Vein Pattern Locating Technology for Cannulation: A Review of the Low-Cost Vein Finder Prototypes Utilizing near Infrared (NIR) Light to Improve Peripheral Subcutaneous Vein Selection for Phlebotomy
Open accessThe DOI matches its Crossref record; 67 of the 67 references it cites were found in a registry or an index.
Abstract
One of the most common means for diagnosis is through medical laboratory testing, which primarily uses venous blood as a sample. This requires an invasive method by cannulation that needs proper vein selection. The use of a vein finder would help the phlebotomist to easily locate the vein, preventing possible pre-analytical error in the specimen collection and even more discomfort and pain to the patient. This paper is a review of the scientific publications on the different developed low-cost vein finder prototypes utilizing camera assisted near infrared (NIR) light technology. Methods: Electronic databases were searched online, these included PubMed (PMC), MEDLINE, Science Direct, ResearchGate, and Institute of Electrical and Electronics Engineers (IEEE) Xplore digital library. Specifically, publications with the terms vein finder prototype, NIR technology, vein detection, and infrared imaging were screened. In addition, reference lists were used to further review related publications. Results: Cannulation challenges medical practitioners because of the different factors that can be reduced by the utilization of a vein finder. A limited number of publications regarding the assessment of personnel performing cannulation were observed. Moreover, variations in methodology, number of patients, type of patients according to their demographics and materials used in the assessment of the developed prototypes were noted. Some studies were limited with regard to the actual human testing of the prototype. Conclusions: The development of a low-cost effective near infrared (NIR) vein finder remains in the phase of improvement. Since, it is being challenged by different human factors, increasing the number of parameters and participants/human for actual testing of the prototypes must also be taken into consideration for possible commercialization. Finally, it was noted that publications regarding the assessment of the performance of phlebotomists using vein finders were limited.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures



![Figure 4 Sample veins image from different wavelengths: (a) 850 nm [50] and (b) 696 nm [43].](https://figures.cochat.ai/paper-figures/ec160c930a31485bbb9f52bba331aa9d/vein-pattern-locating-technology-for-cannulation-a-review-of-the-low-cost-vein-figure-4.jpg)

![Figure 6 (a) Selecting a point from fused images to extract values of: (b) Depth and (c) thickness using the proposed imaging system [71].](https://figures.cochat.ai/paper-figures/38fc746242854d4f80d73ff2b8bb4574/vein-pattern-locating-technology-for-cannulation-a-review-of-the-low-cost-vein-figure-6.jpg)
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“First-attempt success rates improved by 28% (RR 1.28, 95% CI 1.14–1.42, p < 0.001) [15,22,23], with a number needed to treat of 4.7.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The 42% relative improvement in first-attempt success [13,14,15] translates to substantially reduced patient discomfort, particularly important in vulnerable populations such as children and cancer patients who may require frequent blood sampling.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Fifteen studies encompassing 2468 patients [15,22,23] examined near-infrared visualization devices compared to traditional venipuncture.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Complication rates were significantly reduced (RR 0.41, 95% CI 0.33–0.52, p < 0.001) [13,15], primarily reflecting decreased hematoma formation and patient discomfort.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“First-attempt success rates showed a dramatic improvement with ultrasound guidance (RR 1.42, 95% CI 1.26–1.58, p < 0.001) [13,14,15], corresponding to a number needed to treat of 3.2 patients.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measured
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Prevalence of Obesity Among Adults and Youth: United States, 2015-2016.
Hales, C. M., Carroll, M. D., Fryar, C. D., & Ogden, C. L. (2017) · PubMed · Cited by 3,884
Preanalytical quality improvement: from dream to reality
Lippi, G., Chance, J. J., Church, S., Dazzi, P., Fontana, R., Giavarina, D., Grankvist, K., Huisman, W., Kouri, T., Palicka, V., et al. (2011) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2011.600 · Cited by 388
Hemoglobin Variants: Biochemical Properties and Clinical Correlates
Thom, C. S., Dickson, C. F., Gell, D. A., & Weiss, M. J. (2013) · Cold Spring Harbor Perspectives in Medicine · doi:10.1101/cshperspect.a011858 · Cited by 336 · cited 1×
“As to its major function, it is able to transport oxygen (O2) rich blood (oxyhemoglobin) from the lungs to the peripheral tissues and carbon dioxide (CO2) rich blood (deoxyhemoglobin) with the opposite route [9].”
BackgroundRising Burden of Obesity in Asia
Ramachandran, A., & Snehalatha, C. (2010) · Journal of Obesity · doi:10.1155/2010/868573 · Cited by 301 · cited 4×
“More so, obesity levels have reached epidemic proportions in many Asian countries and many of them have rates which are not far from that in the USA [63] with the Centers for Disease Control and Prevention (CDC) report in 2015–2016, of the obesity prevalence in US of 39.8% in adults and 18.5% in youth [64].”
