Vein Pattern Locating Technology for Cannulation: A Review of the Low-Cost Vein Finder Prototypes Utilizing near Infrared (NIR) Light to Improve Peripheral Subcutaneous Vein Selection for Phlebotomy
Open accessThe DOI matches its Crossref record; 67 of the 67 references it cites were found in a registry or an index.
Abstract
One of the most common means for diagnosis is through medical laboratory testing, which primarily uses venous blood as a sample. This requires an invasive method by cannulation that needs proper vein selection. The use of a vein finder would help the phlebotomist to easily locate the vein, preventing possible pre-analytical error in the specimen collection and even more discomfort and pain to the patient. This paper is a review of the scientific publications on the different developed low-cost vein finder prototypes utilizing camera assisted near infrared (NIR) light technology. Methods: Electronic databases were searched online, these included PubMed (PMC), MEDLINE, Science Direct, ResearchGate, and Institute of Electrical and Electronics Engineers (IEEE) Xplore digital library. Specifically, publications with the terms vein finder prototype, NIR technology, vein detection, and infrared imaging were screened. In addition, reference lists were used to further review related publications. Results: Cannulation challenges medical practitioners because of the different factors that can be reduced by the utilization of a vein finder. A limited number of publications regarding the assessment of personnel performing cannulation were observed. Moreover, variations in methodology, number of patients, type of patients according to their demographics and materials used in the assessment of the developed prototypes were noted. Some studies were limited with regard to the actual human testing of the prototype. Conclusions: The development of a low-cost effective near infrared (NIR) vein finder remains in the phase of improvement. Since, it is being challenged by different human factors, increasing the number of parameters and participants/human for actual testing of the prototypes must also be taken into consideration for possible commercialization. Finally, it was noted that publications regarding the assessment of the performance of phlebotomists using vein finders were limited.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures



![Figure 4 Sample veins image from different wavelengths: (a) 850 nm [50] and (b) 696 nm [43].](https://figures.cochat.ai/paper-figures/ec160c930a31485bbb9f52bba331aa9d/vein-pattern-locating-technology-for-cannulation-a-review-of-the-low-cost-vein-figure-4.jpg)

![Figure 6 (a) Selecting a point from fused images to extract values of: (b) Depth and (c) thickness using the proposed imaging system [71].](https://figures.cochat.ai/paper-figures/38fc746242854d4f80d73ff2b8bb4574/vein-pattern-locating-technology-for-cannulation-a-review-of-the-low-cost-vein-figure-6.jpg)
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“First-attempt success rates improved by 28% (RR 1.28, 95% CI 1.14–1.42, p < 0.001) [15,22,23], with a number needed to treat of 4.7.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The 42% relative improvement in first-attempt success [13,14,15] translates to substantially reduced patient discomfort, particularly important in vulnerable populations such as children and cancer patients who may require frequent blood sampling.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Fifteen studies encompassing 2468 patients [15,22,23] examined near-infrared visualization devices compared to traditional venipuncture.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Complication rates were significantly reduced (RR 0.41, 95% CI 0.33–0.52, p < 0.001) [13,15], primarily reflecting decreased hematoma formation and patient discomfort.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“First-attempt success rates showed a dramatic improvement with ultrasound guidance (RR 1.42, 95% CI 1.26–1.58, p < 0.001) [13,14,15], corresponding to a number needed to treat of 3.2 patients.”
From Strategies to Address Difficult Venous Access in Blood Sampling: A Comprehensive Meta-Analysis (2026) · cites this paper as [15]
Experimental — accuracy not yet measured
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Prevalence of Obesity Among Adults and Youth: United States, 2015-2016.
