Development of Magnetocardiograph without Magnetically Shielded Room Using High-Detectivity TMR Sensors

Open access
Kurashima, Koshi, Kataoka, Makoto, Nakano, Takafumi, Fujiwara, Kosuke, Kato, Seiichi, Nakamura, Takenobu, Yuzawa, Masaki, Masuda, Masanori, Ichimura, Kakeru, Okatake, Shigeki, Moriyasu, Yoshitaka, Sugiyama, Kazuhiro, Oogane, Mikihiko, Ando, Yasuo, Kumagai, Seiji, Matsuzaki, Hitoshi, Mochizuki, Hidenori · 2023 · Sensors

The DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index.

Abstract

A magnetocardiograph that enables the clear observation of heart magnetic field mappings without magnetically shielded rooms at room temperatures has been successfully manufactured. Compared to widespread electrocardiographs, magnetocardiographs commonly have a higher spatial resolution, which is expected to lead to early diagnoses of ischemic heart disease and high diagnostic accuracy of ventricular arrhythmia, which involves the risk of sudden death. However, as the conventional superconducting quantum interference device (SQUID) magnetocardiographs require large magnetically shielded rooms and huge running costs to cool the SQUID sensors, magnetocardiography is still unfamiliar technology. Here, in order to achieve the heart field detectivity of 1.0 pT without magnetically shielded rooms and enough magnetocardiography accuracy, we aimed to improve the detectivity of tunneling magnetoresistance (TMR) sensors and to decrease the environmental and sensor noises with a mathematical algorithm. The magnetic detectivity of the TMR sensors was confirmed to be 14.1 pTrms on average in the frequency band between 0.2 and 100 Hz in uncooled states, thanks to the original multilayer structure and the innovative pattern of free layers. By constructing a sensor array using 288 TMR sensors and applying the mathematical magnetic shield technology of signal space separation (SSS), we confirmed that SSS reduces the environmental magnetic noise by −73 dB, which overtakes the general triple magnetically shielded rooms. Moreover, applying digital processing that combined the signal average of heart magnetic fields for one minute and the projection operation, we succeeded in reducing the sensor noise by about −23 dB. The heart magnetic field resolution measured on a subject in a laboratory in an office building was 0.99 pTrms and obtained magnetocardiograms and current arrow maps as clear as the SQUID magnetocardiograph does in the QRS and ST segments. Upon utilizing its superior spatial resolution, this magnetocardiograph has the potential to be an important tool for the early diagnosis of ischemic heart disease and the risk management of sudden death triggered by ventricular arrhythmia.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3390/s23020646

