Development of Magnetocardiograph without Magnetically Shielded Room Using High-Detectivity TMR Sensors
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index.
Abstract
A magnetocardiograph that enables the clear observation of heart magnetic field mappings without magnetically shielded rooms at room temperatures has been successfully manufactured. Compared to widespread electrocardiographs, magnetocardiographs commonly have a higher spatial resolution, which is expected to lead to early diagnoses of ischemic heart disease and high diagnostic accuracy of ventricular arrhythmia, which involves the risk of sudden death. However, as the conventional superconducting quantum interference device (SQUID) magnetocardiographs require large magnetically shielded rooms and huge running costs to cool the SQUID sensors, magnetocardiography is still unfamiliar technology. Here, in order to achieve the heart field detectivity of 1.0 pT without magnetically shielded rooms and enough magnetocardiography accuracy, we aimed to improve the detectivity of tunneling magnetoresistance (TMR) sensors and to decrease the environmental and sensor noises with a mathematical algorithm. The magnetic detectivity of the TMR sensors was confirmed to be 14.1 pTrms on average in the frequency band between 0.2 and 100 Hz in uncooled states, thanks to the original multilayer structure and the innovative pattern of free layers. By constructing a sensor array using 288 TMR sensors and applying the mathematical magnetic shield technology of signal space separation (SSS), we confirmed that SSS reduces the environmental magnetic noise by −73 dB, which overtakes the general triple magnetically shielded rooms. Moreover, applying digital processing that combined the signal average of heart magnetic fields for one minute and the projection operation, we succeeded in reducing the sensor noise by about −23 dB. The heart magnetic field resolution measured on a subject in a laboratory in an office building was 0.99 pTrms and obtained magnetocardiograms and current arrow maps as clear as the SQUID magnetocardiograph does in the QRS and ST segments. Upon utilizing its superior spatial resolution, this magnetocardiograph has the potential to be an important tool for the early diagnosis of ischemic heart disease and the risk management of sudden death triggered by ventricular arrhythmia.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, very promising results for the development of a low-cost device for unshielded multichannel MCG were recently achieved with an innovative microfabricated tunneling magnetoresistance (TMR) sensor technology featuring 14.1 pTrms sensitivity in the frequency band between 0.2 and 100 Hz combined to digital decrease of the environmental and sensor noises (299). The MCG recorded in an unshield”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [299]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Moving magnetoencephalography towards real-world applications with a wearable system
Boto, E., Holmes, N., Leggett, J., Roberts, G., Shah, V., Meyer, S. S., Muñoz, L. D., Mullinger, K. J., Tierney, T. M., Bestmann, S., et al. (2018) · Nature · doi:10.1038/nature26147 · Cited by 1,377 · cited 1×
“As magnetic sensors are suitable for detecting minute magnetic fields such as heart magnetic fields, optically pumped magnetometers [13,14] and nitrogen-vacancy center magnetometers [15,16] have recently been attracting interest.”
BackgroundDemagnetizing factors for rectangular ferromagnetic prisms
Aharoni, A. (1998) · Journal of Applied Physics · doi:10.1063/1.367113 · Cited by 1,255 · cited 1×
“In general, the diamagnetic field, which occurs with the magnetic field in ferromagnetic materials, is decreased as the material is longer along the field [31].”
BackgroundPresentation of electromagnetic multichannel data: The signal space separation method
Taulu, S., & Kajola, M. (2005) · Journal of Applied Physics · doi:10.1063/1.1935742 · Cited by 390 · cited 2×
“It is known that the sensor errors affect the reduction in environmental noise [27].”
BackgroundMethodClinical application of magnetocardiography
Fenici, R., Brisinda, D., & Meloni, A. M. (2005) · Expert Review of Molecular Diagnostics · doi:10.1586/14737159.5.3.291 · Cited by 174 · cited 1×
“Magnetocardiography, which detects heart magnetic fields outside the body, has a higher spatial resolution than electrocardiography, which measures the electric potential on the body surface produced by heart activities [1].”
BackgroundCritical challenges for picoTesla magnetic-tunnel-junction sensors
Egelhoff, W. F., Pong, P. W. T., Unguris, J., McMichael, R. D., Nowak, E. R., Edelstein, A. S., Burnette, J. E., & Fischer, G. (2009) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2009.08.016 · Cited by 138 · cited 1×
“Moreover, the degrading of the flatness leads to magnetization domain hopping between the metastable ripple states, causing the enhanced magnetic 1/f noise [30].”
BackgroundFujiwara, K., Oogane, M., Kanno, A., Imada, M., Jono, J., Terauchi, T., Okuno, T., Aritomi, Y., Morikawa, M., Tsuchida, M., et al. (2018) · Applied Physics Express · doi:10.7567/apex.11.023001 · Cited by 133 · cited 1×
“5Si15B10 with excellent flatness and a soft magnetic property [23,25,26], which has been reported to lead the small anisotropy field and high sensitivity.”
BackgroundSelf-citationSub-pT magnetic field detection by tunnel magneto-resistive sensors
Oogane, M., Fujiwara, K., Kanno, A., Nakano, T., Wagatsuma, H., Arimoto, T., Mizukami, S., Kumagai, S., Matsuzaki, H., Nakasato, N., et al. (2021) · Applied Physics Express · doi:10.35848/1882-0786/ac3809 · Cited by 110 · cited 1×
“5Si15B10 with excellent flatness and a soft magnetic property [23,25,26], which has been reported to lead the small anisotropy field and high sensitivity.”