Hales, C. M., Carroll, M. D., Fryar, C. D., & Ogden, C. L. (2017) · PubMed · Cited by 3,884
Preanalytical quality improvement: from dream to reality
Lippi, G., Chance, J. J., Church, S., Dazzi, P., Fontana, R., Giavarina, D., Grankvist, K., Huisman, W., Kouri, T., Palicka, V., et al. (2011) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2011.600 · Cited by 388
Hemoglobin Variants: Biochemical Properties and Clinical Correlates
Thom, C. S., Dickson, C. F., Gell, D. A., & Weiss, M. J. (2013) · Cold Spring Harbor Perspectives in Medicine · doi:10.1101/cshperspect.a011858 · Cited by 336 · cited 1×
“As to its major function, it is able to transport oxygen (O2) rich blood (oxyhemoglobin) from the lungs to the peripheral tissues and carbon dioxide (CO2) rich blood (deoxyhemoglobin) with the opposite route [9].”
BackgroundRising Burden of Obesity in Asia
Ramachandran, A., & Snehalatha, C. (2010) · Journal of Obesity · doi:10.1155/2010/868573 · Cited by 301 · cited 4×
“More so, obesity levels have reached epidemic proportions in many Asian countries and many of them have rates which are not far from that in the USA [63] with the Centers for Disease Control and Prevention (CDC) report in 2015–2016, of the obesity prevalence in US of 39.8% in adults and 18.5% in youth [64].”
BackgroundInfluentialSebbane, M., Claret, P.-G., Lefebvre, S., Mercier, G., Rubenovitch, J., Jreige, R., Eledjam, J.-J., & de La Coussaye, J.-E. (2013) · The Journal of Emergency Medicine · doi:10.1016/j.jemermed.2012.07.051 · Cited by 179 · cited 2×
“Moreover, cannulation is being used in performing technical procedures in emergency medicine [10].”
BackgroundMethodLippi, G., Banfi, G., Church, S., Cornes, M., De Carli, G., Grankvist, K., Kristensen, G. B., Ibarz, M., Panteghini, M., Plebani, M., et al. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2014-1051 · Cited by 151 · cited 1×
“The said processes are part of the pre-analytical phase [2].”
BackgroundA Review of Medical Errors in Laboratory Diagnostics and Where We Are Today: Table 1
Hammerling, J. A. (2012) · Laboratory Medicine · doi:10.1309/lm6er9wjr1ihqauy · Cited by 146
Blood venous sample collection: Recommendations overview and a checklist to improve quality
Giavarina, D., & Lippi, G. (2017) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2017.02.021 · Cited by 136 · cited 1×
“With these challenges, the integrity of the biological samples that should represent the biological status in vivo was affected leading to unreliable diagnostic information [3].”
BackgroundEvaluation of errors in a clinical laboratory: a one-year experience
Goswami, B., Singh, B. M., Chawla, R., & Mallika, V. (2010) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.006 · Cited by 121
Lima-Oliveira, G., Volanski, W., Lippi, G., Picheth, G., & Guidi, G. C. (2017) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2017.1295317 · Cited by 117 · cited 2×
“Moreover, it was mentioned that the transilluminator (vein finder) can make for more accurate visualization of the vein course, unfortunately, it is not widely used mainly due to the cost of the commercially available devices [19].”
BackgroundMIYAKE, R. K., ZEMAN, H. D., DUARTE, F. H., KIKUCHI, R., RAMACCIOTTI, E., LOVHOIDEN, G., & VRANCKEN, C. (2006) · Dermatologic Surgery · doi:10.1111/j.1524-4725.2006.32226.x · Cited by 110 · cited 1×
“Aside from venipuncture, the near infrared (NIR) transilluminator imaging was used to enhance vein treatment by helping in the identification of veins which were noted to be invisible to the naked eye but too shallow for ultrasound detection [22].”
MethodChiao, F., Resta-Flarer, F., Lesser, J., Ng, J., Ganz, A., Pino-Luey, D., Bennett, H. L., Perkins, C., & Witek, B. (2013) · British Journal of Anaesthesia · doi:10.1093/bja/aet003 · Cited by 102 · cited 2×
“, 2013 [25] study showed that the possible cannulation site with the use of a vein finder device has better results with 9.”
