Assessment of the Mechanical Properties of Soft Tissue Phantoms Using Impact Analysis

Bouffandeau, Arthur, Poudrel, Anne-Sophie, Brossier, Chloé, Rosi, Giuseppe, Nguyen, Vu-Hieu, Flouzat-Lachaniette, Charles-Henri, Meningaud, Jean-Paul, Haïat, Guillaume · 2025 · Sensors
Cited by 8

The DOI matches its Crossref record.

Abstract

Skin physiopathological conditions have a strong influence on its biomechanical properties. However, it remains difficult to accurately assess the surface stiffness of soft tissues. The aim of this study was to evaluate the performances of an impact-based analysis method (IBAM) and to compare them with those of an existing digital palpation device, MyotonPro®. The IBAM is based on the impact of an instrumented hammer equipped with a force sensor on a cylindrical punch in contact with agar-based phantoms mimicking soft tissues. The indicator Δt is estimated by analyzing the force signal obtained from the instrumented hammer. Various phantom geometries, stiffnesses and structures (homogeneous and bilayer) were used to estimate the performances of both methods. Measurements show that the IBAM is sensitive to a volume of interest equivalent to a sphere approximately twice the punch diameter. The sensitivity of the IBAM to changes in Young’s modulus is similar to that of dynamic mechanical analysis (DMA) and significantly better compared to MyotonPro. The axial (respectively, lateral) resolution is two (respectively, five) times lower with the IBAM than with MyotonPro. The present study paves the way for the development of a simple, quantitative and non-invasive method to measure skin biomechanical properties.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3390/s25051344

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

5 more claims about this paper

Create a free account to see every claim other papers make about this one.

See all claims — free account

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

No references could be resolved yet.

