Towards long-term stable polyimide-based flexible electrical insulation for chronically implanted neural electrodes
Open accessAbstract
For chronic applications of flexible neural implants, e.g., intracortical probes, the flexible substrate material has to encapsulate the electrical conductors with a long-term stability against the saline environment of the neural tissue. The biocompatible polymer polyimide is often used for this purpose. Due to its chemical inertness, the adhesion between two polyimide layers is, however, a challenge, which can lead to delamination and, finally, to short circuits. The state-of-the-art method to improve the adhesion strength is activating the polyimide surface using oxygen reactive ion etching (O2 RIE). However, the influence of the process variations (etching time, bias power) on the long-term stability is still unclear. Therefore, we establish a test method, where the aging of a gold interdigital structure embedded in two polyimide layers and immersed in saline solution is accelerated using an elevated temperature, mechanical stress and an electrical field. A continuous measurement of a leakage current is used to define the failure state. The results show that the variation of the O2 RIE plasma process has a significant effect on the long-term stability of the test samples. Comparing the two different plasma treatments 0.5 min at 25 W and 1 min at 50 W, the long-term stability could be increased from 20.9 ± 19.1 days to 44.9 ± 18.9 days. This corresponds to more than a doubled lifetime. An ideal solution for the delamination problem is still not available; however, the study shows that the fine-tuning of the fabrication processes can improve the long-term stability of chronically implanted neural electrodes.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Leaked substances can disrupt the ionic balance of sodium, potassium, and calcium, leading to excitotoxicity.
“These leaked substances can also disrupt the ionic balance of essential elements like sodium, potassium, and calcium, leading to excitotoxicity—a pathological process causing neuron damage due to excessive neurotransmitter stimulation.”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Schander et al., 2021
The Introduction (page unavailable) says that leachables from inadequately processed SU-8 can cause cytotoxicity. It does not say that leaked substances disrupt sodium, potassium, or calcium balance, or that they cause excitotoxicity; the Methods’ discussion of electrolyte concerns an insulation test, not that mechanism. (LLM, conf=0.78)
Experimental — accuracy not yet measuredNeural tissue might undergo fibrosis in response to chemical exposure.
“Lastly, in response to chemical exposure, neural tissue might undergo fibrosis, a defensive yet functionally compromising mechanism (Grisold and Carozzi, 2021; Schander et al., 2021; Yin et al., 2021; World Health Organization: WHO, 2023).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Schander et al., 2021
The Introduction discusses foreign-body responses to neural implants and possible toxic reactions from leachable materials, but it does not say that chemical exposure causes fibrosis in neural tissue. A search of our full copy for fibrosis and chemical exposure found no passage addressing that specific response. (LLM, conf=0.74)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Novel electrode technologies for neural recordings
Hong, G., & Lieber, C. M. (2019) · Nature Reviews Neuroscience · doi:10.1038/s41583-019-0140-6 · Cited by 695 · see citing works
Hassler, C., Boretius, T., & Stieglitz, T. (2011) · Journal of Polymer Science Part B: Polymer Physics · doi:10.1002/polb.22169 · Cited by 501 · see citing works
Accelerated aging for testing polymeric biomaterials and medical devices
Hukins, D. W. L., Mahomed, A., & Kukureka, S. N. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2008.06.001 · Cited by 251 · see citing works
In vitro evaluation of the long-term stability of polyimide as a material for neural implants
Rubehn, B., & Stieglitz, T. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.01.053 · Cited by 242 · see citing works
Takmakov, P. A., Ruda, K., Phillips, K. S., Isayeva, I. S., Krauthamer, V., & Welle, C. G. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/2/026003 · Cited by 179 · see citing works
Lee, J. H., Kim, H., Kim, J. H., & Lee, S. (2016) · Lab on a Chip · doi:10.1039/c5lc00842e · Cited by 124 · see citing works
Conformable polyimide-based μECoGs: Bringing the electrodes closer to the signal source
Vomero, M., Porto Cruz, M. F., Zucchini, E., Ciarpella, F., Delfino, E., Carli, S., Boehler, C., Asplund, M., Ricci, D., Fadiga, L., et al. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.120178 · Cited by 95 · see citing works