BackgroundMethodInfluentialPhlebotomy, a bridge between laboratory and patient
Ialongo, C., & Bernardini, S. (2016) · Biochemia Medica · doi:10.11613/bm.2016.002 · Cited by 97 · cited 6×
“In venipuncture, there are means on how to help the phlebotomist view the vein properly, among those were by slightly tapping the site [5], use of tourniquet, and applying cold vibration close to the site for venipuncture or fist clenching, which are considered and studied to cause physiological disturbance on veins a…”
BackgroundInfluentialNajat, D. (2017) · PLoS ONE · doi:10.1371/journal.pone.0170211 · Cited by 75
West, J., Atherton, J., Costelloe, S. J., Pourmahram, G. E., Stretton, A., & Cornes, M. P. (2016) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/0004563216669384 · Cited by 70
II. Difficult peripheral veins: turn on the lights
Lamperti, M., & Pittiruti, M. (2013) · British Journal of Anaesthesia · doi:10.1093/bja/aet078 · Cited by 66
Parker, S. I. A., Benzies, K., Hayden, A., & Lang, E. (2016) · International Emergency Nursing · doi:10.1016/j.ienj.2016.05.004 · Cited by 66 · cited 1×
“It includes peripheral intravenous catherization (PIVC) that remains challenging for many clinicians to complete successfully on the first attempt [11].”
BackgroundBuowari, O. Y. (2013) · Advances in Bioscience and Biotechnology · doi:10.4236/abb.2013.41a018 · Cited by 59
Phipps, K., Modic, A., O’Riordan, M. A., & Walsh, M. C. (2011) · Journal of Perinatology · doi:10.1038/jp.2011.129 · Cited by 56 · cited 1×
“, [26] a study based on successful first attempt for peripherally inserted central catheter shows 59.”
BackgroundTowards a Low-Cost Mobile Subcutaneous Vein Detection Solution Using Near-Infrared Spectroscopy
Jurič, S., Flis, V., Debevc, M., Holzinger, A., & Žalik, B. (2014) · The Scientific World JOURNAL · doi:10.1155/2014/365902 · Cited by 55 · cited 2×
“(2014) [24] 740 Standard (Universal Serial Bus) USB camera Pass-through filter (exposed and developed empty 35 mm camera film BMI 72 subjects”
BackgroundHigh-contrast subcutaneous vein detection and localization using multispectral imaging
Wang, F., Behrooz, A., Morris, M., & Adibi, A. (2013) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.5.050504 · Cited by 53 · cited 8×
“l t t r ti res lti t al ost vis a ly unnoticeable in the IR i age [50].”
BackgroundMethodInfluentialLippi, G., Baird, G. S., Banfi, G., Bölenius, K., Cadamuro, J., Church, S., Cornes, M. P., Dacey, A., Guillon, A., HOFFMANN, G. E., et al. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0107 · Cited by 53 · cited 1×
“It must be with advancement in order to achieve a greater degree in quality healthcare, especially from activities starting with ordering, obtaining and handling of biological specimens [1].”
BackgroundVisualizing Veins With Near-Infrared Light to Facilitate Blood Withdrawal in Children
Cuper, N. J., Verdaasdonk, R. M., de Roode, R., de Vooght, K. M. K., Viergever, M. A., Kalkman, C. J., & de Graaff, J. C. (2011) · Clinical Pediatrics · doi:10.1177/0009922810395932 · Cited by 48 · cited 7×
“, 2011 [29] has the highest number of subjects with a total of 125 male and female children followed by Chen A et al.”
BackgroundMethodResultInfluentialvan der Woude, O. C. P., Cuper, N. J., Getrouw, C., Kalkman, C. J., & de Graaff, J. C. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31828e5bde · Cited by 47 · cited 1×
“Moreover, a study showed that a commercial NIR vascular imaging device was reported with limited value in improving success at first attempt of intravenous (IV) cannulation in children with dark skin color [66].”
