Flexible Carbon Nanotube-Based Polymer Electrode for Long-Term Electrocardiographic Recording
Open accessAbstract
The long-term monitoring of electrocardiogram (ECG) is critical for the accurate diagnosis and tracking of cardiovascular diseases (CVDs). However, the commercial Ag/AgCl electrode is not suitable for long-term monitoring due to skin allergies and signal degradation, caused by the conductive gel drying over time. In this paper, a flexible gel-free electrode, composed of a multi-wall carbon nanotube (MWCNT) and polydimethylsiloxane (PDMS), is proposed for long-term wearable ECG monitoring. To achieve uniform dispersion of MWCNTs in viscous PDMS, we developed a novel parallel solvent-assisted ultrasonic dispersion method, wherein the organic solvent n–Hexane served as a dispersion to avoid MWCNT aggregates. The properties of the MWCNT/PDMS electrode were assessed through structural characterization, contact impedance tests, ECG measurements, and biocompatibility tests. When the MWCNT weight fraction reached 5.5 wt%, the skin-electrode contact impedance of the MWCNT/PDMS electrode was lower than that of the Ag/AgCl electrode below 100 Hz. In daily ECG monitoring, the MWCNT/PDMS electrodes showed superior performance against motion artifact compared to the Ag/AgCl electrode. After seven days of wearing the MWCNT/PDMS electrode, ECG signals did not degrade and no side effects, such as skin redness and swelling, were observed. Thus, this electrode could enable long-term ECG monitoring in wearable healthcare systems.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Previous work has shown the disadvantage of using gel electrodes for long durations.
“Previous work has already shown the disadvantage of using gel electrodes for long durations. [35,36]”
From A Chest‐Conformable, Wireless Electro‐Mechanical E‐Tattoo for Measuring Multiple Cardiac Time Intervals (2023) · cites this paper as [35]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A wearable and highly sensitive pressure sensor with ultrathin gold nanowires
Gong, S., Schwalb, W., Wang, Y., Chen, Y., Tang, Y., Si, J., Shirinzadeh, B., & Cheng, W. (2014) · Nature Communications · doi:10.1038/ncomms4132 · Cited by 2,064 · see citing works
CNT/PDMS Composite Flexible Dry Electrodesfor Long-Term ECG Monitoring
Jung, H.-C., Moon, J.-H., Baek, D.-H., Lee, J.-H., Choi, Y.-Y., Hong, J.-S., & Lee, S.-H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2190288 · Cited by 434 · see citing works
A 3D printed dry electrode for ECG/EEG recording
Salvo, P., Raedt, R., Carrette, E., Schaubroeck, D., Vanfleteren, J., & Cardon, L. (2011) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2011.12.017 · Cited by 265 · see citing works
Micromachined electrodes for biopotential measurements
Griss, P., Enoksson, P., Tolvanen-Laakso, H., Meriläinen, P., Ollmar, S., & Stemme, G. (2001) · Journal of Microelectromechanical Systems · doi:10.1109/84.911086 · Cited by 240 · see citing works
Wearable silver nanowire dry electrodes for electrophysiological sensing
Myers, A. C., Huang, H., & Zhu, Y. (2015) · RSC Advances · doi:10.1039/c4ra15101a · Cited by 231 · see citing works
A Wearable Hydration Sensor with Conformal Nanowire Electrodes
Yao, S., Myers, A., Malhotra, A., Lin, F., Bozkurt, A., Muth, J. F., & Zhu, Y. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201601159 · Cited by 213 · see citing works
Lee, S. M., Byeon, H. J., Lee, J. H., Baek, D. H., Lee, K. H., Hong, J. S., & Lee, S.-H. (2014) · Scientific Reports · doi:10.1038/srep06074 · Cited by 181 · see citing works
Capacitive Biopotential Measurement for Electrophysiological Signal Acquisition: A Review
Sun, Y., & Yu, X. B. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2519392 · Cited by 174 · see citing works
Flexible dry surface-electrodes for ECG long-term monitoring
Hoffmann, K., & Ruff, R. (2007) · doi:10.1109/iembs.2007.4353650 · Cited by 129 · see citing works
Non-contact Low Power EEG/ECG Electrode for High Density Wearable Biopotential Sensor Networks
Yu, M., Deiss, S. R., & Cauwenberghs, G. (2009) · International Workshop on Wearable and Implantable Body Sensor Networks (BSN) · doi:10.1109/bsn.2009.52 · Cited by 128 · see citing works
Novel Electrodes for Underwater ECG Monitoring
Reyes, B. A., Posada-Quintero, H. F., Bales, J. R., Clement, A. L., Pins, G. D., Swiston, A. J., Riistama, J. M., Florian, J. P., Shykoff, B. E., Qin, M. K., et al. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2309293 · Cited by 114 · see citing works
CNT/PDMS composite membranes for H2 and CH4 gas separation
Nour, M., Berean, K. J., Balendhran, S., et al. (2013) · International Journal of Hydrogen Energy · doi:10.1016/j.ijhydene.2013.05.162 · Cited by 113 · see citing works
