Non-Contact Vital Signs Monitoring of Dog and Cat Using a UWB Radar
Abstract
As pets are considered members of the family, their health has received widespread attention. Since pets cannot talk and complain when they feel uncomfortable, monitoring vital signs becomes very helpful in disease detection, as well as observing their progression and response to treatment. In this study, we proposed an ultra-wideband radar-based, non-contact animal vital sign monitoring scheme that could monitor the breathing and heart rate of a pet in real-time. The primary advantage of the ultra-wideband radar was its ability to operate remotely without electrodes or wires and through any clothing or fur. Because of the existing noise and clutter in non-contact detection, background noise removal was applied. Furthermore, the respiration rate was directly obtained through spectrum analysis, while the heartbeat signal was extracted by the variational mode decomposition algorithm. By using electrocardiogram measurements, we verified the accuracy of the radar technology in detecting the anesthetized animals’ respiratory rate and heart rate. Besides, three beagles and five cats in a non-sedated state were measured by radar and contact pressure sensors simultaneously; the experimental results showed that radar could effectively measure the respiration of cats and dogs, and the accuracy rate was over 95%. Due to its excellent performance, the proposed method has the potential to become a new choice in application scenarios, such as pet sleep monitoring and health assessment.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The respiratory rate has been used to monitor different pathological conditions and stressors, including emotional stress, cognitive load, heat and cold.
“The respiratory rate has been used to monitor different pathological conditions and stressors, including emotional stress, cognitive load, heat and cold [103,104].”
From Therapy Dog Welfare Revisited: A Review of the Literature (2021) · cites this paper as [104]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
Flagged first, then by citations
Research works printed but not linked: 40
- Langbein, 2007, Heart rate variability as a measure of autonomic regulation of cardiac activity for assessing stress and welfare in farm animals—A review, Physiol. Behav., 92, 293, 10.1016/j.physbeh.2007.01.007 · DOI
- Varga, B., Gergely, A., Galambos, Á., and Kis, A. (2018). Heart Rate and Heart Rate Variability during Sleep in Family Dogs (Canis familiaris). Moderate Effect of Pre-Sleep Emotions. Animals, 8. · DOI
- Wiepkema, 1992, The emotional brain, Anim. Welf., 1, 13, 10.1017/S0962728600014688 · DOI
- Rosselot, P., Mendonça, T., González, I., and Tadich, T. (2019). Behavioral and Physiological Differences between Working Horses and Chilean Rodeo Horses in a Handling Test. Animals, 9. · DOI
- Marocolo, 2010, Noninvasive method for electrocardiogram recording in conscious rats: Feasibility for heart rate variability analysis, An. Acad. Bras. Ciências, 82, 431, 10.1590/S0001-37652010000200019 · DOI
- Bailey, 1998, Anesthetic monitoring and monitoring equipment: Application in small exotic pet practice, Seminars in Avian and Exotic Pet Medicine, Volume 7, 53, 10.1016/S1055-937X(98)80058-3 · DOI
- Detweiler, D.K. (2010). The dog electrocardiogram: A critical review. Compr. Electrocardiol., 1861–1908. · DOI
- Farmer, 1968, A simple method for recording the electrocardiogram and heart rate from conscious animals, Br. J. Pharmacol. Chemother., 32, 193, 10.1111/j.1476-5381.1968.tb00443.x · DOI
- Gonder, J.C., Gard, E.A., and Lott, N.E. (2020, January 21). Electrocardiograms of Nine Species of Nonhuman Primates Sedated with Ketamine. Available online: https://www.ncbi.nlm.nih.gov/pubmed/6776852.
- Eckenfels, 1979, The normal electrocardiogram of the conscious beagle dog, Toxicol. Appl. Pharmacol., 47, 567, 10.1016/0041-008X(79)90527-1 · DOI
- Kindig, 1982, Acoustical performance of the stethoscope: A comparative analysis, Am. Heart J., 104, 269, 10.1016/0002-8703(82)90203-4 · DOI
- Liang, H., Lukkarinen, S., and Hartimo, I. (1997). Heart sound segmentation algorithm based on heart sound envelogram. Computers in Cardiology 1997, IEEE. · DOI
- Nyland, T.G., and Mattoon, J.S. (2002). Small Animal Diagnostic Ultrasound, Elsevier Health Sciences.