Research works printed but not linked: 53

  1. Krouskop, 1998, Elastic moduli of breast and prostate tissues under compression, Ultrason. Imaging, 20, 260, 10.1177/016173469802000403 · DOI · Paper page
  2. Shiina, 2015, WFUMB Guidelines and Recommendations for Clinical Use of Ultrasound Elastography: Part 1: Basic Principles and Terminology, Ultrasound Med. Biol., 41, 1126, 10.1016/j.ultrasmedbio.2015.03.009 · DOI · Paper page
  3. Skovoroda, 1995, Quantitative analysis of the mechanical characteristics of pathologically changed soft biological tissues, Biophysics, 40, 1359
  4. Dobrev, 1999, In vivo study of skin mechanical properties in patients with systemic sclerosis, J. Am. Acad. Dermatol., 40, 436, 10.1016/S0190-9622(99)70494-9 · DOI
  5. Gennisson, 2004, Assessment of elastic parameters of human skin using dynamic elastography, IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 51, 980, 10.1109/TUFFC.2004.1324402 · DOI · Paper page
  6. Dobrev, 2005, Application of Cutometer area parameters for the study of human skin fatigue, Skin Res. Technol., 11, 120, 10.1111/j.1600-0846.2005.00090.x · DOI · Paper page
  7. Stroumza, 2015, Intérêt du cutomètre pour l’évaluation de l’efficacité des traitements cutanés en chirurgie plastique et maxillo-faciale, Rev. Stomatol. Chir. Maxillo Faciale Et Chir. Orale, 116, 77, 10.1016/j.revsto.2015.02.002 · DOI · Paper page
  8. Bayat, 2016, Non-invasive objective devices for monitoring the inflammatory, proliferative and remodelling phases of cutaneous wound healing and skin scarring, Exp. Dermatol., 25, 579, 10.1111/exd.13027 · DOI
  9. Gennisson, 2013, Ultrasound elastography: Principles and techniques, Diagn. Interv. Imaging, 94, 487, 10.1016/j.diii.2013.01.022 · DOI · Paper page
  10. Catheline, 2003, Measurement of elastic nonlinearity of soft solid with transient elastography, J. Acoust. Soc. Am., 114, 3087, 10.1121/1.1610457 · DOI
  11. Sandrin, 2003, Transient elastography: A new noninvasive method for assessment of hepatic fibrosis, Ultrasound Med. Biol., 29, 1705, 10.1016/j.ultrasmedbio.2003.07.001 · DOI · Paper page
  12. Ryu, 2008, Influence of age and regional differences on skin elasticity as measured by the Cutometer, Skin Res. Technol., 14, 354, 10.1111/j.1600-0846.2008.00302.x · DOI · Paper page
  13. Dellalana, 2019, Reproducibility of the durometer and myoton devices for skin stiffness measurement in healthy subjects, Skin Res. Technol., 25, 289, 10.1111/srt.12646 · DOI
  14. Lee, Y., Kim, M., and Lee, H. (2021). The Measurement of Stiffness for Major Muscles with Shear Wave Elastography and Myoton: A Quantitative Analysis Study. Diagnostics, 11. · doi:10.3390/diagnostics11030524
  15. Poudrel, A.-S., Bouffandeau, A., Demeet, O.L., Rosi, G., Nguyen, V.-H., and Haiat, G. (2024). Characterization of the concentration of agar-based soft tissue mimicking phantoms by impact analysis. J. Mech. Behav. Biomed. Mater., 152. · DOI · Paper page
  16. Michel, A., Bosc, R., Vayron, R., and Haiat, G. (2015). In vitro evaluation of the acetabular cup primary stability by impact analysis. J. Biomech. Eng., 137. · DOI
  17. Michel, A., Bosc, R., Meningaud, J.-P., Hernigou, P., and Haiat, G. (2016). Assessing the Acetabular Cup Implant Primary Stability by Impact Analyses: A Cadaveric Study. PLoS ONE, 11. · DOI
  18. Michel, 2017, Finite element model of the impaction of a press-fitted acetabular cup, Med. Biol. Eng. Comput., 55, 781, 10.1007/s11517-016-1545-2 · doi:10.1007/s11517-016-1545-2
  19. Albini Lomami, H., Damour, C., Rosi, G., Poudrel, A.-S., Dubory, A., Flouzat-Lachaniette, C.-H., and Haiat, G. (2020). Ex vivo estimation of cementless femoral stem stability using an instrumented hammer. Clin. Biomech., 76. · DOI
  20. Dubory, A., Rosi, G., Tijou, A., Lomami, H.A., Flouzat-Lachaniette, C.-H., and Haïat, G. (2020). A cadaveric validation of a method based on impact analysis to monitor the femoral stem insertion. J. Mech. Behav. Biomed. Mater., 103. · DOI
  21. Tijou, 2018, Monitoring cementless femoral stem insertion by impact analyses: An in vitro study, J. Mech. Behav. Biomed. Mater., 88, 102, 10.1016/j.jmbbm.2018.08.009 · doi:10.1016/j.jmbbm.2018.08.009
  22. Hubert, A., Rosi, G., Bosc, R., and Haiat, G. (2020). Using an Impact Hammer to Estimate Elastic Modulus and Thickness of a Sample During an Osteotomy. J. Biomech. Eng., 142. · DOI
  23. Bas Dit Nugues, M., Rosi, G., Hériveaux, Y., and Haïat, G. (2023). Using an Instrumented Hammer to Predict the Rupture of Bone Samples Subject to an Osteotomy. Sensors, 23. · DOI
  24. Lamassoure, 2021, Anatomical subject validation of an instrumented hammer using machine learning for the classification of osteotomy fracture in rhinoplasty, Med. Eng. Phys., 95, 111, 10.1016/j.medengphy.2021.08.004 · DOI
  25. Lamassoure, 2021, Using an impact hammer to perform biomechanical measurements during osteotomies: Study of an animal model, Proc. Inst. Mech. Eng. H., 235, 838, 10.1177/09544119211011824 · DOI
  26. Manickam, 2014, Characterization of biomechanical properties of agar based tissue mimicking phantoms for ultrasound stiffness imaging techniques, J. Mech. Behav. Biomed. Mater., 35, 132, 10.1016/j.jmbbm.2014.03.017 · doi:10.1016/j.jmbbm.2014.03.017
