Ballistocardiography and Seismocardiography: A Review of Recent Advances
Open accessAbstract
In the past decade, there has been a resurgence in the field of unobtrusive cardiomechanical assessment, through advancing methods for measuring and interpreting ballistocardiogram (BCG) and seismocardiogram (SCG) signals. Novel instrumentation solutions have enabled BCG and SCG measurement outside of clinical settings, in the home, in the field, and even in microgravity. Customized signal processing algorithms have led to reduced measurement noise, clinically relevant feature extraction, and signal modeling. Finally, human subjects physiology studies have been conducted using these novel instruments and signal processing tools with promising results. This paper reviews the recent advances in these areas of modern BCG and SCG research.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The z-axis accelerometer data from the chest sensor of a 1-wk-old subject contains clear information on cardiac activity, as a seismocardiogram with prominent peaks induced by Aortic valve opening and Mitral valve opening.
“For example, the z-axis accelerometer data from the chest sensor of a 1-wk-old subject (Top in Fig. 6A) contains clear information on cardiac activity, as a seismocardiogram with prominent peaks induced by Aortic valve opening (black arrow) and Mitral valve opening (gray arrow) (30–32).”
From Miniaturized wireless, skin-integrated sensor networks for quantifying full-body movement behaviors and vital signs in infants (2021) · cites this paper as [31]
Conventional accelerometers are uncomfortable to wear and susceptible to the global inertial motion of the body.
“Although SCG can be measured by mounting commercial accelerometers on the human chest,6 the thickness and rigidity of conventional accelerometers make them uncomfortable to wear and susceptible to the global inertial motion of the body.7”
From A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals (2019) · cites this paper as [6]
Another patch that uses seismocardiography techniques has been able to detect congestion on the basis of measurement of chest wall vibration correlating with the movement of the heart in the chest and blood flow.
“Another patch that uses seismocardiography techniques has been able to detect congestion on the basis of measurement of chest wall vibration correlating with the movement of the heart in the chest and blood flow.38”
From Unlocking Tomorrow’s Health Care: Expanding the Clinical Scope of Wearables by Applying Artificial Intelligence (2024) · cites this paper as [38]
Cites
Telemonitoring in patients with heart failure
Chaudhry, S. I., Mattera, J. A., Curtis, J. P., Spertus, J. A., Herrin, J., Lin, Z., Phillips, C. O., Hodshon, B. V., Cooper, L. S., & Krumholz, H. M. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1010029 · Cited by 1,232 · see citing works
Systolic Time Intervals in Heart Failure in Man
WEISSLER, A. M., HARRIS, W. S., & SCHOENFELD, C. D. (1968) · Circulation · doi:10.1161/01.cir.37.2.149 · Cited by 1,217 · see citing works
Analog studies of the human systemic arterial tree
Westerhof, N., Bosman, F., De Vries, C. J., & Noordergraaf, A. (1969) · Journal of Biomechanics · doi:10.1016/0021-9290(69)90024-4 · Cited by 684 · cited 1× · see citing works
“[86].”
Detecting Pulse from Head Motions in Video
Balakrishnan, G., Durand, F., & Guttag, J. V. (2013) · IEEE Computer Society Conference on Computer Vision and Pattern Recognition/Proceedings - IEEE Computer Society Conference on Computer Vision and Pattern Recognition/Proceedings · doi:10.1109/cvpr.2013.440 · Cited by 572 · see citing works
Garrard, C. L., Weissler, A. M., & Dodge, H. T. (1970) · Circulation · doi:10.1161/01.cir.42.3.455 · Cited by 455 · see citing works
Patterns of Weight Change Preceding Hospitalization for Heart Failure
Chaudhry, S. I., Wang, Y., Concato, J., Gill, T. M., & Krumholz, H. M. (2007) · Circulation · doi:10.1161/circulationaha.107.690768 · Cited by 437 · see citing works
Starr, I., Rawson, A. J., Schroeder, H. A., & Joseph, N. (1939) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1939.127.1.1 · Cited by 398 · see citing works
A new method for long-term monitoring of the ballistocardiogram, heart rate, and respiration
Alihanka, J., Vaahtoranta, K., & Saarikivi, I. (1981) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1981.240.5.r384 · Cited by 272 · see citing works
Watanabe, K., Watanabe, T., Watanabe, H., Ando, H., Ishikawa, T., & Kobayashi, K. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.857637 · Cited by 239 · see citing works
Robust ballistocardiogram acquisition for home monitoring
Inan, O. T., Etemadi, M., Wiard, R. M., Giovangrandi, L., & Kovacs, G. T. A. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/2/005 · Cited by 234 · see citing works
Self-citationAdaptive Beat-to-Beat Heart Rate Estimation in Ballistocardiograms
Brüser, C., Stadlthanner, K., de Waele, S., & Leonhardt, S. K. (2011) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2128337 · Cited by 219 · see citing works
Sleep Staging Based on Signals Acquired Through Bed Sensor
Kortelainen, J., Mendez, M. O., Bianchi, A. M., Matteucci, M., & Cerutti, S. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2010.2044797 · Cited by 202 · see citing works
Mack, D. C., Patrie, J. T., Suratt, P. M., Felder, R. A., & Alwan, M. (2008) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2008.2007194 · Cited by 196 · see citing works
Certain Molar Movements of the Human Body produced by the Circulation of the Blood.