The Anodic Corrosion of Gold in Concentrated Chloride Solutions
Frankenthal, R. P., & Siconolfi, D. J. (1982) · Journal of The Electrochemical Society · doi:10.1149/1.2124085 · Cited by 84 · see citing works
Physical characteristics of polyimide films for flexible sensors
Chang, W., Fang, T., & Lin, Y.-C. (2008) · Applied Physics A · doi:10.1007/s00339-008-4623-y · Cited by 76 · see citing works
Smart Polymers for Neural Interfaces
Ware, T. H., Simon, D., Rennaker, R. L., & Voit, W. (2013) · Journal of macromolecular science. Part C, Reviews in macromolecular chemistry and physics/Journal of macromolecular science. Reviews in macromolecular chemistry and physics · doi:10.1080/15583724.2012.751924 · Cited by 74 · see citing works
Insulation lifetime improvement of polyimide thin film neural implants
Ceyssens, F., & Puers, R. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/5/054001 · Cited by 44 · see citing works
An Assessment of Surface Treatments for Adhesion of Polyimide Thin Films
Cen‐Puc, M., Schander, A., Gleason, M. G. V., & Lang, W. (2021) · Polymers · doi:10.3390/polym13121955 · Cited by 43 · see citing works
Self-citationSchander, A., Stemmann, H., Tolstosheeva, E., Roese, R., Biefeld, V., Kempen, L., Kreiter, A. K., & Lang, W. (2016) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2016.05.034 · Cited by 39 · see citing works
Self-citationSchander, A., Strokov, S., Stemmann, H., Tebmann, T., Kreiter, A. K., & Lang, W. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2880833 · Cited by 32 · see citing works
Self-citationSchander, A., Stemmann, H., Kreiter, A. K., & Lang, W. (2018) · Micromachines · doi:10.3390/mi9030131 · Cited by 16 · see citing works
Self-citationAccelerated soak performance of BPDA-PPD polyimide for implantable MEAs
Tolstosheeva, E., Biefeld, V., & Lang, W. (2015) · Procedia Engineering · doi:10.1016/j.proeng.2015.08.560 · Cited by 16 · see citing works
Self-citationSchander, A., Stemmann, H., Claussen, K. C., Kreiter, A. K., & Lang, W. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8717089 · Cited by 6 · see citing works
Self-citationDevelopment of a process for the manufacturing of SU-8 100 for the use in cell culture
Blume, G., Mielke, G., Kohnert, J., Pörtner, R., & Trieu, K. H. (2018) · Journal of Bioactive and Compatible Polymers · doi:10.1177/0883911518765216 · Cited by 3 · see citing works
Li, C., Cauwe, M., Mader, L., Schaubroeck, D., & de Beeck, M. O. (2019) · Coatings · doi:10.3390/coatings10010019 · Cited by 2 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wu, Z., He, J., Yang, H., & Yang, S. (2022) · Polymers · doi:10.3390/polym14061269 · Cited by 157 · see citing works
Tang, X., Lu, Y., Li, S., Zhu, M., Wang, Z., Li, Y., Hu, Z., Zheng, P., Wang, Z., & Liu, T. (2024) · Nano-Micro Letters · doi:10.1007/s40820-024-01590-3 · Cited by 103 · see citing works
Applications of flexible polyimide: barrier material, sensor material, and functional material
Lin, J., Su, J., Weng, M., Xu, W., Huang, J., Fan, T., Liu, Y., & Min, Y. (2022) · Soft Science · doi:10.20517/ss.2022.24 · Cited by 45 · see citing works
Flexible high-density microelectrode arrays for closed-loop brain–machine interfaces: a review
Liu, X., Gong, Y., Jiang, Z., Stevens, T., & Li, W. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2024.1348434 · Cited by 42 · see citing works
Wang, X., Wu, S. P., Yang, H., Bao, Y., Li, Z., Gan, C., Deng, Y., Cao, J., Li, X., Wang, Y., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-53720-5 · Cited by 28 · see citing works
The Trade‐Off between Transconductance and Speed for Vertical Organic Electrochemical Transistors
Skowrons, M., Schander, A., Negron, A. G. P., & Lüssem, B. (2024) · Advanced Electronic Materials · doi:10.1002/aelm.202300673 · Cited by 27 · see citing works
Self-citationComparative analysis of energy transfer mechanisms for neural implants
Miziev, S., Pawlak, W. A., & Howard, N. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2023.1320441 · Cited by 21 · see citing works
Lifetime engineering of bioelectronic implants with mechanically reliable thin film encapsulations
Niemiec, M., & Kim, K. (2023) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/ad0b19 · Cited by 21 · see citing works
Tintelott, M., Schander, A., & Lang, W. (2022) · Polymers · doi:10.3390/polym14183702 · Cited by 20 · see citing works
Self-citationKullmann, F. A., Kridner, D., Mertens, S., Christianson, M., Rosa, D., & Diaz-Botia, C. A. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.876877 · Cited by 18 · see citing works
Li, W., Cao, Y., & Sepúlveda, N. (2022) · Micromachines · doi:10.3390/mi14010099 · Cited by 14 · see citing works
