Application of Polymethylpentene, an Oxygen Permeable Thermoplastic, for Long-Term on-a-Chip Cell Culture and Organ-on-a-Chip Devices
Open accessThe DOI matches its Crossref record; 34 of the 34 references it cites were found in a registry or an index.
Abstract
The applicability of a gas-permeable, thermoplastic material polymethylpentene (PMP) was investigated, experimentally and analytically, for organ-on-a-chip (OoC) and long-term on-a-chip cell cultivation applications. Using a sealed culture chamber device fitted with oxygen sensors, we tested and compared PMP to commonly used glass and polydimethylsiloxane (PDMS). We show that PMP and PDMS have comparable performance for oxygen supply during 4 days culture of epithelial (A549) cells with oxygen concentration stabilizing at 16%, compared with glass control where it decreases to 3%. For the first time, transmission light images of cells growing on PMP were obtained, demonstrating that the optical properties of PMP are suitable for non-fluorescent, live cell imaging. Following the combined transmission light imaging and calcein-AM staining, cell adherence, proliferation, morphology, and viability of A549 cells were shown to be similar on PMP and glass coated with poly-L-lysine. In contrast to PDMS, we demonstrate that a film of PMP as thin as 0.125 mm is compatible with high-resolution confocal microscopy due to its excellent optical properties and mechanical stiffness. PMP was also found to be fully compatible with device sterilization, cell fixation, cell permeabilization and fluorescent staining. We envision this material to extend the range of possible microfluidic applications beyond the current state-of-the-art, due to its beneficial physical properties and suitability for prototyping by different methods. The integrated device and measurement methodology demonstrated in this work are transferrable to other cell-based studies and life-sciences applications.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures







Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“When materials other than PDMS are used, factors like oxygen permeability and cell adhesion properties may also impact cell behaviour (Sønstevold et al., 2023).”
From Considering future qualification for regulatory science in the early development of microphysiological systems: a case study of microthrombosis in a Vessel-on-Chip (2024) · cites this paper as Sønstevold et al., 2023
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A guide to the organ-on-a-chip
Leung, C. M., de Haan, P., Ronaldson-Bouchard, K., Kim, G.-A., Ko, J., Rho, H. S., Chen, Z., Habibovic, P., Jeon, N. L., Takayama, S., et al. (2022) · Nature Reviews Methods Primers · doi:10.1038/s43586-022-00118-6 · Cited by 1,116 · cited 6×
“For OoC and cell culture devices fabricated in gas-impermeable materials such as glass and common thermoplastics such as PMMA, PC, COC and PS [2], perfusion of the culture medium is the only source of supply of both nutrients and gas, as well as the removal of waste and secreted soluble factors [2,10].”
BackgroundInfluentialEngineers are from PDMS-land, Biologists are from Polystyrenia
Berthier, E., Young, E. W. K., & Beebe, D. (2012) · Lab on a Chip · doi:10.1039/c2lc20982a · Cited by 799 · cited 8×
“PDMS has some favourable properties as a research tool as it allows for rapid and cheap prototyping, facile bonding to glass, as well as favourable optical properties and deformability allowing integration of on-chip pumps and valves [14].”
BackgroundMethodInfluentialBiological implications of polydimethylsiloxane-based microfluidic cell culture
Regehr, K. J., Domenech, M., Koepsel, J. T., Carver, K. C., Ellison-Zelski, S. J., Murphy, W. L., Schuler, L. A., Alarid, E. T., & Beebe, D. J. (2009) · Lab on a Chip · doi:10.1039/b903043c · Cited by 719 · cited 2×
“However, it also absorbs small hydrophobic molecules due to its porosity and hydrophobic nature, which affects the concentration of important solutes such as hydrophobic drugs and hormones [15].”
BackgroundPhototoxicity in live fluorescence microscopy, and how to avoid it
Icha, J., Weber, M., Waters, J. C., & Norden, C. (2017) · BioEssays · doi:10.1002/bies.201700003 · Cited by 622 · cited 1×
“As fluorescence staining and imaging may be detrimental to cells over time due to phototoxicity [32] and only make parts of the cells visible, transmission light microscopy is a preferred cell monitoring method.”
MethodThe effect of HIF on metabolism and immunity
Taylor, C. T., & Scholz, C. C. (2022) · Nature Reviews Nephrology · doi:10.1038/s41581-022-00587-8 · Cited by 517 · cited 2×
“Lack of proper oxygen supply can have a major impact on experimental results through the activation of proteins from the hypoxia-inducible factor (HIF) family, which can rewrite cellular transcription programs in response to lowoxygen conditions [11,12].”
BackgroundPlace, T. L., Domann, F. E., & Case, A. J. (2017) · Free Radical Biology and Medicine · doi:10.1016/j.freeradbiomed.2017.10.003 · Cited by 434 · cited 1×
“Gas supply is provided by diffusion from the surrounding air through the cell culture medium and is thereby continuous [9].”
BackgroundCommercialization of microfluidic devices
Volpatti, L. R., & Yetisen, A. K. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.010 · Cited by 432 · cited 1×
“Further, it has become evident that large-scale production and commercialization of PDMS-based devices are hindered by up-scaling difficulties and low fabrication throughput [19].”