Jw, G. (1877) · PubMed · Cited by 193 · see citing works
Monitoring Respiration and Cardiac Activity Using Fiber Bragg Grating-Based Sensor
Dziuda, Ł., Skibniewski, F. W., Krej, M., & Lewandowski, J. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2194145 · Cited by 186 · see citing works
Relationship between Seismocardiogram and Echocardiogram for Events in the Cardiac Cycle
Crow, R. S., Hannan, P., Jacobs, D., Hedquist, L., & Salerno, D. M. (2017) · American Journal of Noninvasive Cardiology · doi:10.1159/000470156 · Cited by 186 · cited 3× · see citing works
“Significant features of the SCG waveform were identified and associated with key events in the cardiac cycle [17].”
BackgroundMethodInfluentialTheory and Developments in an Unobtrusive Cardiovascular System Representation: Ballistocardiography
Pinheiro, E., Postolache, O. A., & Girão, P. S. (2010) · The Open Biomedical Engineering Journal · doi:10.2174/1874120701004010201 · Cited by 176 · see citing works
Real-Time Monitoring of Respiration Rhythm and Pulse Rate During Sleep
Zhu, X., Chen, W., Nemoto, T., Kanemitsu, Y., Kitamura, K., Yamakoshi, K., & Wei, D. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.884641 · Cited by 171 · see citing works
Ballistocardiography — A method worth revisiting
Giovangrandi, L., Inan, O. T., Wiard, R. M., Etemadi, M., & Kovacs, G. T. A. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091062 · Cited by 165 · see citing works
Self-citationDi Rienzo, M., Vaini, E., Castiglioni, P., Merati, G., Meriggi, P., Parati, G., Faini, A., & Rizzo, F. (2013) · Autonomic Neuroscience · doi:10.1016/j.autneu.2013.04.005 · Cited by 163 · see citing works
Castiglioni, P., Faini, A., Parati, G., & Rienzo, M. D. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353199 · Cited by 156 · see citing works
Seismocardiography: Past, present and future
Zanetti, J. M., & Tavakolian, K. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6611170 · Cited by 152 · see citing works
Self-citationSeismocardiography for Monitoring Changes in Left Ventricular Function during Ischemia
Salerno, D. M., & Zanetti, J. (1991) · Chest · doi:10.1378/chest.100.4.991 · Cited by 149 · see citing works
Self-citationNoncontact Method for Sleep Stage Estimation
Watanabe, T., & Watanabe, K. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.828037 · Cited by 148 · see citing works
The smart car seat: personalized monitoring of vital signs in automotive applications
Walter, M., Eilebrecht, B., Wartzek, T., & Leonhardt, S. K. (2011) · Personal and Ubiquitous Computing · doi:10.1007/s00779-010-0350-4 · Cited by 139 · see citing works
Robust inter-beat interval estimation in cardiac vibration signals
Brüser, C., Winter, S., & Leonhardt, S. K. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/2/123 · Cited by 138 · see citing works
Adaptive Heartbeat Modeling for Beat-to-Beat Heart Rate Measurement in Ballistocardiograms
Paalasmaa, J., Toivonen, H. T. T., & Partinen, M. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2314144 · Cited by 129 · see citing works
Automatic Detection of Atrial Fibrillation in Cardiac Vibration Signals
Brüser, C., Diesel, J., Zink, M. D., Winter, S., SCHAUERTE, P. N., & Leonhardt, S. K. (2012) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/titb.2012.2225067 · Cited by 126 · see citing works
Salerno, D. M., & Zanetti, J. M. (1990) · Cited by 123 · see citing works
Self-citationDevelopment of a PVDF Piezopolymer Sensor for Unconstrained In-Sleep Cardiorespiratory Monitoring
Wang, F., Tanaka, M., & CHONAN, S. (2003) · Journal of Intelligent Material Systems and Structures · doi:10.1177/1045389x03014003006 · Cited by 114 · see citing works
Pandia, K., Inan, O. T., Kovacs, G. T. A., & Giovangrandi, L. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1643 · Cited by 114 · see citing works