BackgroundPortable robot for autonomous venipuncture using 3D near infrared image guidance
Chen, A., Nikitczuk, K., Nikitczuk, J., Maguire, T. J., & Yarmush, M. L. (2013) · TECHNOLOGY · doi:10.1142/s2339547813500064 · Cited by 46 · cited 6×
“(2013) [49] 940 Two monochrome FireWire cameras with high sensitivity CCD sensors in the near-infrared range (Point Grey Firefly MV) 850–1060 nm band pass filters (Edmund Optics) Age, gender, BMI and skin color (Fitzpatrick skin type) 101 patients”
BackgroundResultInfluentialShahzad, A., Saad, M. N. M., Walter, N., Malik, A. S., & Mériaudeau, F. (2014) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-13-109 · Cited by 44 · cited 7×
“The imaging, through the use of trans-illuminating devices, allows a deeper look below the skin with best wavelength range for better contrast between skin and veins image [7].”
BackgroundInfluentialLima-Oliveira, G., Lippi, G., Salvagno, G. L., Montagnana, M., Picheth, G., & Guidi, G. C. (2013) · Biochemia Medica · doi:10.11613/bm.2013.037 · Cited by 40 · cited 2×
“The procedure for venipuncture cannot be considered as a gold standard yet, since errors are still being detected [20] with continuous recommendations for improvements were supplemented.”
BackgroundResultJurič, S., & Žalik, B. (2014) · BMC Medical Informatics and Decision Making · doi:10.1186/s12911-014-0100-z · Cited by 39 · cited 3×
“Moreover, the study of Juric S and Zalik B, 2014 [58] included the participation of 43 trainee clinicians (nursing students) in performing the procedure to evaluate their prototype through a non-invasive, observational study having the highest result of 4.”
BackgroundMethodPreliminary Study for Designing a Novel Vein-Visualizing Device
Kim, D., Kim, Y.-J., Yoon, S., & Lee, D. (2017) · Sensors · doi:10.3390/s17020304 · Cited by 33
Ogden-Grable, H., & Gill, G. W. (2005) · Laboratory Medicine · doi:10.1309/extw9lbm0cd7p9ev · Cited by 31 · cited 1×
“Compared to the two other sites it is well anchored vein, usually large and prominent [69].”
BackgroundLeitch, A., McCormick, I., Gunn, I., & Gillespie, T. A. (2006) · Journal of Hospital Infection · doi:10.1016/j.jhin.2006.03.006 · Cited by 29 · cited 1×
“, [72] it was showed that the rate of contamination with Methicillin-resistant Staphylococcus aureus (MRSA) was 32 of 131 (25%) of the tourniquets tested, which leads to the conclusion that the tourniquets being used for phlebotomy may be a potential vectors for transferring bacteria including MRSA.”
BackgroundShedding Light on the Veins - Reflected Light or Transillumination in Hand-Vein Recognition
Kauba, C., & Uhl, A. (2018) · International Conference on Biometrics · doi:10.1109/icb2018.2018.00050 · Cited by 29 · cited 4×
“While transilluminati n devices is the opposite, which require higher light intensity that uses more power and non-compact with the arm or hand b ing sc nned were plac d in between the light sou ce and the camera shown in Figure 2b [34] (see Figure 3).”
BackgroundInfluentialPrototype design for wearable veins localization system using near infrared imaging technique
Meng, G. C., Shahzad, A., Saad, M. N. M., Malik, A. S., & Mériaudeau, F. (2015) · IEEE International Colloquium on Signal Processing & Its Applications (CSPA) · doi:10.1109/cspa.2015.7225628 · Cited by 28 · cited 4×
“(2015) [48] 830 and 850 Vuzix STAR 1200XL eyewear system IR CCD camera NA NA—No patient testing done”
BackgroundInfluentialObesity significantly increases the difficulty of patient management in the emergency department
Kam, J. K. T., & Taylor, D. M. (2010) · Emergency Medicine Australasia · doi:10.1111/j.1742-6723.2010.01307.x · Cited by 27 · cited 1×
“A 2010 study by Kam J and Taylor D [62] showed that patient management difficulty did not increase until the BMI was in the obese or morbidly obese range, especially in cannulation and venipuncture.”