  27. Manickam, 2014, Study of ultrasound stiffness imaging methods using tissue mimicking phantoms, Ultrasonics, 54, 621, 10.1016/j.ultras.2013.08.018 · DOI
  28. Ismail, H.M., Pretty, C.G., Signal, M.K., Haggers, M., Zhou, C., and Chase, J.G. (2017). Mechanical behaviour of tissue mimicking breast phantom materials. Biomed. Phys. Eng. Express, 3. · DOI
  29. Hall, 1997, Phantom materials for elastography, IEEE Trans. Ultrason. Ferroelectr. Freq. Control., 44, 1355, 10.1109/58.656639 · DOI · Paper page
  30. Oyen, 2014, Mechanical characterisation of hydrogel materials, Int. Mater. Rev., 59, 44, 10.1179/1743280413Y.0000000022 · DOI · Paper page
  31. Nayar, 2012, Elastic and viscoelastic characterization of agar, J. Mech. Behav. Biomed. Mater., 7, 60, 10.1016/j.jmbbm.2011.05.027 · DOI
  32. Ebenstein, 2004, Nanoindentation of soft hydrated materials for application to vascular tissues, J. Biomed. Mater. Res. A, 69, 222, 10.1002/jbm.a.20096 · doi:10.1002/jbm.a.20096
  33. Bizzini, 2003, Reliability of a new, hand-held device for assessing skeletal muscle stiffness, Clin. Biomech., 18, 459, 10.1016/S0268-0033(03)00042-1 · DOI · Paper page
  34. John, 2023, An evaluation of mechanical and biophysical skin parameters at different body locations, Skin Res. Technol., 29, e13292, 10.1111/srt.13292 · DOI · Paper page
  35. Menard, K.P., and Noah, M. (2020). Dynamic Mechanical Analysis, CRC Press. [3rd ed.]. · doi:10.1201/9780429190308
  36. Ewins, D.J. (2000). Modal Testing: Theory, Practice and Application, Research Studies Press Ltd.. [2nd ed.].
  37. Silva, 2000, Cancellation of Mass-Loading Effects of Transducers and Evaluation of Unmeasured Frequency Response Functions, J. Sound Vib., 236, 761, 10.1006/jsvi.1999.2993 · doi:10.1006/jsvi.1999.2993
  38. Cakar, 2005, Elimination of transducer mass loading effects from frequency response functions, Mech. Syst. Signal Process., 19, 87, 10.1016/S0888-3270(03)00086-4 · DOI
  39. Keswick, 1987, A comparison of modal density measurement techniques, Appl. Acoust., 20, 137, 10.1016/0003-682X(87)90059-4 · doi:10.1016/0003-682x(87)90059-4
  40. Rao, S.S. (2019). Vibration of Continuous Systems, John Wiley & Sons. [2nd ed.]. · doi:10.1002/9781119424284
  41. Hamhaber, 2003, Comparison of quantitative shear wave MR-elastography with mechanical compression tests, Magn. Reson. Med., 49, 71, 10.1002/mrm.10343 · DOI
  42. Han, 2015, Quantitative methods for reconstructing tissue biomechanical properties in optical coherence elastography: A comparison study, Phys. Med. Biol., 60, 3531, 10.1088/0031-9155/60/9/3531 · doi:10.1088/0031-9155/60/9/3531
  43. Ringleb, 2005, Quantitative shear wave magnetic resonance elastography: Comparison to a dynamic shear material test, Magn. Reson. Med., 53, 1197, 10.1002/mrm.20449 · DOI
  44. Dougherty, J., Schaefer, E., Nair, K., Kelly, J., and Masi, A. (2013, January 26–29). Repeatability, Reproducibility, and Calibration of the MyotonPro® on Tissue Mimicking Phantoms. American Society of Mechanical Engineers Digital Collection. Proceedings of the ASME 2013 Summer Bioengineering Conference, Sunriver, OR, USA. · doi:10.1115/sbc2013-14622
  45. Kelly, 2018, Characterization of tissue stiffness of the infraspinatus, erector spinae, and gastrocnemius muscle using ultrasound shear wave elastography and superficial mechanical deformation, J. Electromyogr. Kinesiol., 38, 73, 10.1016/j.jelekin.2017.11.001 · DOI
  46. Feng, Y.N., Li, Y.P., Liu, C.L., and Zhang, Z.J. (2018). Assessing the elastic properties of skeletal muscle and tendon using shearwave ultrasound elastography and MyotonPRO. Sci. Rep., 8. · DOI · Paper page
  47. Rosicka, 2021, Comparison of different MyotonPRO probes for skin stiffness evaluation in young women, Skin Res. Technol., 27, 332, 10.1111/srt.12946 · DOI
  48. Junker, 2023, A quantitative comparison of devices for in vivo biomechanical characterization of human Skin, Mech. Soft Mater., 5, 5, 10.1007/s42558-023-00053-w · DOI · Paper page
  49. Vayron, 2014, Assessment of in vitro dental implant primary stability using an ultrasonic method, Ultrasound Med. Biol., 40, 2885, 10.1016/j.ultrasmedbio.2014.03.035 · doi:10.1016/j.ultrasmedbio.2014.03.035
  50. Vayron, 2018, Evaluation of dental implant stability in bone phantoms: Comparison between a quantitative ultrasound technique and resonance frequency analysis, Clin. Implant. Dent. Relat. Res., 20, 470, 10.1111/cid.12622 · doi:10.1111/cid.12622
  51. Bartsch, K., Brandl, A., Weber, P., Wilke, J., Bensamoun, S.F., Bauermeister, W., Klingler, W., and Schleip, R. (2023). Assessing reliability and validity of different stiffness measurement tools on a multi-layered phantom tissue model. Sci. Rep., 13. · DOI · Paper page
  52. Kissin, 2006, Durometry for the assessment of skin disease in systemic sclerosis, Arthritis Rheum., 55, 603, 10.1002/art.22093 · DOI
  53. Koch, V., and Wilke, J. (2022). Reliability of a New Indentometer Device for Measuring Myofascial Tissue Stiffness. J. Clin. Med., 11. · DOI