Self-citationChee, Y., Han, J., Youn, J., & Park, K. (2005) · Physiological Measurement · doi:10.1088/0967-3334/26/4/007 · Cited by 111 · cited 1× · see citing works
“[40], and more recently the following implementations: Pressure sensor in the air mattress [41] or in pad [42], film-type force sensors [43] or load cells in the legs of bed [44], microbend fiber optic BCG sensor [45]–[47], EMFi sensors [48], piezoelectric film sensors [49] or polyvinylidene fluoride (PVDF) sensors [5…”
BackgroundSelf-citationA new method for measuring the ballistocardiogram using EMFi sensors in a normal chair
Koivistoinen, T., Junnila, S., Värri, A. O., et al. (2005) · doi:10.1109/iembs.2004.1403596 · Cited by 104 · see citing works
Chuo, Y., Marzencki, M., Hung, B., Jaggernauth, C., Tavakolian, K., Lin, P., & Kaminska, B. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2052616 · Cited by 102 · see citing works
Self-citationHe, D. D., Winokur, E. S., & Sodini, C. G. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091171 · Cited by 98 · see citing works
Unobtrusive online monitoring of sleep at home
Paalasmaa, J., Waris, M., Toivonen, H. T. T., Leppakorpi, L., & Partinen, M. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346791 · Cited by 98 · see citing works
Rapid Assessment of Cardiac Contractility on a Home Bathroom Scale
Etemadi, M., Inan, O. T., Giovangrandi, L., & Kovacs, G. T. A. (2011) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2161998 · Cited by 97 · see citing works
Self-citationProposals for Ballistocardiographic Nomenclature and Conventions: Revised and Extended
SCARBOROUGH, W. R., Talbot, S. A., Braunstein, J. R., et al. (1956) · Circulation · doi:10.1161/01.cir.14.3.435 · Cited by 94 · see citing works
Inan, O. T., Etemadi, M., Paloma, A., Giovangrandi, L., & Kovacs, G. T. A. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/3/003 · Cited by 94 · see citing works
Self-citationKortelainen, J. M., & Virkkala, J. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353894 · Cited by 90 · see citing works
Automatic ballistocardiogram (BCG) beat detection using a template matching approach
Shin, J., Choi, B. H., Lim, Y. G., Jeong, D. U., & Park, K. S. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4649363 · Cited by 90 · see citing works
Self-citationNon-constrained monitoring of systolic blood pressure on a weighing scale
Shin, J. H., Lee, K. M., & Park, K. S. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/7/011 · Cited by 86 · see citing works
Self-citationMonitoring physiological signals using nonintrusive sensors installed in daily life equipment
Lim, Y. G., Hong, K. H., Kim, K. K., Shin, J. H., Lee, S. M., Chung, G. S., Baek, H. J., Jeong, D., & Park, K. S. (2011) · Biomedical Engineering Letters · doi:10.1007/s13534-011-0012-0 · Cited by 81 · see citing works
Self-citationSlow-wave sleep estimation on a load-cell-installed bed: a non-constrained method
Choi, B., Chung, G. S., Lee, J.-S., et al. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/11/002 · Cited by 77 · see citing works
THE MASS-MOVEMENTS OF THE CIRCULATION AS SHOWN BY A RECOIL CURVE
Henderson, Y. (1905) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1905.14.3.287 · Cited by 74 · see citing works
Monitoring of the ballistocardiogram with the static charge sensitive bed
Jansen, B. M. P., Larson, B. H., & Shankar, K. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83586 · Cited by 73 · see citing works
Improvement of Force-Sensor-Based Heart Rate Estimation Using Multichannel Data Fusion
Brüser, C., Kortelainen, J. M., Winter, S., Tenhunen, M. L., Pärkkä, J., & Leonhardt, S. K. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2311582 · Cited by 72 · see citing works
Nocturnal Awakening and Sleep Efficiency Estimation Using Unobtrusively Measured Ballistocardiogram