- Frinking, 2000, Ultrasound contrast imaging: Current and new potential methods, Ultrasound Med. Biol., 26, 965, 10.1016/S0301-5629(00)00229-5 · DOI
- Allen, 2007, Photoplethysmography and its application in clinical physiological measurement, Physiol. Meas., 28, R1, 10.1088/0967-3334/28/3/R01 · DOI · Paper page
- Poh, 2011, Advancements in Noncontact Multiparameter Physiological Measurements Using a Webcam, IEEE Trans. Biomed. Eng., 58, 7, 10.1109/TBME.2010.2086456 · DOI
- Wang, 2017, Robust heart rate from fitness videos, Physiol. Meas., 38, 1023, 10.1088/1361-6579/aa6d02 · DOI
- Wang, 2018, A Comparative Survey of Methods for Remote Heart Rate Detection From Frontal Face Videos, Front. Bioeng. Biotechnol., 6, 33, 10.3389/fbioe.2018.00033 · DOI
- Li, 2013, A review on recent advances in Doppler radar sensors for noncontact healthcare monitoring, IEEE Trans. Microw. Theory Tech., 61, 2046, 10.1109/TMTT.2013.2256924 · DOI
- Li, 2017, A review on recent progress of portable short-range noncontact microwave radar systems, IEEE Trans. Microw. Theory Tech., 65, 1692, 10.1109/TMTT.2017.2650911 · DOI
- Zhang, 2019, Bioradar Technology: Recent Research and Advancements, IEEE Microw. Mag., 20, 58, 10.1109/MMM.2019.2915491 · DOI
- Granhaug, 2017, A 118-mW pulse-based radar soc in 55-nm CMOS for non-contact human vital signs detection, IEEE J. Solid-State Circuits, 52, 3421, 10.1109/JSSC.2017.2764051 · DOI
- Wang, P., Zhang, Y., Ma, Y., Liang, F., An, Q., Xue, H., Yu, X., Lv, H., and Wang, J. (2019). Method for Distinguishing Humans and Animals in Vital Signs Monitoring Using IR-UWB Radar. Int. J. Environ. Res. Public Health, 16. · DOI
- Wang, 2019, Noncontact Heart Rate Measurement Based on an Improved Convolutional Sparse Coding Method Using IR-UWB Radar, IEEE Access, 7, 158492, 10.1109/ACCESS.2019.2950423 · DOI
- Huang, T.Y., Lin, J., and Hayward, L. (2015). Non-invasive measurement of laboratory rat’s cardiorespiratory movement using a 60-GHz radar and nonlinear Doppler phase modulation. 2015 IEEE MTT-S 2015 International Microwave Workshop Series on RF and Wireless Technologies for Biomedical and Healthcare Applications (IMWS-BIO), IEEE. · DOI
- Huang, 2016, Noninvasive measurement and analysis of laboratory rat’s cardiorespiratory movement, IEEE Trans. Microw. Theory Tech., 65, 574, 10.1109/TMTT.2016.2616870 · DOI
- Churkin, S., and Anishchenko, L. (2015). Millimeter-wave radar for vital signs monitoring. 2015 IEEE International Conference on Microwaves, Communications, Antennas and Electronic Systems (COMCAS), IEEE. · DOI
- Hui, 2019, No-touch measurements of vital signs in small conscious animals, Sci. Adv., 5, eaau0169, 10.1126/sciadv.aau0169 · DOI
- Lazaro, 2010, Analysis of vital signs monitoring using an IR-UWB radar, Prog. Electromagn. Res., 100, 265, 10.2528/PIER09120302 · DOI
- Dragomiretskiy, 2014, Variational Mode Decomposition, IEEE Trans. Signal Process., 62, 531, 10.1109/TSP.2013.2288675 · DOI · Paper page
- Wang, 2015, Research on variational mode decomposition and its application in detecting rub-impact fault of the rotor system, Mech. Syst. Signal Process., 60–61, 243, 10.1016/j.ymssp.2015.02.020 · DOI
- Ming, 2017, Research on variational mode decomposition in rolling bearings fault diagnosis of the multistage centrifugal pump, Mech. Syst. Signal Process., 93, 460, 10.1016/j.ymssp.2017.02.013 · DOI
- Lahmiri, 2014, Comparative study of ECG signal denoising by wavelet thresholding in empirical and variational mode decomposition domains, Healthc. Technol. Lett., 1, 104
- Shen, 2018, Respiration and Heartbeat Rates Measurement Based on Autocorrelation Using IR-UWB Radar, IEEE Trans. Circuits Syst. II Express Briefs, 65, 1470
- Xue, 2017, Application of the Variational-Mode Decomposition for Seismic Time–frequency Analysis, IEEE J. Sel. Top. Appl. Earth Obs. Remote Sens., 9, 3821, 10.1109/JSTARS.2016.2529702 · DOI
- Pinheiro, 2010, Theory and developments in an unobtrusive cardiovascular system representation: Ballistocardiography, Open Biomed. Eng. J., 4, 201, 10.2174/1874120701004010201 · DOI
- Inan, 2014, Ballistocardiography and seismocardiography: A review of recent advances, IEEE J. Biomed. Health Inform., 19, 1414, 10.1109/JBHI.2014.2361732 · DOI · Paper page
- Dong, S., Zhang, Y., Ma, C., Zhu, C., Gu, Z., Lv, Q., Zhang, B., Li, C., and Ran, L. (2019). Doppler Cardiogram: A Remote Detection of Human Heart Activities. IEEE Trans. Microw. Theory Tech., 1–10.
- Ferris, 1998, Microwave and millimeter-wave systems for wall penetration, Targets and Backgrounds: Characterization and Representation IV, Volume 3375, 269
- Rong, Y., and Bliss, D.W. (2019). Smart Homes: See Multiple Heartbeats Through Wall Using Wireless Signals. 2019 IEEE Radar Conference (RadarConf), IEEE. · DOI
Cited by
Sorted by citations
Therapy Dog Welfare Revisited: A Review of the Literature
Glenk, L. M., & Foltin, S. (2021) · Veterinary Sciences · doi:10.3390/vetsci8100226 · Cited by 71
Metadata
- Authors
- Wang, P., Ma, Y., Liang, F., Zhang, Y., Yu, X., Li, Z., An, Q., Lv, H., & Wang, J.
- Affiliations
- Department of Medical Electronics, School of Biomedical Engineering, The Fourth Military Medical University, Xi’an 710032, China
- Venue
- Animals
- Publisher
- MDPI AG
- Published
- 2020-01-25
- Volume
- 10
- Issue
- 2
- Pages
- 205
- ISSN
- 2076-2615
- PMC
- PMC7070589
- Cited by
- 45
- References
- 1
- Language
- en
- Funders
- Natural Science Basic Research Project of Shaanxi (2019JZ-35)
- National Natural Science Foundation of China (81601567, 31800822)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review