/fnins.2022.1034391

Cites

66 of 66 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 (A) Energy level diagram of negatively charged NV center in diamond. The bias field B0 degenerate the ms = ±1 states by 2γeB0. The zero-field splitting is D ≈ 2.87 GHz. The numbers are used to label the different levels. (B) Structure of NV center in diamond. The applied B0 field is parallel to the N-V orientation, and the applied driving field B1 is supposed to be perpendicular to the N-V axis. (C) A typical setup of the NV magnetometer. The green laser is illuminated on the diamond, which is assembled with the optics to collect and guide the fluorescences for detection. The microwave field is used for resonant driving of the spin states. (D) Fluorescence spectrum (the highest black line) emitted by NV center ensembles in a 0.53mm3 diamond. The signals from the two different charge states, i.e., NV− (the middle red line) and NV0 (the lowest yellow line) centers, are estimated in the figure.
Figure 1 (A) Energy level diagram of negatively charged NV center in diamond. The bias field B0 degenerate the ms = ±1 states by 2γeB0. The zero-field splitting is D ≈ 2.87 GHz. The numbers are used to label the different levels. (B) Structure of NV center in diamond. The applied B0 field is parallel to the N-V orientation, and the applied driving field B1 is supposed to be perpendicular to the N-V axis. (C) A typical setup of the NV magnetometer. The green laser is illuminated on the diamond, which is assembled with the optics to collect and guide the fluorescences for detection. The microwave field is used for resonant driving of the spin states. (D) Fluorescence spectrum (the highest black line) emitted by NV center ensembles in a 0.53mm3 diamond. The signals from the two different charge states, i.e., NV− (the middle red line) and NV0 (the lowest yellow line) centers, are estimated in the figure.CC BY · PubMed Central
Figure 2 (A) CW-ODMR method uses continuous wave laser pumping and MW field driving. The MW field is modulated in frequency or phase to get a modulated fluorescence signal. (B) CW-ODMR spectrum demodulated by the lock-in amplifier. MW frequency shift is used to simulate the line shift induced by an external field. B × (2.8 MHz/G) is used as the description of the horizontal axis to show the relationship between the external field (in Gauss) and the line shift (in MHz). (C) Most of the interferometry methods consists of three parts, i.e., initialization, sensing, and detection. In the initialization part, a green laser (532 nm) is used to polarize spin states into ms = 0, regardless of the fluorescence emitted from the diamond. In the sensing part, the two blue blocks represent two MW pulses in a Ramsey sequence, and the yellow block depicts the magnetic field sensing time. The Ramsey sequence can be replaced by different interferometry sequences to measure dc/ac magnetic fields. In the detection part, fluorescences are collected to readout the spin population which indicates the spin-detected magnetic field information. The wedge red indicates that the fluorescence drops at the beginning of the detection window and continuously increases due to the repolarization induced by the detection-laser pulse. (D) The sensor response of Ramsey measurement to the MW frequency shift, which is equivalent to the response to the magnetic field.
Figure 2 (A) CW-ODMR method uses continuous wave laser pumping and MW field driving. The MW field is modulated in frequency or phase to get a modulated fluorescence signal. (B) CW-ODMR spectrum demodulated by the lock-in amplifier. MW frequency shift is used to simulate the line shift induced by an external field. B × (2.8 MHz/G) is used as the description of the horizontal axis to show the relationship between the external field (in Gauss) and the line shift (in MHz). (C) Most of the interferometry methods consists of three parts, i.e., initialization, sensing, and detection. In the initialization part, a green laser (532 nm) is used to polarize spin states into ms = 0, regardless of the fluorescence emitted from the diamond. In the sensing part, the two blue blocks represent two MW pulses in a Ramsey sequence, and the yellow block depicts the magnetic field sensing time. The Ramsey sequence can be replaced by different interferometry sequences to measure dc/ac magnetic fields. In the detection part, fluorescences are collected to readout the spin population which indicates the spin-detected magnetic field information. The wedge red indicates that the fluorescence drops at the beginning of the detection window and continuously increases due to the repolarization induced by the detection-laser pulse. (D) The sensor response of Ramsey measurement to the MW frequency shift, which is equivalent to the response to the magnetic field.CC BY · PubMed Central
Figure 3 (A) Optimization of parameters for the sensitivity of CW-ODMR NV-magnetometry. The three curves (dash, dot, and dash-dot) indicate the Rabi frequencies for the optimized sensitivities with different laser powers. The different curves use different T2^(*) for the calculations. The solid line plots both the laser powers and the Rabi frequencies of the optimized sensitivities for different T2^(*), and the dots are the crossing spots with the other three curves. (B) The upper graph shows that the optimal Rabi frequency is logarithmically linear to the T2^(*) of the diamond. The lower graph shows that the calculated optimized sensitivity is also logarithmically linear to the T2^(*). (C) Optimization of parameters for sensitivities of diamonds with different T1. The curves for different T1 are calculated with T2^(*)=2μs. The solid line indicates the parameters for optimized sensitivity with different T1. (D) The upper graph shows the relationship of the optimal Rabi frequency for different T1, and the lower graph shows the best sensitivities that can be achieved by the diamonds with different T1.
Figure 3 (A) Optimization of parameters for the sensitivity of CW-ODMR NV-magnetometry. The three curves (dash, dot, and dash-dot) indicate the Rabi frequencies for the optimized sensitivities with different laser powers. The different curves use different T2^(*) for the calculations. The solid line plots both the laser powers and the Rabi frequencies of the optimized sensitivities for different T2^(*), and the dots are the crossing spots with the other three curves. (B) The upper graph shows that the optimal Rabi frequency is logarithmically linear to the T2^(*) of the diamond. The lower graph shows that the calculated optimized sensitivity is also logarithmically linear to the T2^(*). (C) Optimization of parameters for sensitivities of diamonds with different T1. The curves for different T1 are calculated with T2^(*)=2μs. The solid line indicates the parameters for optimized sensitivity with different T1. (D) The upper graph shows the relationship of the optimal Rabi frequency for different T1, and the lower graph shows the best sensitivities that can be achieved by the diamonds with different T1.CC BY · PubMed Central