A microfabricated electrode with hollow microneedles for ECG measurement
Yu, L., Tay, F. E. H., Guo, D., Xu, L., & Yap, K. L. (2009) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2009.01.020 · Cited by 113 · see citing works
Wireless and Non-contact ECG Measurement System – the “Aachen SmartChair”
Aleksandrowicz, A., & Leonhardt, S. K. (2007) · Acta Polytechnica · doi:10.14311/974 · Cited by 105 · see citing works
Microneedle-based electrodes with integrated through-silicon via for biopotential recording
O’Mahony, C., Pini, F., Blake, A., et al. (2012) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2012.04.037 · Cited by 95 · see citing works
Forvi, E., Bedoni, M., Carabalona, R., Soncini, M., Mazzoleni, P., Rizzo, F., O’Mahony, C., Morasso, C., Cassarà, D. G., & Gramatica, F. (2012) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2012.04.019 · Cited by 91 · see citing works
Chi, Y. M., & Cauwenberghs, G. (2009) · doi:10.1109/iembs.2009.5333527 · Cited by 89 · see citing works
Peng, H., Liu, J., Tian, H., Xu, B. B., Dong, Y.-Z., Yang, B. T., Chen, X., & Yang, C. (2015) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2015.09.024 · Cited by 78 · see citing works
Self-citationDynamic Behavior of Nanocomposites Reinforced with Multi-Walled Carbon Nanotubes (MWCNTs)
Her, S.-C., & Lai, C.-Y. (2013) · Materials · doi:10.3390/ma6062274 · Cited by 72 · see citing works
Developing Barbed Microtip-Based Electrode Arrays for Biopotential Measurement
Hsu, L.-S., Tung, S.-W., Kuo, C.-H., et al. (2014) · Sensors · doi:10.3390/s140712370 · Cited by 67 · see citing works
Flexible PDMS -based dry electrodes for electro-optic acquisition of ECG signals in wearable devices
Fernandes, M. S., Lee, K. S., Ram, R. J., Correia, J. H., & Mendes, P. M. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5627799 · Cited by 44 · see citing works
New insight on the interfacial interaction between multiwalled carbon nanotubes and elastomers
Ning, N., Cheng, D., Yang, J., et al. (2017) · Composites Science and Technology · doi:10.1016/j.compscitech.2017.02.012 · Cited by 42 · see citing works
Sankar, N., REDDY, M. N., & Prasad, R. K. (2016) · Bulletin of Materials Science · doi:10.1007/s12034-015-1117-3 · Cited by 41 · see citing works
Liu, B., Luo, Z., Zhang, W., Tu, Q., & Jin, X. (2016) · Journal of Biomaterials Science Polymer Edition · doi:10.1080/09205063.2016.1239951 · Cited by 40 · see citing works
A flexible dry electrode based on APTES-anchored PDMS substrate for portable ECG acquisition system
Meng, Y., Li, Z., & Chen, J. (2015) · Microsystem Technologies · doi:10.1007/s00542-015-2490-y · Cited by 37 · see citing works
Yan, J., Kim, B., & Jeong, Y. G. (2015) · Journal of Materials Science · doi:10.1007/s10853-015-9110-1 · Cited by 34 · see citing works
A flexible dry micro-dome electrode for ECG monitoring
Meng, Y., Li, Z. B., Chen, X., & Chen, J. P. (2014) · Microsystem Technologies · doi:10.1007/s00542-014-2235-3 · Cited by 33 · see citing works
Non-contact, non-obtrusive electrocardiography in clinical environements
Chamadiya, B., Mankodiya, K., Wagner, M., et al. (2011) · doi:10.4108/icst.pervasivehealth.2011.246051 · Cited by 7 · see citing works
Study of nano mechanical properties polydimethylsiloxane (PDMS)/MWCNT composites
Murudkar, V. V., Gaonkar, A. A., Deshpande, V. D., et al. (2018) · AIP conference proceedings · doi:10.1063/1.5032902 · Cited by 5 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Bhattacharya, S., Nikbakht, M., Alden, A., Tan, P., Wang, J., Alhalimi, T. A., Kim, S., Wang, P., Tanaka, H., Tandon, A., et al. (2023) · Advanced Electronic Materials · doi:10.1002/aelm.202201284 · Cited by 42 · see citing works
- Authors
- Chi, M., Zhao, J., Dong, Y., & Wang, X.
- Affiliations
- University Town of Shenzhen (CN)
- Corresponding author
- Dong, Ying
- Venue
- Materials
- Publisher
- MDPI AG
- Published
- 2019-03-23
- Volume
- 12
- Issue
- 6
- Pages
- 971
- ISSN
- 1996-1944 (ISSN-L)
- PMC
- PMC6470748
- OpenAlex
- W2923735548
- Cited by
- 68
- References
- 29
- Citations by year
- 2026: 6 · 2025: 11 · 2024: 9 · 2023: 11 · 2022: 9 · 2021: 10 · 2020: 8 · 2019: 4
- FWCI
- 2.73
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Materials (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- Shenzhen Fundamental Research Funds ((JCYJ20150831192244849)
- Shenzhen Technology Research Funds (JSGG20170414094227487)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- multi-walled carbon nanotubes, polydimethylsiloxane, flexible electrodes, ECG monitoring, ultrasonic dispersion, hexane solvent, motion artifacts, biocompatibility, cardiovascular disease monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review