MethodFukuroku, K., Narita, Y., Taneda, Y., Kobayashi, S., & Gayle, A. A. (2016) · Nurse Education in Practice · doi:10.1016/j.nepr.2016.02.005 · Cited by 27 · cited 2×
“, 2016 [28] study on validating the effectiveness of vein infrared visualization performed by second-year nursing students, confirmed that the vein viewer contribute immensely to the improvement of their practical skills in venipuncture, specifically, in the site selection time with the context for the elderly patient…”
BackgroundVein detection system using infrared camera
Ghozali, H. K. A., Setiawardhana, S., & Sigit, R. (2016) · International Electronics Symposium (IES) · doi:10.1109/elecsym.2016.7860987 · Cited by 27 · cited 2×
“It often resu ts i multipl attempts in needle insertion eading to pain, di comf rt, dissatisfaction, de ay d treatment [37], h matoma, swell in th skin, cuts on the bone [39], bleeding, d even in ec ion.”
BackgroundTen years of preanalytical monitoring and control: Synthetic Balanced Score Card Indicator
Salinas, M., López‐Garrigós, M., Flores, E., Santo-Quiles, A., Gutiérrez, M., Lugo, J., Lillo, R., & Leiva‐Salinas, C. (2015) · Biochemia Medica · doi:10.11613/bm.2015.005 · Cited by 26
Venipuncture-related lateral antebrachial cutaneous nerve injury: what to know?
Ramos, J. A. (2014) · Brazilian Journal of Anesthesiology (English Edition) · doi:10.1016/j.bjane.2013.06.003 · Cited by 26 · cited 1×
“Moreover, failure in venipuncture can cause thrombosis of the vein [4], hematoma, or even nerve injury involving the lateral antebrachial cutaneous nerve (LACN), which can lead to the so-called “causalgia” or complex regional pain syndrome (CRPS) [5,6].”
Backgroundde Graaff, J. C., Cuper, N. J., van Dijk, A. T. H., Timmers-Raaijmaakers, B. C. M. S., van der Werff, D. B. M., & Kalkman, C. J. (2014) · Pediatric Anesthesia · doi:10.1111/pan.12501 · Cited by 23
PRE-ANALYTICAL ERRORS IN THE CLINICAL LABORATORY AND HOW TO MINIMIZE THEM
Patra, M. D. S., Mukherjee, B., & Das, A. K. (2013) · International Journal of Bioassays · doi:10.21746/ijbio.2013.03.0017 · Cited by 21
A fast vein display device based on the camera-projector system
Dai, X., Zhou, Y., Hu, X., Liu, M., Zhu, X., & Wu, Z. (2013) · IEEE International Conference on Imaging Systems and Techniques (IST) · doi:10.1109/ist.2013.6729680 · Cited by 21 · cited 5×
“This is usually due to the following reasons; veins are either very small, and/or deep, the patient’s age (elderly or pediatrics), skin color [37], and obesity level [32,35].”
BackgroundMethodInfluentialMarathe, M. S., Bhatt, N., & Sundararajan, R. (2014) · doi:10.1109/cimca.2014.7057805 · Cited by 21 · cited 4×
“This is composed o two stages, firstly; the infrared (IR) light penetrat s the human tissue and secondly, de xyhemoglobin in the venous blood absorbs more of th incident IR light than the surrounding tissue [31].”
BackgroundMethodInfluentialLow-cost subcutaneous vein detection system using ARM9 based single board computer
Chakravorty, T., Sonawane, D., Sharma, S. D., & Patil, T. (2011) · International Conference on Electronics Computer Technology · doi:10.1109/icectech.2011.5941714 · Cited by 21 · cited 5×
“This is usually due to the following reasons; veins are either very small, and/or deep, the patient’s age (elderly or pediatrics), skin color [37], and obesity level [32,35].”