Figure 4 (A) Filter functions of different pulsed sequences that can be used in NV-magnetometry. In the calculation of the filter function of the Ramsey sequence, the sensing time Tϕ = 6.5μs. In the calculations of the sequences for ac field sensing, i.e., Hahn-echo, CPMG2, and XY8, the sensing time Tϕ = 50μs. (B) An overview of the measured frequency responses of NV-magnetometry using the different sequences. The QPSD technique is used to read out the field strength by extracting the quantum phase measured by the sequences. The “×” shows the response of the Ramsey sequence, of which the bandwidth is limited by the sampling rate in the experiment. The “+” and the “◦” shows the ac field responses regarding different Tϕ when Hahn-echo and XY8 are applied for the measurements. The “△” shows the measurements of a 200 kHz signal by CPMG sequences with different numbers of the π-pulses. One of the primary goals of the NV-magnetometry development is to develop concatenation of the sequences that can ensure a flat frequency response for ac field sensing.
Figure 4 (A) Filter functions of different pulsed sequences that can be used in NV-magnetometry. In the calculation of the filter function of the Ramsey sequence, the sensing time Tϕ = 6.5μs. In the calculations of the sequences for ac field sensing, i.e., Hahn-echo, CPMG2, and XY8, the sensing time Tϕ = 50μs. (B) An overview of the measured frequency responses of NV-magnetometry using the different sequences. The QPSD technique is used to read out the field strength by extracting the quantum phase measured by the sequences. The “×” shows the response of the Ramsey sequence, of which the bandwidth is limited by the sampling rate in the experiment. The “+” and the “◦” shows the ac field responses regarding different Tϕ when Hahn-echo and XY8 are applied for the measurements. The “△” shows the measurements of a 200 kHz signal by CPMG sequences with different numbers of the π-pulses. One of the primary goals of the NV-magnetometry development is to develop concatenation of the sequences that can ensure a flat frequency response for ac field sensing.CC BY · PubMed Central
Figure 5 (A) ODMR spectra with different local gradients at the diamond. (B) Magnetic field detected by the NV gradiometer in the unshielded environment with the comparison of a fluxgate magnetometer. The schematic of the setup is shown as the inset drawing. The step signals are generated by an elevator nearby. From top to bottom, the lines are signals from the fluxgate sensor, NV gradiometer channel 1 and channel 2, and the output of the NV gradiometer, i.e., subtraction of the two channels. The dashed lines and arrows are the eye-guide for comparing the noise of NV channels and the fluxgate magnetometer. (C) The action potential signal (generated by a phantom) detected by the NV magnetometer. The original input of the action potential is as the lower waveform shows. The detected magnetic field amplitude is about 2 nT. (D) is the Fourier transform of the detected action potential signal. The principle noise peaks are at dc, 50 Hz, and 100 Hz. (E) The phantom used for generating magnetic field signals to simulate the MMG signal.
Figure 5 (A) ODMR spectra with different local gradients at the diamond. (B) Magnetic field detected by the NV gradiometer in the unshielded environment with the comparison of a fluxgate magnetometer. The schematic of the setup is shown as the inset drawing. The step signals are generated by an elevator nearby. From top to bottom, the lines are signals from the fluxgate sensor, NV gradiometer channel 1 and channel 2, and the output of the NV gradiometer, i.e., subtraction of the two channels. The dashed lines and arrows are the eye-guide for comparing the noise of NV channels and the fluxgate magnetometer. (C) The action potential signal (generated by a phantom) detected by the NV magnetometer. The original input of the action potential is as the lower waveform shows. The detected magnetic field amplitude is about 2 nT. (D) is the Fourier transform of the detected action potential signal. The principle noise peaks are at dc, 50 Hz, and 100 Hz. (E) The phantom used for generating magnetic field signals to simulate the MMG signal.CC BY · PubMed Central
Figure 6 (A) Sketch of relevant parameters that have to be taken into account when compromising the noise and signal suppression in a linear gradiometer configuration. (B) Noise suppression ratio for different baselengths as a function of the distance to the noise sources. (C) Relative differential signal of the gradiometers for different distances to the signal source as a function of baselength. For both cases, it is assumed that the magnetic field strength of the noise source and signal sources decay as B(r) ∝ r−2 with distance to the different sources.
Figure 6 (A) Sketch of relevant parameters that have to be taken into account when compromising the noise and signal suppression in a linear gradiometer configuration. (B) Noise suppression ratio for different baselengths as a function of the distance to the noise sources. (C) Relative differential signal of the gradiometers for different distances to the signal source as a function of baselength. For both cases, it is assumed that the magnetic field strength of the noise source and signal sources decay as B(r) ∝ r−2 with distance to the different sources.CC BY · PubMed Central
Figure 7 (A) For a cube-shaped muscle compound, muscle action potentials are simulated when muscle fibers in different depths are stimulated. A 2 mm thick fat tissue layer is added on top of the muscle, and the magnetic field is observed midway between the innervation zone (IZ) and the myotendinous junction (MTJ). For obtaining a differential signal, the magnetic field is sampled at the body surface and in a line with distance Δxt^(⊥) above the muscle surface. (B) Exemplary MMG signal measured at one point with a distance of 0.5 mm to the body surface when varying the depth of the active muscle fibers. (C) Signal amplitude on the body surface in a line perpendicular to the muscle fibers (cf. A). The left column shows the root mean square (RMS) of the raw MMG signal, depending on the spatial coordinate and the depth of the active muscle fibers. The middle and the right column show the RMS distribution corresponding to differential signals with inter-sensor distances of 5 and 0.5 mm, respectively. Thereby, it can be observed that decreasing the distance between the sensors narrows the amplitude distributions and hence reduces the detection volume of measurement.
Figure 7 (A) For a cube-shaped muscle compound, muscle action potentials are simulated when muscle fibers in different depths are stimulated. A 2 mm thick fat tissue layer is added on top of the muscle, and the magnetic field is observed midway between the innervation zone (IZ) and the myotendinous junction (MTJ). For obtaining a differential signal, the magnetic field is sampled at the body surface and in a line with distance Δxt^(⊥) above the muscle surface. (B) Exemplary MMG signal measured at one point with a distance of 0.5 mm to the body surface when varying the depth of the active muscle fibers. (C) Signal amplitude on the body surface in a line perpendicular to the muscle fibers (cf. A). The left column shows the root mean square (RMS) of the raw MMG signal, depending on the spatial coordinate and the depth of the active muscle fibers. The middle and the right column show the RMS distribution corresponding to differential signals with inter-sensor distances of 5 and 0.5 mm, respectively. Thereby, it can be observed that decreasing the distance between the sensors narrows the amplitude distributions and hence reduces the detection volume of measurement.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 66 references