Cites

29 of 29 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure A1 (a–c) Cross-sectional STEM image of TMR multilayers. The stacking structure of (a,b) is Sub. Si, SiO2/Ta 5/Ru 20/Ta 5/free-layer 130/Ru 0.9/Co40Fe40B20 3/MgO/Co40Fe40B20 3/Ru 0.85/Co75Fe25 2/ Ir22Mn78 10/ Ta 5 / Ru 10 in nm. The table shows the configuration of the sputtering system and the material of the free layer. (d,e) Average surface roughness, Ra, dependence of magnetic and electronic noises, respectively. The longitudinal error bars show the difference between the maximum or minimum values and the average value in three samples prepared in the same condition. The transverse error bars show the empirical measurement errors.
Figure A1 (a–c) Cross-sectional STEM image of TMR multilayers. The stacking structure of (a,b) is Sub. Si, SiO2/Ta 5/Ru 20/Ta 5/free-layer 130/Ru 0.9/Co40Fe40B20 3/MgO/Co40Fe40B20 3/Ru 0.85/Co75Fe25 2/ Ir22Mn78 10/ Ta 5 / Ru 10 in nm. The table shows the configuration of the sputtering system and the material of the free layer. (d,e) Average surface roughness, Ra, dependence of magnetic and electronic noises, respectively. The longitudinal error bars show the difference between the maximum or minimum values and the average value in three samples prepared in the same condition. The transverse error bars show the empirical measurement errors.CC BY · PubMed Central
Figure A2 Magnetic domain structure observed in patterned deposited wafer of (a) Sub. Si, SiO2/Ta 5/Ru 20/Ta 5/Co70.5Fe4.5Si15B10 170/Ru 0.4/Co40Fe40B20 3/Ta 2.5 and (b) Sub. Si, SiO2/Ta 5/Ru 20/Ta 5/Co70.5Fe4.5Si15B10 100/Ru 10/Co70.5Fe4.5Si15B10 100/Ru 0.4/Co40Fe40B20 3/Ta 2.5. The pattern sizes are 100 × 140 μm in both. The broken lines on the right-hand diagram show the observed area.
Figure A2 Magnetic domain structure observed in patterned deposited wafer of (a) Sub. Si, SiO2/Ta 5/Ru 20/Ta 5/Co70.5Fe4.5Si15B10 170/Ru 0.4/Co40Fe40B20 3/Ta 2.5 and (b) Sub. Si, SiO2/Ta 5/Ru 20/Ta 5/Co70.5Fe4.5Si15B10 100/Ru 10/Co70.5Fe4.5Si15B10 100/Ru 0.4/Co40Fe40B20 3/Ta 2.5. The pattern sizes are 100 × 140 μm in both. The broken lines on the right-hand diagram show the observed area.CC BY · PubMed Central
Figure 1 (a) Diagram of sensors with MFCs bonded on a print circuit board. The Si spacers and the sensor chip on the board and the second MFCs on the Si spaces are fixed with thermosetting resin. (b) Enlarged diagram of the sensor chip in (a). The wiring between the sensors and electrodes is omitted for simplicity. (c) Enlarged diagram of TMR sensors. (d) Transfer curve of a TMR sensor with the first and second MFCs. (e) Detectivity spectrum of TMR sensors with/without the first and second MFCs measured in an open-loop system with the AC magnetic field of 1 nTp-p at 10 Hz along the sensing axis. (f) Detectivity histogram measured with 108 sensors of Sample-A and 124 sensors of Sample-B. Red and gray data correspond to Sample-A and Sample-B, respectively.
Figure 1 (a) Diagram of sensors with MFCs bonded on a print circuit board. The Si spacers and the sensor chip on the board and the second MFCs on the Si spaces are fixed with thermosetting resin. (b) Enlarged diagram of the sensor chip in (a). The wiring between the sensors and electrodes is omitted for simplicity. (c) Enlarged diagram of TMR sensors. (d) Transfer curve of a TMR sensor with the first and second MFCs. (e) Detectivity spectrum of TMR sensors with/without the first and second MFCs measured in an open-loop system with the AC magnetic field of 1 nTp-p at 10 Hz along the sensing axis. (f) Detectivity histogram measured with 108 sensors of Sample-A and 124 sensors of Sample-B. Red and gray data correspond to Sample-A and Sample-B, respectively.CC BY · PubMed Central
Figure 2 (a) Diagram of measurement using the closed-loop system. The amplified differential output from the bridge circuit with a TMR sensor used as a resistive element was induced to the feedback coil. (b) Photograph of TMR sensor covered by feedback coil. (c) Sensor output and non-linearity in the range of +/−50 μT. Non-linearity was calculated from Vout after AD conversion, which was settled within the error of +/−0.1%.
Figure 2 (a) Diagram of measurement using the closed-loop system. The amplified differential output from the bridge circuit with a TMR sensor used as a resistive element was induced to the feedback coil. (b) Photograph of TMR sensor covered by feedback coil. (c) Sensor output and non-linearity in the range of +/−50 μT. Non-linearity was calculated from Vout after AD conversion, which was settled within the error of +/−0.1%.CC BY · PubMed Central
Figure 3 (a) Diagram of sensor array. (b) Diagram of magnetocardiography measurement. Measured information of three-dimensional magnetic field distribution is stored in the memory via 0.2 Hz high-pass filter, low-noise amplifier, 194 kHz low-pass filter and 24-bit A/D converter. Collected data in memory were sent to a PC through FPGA after being stored for a definite period of time, where digital processing took place. (c) Power spectrum density before and after applying SSS measured with the environmental magnetic field at 7 Hz along the Y-axis.
Figure 3 (a) Diagram of sensor array. (b) Diagram of magnetocardiography measurement. Measured information of three-dimensional magnetic field distribution is stored in the memory via 0.2 Hz high-pass filter, low-noise amplifier, 194 kHz low-pass filter and 24-bit A/D converter. Collected data in memory were sent to a PC through FPGA after being stored for a definite period of time, where digital processing took place. (c) Power spectrum density before and after applying SSS measured with the environmental magnetic field at 7 Hz along the Y-axis.CC BY · PubMed Central
Figure 4 (a) Photograph of magnetocardiography measurement. Cells underneath show the channel number at the corresponding position. (b) Time series of magnetocardiogram in the digital processing measured by Z-sensors in the lower layer in ch27: (i) Raw data (Φ), (ii) After SSS and Digital Filter and (iii) After signal average for one minute (50 heart beats).
Figure 4 (a) Photograph of magnetocardiography measurement. Cells underneath show the channel number at the corresponding position. (b) Time series of magnetocardiogram in the digital processing measured by Z-sensors in the lower layer in ch27: (i) Raw data (Φ), (ii) After SSS and Digital Filter and (iii) After signal average for one minute (50 heart beats).CC BY · PubMed Central
Figure 5 (a) Processing flow of projection operation. (b) Time series of Z-sensors in the lower layer in ch2 measured without the subject to evaluate the change of noises before and after the projection operation. The standard deviation before and after the projection operation was calculated to be 1.77 pTrms and 0.99 pTrms, respectively. (c) Magnetocardiogram map obtained with the Z-sensors in the lower layers in 48 channels after SSS, digital filter, signal average and projection operation.
Figure 5 (a) Processing flow of projection operation. (b) Time series of Z-sensors in the lower layer in ch2 measured without the subject to evaluate the change of noises before and after the projection operation. The standard deviation before and after the projection operation was calculated to be 1.77 pTrms and 0.99 pTrms, respectively. (c) Magnetocardiogram map obtained with the Z-sensors in the lower layers in 48 channels after SSS, digital filter, signal average and projection operation.CC BY · PubMed Central
Figure 6 (a) Top: magnetocardiogram map reconstructed in the area of 200 × 200 mm at 40 mm above the body surface with 25 mm intervals; lower left: reconstructed area; lower right: expanded magnetocardiograph of data surrounded with the red square on the magnetocardiogram map. (b) Magnetic field strength maps in contour lines and current arrow maps at peaks of R and T waves.
Figure 6 (a) Top: magnetocardiogram map reconstructed in the area of 200 × 200 mm at 40 mm above the body surface with 25 mm intervals; lower left: reconstructed area; lower right: expanded magnetocardiograph of data surrounded with the red square on the magnetocardiogram map. (b) Magnetic field strength maps in contour lines and current arrow maps at peaks of R and T waves.CC BY · PubMed Central
Figure 7 (a) Time series of current arrow maps in the QRS domain (300~390 msec in the lower left image in Figure 6a) at each time point referencing the beginning of the QRS wave (300 msec). SSS was performed with Lin = 6 and Lout = 4 using sensors in the lower layer of each channel. (b) Reference of current arrow maps measured by 64 channels of SQUID sensors with magnetically shielded rooms (Kimura et al., Circ. J. 2018).
Figure 7 (a) Time series of current arrow maps in the QRS domain (300~390 msec in the lower left image in Figure 6a) at each time point referencing the beginning of the QRS wave (300 msec). SSS was performed with Lin = 6 and Lout = 4 using sensors in the lower layer of each channel. (b) Reference of current arrow maps measured by 64 channels of SQUID sensors with magnetically shielded rooms (Kimura et al., Circ. J. 2018).CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 29 references

