Analysis and Optimization of Spiral Circular Inductive Coupling Link for Bio-Implanted Applications on Air and within Human Tissue
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
The use of wireless communication using inductive links to transfer data and power to implantable microsystems to stimulate and monitor nerves and muscles is increasing. This paper deals with the development of the theoretical analysis and optimization of an inductive link based on coupling and on spiral circular coil geometry. The coil dimensions offer 22 mm of mutual distance in air. However, at 6 mm of distance, the coils offer a power transmission efficiency of 80% in the optimum case and 73% in the worst case via low input impedance, whereas, transmission efficiency is 45% and 32%, respectively, via high input impedance. The simulations were performed in air and with two types of simulated human biological tissues such as dry and wet-skin using a depth of 6 mm. The performance results expound that the combined magnitude of the electric field components surrounding the external coil is approximately 98% of that in air, and for an internal coil, it is approximately 50%, respectively. It can be seen that the gain surrounding coils is almost constant and confirms the omnidirectional pattern associated with such loop antennas which reduces the effect of non-alignment between the two coils. The results also show that the specific absorption rate (SAR) and power loss within the tissue are lower than that of the standard level. Thus, the tissue will not be damaged anymore.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Visual perception in a blind subject with a chronic microelectronic retinal prosthesis
Humayun, M. S., Weiland, J. D., Fujii, G. Y., Greenberg, R., Williamson, R., Little, J., Mech, B., Cimmarusti, V., Van Boemel, G., Dagnelie, G., et al. (2003) · Vision Research · doi:10.1016/s0042-6989(03)00457-7 · Cited by 856 · cited 1×
“For example, in general implanted micro-system stimulators depth is 1–4 mm, for retinal implants it is 5 mm and for cochlear implants depth is 3–6 mm, respectively [2–4].”
BackgroundA low-power integrated circuit for a wireless 100-electrode neural recording system.
Harrison, R. R., Watkins, P. T., Kier, R. J., Lovejoy, R. O., Black, D. J., Greger, B., & Solzbacher, F. (2007) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2006.886567 · Cited by 782
Design of radio-frequency powered coils for implant instruments
Ko, W. H., Liang, S. P., & Fung, C. D. F. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457921 · Cited by 374
Cited by
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-0792-9 · Cited by 482
Wireless power transfer techniques for implantable medical devices: a review
Khan, S. R., Pavuluri, S. K., Cummins, G., & Desmulliez, M. P. Y. (2020) · Sensors · doi:10.3390/s20123487 · Cited by 355
“As shown in Figure 5, an iterative optimization flow diagram has been adopted in most of the research literature to achieve maximum PTE [36,66,70,75,76].”
MethodWireless power transfer approaches for medical implants: a review
Haerinia, M., & Shadid, R. (2020) · Signals · doi:10.3390/signals1020012 · Cited by 174
“Mainly, determining SAR can be achieved by using numerical techniques and empirical models using fabricated tissue phantoms [4,34–38].”
BackgroundMethodInfluentialNear-field wireless power transfer used in biomedical implants: a comprehensive review
Mahmood, A. I., Gharghan, S. K., Eldosoky, M. A., & Soliman, A. M. (2022) · IET Power Electronics · doi:10.1049/pel2.12351 · Cited by 59
Metadata
- Authors
- Mutashar, S., Hannan, M., Samad, S., & Hussain, A.
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2014-06-30
- Volume
- 14
- Issue
- 7
- Pages
- 11522-11541
- ISSN
- 1424-8220 (ISSN-L)
- OpenAlex
- W2024498187
- Cited by
- 74
- References
- 3
- FWCI
- 2.65
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.