Jung, D. W., Hwang, S. H., Yoon, H. N., Lee, Y.-J. G., Jeong, D. U., & Park, K. S. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2278020 · Cited by 71 · see citing works
Self-citationMotion artifact cancellation to obtain heart sounds from a single chest-worn accelerometer
Pandia, K., Ravindran, S., Cole, R., et al. (2010) · IEEE International Conference on Acoustics Speech and Signal Processing · doi:10.1109/icassp.2010.5495553 · Cited by 69 · see citing works
Salerno, D. M., Zanetti, J. M., Green, L. A., Mooney, M. R., Madison, J. D., & Tassel, R. A. V. (1991) · The American Journal of Cardiology · doi:10.1016/0002-9149(91)90744-6 · Cited by 68 · see citing works
Self-citationStudies with the ballistocardiograph in acute cardiac infarction and chronic angina pectoris
Starr, I., & Wood, F. C. (1943) · American Heart Journal · doi:10.1016/s0002-8703(43)90383-7 · Cited by 65 · see citing works
Multichannel bed pressure sensor for sleep monitoring
Kortelainen, J. M., van Gils, M., & Pärkkä, J. (2012) · Cited by 61 · see citing works
Wilson, R. A., Bamrah, V. S., Lindsay, J., Schwaiger, M., & Morganroth, J. (1993) · The American Journal of Cardiology · doi:10.1016/0002-9149(93)90508-a · Cited by 61 · see citing works
A wearable system for the seismocardiogram assessment in daily life conditions
Rienzo, M. D., Meriggi, P., Rizzo, F., Vaini, E., Faini, A., Merati, G., Parati, G., & Castiglioni, P. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091058 · Cited by 61 · see citing works
[Seismocardiography--a new method in the study of functional conditions of the heart].
Bs, B. (1961) · PubMed · Cited by 60 · see citing works
Winokur, E. S., He, D. D., & Sodini, C. G. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346527 · Cited by 59 · see citing works
Heart rate detection from an electronic weighing scale
González-Landaeta, R., Casas, Ó., & Pallas-Areny, R. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/8/009 · Cited by 59 · see citing works
Šprager, S., & Zazula, D. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2213302 · Cited by 57 · see citing works
Precordial Vibrations Provide Noninvasive Detection of Early-Stage Hemorrhage
Tavakolian, K., Dumont, G. A., Houlton, G., & Blaber, A. P. (2013) · Shock · doi:10.1097/shk.0000000000000077 · Cited by 57 · see citing works
Self-citationShin, J. H., Hwang, S. H., Chang, M. H., & Park, K. S. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/8/015 · Cited by 55 · see citing works
Self-citationAutomatic sleep staging based on ballistocardiographic signals recorded through bed sensors
Migliorini, M., Bianchi, A. M., Nistico, D., Kortelainen, J., Arce-Santana, E. R., Cerutti, S., & Mendez, M. O. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5627217 · Cited by 53 · see citing works
Heart rate estimation on a beat-to-beat basis via ballistocardiography - a hybrid approach
Friedrich, D., Aubert, X. L., Führ, H., & Brauers, A. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5627626 · Cited by 46 · see citing works
Multi-channel optical sensor-array for measuring ballistocardiograms and respiratory activity in bed
Brüser, C., Kerekes, A., Winter, S., & Leonhardt, S. K. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6347126 · Cited by 44 · see citing works
Unobstrusive heart rate and respiratory rate monitor embedded on a wheelchair
Postolache, O. A., Girão, P. S., Mendes, J. G., et al. (2009) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2009.5167960 · Cited by 41 · see citing works
Marcus, G. M., Sorrell, V. L., Zanetti, J. M., Bosnos, M., Baweja, G., PERLICK, D., Ott, P., Indik, J. H., He, D. S., & Gear, K. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00894.x · Cited by 41 · see citing works
Self-citationThree-dimensional ballistocardiography and respiratory motion in sustained microgravity.