Research works printed but not linked: 3

  1. Zhao, 2019, Comparison betweenmagnetocardiography and myocardial perfusion imaging in the diagnosis of myocardial ischemia in patients with coronary artery disease, Int. J. Clin. Exp. Med., 12, 13127
  2. Wu, 2014, QTc Heterogeneity in Restmagnetocardiography in Sensitive to Detet Coronary Artery Disease: In Comparison with Stress Myocardial Perfusion Imaging, Acta Cardiol. Sin., 30, 445
  3. Hailer, B., and Leeuwen, P.-V. (2000, January 13–17). Prediction of malignant arrhythmias after myocardial infarction on the basis of MCG. Proceedings of the 12th International Conference on Biomagnetism, Espoo, Finland.

Cited by

See all 58

Metadata

Authors
Kurashima, K., Kataoka, M., Nakano, T., Fujiwara, K., Kato, S., Nakamura, T., Yuzawa, M., Masuda, M., Ichimura, K., Okatake, S., Moriyasu, Y., Sugiyama, K., Oogane, M., Ando, Y., Kumagai, S., Matsuzaki, H., & Mochizuki, H.
Affiliations
  • Asahi Kasei (Japan) (JP)
  • Tohoku University (JP)
  • Research Center for Rare Metal and Green Innovation, Tohoku University (JP)
Corresponding author
Kurashima, Koshi
Venue
Sensors
Publisher
MDPI AG
Published
2023-01-06
Volume
23
Issue
2
Pages
646
ISSN
1424-8220 (ISSN-L)
OpenAlex
W4313646746
Cited by
56
References
29
Citations by year
2026: 14 · 2025: 18 · 2024: 16 · 2023: 8
FWCI
4.15
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
  • Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
Keywords
magnetocardiography, tunneling magnetoresistance sensors, sensor arrays, ischemic heart disease diagnosis, ventricular arrhythmias, sudden cardiac death risk
MeSH terms
Arrhythmias, Cardiac, Death, Sudden, Heart, Humans, Myocardial Ischemia, Magnetocardiography

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Development of Magnetocardiograph without Magnetically Shielded Room Using High-Detectivity TMR Sensors. https://app.cochat.ai/papers/10.3390/s23020646
The paper
Kurashima, K., Kataoka, M., Nakano, T., Fujiwara, K., Kato, S., Nakamura, T., Yuzawa, M., Masuda, M., Ichimura, K., Okatake, S., Moriyasu, Y., Sugiyama, K., Oogane, M., Ando, Y., Kumagai, S., Matsuzaki, H., & Mochizuki, H. (2023). Development of Magnetocardiograph without Magnetically Shielded Room Using High-Detectivity TMR Sensors. Sensors. https://doi.org/10.3390/s23020646

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.