BackgroundSelf-citationMagnetocardiography Predicts Coronary Artery Disease in Patients with Acute Chest Pain
Park, J., Hill, P. M., Chung, N., Hugenholtz, P. G., & Jung, F. (2005) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2005.00634.x · Cited by 109
Spatial location of electron trapping defects on silicon by scanning tunneling microscopy
Welland, M. E., & Koch, R. H. (1986) · Applied Physics Letters · doi:10.1063/1.96702 · Cited by 98 · cited 1×
“When the flatness of the tunnel barrier layer in the TMR thin films is worsened, the electronic 1/f noise is increased due to the charge trapping in the tunnel barrier layer and at the interfaces of the tunnel junctions [29].”
BackgroundMillimetre-scale magnetocardiography of living rats with thoracotomy
Arai, K., Kuwahata, A., Nishitani, D., Fujisaki, I., Matsuki, R., Nishio, Y., Xin, Z., Cao, X., Hatano, Y., Onoda, S., et al. (2022) · Communications Physics · doi:10.1038/s42005-022-00978-0 · Cited by 85
Research works printed but not linked: 3
- Zhao, 2019, Comparison betweenmagnetocardiography and myocardial perfusion imaging in the diagnosis of myocardial ischemia in patients with coronary artery disease, Int. J. Clin. Exp. Med., 12, 13127
- Wu, 2014, QTc Heterogeneity in Restmagnetocardiography in Sensitive to Detet Coronary Artery Disease: In Comparison with Stress Myocardial Perfusion Imaging, Acta Cardiol. Sin., 30, 445
- Hailer, B., and Leeuwen, P.-V. (2000, January 13–17). Prediction of malignant arrhythmias after myocardial infarction on the basis of MCG. Proceedings of the 12th International Conference on Biomagnetism, Espoo, Finland.
Cited by
Preprocessing and Denoising Techniques for Electrocardiography and Magnetocardiography: A Review
Jia, Y., Pei, H., Liang, J., Zhou, Y., Yang, Y., Cui, Y., & Xiang, M. (2024) · Bioengineering · doi:10.3390/bioengineering11111109 · Cited by 92
Clinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Spintronic devices for biomedical applications
Mostufa, S., Liang, S. A., Chugh, V. K., Wang, J., & Wu, K. (2024) · npj Spintronics · doi:10.1038/s44306-024-00031-6 · Cited by 68
“A reference design based on the MTJ stack developed by Amaral etal. for magnetic measurements of neuron signals 118 To achievehighdevicesensitivitiesand improvedsensorresilience, the MTJ-based sensors are often con fi gured in a Wheatstone bridge implementation with one variable arm 115,116 .”
Enhanced performance and functionality in spintronic sensors
Leitão, D. C., van Riel, F. J. F., Rasly, M., Araujo, P. D. R., Salvador, M., Paz, E., & Koopmans, B. (2024) · npj Spintronics · doi:10.1038/s44306-024-00058-9 · Cited by 41
“Nevertheless, given the minute magnitude of the fi eld to be detected, fi nal systems used outside controlled laboratory environments must consider other effects such as temperature fl uctuations 117 , environmental noise 118 , and humidity on measurement accuracy.”
Magnetoelectric BAW and SAW Devices: A Review
Luo, B., Velvaluri, P., Liu, Y., & Sun, N. X. (2024) · Micromachines · doi:10.3390/mi15121471 · Cited by 28
Metadata
- Authors
- Kurashima, K., Kataoka, M., Nakano, T., Fujiwara, K., Kato, S., Nakamura, T., Yuzawa, M., Masuda, M., Ichimura, K., Okatake, S., Moriyasu, Y., Sugiyama, K., Oogane, M., Ando, Y., Kumagai, S., Matsuzaki, H., & Mochizuki, H.
- Affiliations
- Asahi Kasei (Japan) (JP)
- Tohoku University (JP)
- Research Center for Rare Metal and Green Innovation, Tohoku University (JP)
- Corresponding author
- Kurashima, Koshi
- ORCID
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2023-01-06
- Volume
- 23
- Issue
- 2
- Pages
- 646
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC9866167
- OpenAlex
- W4313646746
- Cited by
- 56
- References
- 29
- Citations by year
- 2026: 14 · 2025: 18 · 2024: 16 · 2023: 8
- FWCI
- 4.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Keywords
- magnetocardiography, tunneling magnetoresistance sensors, sensor arrays, ischemic heart disease diagnosis, ventricular arrhythmias, sudden cardiac death risk
- MeSH terms
- Arrhythmias, Cardiac, Death, Sudden, Heart, Humans, Myocardial Ischemia, Magnetocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Development of Magnetocardiograph without Magnetically Shielded Room Using High-Detectivity TMR Sensors. https://app.cochat.ai/papers/10.3390/s23020646
- The paper
- Kurashima, K., Kataoka, M., Nakano, T., Fujiwara, K., Kato, S., Nakamura, T., Yuzawa, M., Masuda, M., Ichimura, K., Okatake, S., Moriyasu, Y., Sugiyama, K., Oogane, M., Ando, Y., Kumagai, S., Matsuzaki, H., & Mochizuki, H. (2023). Development of Magnetocardiograph without Magnetically Shielded Room Using High-Detectivity TMR Sensors. Sensors. https://doi.org/10.3390/s23020646
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.