BackgroundInfluentialA low cost vein detection system using integrable mobile camera devices
Nundy, K. K., & Sanyal, S. (2010) · IEEE India Conference · doi:10.1109/indcon.2010.5712670 · Cited by 20 · cited 3×
“(2010) [52] 740–760 Ordinary camera phones with even VGA quality pictures Optical filter using butter paper and filter made from exposed and developed film strips NA NA—No patient testing done”
BackgroundMethodSubcutaneous veins depth measurement using diffuse reflectance images
Goh, C. M., Subramaniam, R., Saad, M. N. M., Ali, S. A., & Mériaudeau, F. (2017) · Optics Express · doi:10.1364/oe.25.025741 · Cited by 20 · cited 4×
“The obtained OD ratio is referred to the 3000 diffuse reflectance models to produce depth and thickness, as shown in Figure 6 [71].”
BackgroundMethodInfluentialMultisensory System for the Detection and Localization of Peripheral Subcutaneous Veins
Fernández, R., & Armada, M. (2017) · Sensors · doi:10.3390/s17040897 · Cited by 18 · cited 2×
“(2017) [44] 940 GoldEye P-032 SWIR camera (AlliedVision, Stradtroda, Germany), a Swiss Ranger SR-400011 TOF 3D camera (MesaImaging, Zürich, Switzerland) NA NA—No patient testing done”
BackgroundPeripheral vein locating techniques
Cantor-Peled, G., & Ovadia-Blechman, M. H. Z. (2016) · Imaging in Medicine · Cited by 17
Design of vein finder with multi tuning wavelength using RGB LED
Chandra, F., Wahyudianto, A., & Yasin, M. (2017) · Journal of Physics Conference Series · doi:10.1088/1742-6596/853/1/012019 · Cited by 17 · cited 6×
“(2017) [43] 600–696 NA (*) BMI, age, and skin color Tested to 10 patients”
BackgroundResultInfluentialDepth and Size Limits for the Visibility of Veins Using the VeinViewer Imaging System
Ganesh, S. (2007) · doi:10.21007/etd.cghs.2007.0103 · Cited by 17
Diagnostic Superficial Vein Scanner
Ayoub, Y., Serhal, S., Farhat, B., Ali, A., Amatoury, J., Nasser, H., & Ali, M. A. (2018) · International Conference on Computer and Applications (ICCA) · doi:10.1109/comapp.2018.8460229 · Cited by 16 · cited 3×
“(2018) [33] 850 and 940 Nikon D810 camera (resolution 36.”
BackgroundMethodInfluentialAsh, C., Town, G., Bjerring, P., & Webster, S. (2014) · Photonics and Lasers in Medicine · doi:10.1515/plm-2013-0056 · Cited by 15 · cited 2×
“Various studies showed that darker skin color causes difficulty in viewing the vein [7,46,47,49,65].”
BackgroundEnhanced vision based vein detection system
Ahmed, K. I., Habaebi, M. H., Islam, M. R., & Zainal, N. A. (2017) · IEEE International Conference on Smart Instrumentation, Measurement and Applications (ICSIMA) · doi:10.1109/icsima.2017.8311990 · Cited by 15 · cited 1×
“CLAHE was developed in 1994, an algorithm which cut the produced histogram of the dark level of a picture then redistributed the pixels over the entire histogram, resulting into an enhanced image [59].”
MethodBarreras, J., & Chang, T. P. (2017) · Journal of the Association for Vascular Access · doi:10.1016/j.java.2016.12.005 · Cited by 11 · cited 2×
“Recently, a 2016 study by Barreras and Chang [27] revealed that the use of a near infrared (NIR) device significantly increased intravenous access success rates compared with visualization or palpation 26% vs.”