Prisk, G. K., Verhaeghe, S. T. L., Padeken, D., Hamacher, H., & Paiva, M. (2001) · PubMed · Cited by 40 · see citing works
Self-citationWiard, R. M., Inan, O. T., Argyres, B., et al. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0722-y · Cited by 38 · see citing works
Self-citationInan, O. T., Etemadi, M., Widrow, B. C., et al. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2018831 · Cited by 38 · see citing works
Self-citationPulse arrival time and ballistocardiogram application to blood pressure variability estimation
Pinheiro, E., Postolache, O. A., & Girão, P. S. (2009) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2009.5167970 · Cited by 37 · see citing works
Inan, O. T., Etemadi, M., Wiard, R. M., et al. (2009) · doi:10.1109/iembs.2009.5333709 · Cited by 36 · see citing works
Self-citationVehkaoja, A. T., Rajala, S., Kumpulainen, P. S., & Lekkala, J. O. (2013) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2013.807523 · Cited by 35 · see citing works
Comparative analysis of three different modalities for characterization of the seismocardiogram
Akhbardeh, A., Tavakolian, K., Gurev, V., et al. (2009) · doi:10.1109/iembs.2009.5334444 · Cited by 35 · see citing works
Self-citationExercise Seismocardiography for Detection of Coronary Artery Disease
Salerno, D. M., Zanetti, J. M., Poliac, L. C., Crow, R. S., Hannan, P. J., Wang, K., Goldenberg, I. F., & Tassel, R. A. V. (2017) · American Journal of Noninvasive Cardiology · doi:10.1159/000470383 · Cited by 35 · see citing works
Self-citationPreliminary results from BCG and ECG measurements in the heart failure clinic
Giovangrandi, L., Inan, O. T., Banerjee, D., et al. (2012) · doi:10.1109/embc.2012.6346790 · Cited by 34 · see citing works
Self-citationJerosch‐Herold, M., Zanetti, J. M., Merkle, H., Poliac, L. C., Huang, H., Mansoor, A., Zhao, F., & Wilke, N. M. (1999) · International journal of cardiac imaging · doi:10.1023/a:1006364518204 · Cited by 33 · see citing works
Self-citationRienzo, M. D., Vaini, E., Castiglioni, P., Meriggi, P., & Rizzo, F. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6611173 · Cited by 33 · see citing works
Zhu, Y., Zhang, H., Jayachandran, M., et al. (2013) · doi:10.1109/embc.2013.6610721 · Cited by 31 · see citing works
Inan, O. T., Kovacs, G. T. A., & Giovangrandi, L. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2010.2044185 · Cited by 31 · see citing works
Self-citationCharacterization and analysis of seismocardiogram for estimation of hemodynamic parameters
Tavakolian, K. (2010) · Summit (Simon Fraser University) · Cited by 30 · see citing works
Self-citationMigeotte, P.-F., Ridder, S. D., Tank, J., et al. (2012) · doi:10.1109/embc.2012.6347127 · Cited by 27 · see citing works
Self-citationFranzblau, S. A. (1968) · JAMA · doi:10.1001/jama.1968.03140020085031 · Cited by 26 · see citing works
Studies Utilizing the Portable Electromagnetic Ballistocardiograph
Mandelbaum, H., & MANDELBAUM, R. A. (1953) · Circulation · doi:10.1161/01.cir.7.6.910 · Cited by 26 · see citing works
Comparative analysis of seismocardiogram waves with the ultra-low frequency ballistocardiogram
Ngai, B., Tavakolian, K., Akhbardeh, A., Blaber, A. P., Kamińska, B., & Noordergraaf, A. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5333649 · Cited by 25 · see citing works
Self-citationWilliams, J. (2011) · Elsevier eBooks · doi:10.1016/b978-0-12-385185-7.00032-9 · Cited by 20 · see citing works
Sensor Layer of a Multiparameter Single-Point Integrated System
Chuo, Y., & Kamińska, B. (2009) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2009.2021769 · Cited by 20 · see citing works
Starr, I., & Noordergraaf, A. (1967) · North-Holland eBooks · Cited by 19 · see citing works
Autonomic function testing aboard the ISS using “PNEUMOCARD”
Baevsky, R. M., Funtova, I. I., Diedrich, A., et al. (2009) · Acta Astronautica · doi:10.1016/j.actaastro.2009.03.029 · Cited by 19 · see citing works
Self-citationTanida, K., Shibata, M., & Heitkemper, M. M. (2012) · Biological Research For Nursing · doi:10.1177/1099800412440498 · Cited by 17 · see citing works
On time interval measurements using BCG
Casanella, R., Gomez-Clapers, J., & Pallas-Areny, R. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6347124 · Cited by 16 · see citing works
Self-citationUnsupervised Heart Rate Variability Estimation from Ballistocardiograms
Brüser, C., Winter, S., & Leonhardt, S. K. (2012) · RWTH Publications (RWTH Aachen) · Cited by 16 · see citing works
Estimation of central aortic forces in the ballistocardiogram under rest and exercise conditions
Wiard, R. M., Kim, H. J., Figueroa, C. A., et al. (2009) · doi:10.1109/iembs.2009.5333577 · Cited by 14 · see citing works
The Tei index and exercise capacity.
