Active Magnetic-Field Stabilization with Atomic Magnetometer
Open accessThe DOI matches its Crossref record; 58 of the 58 references it cites were found in a registry or an index.
Abstract
A magnetically-quiet environment is important for detecting faint magnetic-field signals or nonmagnetic spin-dependent interactions. Passive magnetic shielding using layers of large magnetic-permeability materials is widely used to reduce the magnetic-field noise. The magnetic-field noise can also be actively monitored with magnetometers and then compensated, acting as a complementary method to the passive shielding. We present here a general model to quantitatively depict and optimize the performance of active magnetic-field stabilization and experimentally verify our model using optically-pumped atomic magnetometers. We experimentally demonstrate a magnetic-field noise rejection ratio of larger than ∼800 at low frequencies and an environment with a magnetic-field noise floor of ∼40 fT/Hz1/2 in unshielded Earth’s field. The proposed model provides a general guidance on analyzing and improving the performance of active magnetic-field stabilization with magnetometers. This work offers the possibility of sensitive detections of magnetic-field signals in a variety of unshielded natural environments.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 Optimization of the active magnetic-field stabilization system. (a) Amplitude–frequency responses of the feedback controller and the magnetometer. (b) Phase–frequency responses of the feedback controller and the magnetometer. (c) The calculated noise rejection ratios under different parameter settings of the feedback controller. The dashed black lines in (a,b) are the calculated amplitude- and phase- frequency responses of the magnetometer. The solid lines in (a,b) are the calculated amplitude- and phase- frequency responses of −1/[FB(s)×MG(s)]. Different colors of the lines indicate different parameters of the feedback controllers, i.e., orange (blue) for a PID module with a small (big) integration coefficient KI, and green for a PI²D module.](https://figures.cochat.ai/paper-figures/636fb654b51e44e791413fe9287859d8/active-magnetic-field-stabilization-with-atomic-magnetometer-figure-2.jpg)


Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“If we want to realize a high CMRR in a low-frequency band for MCG signals, then the OPMs should have wide bandwidths, which usually benefit the consistency of OPMs (51).”
From A movable unshielded magnetocardiography system (2023) · cites this paper as [51]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Moving magnetoencephalography towards real-world applications with a wearable system
Boto, E., Holmes, N., Leggett, J., Roberts, G., Shah, V., Meyer, S. S., Muñoz, L. D., Mullinger, K. J., Tierney, T. M., Bestmann, S., et al. (2018) · Nature · doi:10.1038/nature26147 · Cited by 1,377
Improved Experimental Limit on the Electric Dipole Moment of the Neutron
Baker, C. A., Doyle, D. D., Geltenbort, P., Green, K., van der Grinten, M. G. D., Harris, P. G., Iaydjiev, P., Ivanov, S. N., May, D. J. R., Pendlebury, J. M., et al. (2006) · Physical Review Letters · doi:10.1103/physrevlett.97.131801 · Cited by 1,311
Cohen, D. (1972) · Science · doi:10.1126/science.175.4022.664 · Cited by 785
Magnetoencephalography with an atomic magnetometer
Xia, H., Ben-Amar Baranga, A., Hoffman, D., & Romalis, M. V. (2006) · Applied Physics Letters · doi:10.1063/1.2392722 · Cited by 441
Optically Driven Spin Precession
Bell, W. E., & Bloom, A. L. (1961) · Physical Review Letters · doi:10.1103/physrevlett.6.280 · Cited by 366
Magnetocardiograms taken inside a shielded room with a superconducting point-contact magnetometer
Cohen, D., Edelsack, E. A., & Zimmerman, J. E. (1970) · Applied Physics Letters · doi:10.1063/1.1653195 · Cited by 361
Developments in alkali -metal atomic magnetometry
Seltzer, S. J. (2008) · Cited by 308
New Limit on Lorentz- andCPT-Violating Neutron Spin Interactions
Brown, J. M., Smullin, S. J., Kornack, T. W., & Romalis, M. V. (2010) · Physical Review Letters · doi:10.1103/physrevlett.105.151604 · Cited by 295
Light Shifts in the Alkali Atoms
Mathur, B. S., Tang, H., & Happer, W. (1968) · Physical Review · doi:10.1103/physrev.171.11 · Cited by 286 · cited 1×
“Such a phenomenon is quite interesting since it proves that, with the active magnetic-field stabilization system, some of the magnetic-like-noise of the magnetometer for field stabilization, for example, the noise originating from the vector light shift [51], are taken as the magnetic-field noise and is actively stabi…”
Background(2013) · doi:10.1017/cbo9780511846380 · Cited by 286
Research works printed but not linked: 3
- Bork, J., Hahlbohm, H., Klein, R., and Schnabel, A. (2000, January 13–17). The 8-layered magnetically shielded room of the PTB: Design and construction. Proceedings of the 12th International Conference on Biomagnetism (Biomag2000), Espoo, Finland.