Research works printed but not linked: 2

  1. Kandel, 2000, Principles of Neural Science, Vol. 4
  2. Wolf, 2018, On magnetometry using ensembles of NV defect centers in diamond

Cited by

See all 41

Metadata

Authors
Zhang, C., Zhang, J., Widmann, M., Benke, M., Kübler, M., Dasari, D., Klotz, T., Gizzi, L., Röhrle, O., Brenner, P., & Wrachtrup, J.
Affiliations
University of Stuttgart (DE)
Corresponding author
  • Zhang, Chen
  • Wrachtrup, Jörg
Venue
Frontiers in Neuroscience
Published
2023-01-12
Volume
16
Pages
1034391-1034391
ISSN
1662-453X (ISSN-L)
OpenAlex
W4315640927
Cited by
34
References
66
Citations by year
2026: 15 · 2025: 7 · 2024: 9 · 2023: 3
FWCI
2.51
Type
journal article
Language
en
Open access
Locations
  • Frontiers in Neuroscience (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • OPUS Publication Server of the University of Stuttgart (University of Stuttgart) (submittedVersion, cc-by)
  • Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
  • Universitätsbibliothek Stuttgart (cc-by)
Funders
  • European Commission (742610, 820394, HORIZON2020)
  • Bundesministerium für Bildung und Forschung
  • H2020 European Research Council (742610)
  • HORIZON EUROPE European Research Council (101055186)
Topics
  • Physical Sciences › Materials Science › Materials Chemistry › Diamond and Carbon-based Materials Research
  • Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
Keywords
diamond color centers, magnetomyography, magnetometer sensitivity, dynamic range, bandwidth, room-temperature operation

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Optimizing NV magnetometry for Magnetoneurography and Magnetomyography applications. https://app.cochat.ai/papers/10.3389/fnins.2022.1034391
The paper
Zhang, C., Zhang, J., Widmann, M., Benke, M., Kübler, M., Dasari, D., Klotz, T., Gizzi, L., Röhrle, O., Brenner, P., & Wrachtrup, J. (2023). Optimizing NV magnetometry for Magnetoneurography and Magnetomyography applications. Frontiers in Neuroscience. https://doi.org/10.3389/fnins.2022.1034391

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.