Libonati, J. R., Ciccolo, J. T., & GLASSBERG, H. L. (2001) · PubMed · Cited by 13 · see citing works
A Portable Measurement System for the Superior-Inferior Axis of the Seismocardiogram
Paukkunen, M., & Linnavuo, M. (2013) · Journal of Bioengineering and Biomedical Sciences · doi:10.4172/2155-9538.1000123 · Cited by 13 · see citing works
Detection of the seismocardiogram W complex based on multiscale edges
Xu, W., Sandham, W. A., Fisher, A. C., et al. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.652691 · Cited by 13 · see citing works
Starr, I., & Ogawa, S. (1961) · The American Journal of the Medical Sciences · doi:10.1097/00000441-196110000-00001 · Cited by 12 · see citing works
Cardiovascular changes in parabolic flights assessed by ballistocardiography
Delière, Q., Migeotte, P., Neyt, X., et al. (2013) · doi:10.1109/embc.2013.6610372 · Cited by 12 · see citing works
Self-citationŠprager, S., & Zazula, D. (2013) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2282403 · Cited by 12 · see citing works
BIOTELEMETRY OF THE TRIAXIAL BALLISTOCARDIOGRAM AND ELECTROCARDIOGRAM IN A WEIGHTLESS ENVIRONMENT
Beischer, D. E., & Hixson, W. C. (1964) · NASA Technical Reports Server (NASA) · Cited by 11 · see citing works
Uncertainty factors in time-interval measurements in ballistocardiography
Gomez-Clapers, J., Serra-Rocamora, A., Casanella, R., et al. (2013) · Instrumentation viewpoint · doi:10.5821/iwp.2013.14.15476 · Cited by 11 · see citing works
Self-citation3D-ballistocardiography in microgravity: Comparison with ground based recordings
Migeotte, P., Delière, Q., Tank, J., et al. (2013) · doi:10.1109/embc.2013.6611172 · Cited by 11 · see citing works
Self-citationA low-noise ac-bridge amplifier for ballistocardiogram measurement on an electronic weighing scale
Inan, O. T., & Kovacs, G. T. A. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/7/n02 · Cited by 10 · see citing works
Self-citationPatch type integrated sensor system for measuring electrical and mechanical cardiac activities
Hyun, E., Noh, S., Yoon, C., et al. (2014) · IEEE Sensors Applications Symposium (SAS) · doi:10.1109/sas.2014.6798924 · Cited by 10 · see citing works
Hernandez, A. I., Ziglio, F., Amblard, A., et al. (2013) · doi:10.1109/embc.2013.6611169 · Cited by 9 · see citing works
Preliminary results from standing ballistocardiography measurements in microgravity
Wiard, R. M., Inan, O. T., Giovangrandi, L., et al. (2013) · doi:10.1109/embc.2013.6611241 · Cited by 8 · see citing works
Self-citationLuchitskaya, E. S., Delière, Q., Diedrich, A., et al. (2013) · doi:10.1109/embc.2013.6611242 · Cited by 6 · see citing works
Comparison of pulse rate variability indices estimated from pressure signal and photoplethysmogram
Zhu, X., Chen, W., Kitamura, K., et al. (2012) · doi:10.1109/bhi.2012.6211725 · Cited by 4 · see citing works
Devices in heart failure: potential methods for device-based monitoring of congestive heart failure.