- Keviczky, L., Bars, R., Hetthéssy, J., and Bányász, C. (2019). Control Engineering, Springer. · DOI
- Keviczky, L., Bars, R., Hetthéssy, J., and Bányász, C. (2019). Proofs and derivations (by chapters). Control Engineering, Springer. · DOI
Cited by
A movable unshielded magnetocardiography system
Xiao, W., Sun, C., Shen, L., Feng, Y., Liu, M., Wu, Y., Liu, X., Wu, T., Peng, X., & Guo, H. (2023) · Science Advances · doi:10.1126/sciadv.adg1746 · Cited by 116
“If we want to realize a high CMRR in a low-frequency band for MCG signals, then the OPMs should have wide bandwidths, which usually benefit the consistency of OPMs ( 51 ).”
Self-citationSensitive magnetometry in challenging environments
Fu, K. C., Iwata, G. Z., Wickenbrock, A., & Budker, D. (2020) · AVS Quantum Science · doi:10.1116/5.0025186 · Cited by 132
Two-photon electromagnetic induction imaging with an atomic magnetometer
Maddox, B., & Renzoni, F. (2023) · Applied Physics Letters · doi:10.1063/5.0147291 · Cited by 19
VCSEL frequency stabilization for optically pumped magnetometers
Yan, Y., Liu, G., Lin, H., Yin, K., Wang, K., & Lu, J. (2021) · Chinese Optics Letters · doi:10.3788/col202119.121407 · Cited by 16
Tian, P., Le, Y., Zhang, H., & Zheng, S. (2025) · IEEE Transactions on Industrial Informatics · doi:10.1109/tii.2025.3538134 · Cited by 14
Metadata
- Authors
- Zhang, R., Ding, Y., Yang, Y., Zheng, Z., Chen, J., Peng, X., Wu, T., & Guo, H.
- Affiliations
- National University of Defense Technology (CN)
- Peking University (CN)
- Corresponding author
- Guo, Hong
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-07-30
- Volume
- 20
- Issue
- 15
- Pages
- 4241
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC7435849
- OpenAlex
- W3046171984
- Cited by
- 32
- References
- 58
- Citations by year
- 2026: 3 · 2025: 6 · 2024: 5 · 2023: 6 · 2022: 6 · 2021: 4 · 2020: 2
- FWCI
- 1.83
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- the National Natural Science Foundation of China (No. 61571018, No. 61531003, No. 91436210)
- the National Science Fund for Distinguished Young Scholars of China (No. 61225003)
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Physical Sciences › Engineering › Biomedical Engineering › Characterization and Applications of Magnetic Nanoparticles
- Keywords
- atomic magnetometer, magnetic noise, optically pumped magnetometers, magnetic shielding, magnetic field compensation, spin-dependent interactions
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Active Magnetic-Field Stabilization with Atomic Magnetometer. https://app.cochat.ai/papers/10.3390/s20154241
- The paper
- Zhang, R., Ding, Y., Yang, Y., Zheng, Z., Chen, J., Peng, X., Wu, T., & Guo, H. (2020). Active Magnetic-Field Stabilization with Atomic Magnetometer. Sensors. https://doi.org/10.3390/s20154241
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.