Ji, B., Wang, T., Li, M., Shi, L., You, X., Sun, F., & Luan, H. (2022) · Micromachines · doi:10.3390/mi13111853 · Cited by 11 · see citing works
Thermally stable benzimidazole based co-polyimides and derived LIG for flexible supercapacitors
Iqbal, A., Amna, B., Islam, I. U., Zhang, Y., Siddiqi, H. M., Zai, J., & Qian, X. (2025) · Polymer · doi:10.1016/j.polymer.2025.128016 · Cited by 8 · see citing works
Fabrication and validation of flexible neural electrodes based on polyimide tape and gold sheet
Jeong, H., Lee, T. K., Kim, J., Jeong, H. S., Jun, S. B., & Seo, J.-M. (2024) · Biomedical Engineering Letters · doi:10.1007/s13534-023-00345-0 · Cited by 7 · see citing works
Emerging trends in the development of flexible optrode arrays for electrophysiology
Almasri, R. M., Ladouceur, F. J., Mawad, D., Esrafilzadeh, D., Firth, J., Lehmann, T., Poole-Warren, L. A., Lovell, N. H., & Abed, A. A. (2023) · APL Bioengineering · doi:10.1063/5.0153753 · Cited by 7 · see citing works
Multi-Channel Microscale Nerve Cuffs for Spatially Selective Neuromodulation
Riley, M., Tala, F., Johnson, K. J., & Johnson, B. C. (2024) · Micromachines · doi:10.3390/mi15081036 · Cited by 5 · see citing works
Failure Reason of PI Test Samples of Neural Implants
Guljakow, J., & Lang, W. (2023) · Sensors · doi:10.3390/s23031340 · Cited by 3 · see citing works
Self-citationHigh-Porosity Sieve-Type Neural Electrodes for Motor Function Recovery and Nerve Signal Acquisition
Choi, W., Park, H., Oh, S., Seok, S., Yoon, D. S., & Kim, J. (2024) · Micromachines · doi:10.3390/mi15070862 · Cited by 2 · see citing works
Wang, H., Yang, G., Zhang, J., Li, S., & Bilal, M. (2022) · Journal of nanostructure in chemistry · doi:10.1007/s40097-022-00489-y · Cited by 2 · see citing works
Zhou, C., Xu, Y., Wang, M., Jiang, H., Wang, W., Zhou, X., Zhao, X., Wang, P., Wang, X., Yang, W., et al. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.180921 · Cited by 1 · see citing works
Innovative ESCS Bridging System: Enhancing Motor Task‐Specific Activation in Spinal Cord Injury
Guo, X., Li, Z., Wang, P., Bai, Y., Ni, L., & Xu, W. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202501055
Advances in Thin-Film Graphene Neurotechnology for Chronic Nerve Stimulation and RecordingPreprint
Ria, N., Rodríguez‐Meana, B., Masvidal‐Codina, E., Andreu, D., Crugeiras, J., William, L., Graf, A., Illa, X., Katirtsidis, G. A., Galceran, A. D., et al. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.01.23.701276
Long-term moisture barrier performance of liquid crystal polymer for implantable medical electronicsPreprint
Thielen, B., Pulicken, C., Aklivanh, E., Sabes, P. N., & Cvitkovic, M. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.02.24.707821
Regenerative Peripheral Nerve Interface and the Future of Intuitive Control
Wang, M. J., Kung, T. A., Cederna, P. S., & Kemp, S. W. P. (2026) · Hand Clinics · doi:10.1016/j.hcl.2026.03.009
Wang, Y., Han, Z., Xu, Y., Wang, X., Yu, J., & Huang, P. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.180690
Role of BSA adsorption in the evolution of polyimide coating performance
Guan, Y., Qi, S., Meng, M., Tang, A., & Li, Y. (2026) · Corrosion Engineering Science and Technology The International Journal of Corrosion Processes and Corrosion Control · doi:10.1177/1478422x261477790
Hybrid Flexible–Soft Neural Interface as a Practical Solution for Long‐Term Implants
Ji, B., Zhang, Z., You, X., Wang, X., Huang, Y., Cao, J., Bai, R., Chen, J., Zhao, H., Chang, H., et al. (2026) · Rare Metals · doi:10.1002/rar2.70619
Zheng, K., Yu, W., Gu, Y., He, Z., Wang, X., & Li, Y. (2026) · ACS Materials Letters · doi:10.1021/acsmaterialslett.6c00685
- Authors
- Schander, A., Gancz, J., Tintelott, M., & Lang, W.
- Affiliations
- University of Bremen (DE)
- Corresponding author
- Schander, Andreas
- Venue
- Micromachines
- Publisher
- MDPI AG
- Published
- 2021-10-20
- Volume
- 12
- Issue
- 11
- Pages
- 1279
- ISSN
- 2072-666X (ISSN-L)
- PMC
- PMC8619170
- OpenAlex
- W3210164395
- Cited by
- 28
- References
- 19
- Citations by year
- 2026: 8 · 2025: 2 · 2024: 8 · 2023: 3 · 2022: 7
- FWCI
- 2.07
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Micromachines (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- RWTH Publications (RWTH Aachen) (submittedVersion, other-oa)
- Funders
- Deutsche Forschungsgemeinschaft (LA 1471/11-2)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
- Physical Sciences › Materials Science › Polymers and Plastics › Synthesis and properties of polymers
- Keywords
- polyimide, electrical insulation, delamination, interlayer bonding, oxygen reactive ion etching, plasma surface treatment, leakage current, accelerated aging test, implant survival
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review