Assessment of the Mechanical Properties of Soft Tissue Phantoms Using Impact Analysis
The DOI matches its Crossref record.
Abstract
Skin physiopathological conditions have a strong influence on its biomechanical properties. However, it remains difficult to accurately assess the surface stiffness of soft tissues. The aim of this study was to evaluate the performances of an impact-based analysis method (IBAM) and to compare them with those of an existing digital palpation device, MyotonPro®. The IBAM is based on the impact of an instrumented hammer equipped with a force sensor on a cylindrical punch in contact with agar-based phantoms mimicking soft tissues. The indicator Δt is estimated by analyzing the force signal obtained from the instrumented hammer. Various phantom geometries, stiffnesses and structures (homogeneous and bilayer) were used to estimate the performances of both methods. Measurements show that the IBAM is sensitive to a volume of interest equivalent to a sphere approximately twice the punch diameter. The sensitivity of the IBAM to changes in Young’s modulus is similar to that of dynamic mechanical analysis (DMA) and significantly better compared to MyotonPro. The axial (respectively, lateral) resolution is two (respectively, five) times lower with the IBAM than with MyotonPro. The present study paves the way for the development of a simple, quantitative and non-invasive method to measure skin biomechanical properties.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Finally, it was proven that the IBAM can accurately predict certain mechanical properties using soft tissue mimicking phantoms (Poudrel et al., 2024; Bouffandeau et al., 2025, 2026).”
From Multimodal assessment of the skin biomechanical properties during wound healing and hyaluronic acid response (2026) · cites this paper as Bouffandeau et al., 2025
In the citing paper
“The stiffness sensitivity of IBAM was shown to be equivalent or better compared to that of Dynamic Mechanical Analysis (DMA) (Bouffandeau et al., 2025), a reference approach for materials characterization, and to that of elastographic techniques (Poudrel et al., 2024; Bouffandeau et al., 2025).”
From Multimodal assessment of the skin biomechanical properties during wound healing and hyaluronic acid response (2026) · cites this paper as Bouffandeau et al., 2025
In the citing paper
“According to the manufacturer (Myoton, 2022), the MyotonPro has a measurement depth of around 20 mm, which has been verified in a previous study using soft tissue phantoms (Bouffandeau et al., 2025).”
From Multimodal assessment of the skin biomechanical properties during wound healing and hyaluronic acid response (2026) · cites this paper as Bouffandeau et al., 2025
In the citing paper
“Based on modeling (Poudrel et al., 2024) and on experimental data (Bouffandeau et al., 2025), previous works have shown that 𝛥𝑡 is roughly inversely proportional to the square root of the tissue Young's modulus (Poudrel et al., 2024; Bouffandeau et al., 2025).”
From Multimodal assessment of the skin biomechanical properties during wound healing and hyaluronic acid response (2026) · cites this paper as Bouffandeau et al., 2025
In the citing paper
“The stiffness sensitivity of IBAM was shown to be equivalent or better compared to that of Dynamic Mechanical Analysis (DMA) (Bouffandeau et al., 2025), a reference approach for materials characterization, and to that of elastographic techniques (Poudrel et al., 2024; Bouffandeau et al., 2025).”
From Multimodal assessment of the skin biomechanical properties during wound healing and hyaluronic acid response (2026) · cites this paper as Bouffandeau et al., 2025
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
No references could be resolved yet.