BackgroundInfluentialSebbane, M., Claret, P.-G., Lefebvre, S., Mercier, G., Rubenovitch, J., Jreige, R., Eledjam, J.-J., & de La Coussaye, J.-E. (2013) · The Journal of Emergency Medicine · doi:10.1016/j.jemermed.2012.07.051 · Cited by 179 · cited 2×
“Moreover, cannulation is being used in performing technical procedures in emergency medicine [10].”
BackgroundMethodLippi, G., Banfi, G., Church, S., Cornes, M., De Carli, G., Grankvist, K., Kristensen, G. B., Ibarz, M., Panteghini, M., Plebani, M., et al. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2014-1051 · Cited by 151 · cited 1×
“The said processes are part of the pre-analytical phase [2].”
BackgroundA Review of Medical Errors in Laboratory Diagnostics and Where We Are Today: Table 1
Hammerling, J. A. (2012) · Laboratory Medicine · doi:10.1309/lm6er9wjr1ihqauy · Cited by 146
Blood venous sample collection: Recommendations overview and a checklist to improve quality
Giavarina, D., & Lippi, G. (2017) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2017.02.021 · Cited by 136 · cited 1×
“With these challenges, the integrity of the biological samples that should represent the biological status in vivo was affected leading to unreliable diagnostic information [3].”
BackgroundEvaluation of errors in a clinical laboratory: a one-year experience
Goswami, B., Singh, B. M., Chawla, R., & Mallika, V. (2010) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.006 · Cited by 121
Lima-Oliveira, G., Volanski, W., Lippi, G., Picheth, G., & Guidi, G. C. (2017) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2017.1295317 · Cited by 117 · cited 2×
“Moreover, it was mentioned that the transilluminator (vein finder) can make for more accurate visualization of the vein course, unfortunately, it is not widely used mainly due to the cost of the commercially available devices [19].”
Background
Research works printed but not linked: 2
- Doshi, 2013, Non-invasive optical sensor for hemoglobin determination, Int. J. Eng. Res. Appl. (Ijera), 3, 559
- Crisan, S., Tarnovan, I., and Crisan, T. (2007, January 19–21). Vein Pattern Recognition. Image Enhancement and feature Extraction Algorithms. Proceedings of the 15th IMEKO TC4 Symposium on Novelties in Electrical Measurements and Instrumentation, Iaşi, Romania.
Cited by
He, X., Liu, Y., Ganesan, K., Ahnood, A., Beckett, P., Eftekhari, F., Smith, D., Uddin, M. H., Skafidas, E. S., Nirmalathas, A. T., et al. (2020) · APL Photonics · doi:10.1063/1.5140215 · Cited by 65
Francisco, M. D., Chen, W., Pan, C.-T., Lin, M., Wen, Z., Liao, C.-F., & Shiue, Y. (2021) · Micromachines · doi:10.3390/mi12040373 · Cited by 53
“These were the veins that were observed during the prototype testing with emphasis on the median cubital vein [9].”
MethodSelf-citationImplementation of Thermal Camera for Non-Contact Physiological Measurement: A Systematic Review
Manullang, M. C. T., Lin, Y., Lai, S.-J., & Chou, N. (2021) · Sensors · doi:10.3390/s21237777 · Cited by 67
“[24] 2019 Focused on vein finder by using near infrared (NIR)”
BackgroundHuang, W., Yoon, S., Wu, B., Lu, K., Lin, C.-M., Yang, H., & Liu, R. (2020) · Chemical Communications · doi:10.1039/d0cc03030a · Cited by 35
Bae, B., Park, M., Lee, D., Sim, I., & Lee, K. (2022) · Advanced Optical Materials · doi:10.1002/adom.202201905 · Cited by 35
Metadata
- Authors
- Pan, C.-T., Francisco, M. D., Yen, C.-K., Wang, S.-Y., & Shiue, Y.-L.
- Affiliations
- National Sun Yat-sen University (TW)
- Department of Mechanical and Electro-Mechanical Engineering, National Sun Yat-sen University (NSYSU), Kaohsiung 80424, Taiwan
- Corresponding author
- Shiue, Yow-Ling
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2019-08-16
- Volume
- 19
- Issue
- 16
- Pages
- 3573
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC6719195
- OpenAlex
- W2967660745
- Semantic Scholar
- e442e7a08b4592a54132c5c81ba54eaa7563059f
- Cited by
- 93
- References
- 67
- Citations by year
- 2026: 10 · 2025: 15 · 2024: 11 · 2023: 16 · 2022: 12 · 2021: 17 · 2020: 12
- FWCI
- 1.54
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Science Council
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Central Venous Catheters and Hemodialysis
- Keywords
- near-infrared imaging, cannulation, phlebotomy, low-cost prototype, vein detection, pre-analytical errors, blood sampling, human factors
- MeSH terms
- Catheterization, Cost-Benefit Analysis, Humans, Semiconductors, Veins, Phlebotomy, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.