BackgroundInfluentialMiyake, R. K., ZEMAN, H. D., Duarte, F. H., et al. (2006) · Dermatologic Surgery · doi:10.1097/00042728-200608000-00008 · Cited by 11
Lee, S. J., Park, S., & Lee, D. (2013) · International Conference on Control, Automation and Systems (ICCAS) · doi:10.1109/iccas.2013.6704026 · Cited by 10 · cited 3×
“(2013) [41] 740 Real-time camera IR long-pass filter 695 nm (*) NA Vein model (plastic tube and dog’s blood)”
BackgroundMethodInfluentialThe image analysis of skin tissue irradiated with difference wavelengths of LED sources
Huang, K., Chang, C., Chang, H.-C., & Chang, C. (2012) · IEEE International Instrumentation and Measurement Technology Conference · doi:10.1109/i2mtc.2012.6229198 · Cited by 9 · cited 1×
“It is even used for optical imaging of the skin to guide dermatologist for proper diagnosis of diseases and skin care, specifically in the study by [23] Huang K et al.”
MethodKimori, K., Sugama, J., Nakatani, T., Nakayama, K., Miyati, T., & Sanada, H. (2015) · SAGE Open Medicine · doi:10.1177/2050312115615365 · Cited by 7 · cited 10×
“, 2015 [46] used 850 transmitted through the subcu aneous tissue and being compared as deep r t an the light s urce of a commercial vein finder.”
BackgroundMethodResultInfluentialDesign and construction of median cubital vein transillumination device by using LED
Amipongongctrch, P., Khaosomboon, K., & Keawgun, T. (2015) · Biomedical Engineering International Conference · doi:10.1109/bmeicon.2015.7399528 · Cited by 7 · cited 13×
“The image in transillumination is du to the pres ce of deoxyhemoglobin in venous blood which absorbs th red light and llumi ates the veins as dark lines n t skin surface, all wing the p lebotomist/m dical practitioners to locate the vein [36,37].”
BackgroundMethodInfluentialEconomically Affordable and Clinically Reliable Vein Finder
Dhakshayani, M. S., & Sikkandar, M. Y. (2015) · Cited by 7
Carlsen, R. W., Zyhier, S., & Şirinterlikçi, A. (2020) · ASEE Annual Conference & Exposition · doi:10.18260/1-2--30903 · Cited by 5 · cited 2×
“Simple materials such as: A plastic sheet of the floppy disks or negative films, as an inexpensive IR filter that can block wavelengths less than 600 nm and tissue paper and frosted window films, as a diffuser of light from light emitting diodes (LEDs) can be used for the said purpose [42].”
BackgroundHigh dynamic range CMOS image sensors in biomedical applications
Vatteroni, M., Covi, D., Stoppa, D., CRESPI, B., & Sartori, A. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4352915 · Cited by 4 · cited 1×
“CMOS image sensors are replacing CCD in many applications, with characteristics related to the possibility to directly interface the photosensitive device with electronic read-out at the pixel level and with better on-chip functionality [55].”
BackgroundNear-Infrared Imaging Approach for in vivo Detecting the Distribution of Human Blood Vessels
Yao, J. F., Zhao, J., & Hu, C. (2012) · doi:10.1109/icbeb.2012.281 · Cited by 4 · cited 1×
“(2012) [51] 940 Camera with CMOS high-transmittance imaging lens Filter passing wavelength of 940 nm Sites: Hand and arm (1) Hand and arm”
Vein detection using vein transillumination and contrast differentiation for practitioner aid
Agnalt, S. K., Canfield, D. M., Perreault, K. M., Legris, J., & McPheron, B. D. (2016) · IEEE MIT Undergraduate Research Technology Conference (URTC) · doi:10.1109/urtc.2016.8284073 · Cited by 4 · cited 2×
“The image in transillumination is due to the presence of deoxyhemoglobin in venous blood which absorbs the red light and illuminates the veins as dark lines on the skin surface, allowing the phlebotomist/medical practitioners to locate the vein [36,37].”