Research works printed but not linked: 53
- Krouskop, 1998, Elastic moduli of breast and prostate tissues under compression, Ultrason. Imaging, 20, 260, 10.1177/016173469802000403 · DOI · Paper page
- Shiina, 2015, WFUMB Guidelines and Recommendations for Clinical Use of Ultrasound Elastography: Part 1: Basic Principles and Terminology, Ultrasound Med. Biol., 41, 1126, 10.1016/j.ultrasmedbio.2015.03.009 · DOI · Paper page
- Skovoroda, 1995, Quantitative analysis of the mechanical characteristics of pathologically changed soft biological tissues, Biophysics, 40, 1359
- Dobrev, 1999, In vivo study of skin mechanical properties in patients with systemic sclerosis, J. Am. Acad. Dermatol., 40, 436, 10.1016/S0190-9622(99)70494-9 · DOI
- Gennisson, 2004, Assessment of elastic parameters of human skin using dynamic elastography, IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 51, 980, 10.1109/TUFFC.2004.1324402 · DOI · Paper page
- Dobrev, 2005, Application of Cutometer area parameters for the study of human skin fatigue, Skin Res. Technol., 11, 120, 10.1111/j.1600-0846.2005.00090.x · DOI · Paper page
- Stroumza, 2015, Intérêt du cutomètre pour l’évaluation de l’efficacité des traitements cutanés en chirurgie plastique et maxillo-faciale, Rev. Stomatol. Chir. Maxillo Faciale Et Chir. Orale, 116, 77, 10.1016/j.revsto.2015.02.002 · DOI · Paper page
- Bayat, 2016, Non-invasive objective devices for monitoring the inflammatory, proliferative and remodelling phases of cutaneous wound healing and skin scarring, Exp. Dermatol., 25, 579, 10.1111/exd.13027 · DOI
- Gennisson, 2013, Ultrasound elastography: Principles and techniques, Diagn. Interv. Imaging, 94, 487, 10.1016/j.diii.2013.01.022 · DOI · Paper page
- Catheline, 2003, Measurement of elastic nonlinearity of soft solid with transient elastography, J. Acoust. Soc. Am., 114, 3087, 10.1121/1.1610457 · DOI
- Sandrin, 2003, Transient elastography: A new noninvasive method for assessment of hepatic fibrosis, Ultrasound Med. Biol., 29, 1705, 10.1016/j.ultrasmedbio.2003.07.001 · DOI · Paper page
- Ryu, 2008, Influence of age and regional differences on skin elasticity as measured by the Cutometer, Skin Res. Technol., 14, 354, 10.1111/j.1600-0846.2008.00302.x · DOI · Paper page
- Dellalana, 2019, Reproducibility of the durometer and myoton devices for skin stiffness measurement in healthy subjects, Skin Res. Technol., 25, 289, 10.1111/srt.12646 · DOI
- Lee, Y., Kim, M., and Lee, H. (2021). The Measurement of Stiffness for Major Muscles with Shear Wave Elastography and Myoton: A Quantitative Analysis Study. Diagnostics, 11. · doi:10.3390/diagnostics11030524
- Poudrel, A.-S., Bouffandeau, A., Demeet, O.L., Rosi, G., Nguyen, V.-H., and Haiat, G. (2024). Characterization of the concentration of agar-based soft tissue mimicking phantoms by impact analysis. J. Mech. Behav. Biomed. Mater., 152. · DOI · Paper page
- Michel, A., Bosc, R., Vayron, R., and Haiat, G. (2015). In vitro evaluation of the acetabular cup primary stability by impact analysis. J. Biomech. Eng., 137. · DOI
- Michel, A., Bosc, R., Meningaud, J.-P., Hernigou, P., and Haiat, G. (2016). Assessing the Acetabular Cup Implant Primary Stability by Impact Analyses: A Cadaveric Study. PLoS ONE, 11. · DOI
- Michel, 2017, Finite element model of the impaction of a press-fitted acetabular cup, Med. Biol. Eng. Comput., 55, 781, 10.1007/s11517-016-1545-2 · doi:10.1007/s11517-016-1545-2
- Albini Lomami, H., Damour, C., Rosi, G., Poudrel, A.-S., Dubory, A., Flouzat-Lachaniette, C.-H., and Haiat, G. (2020). Ex vivo estimation of cementless femoral stem stability using an instrumented hammer. Clin. Biomech., 76. · DOI
- Dubory, A., Rosi, G., Tijou, A., Lomami, H.A., Flouzat-Lachaniette, C.-H., and Haïat, G. (2020). A cadaveric validation of a method based on impact analysis to monitor the femoral stem insertion. J. Mech. Behav. Biomed. Mater., 103. · DOI
- Tijou, 2018, Monitoring cementless femoral stem insertion by impact analyses: An in vitro study, J. Mech. Behav. Biomed. Mater., 88, 102, 10.1016/j.jmbbm.2018.08.009 · doi:10.1016/j.jmbbm.2018.08.009
- Hubert, A., Rosi, G., Bosc, R., and Haiat, G. (2020). Using an Impact Hammer to Estimate Elastic Modulus and Thickness of a Sample During an Osteotomy. J. Biomech. Eng., 142. · DOI