Munir, S. M., Bogaev, R. C., Sobash, E., Shankar, K. J., Gondi, S., Stupin, I. V., Robertson, J., Brewer, M. A., Casscells, S. W., Delgado, R. M., et al. (2008) · PubMed Central · Cited by 2 · see citing works
Physiological insights from gravity-free ballistocardiography
Prisk, G. K., & Migeotte, P. (2013) · doi:10.1109/embc.2013.6611239 · Cited by 2 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Lab-on-Skin: A Review of Flexible and Stretchable Electronics for Wearable Health Monitoring
Liu, Y., Pharr, M., & Salvatore, G. A. (2017) · ACS Nano · doi:10.1021/acsnano.7b04898 · Cited by 1,725 · see citing works
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · see citing works
Liu, Y., Norton, J. J. S., Qazi, R., Zou, Z., Ammann, K. R., Liu, H., Yan, L., Tran, P. L., Jang, K.-I., Lee, J. W., et al. (2016) · Science Advances · doi:10.1126/sciadv.1601185 · Cited by 397 · see citing works
Wu, H., Yang, G., Zhu, K., Liu, S., Guo, W., Jiang, Z., & Li, Z. (2021) · Advanced Science · doi:10.1002/advs.202001938 · Cited by 395 · see citing works
Ha, T., Tran, J., Liu, S., Jang, H., Jeong, H., Mitbander, R., Huh, H., Qiu, Y., Duong, J., Wang, R. L., et al. (2019) · Advanced Science · doi:10.1002/advs.201900290 · Cited by 220 · see citing works
Inan, O. T., Baran Pouyan, M., Javaid, A. Q., Dowling, S., Etemadi, M., Dorier, A., Heller, J. A., Bicen, A. O., Roy, S., De Marco, T., et al. (2018) · Circulation: Heart Failure · doi:10.1161/circheartfailure.117.004313 · Cited by 207 · see citing works
Self-citationCuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180 · see citing works
In-Ear Audio Wearable: Measurement of Heart and Breathing Rates for Health and Safety Monitoring
Martin, A., & Voix, J. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2720463 · Cited by 112 · see citing works
Definition of Fiducial Points in the Normal Seismocardiogram
Sørensen, K., Schmidt, S. E., Jensen, A. S., Søgaard, P., & Struijk, J. J. (2018) · Scientific Reports · doi:10.1038/s41598-018-33675-6 · Cited by 111 · see citing works
Etemadi, M., Inan, O. T., Heller, J. A., Hersek, S., Klein, L., & Roy, S. (2016) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2015.2405480 · Cited by 95 · see citing works
“S EISMOCARDIOGRAPHY is a non-invasive technique based on measuring the micro-vibrations of the chest wall for assessing left ventricular function [1]–[4], systolic and diastolic time intervals [5]–[7], and cardio-respiratory interactions [8].”
Self-citationThe role of wearables in heart failure
Singhal, A., & Cowie, M. R. (2020) · Current Heart Failure Reports · doi:10.1007/s11897-020-00467-x · Cited by 85 · see citing works
A Cardiac Early Warning System with Multi Channel SCG and ECG Monitoring for Mobile Health
Sahoo, P., Thakkar, H., & Lee, M.-Y. (2017) · Sensors · doi:10.3390/s17040711 · Cited by 81 · see citing works
Lin, W.-Y., Tsai, T.-H., Lee, M.-Y., Chou, W.-C., Chang, P.-C., Chou, C.-C., Wen, M.-S., Ho, M.-Y., Lee, W.-C., Hsieh, M.-J., et al. (2018) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2620496 · Cited by 62 · see citing works
Non-invasive human vital signs monitoring based on twin-core optical fiber sensors
Tan, F., Chen, S., Lyu, W., Liu, Z., Yu, C., Lu, C., & Tam, H.-Y. (2019) · Biomedical Optics Express · doi:10.1364/boe.10.005940 · Cited by 62 · see citing works
Jeong, H., Kwak, S. S., Sohn, S., Lee, J. Y., Lee, Y. J., O’Brien, M. K., Park, Y., Avila, R., Kim, J.-T., Yoo, J.-Y., et al. (2021) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2104925118 · Cited by 59 · see citing works
A Unified Framework for Quality Indexing and Classification of Seismocardiogram Signals
Zia, J., Kimball, J., Hersek, S., Shandhi, M. M. H., Semiz, B., & Inan, O. T. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2931348 · Cited by 48 · see citing works
“Predictive performance was analyzed as a function of the number of template sets (n = [1, 9]) and number of signal segments per prediction (m = [1, 9]).”