BackgroundUchida, T., Rossignol, F., Matthay, M. A., Mounier, R., Couette, S., Clottes, E., & Clerici, C. (2004) · Journal of Biological Chemistry · doi:10.1074/jbc.m400461200 · Cited by 428 · cited 2×
“[38], where the level of HIF-1-alpha was shown to rise first at 3% O2 in A549 cells subjected to a range of decreasing oxygen concentrations.”
BackgroundThe rate of oxygen utilization by cells
Wagner, B. A., Venkataraman, S., & Buettner, G. R. (2011) · Free Radical Biology and Medicine · doi:10.1016/j.freeradbiomed.2011.05.024 · Cited by 363
Flow-induced deformation of shallow microfluidic channels
Gervais, T., El-Ali, J., Günther, A., & Jensen, K. F. (2006) · Lab on a Chip · doi:10.1039/b513524a · Cited by 359
Cited by
Cancer-on-chip: a 3D model for the study of the tumor microenvironment
Cauli, E., Polidoro, M. A., Marzorati, S., Bernardi, C., Rasponi, M., & Lleò, A. (2023) · Journal of Biological Engineering · doi:10.1186/s13036-023-00372-6 · Cited by 52
“Solutions can be found using for example additives [196], coatings [197, 198], and completely other materials [199–201].”
BackgroundScalable Processing of Cyclic Olefin Copolymer (COC) Microfluidic Biochips
Rodrigues, R. G., Condelipes, P. G. M., da Rosa, R. R., Chu, V., & Conde, J. P. (2023) · Micromachines · doi:10.3390/mi14101837 · Cited by 25
“With this low gas permeability, the gas bubbles present do not pass through the COC material and will eventually leave the channel carried by the flowing fluid In PDMS, the gas permeability is reported to be between 523 barrer [47] and 733 barrer [48] at 35–37 °C. .”
BackgroundRecent Advances and Future Perspectives in Vascular Organoids and Vessel-on-Chip
Cheruku, G. R., Wilson, C. V., Raviendran, S., & Xiao, Q. (2024) · Organoids · doi:10.3390/organoids3030014 · Cited by 25
Ito, Y., Kawauchi, I., Yanagita, Y., Sakai, Y., Nishikawa, M., Arakawa, H., Kadoguchi, M., Tamai, I., Esashika, K., Takahashi, J., et al. (2024) · Emergent Materials · doi:10.1007/s42247-024-00822-x · Cited by 16
“PMP membranes are now being used in perfusion-type MPS, and Sonstevoid et al. fabricated an MPS with a PMP membrane as a cell attachment surface that enables on-site, real-time oxygen concentration measurement via optical fiber [29].”
Liao, Y., Chai, D., Wang, Q., Wang, X., Yong, Q., Cheng, Z., Zhang, C., Zhang, D., Liu, B., Liu, R., et al. (2025) · Materials Horizons · doi:10.1039/d4mh01174k · Cited by 14
Metadata
- Authors
- Sønstevold, L., Czerkies, M., Escobedo-Cousin, E., Blonski, S., & Vereshchagina, E.
- Affiliations
- SINTEF Digital (NO)
- Institute of Fundamental Technological Research (PL)
- SINTEF Digital, Department of Smart Sensors and Microsystems, Gaustadalléen 23C, 0373 Oslo, Norway
- Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawińskiego St. 5B, 02-106 Warsaw, Poland
- Corresponding author
- Sønstevold, Linda
- Czerkies, Maciej
- ORCID
- Venue
- Micromachines
- Publisher
- MDPI AG
- Published
- 2023-02-24
- Volume
- 14
- Issue
- 3
- Pages
- 532
- ISSN
- 2072-666X (ISSN-L)
- PMC
- PMC10053898
- OpenAlex
- W4322501630
- Cited by
- 25
- References
- 34
- Citations by year
- 2026: 9 · 2025: 5 · 2024: 9 · 2023: 2
- FWCI
- 1.97
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Micromachines (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- NORA - Norwegian Open Research Archives (submittedVersion, cc-by)
- Funders
- Norwegian Financial Mechanism 2014–2021 (2019/34/H/NZ6/00699)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › 3D Printing in Biomedical Research
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Keywords
- polymethylpentene, oxygen permeability, organ-on-a-chip, polydimethylsiloxane, A549 cells, epithelial cells, oxygen supply, live-cell imaging, optical properties, microfluidics, cell viability
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Application of Polymethylpentene, an Oxygen Permeable Thermoplastic, for Long-Term on-a-Chip Cell Culture and Organ-on-a-Chip Devices. https://app.cochat.ai/papers/10.3390/mi14030532
- The paper
- Sønstevold, L., Czerkies, M., Escobedo-Cousin, E., Blonski, S., & Vereshchagina, E. (2023). Application of Polymethylpentene, an Oxygen Permeable Thermoplastic, for Long-Term on-a-Chip Cell Culture and Organ-on-a-Chip Devices. Micromachines. https://doi.org/10.3390/mi14030532
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.