Cited by

Metadata

Authors
Bouffandeau, A., Poudrel, A.-S., Brossier, C., Rosi, G., Nguyen, V.-H., Flouzat-Lachaniette, C.-H., Meningaud, J.-P., & Haïat, G.
Affiliations
  • Laboratoire Modélisation et Simulation Multi Echelle, Centre National de la Recherche Scientifique, MSME UMR 8208 CNRS, 61 Avenue du Général de Gaulle, 94010 Créteil, France
  • Laboratoire Modélisation et Simulation Multi Echelle, Université Paris Est Créteil, MSME UMR 8208 CNRS, 61 Avenue du Général de Gaulle, 94010 Créteil, France
  • Service de Chirurgie Orthopédique et Traumatologique, Hôpital Henri Mondor AP-HP, Laboratoire IMRB INSERM U955—Université Paris—Est Créteil, 8 rue du Général Sarrail, 94010 Créteil, France
  • Service de Chirurgie Plastique, Reconstructrice, Esthétique et Maxillo-Faciale, Hôpital Henri Mondor AP-HP, Laboratoire IMRB INSERM U955—Université Paris—Est Créteil, 8 rue du Général Sarrail, 94010 Créteil, France
Venue
Sensors
Publisher
MDPI AG
Published
2025-02-22
Volume
25
Issue
5
Pages
1344
ISSN
1424-8220
Cited by
8
References
0
Language
en
Funders
  • European Research Council (101147066, ANR-21-CE19-0035-03, ANR-21-CE17-0004)
  • Agence Nationale de la Recherche (101147066, ANR-21-CE19-0035-03, ANR-21-CE17-0004)
  • CNRS Innovation for the Prematuration project CarImpact (101147066, ANR-21-CE19-0035-03, ANR-21-CE17-0004)

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.