BackgroundMethodInfluentialSelf-citationA Ballistographic Approach for Continuous and Non-Obtrusive Monitoring of Movement in Neonates
Joshi, R., Bierling, B. L., Long, X., Weijers, J., Feijs, L., Van Pul, C., & Andriessen, P. (2018) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2018.2875703 · Cited by 43 · see citing works
Bhattacharya, S., Nikbakht, M., Alden, A., Tan, P., Wang, J., Alhalimi, T. A., Kim, S., Wang, P., Tanaka, H., Tandon, A., et al. (2023) · Advanced Electronic Materials · doi:10.1002/aelm.202201284 · Cited by 42 · see citing works
Self-citationNon-Invasive Wearable Patch Utilizing Seismocardiography for Peri-Operative Use in Surgical Patients
Semiz, B., Carek, A. M., Johnson, J. C., Ahmad, S., Heller, J. A., Vicente, F. G., Caron, S., Hogue, C. W., Etemadi, M., & Inan, O. T. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.3032938 · Cited by 35 · see citing works
Self-citationMarvasti, T. B., Gao, Y., Murray, K. R., Hershman, S., McIntosh, C., & Moayedi, Y. (2024) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2024.07.009 · Cited by 35 · see citing works
Detecting Aortic Valve-Induced Abnormal Flow with Seismocardiography and Cardiac MRI
Johnson, E. M. I., Heller, J. A., Garcia Vicente, F., Sarnari, R., Gordon, D., McCarthy, P. M., Barker, A. J., Etemadi, M., & Markl, M. (2020) · Annals of Biomedical Engineering · doi:10.1007/s10439-020-02491-3 · Cited by 14 · see citing works
Self-citationA Synthetic Seismocardiogram and Electrocardiogram Generator Phantom
Nikbakht, M., Lin, D. J., Gazi, A. H., & Inan, O. T. (2022) · 2022 IEEE Sensors · doi:10.1109/sensors52175.2022.9967101 · Cited by 9 · see citing works
Self-citation
- Authors
- Inan, O. T., Migeotte, P.-F., Park, K.-S., Etemadi, M., Tavakolian, K., Casanella, R., Zanetti, J., Tank, J., Funtova, I., Prisk, G. K., & Di Rienzo, M.
- Affiliations
- Georgia Institute of Technology (US)
- University of California, San Francisco (US)
- University of North Dakota (US)
- Universitat Politècnica de Catalunya (ES)
- Medizinische Hochschule Hannover (DE)
- Russian Academy of Sciences (RU)
- University of California San Diego (US)
- ORCID
- Venue
- IEEE Journal of Biomedical and Health Informatics
- Published
- 2015-07
- Volume
- 19
- Issue
- 4
- Pages
- 1414-1427
- ISSN
- 2168-2194 (ISSN-L)
- 2168-2208
- OpenAlex
- W2148325642
- Cited by
- 775
- References
- 108
- Citations by year
- 2026: 60 · 2025: 71 · 2024: 70 · 2023: 68 · 2022: 87 · 2021: 68 · 2020: 89 · 2019: 74 · 2018: 70 · 2017: 56 · 2016: 45 · 2015: 17
- FWCI
- 3.38
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Locations
- IEEE Journal of Biomedical and Health Informatics (publishedVersion)
- PubMed (publishedVersion)
- Dépôt institutionnel de l'Université libre de Bruxelles (Université Libre de Bruxelles) (submittedVersion)
- LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) (submittedVersion, cc-by-nc-nd)
- RECERCAT (Consorci de Serveis Universitaris de Catalunya) (submittedVersion, cc-by-nc-nd)
- UPCommons institutional repository (Universitat Politècnica de Catalunya) (submittedVersion, cc-by-nc-nd)
- Funders
- Belgian Federal Science Policy Office
- European Space Agency PRODEX (ILSRA-2009–0729)
- National Space Biomedical Research Institute (NASA NCC9–58)
- German Space Agency (DLR) (50WB1117)
- Italian Space Agency (ASI 2013–061-I.0, ASI 2013–079-R.0)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Phonocardiography and Auscultation Techniques
- Keywords
- ballistocardiography, seismocardiography, seismocardiogram, signal processing algorithms, noise reduction, feature extraction, signal modeling, microgravity
- MeSH terms
- Ballistocardiography, Humans, Monitoring, Physiologic, Signal Processing, Computer-Assisted
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review