BackgroundCMOS image sensor for recording of intrinsic-optical-signal of the brain
Shishido, S., Oguro, Y., Noda, T., Sasagawa, K., Tokuda, T., & Ohta, J. (2009) · International SoC Design Conference · doi:10.1109/socdc.2009.5423906 · Cited by 3
A design of infrared finger vein image acquisition terminal
Chen, L., Zuojin, L., Ying, W., & Yi, X. (2011) · International Conference on Business Management and Electronic Information (BMEI) · doi:10.1109/icbmei.2011.5921047 · Cited by 3
An On-Chip Interpolation Based Readout Scheme for Low-Power, High-Speed CMOS Image Sensors
Kaur, A., Mishra, D., & Sarkar, M. (2018) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2018.8589890 · Cited by 3 · cited 1×
“Recently, with such applications like motion detection, mobile-phones, wearable devices and especially in bio-medical applications such as in-vivo imaging sensors [56], a sensor device that enables the multi-functional recording of brain activities such as intrinsic optical signals (IOS), electroencephalogram (EEG), i…”
BackgroundOgden-Grable, H., & Gill, G. W. (2005) · Laboratory Medicine · doi:10.1309/extw-9lbm-0cd7-p9ev · Cited by 1
Research works printed but not linked: 2
- Doshi, 2013, Non-invasive optical sensor for hemoglobin determination, Int. J. Eng. Res. Appl. (Ijera), 3, 559
- Crisan, S., Tarnovan, I., and Crisan, T. (2007, January 19–21). Vein Pattern Recognition. Image Enhancement and feature Extraction Algorithms. Proceedings of the 15th IMEKO TC4 Symposium on Novelties in Electrical Measurements and Instrumentation, Iaşi, Romania.
Cited by
He, X., Liu, Y., Ganesan, K., Ahnood, A., Beckett, P., Eftekhari, F., Smith, D., Uddin, M. H., Skafidas, E. S., Nirmalathas, A. T., et al. (2020) · APL Photonics · doi:10.1063/1.5140215 · Cited by 65
Francisco, M. D., Chen, W., Pan, C.-T., Lin, M., Wen, Z., Liao, C.-F., & Shiue, Y. (2021) · Micromachines · doi:10.3390/mi12040373 · Cited by 53
“These were the veins that were observed during the prototype testing with emphasis on the median cubital vein [9].”
MethodSelf-citationImplementation of Thermal Camera for Non-Contact Physiological Measurement: A Systematic Review
Manullang, M. C. T., Lin, Y., Lai, S.-J., & Chou, N. (2021) · Sensors · doi:10.3390/s21237777 · Cited by 67
“[24] 2019 Focused on vein finder by using near infrared (NIR)”
BackgroundHuang, W., Yoon, S., Wu, B., Lu, K., Lin, C.-M., Yang, H., & Liu, R. (2020) · Chemical Communications · doi:10.1039/d0cc03030a · Cited by 35
Bae, B., Park, M., Lee, D., Sim, I., & Lee, K. (2022) · Advanced Optical Materials · doi:10.1002/adom.202201905 · Cited by 35
Metadata
- Authors
- Pan, C.-T., Francisco, M. D., Yen, C.-K., Wang, S.-Y., & Shiue, Y.-L.
- Affiliations
- National Sun Yat-sen University (TW)
- Department of Mechanical and Electro-Mechanical Engineering, National Sun Yat-sen University (NSYSU), Kaohsiung 80424, Taiwan
- Corresponding author
- Shiue, Yow-Ling
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2019-08-16
- Volume
- 19
- Issue
- 16
- Pages
- 3573
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC6719195
- OpenAlex
- W2967660745
- Semantic Scholar
- e442e7a08b4592a54132c5c81ba54eaa7563059f
- Cited by
- 93
- References
- 67
- Citations by year
- 2026: 10 · 2025: 15 · 2024: 11 · 2023: 16 · 2022: 12 · 2021: 17 · 2020: 12
- FWCI
- 1.54
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Science Council
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Central Venous Catheters and Hemodialysis
- Keywords
- near-infrared imaging, cannulation, phlebotomy, low-cost prototype, vein detection, pre-analytical errors, blood sampling, human factors
- MeSH terms
- Catheterization, Cost-Benefit Analysis, Humans, Semiconductors, Veins, Phlebotomy, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.