- Bas Dit Nugues, M., Rosi, G., Hériveaux, Y., and Haïat, G. (2023). Using an Instrumented Hammer to Predict the Rupture of Bone Samples Subject to an Osteotomy. Sensors, 23. · DOI
- Lamassoure, 2021, Anatomical subject validation of an instrumented hammer using machine learning for the classification of osteotomy fracture in rhinoplasty, Med. Eng. Phys., 95, 111, 10.1016/j.medengphy.2021.08.004 · DOI
- Lamassoure, 2021, Using an impact hammer to perform biomechanical measurements during osteotomies: Study of an animal model, Proc. Inst. Mech. Eng. H., 235, 838, 10.1177/09544119211011824 · DOI
- Manickam, 2014, Characterization of biomechanical properties of agar based tissue mimicking phantoms for ultrasound stiffness imaging techniques, J. Mech. Behav. Biomed. Mater., 35, 132, 10.1016/j.jmbbm.2014.03.017 · doi:10.1016/j.jmbbm.2014.03.017
- Manickam, 2014, Study of ultrasound stiffness imaging methods using tissue mimicking phantoms, Ultrasonics, 54, 621, 10.1016/j.ultras.2013.08.018 · DOI
- Ismail, H.M., Pretty, C.G., Signal, M.K., Haggers, M., Zhou, C., and Chase, J.G. (2017). Mechanical behaviour of tissue mimicking breast phantom materials. Biomed. Phys. Eng. Express, 3. · DOI
- Hall, 1997, Phantom materials for elastography, IEEE Trans. Ultrason. Ferroelectr. Freq. Control., 44, 1355, 10.1109/58.656639 · DOI · Paper page
- Oyen, 2014, Mechanical characterisation of hydrogel materials, Int. Mater. Rev., 59, 44, 10.1179/1743280413Y.0000000022 · DOI · Paper page
- Nayar, 2012, Elastic and viscoelastic characterization of agar, J. Mech. Behav. Biomed. Mater., 7, 60, 10.1016/j.jmbbm.2011.05.027 · DOI
- Ebenstein, 2004, Nanoindentation of soft hydrated materials for application to vascular tissues, J. Biomed. Mater. Res. A, 69, 222, 10.1002/jbm.a.20096 · doi:10.1002/jbm.a.20096
- Bizzini, 2003, Reliability of a new, hand-held device for assessing skeletal muscle stiffness, Clin. Biomech., 18, 459, 10.1016/S0268-0033(03)00042-1 · DOI · Paper page
- John, 2023, An evaluation of mechanical and biophysical skin parameters at different body locations, Skin Res. Technol., 29, e13292, 10.1111/srt.13292 · DOI · Paper page
- Menard, K.P., and Noah, M. (2020). Dynamic Mechanical Analysis, CRC Press. [3rd ed.]. · doi:10.1201/9780429190308
- Ewins, D.J. (2000). Modal Testing: Theory, Practice and Application, Research Studies Press Ltd.. [2nd ed.].
- Silva, 2000, Cancellation of Mass-Loading Effects of Transducers and Evaluation of Unmeasured Frequency Response Functions, J. Sound Vib., 236, 761, 10.1006/jsvi.1999.2993 · doi:10.1006/jsvi.1999.2993
- Cakar, 2005, Elimination of transducer mass loading effects from frequency response functions, Mech. Syst. Signal Process., 19, 87, 10.1016/S0888-3270(03)00086-4 · DOI
- Keswick, 1987, A comparison of modal density measurement techniques, Appl. Acoust., 20, 137, 10.1016/0003-682X(87)90059-4 · doi:10.1016/0003-682x(87)90059-4
- Rao, S.S. (2019). Vibration of Continuous Systems, John Wiley & Sons. [2nd ed.]. · doi:10.1002/9781119424284
- Hamhaber, 2003, Comparison of quantitative shear wave MR-elastography with mechanical compression tests, Magn. Reson. Med., 49, 71, 10.1002/mrm.10343 · DOI
- Han, 2015, Quantitative methods for reconstructing tissue biomechanical properties in optical coherence elastography: A comparison study, Phys. Med. Biol., 60, 3531, 10.1088/0031-9155/60/9/3531 · doi:10.1088/0031-9155/60/9/3531
- Ringleb, 2005, Quantitative shear wave magnetic resonance elastography: Comparison to a dynamic shear material test, Magn. Reson. Med., 53, 1197, 10.1002/mrm.20449 · DOI
- Dougherty, J., Schaefer, E., Nair, K., Kelly, J., and Masi, A. (2013, January 26–29). Repeatability, Reproducibility, and Calibration of the MyotonPro® on Tissue Mimicking Phantoms. American Society of Mechanical Engineers Digital Collection. Proceedings of the ASME 2013 Summer Bioengineering Conference, Sunriver, OR, USA. · doi:10.1115/sbc2013-14622
- Kelly, 2018, Characterization of tissue stiffness of the infraspinatus, erector spinae, and gastrocnemius muscle using ultrasound shear wave elastography and superficial mechanical deformation, J. Electromyogr. Kinesiol., 38, 73, 10.1016/j.jelekin.2017.11.001 · DOI
- Feng, Y.N., Li, Y.P., Liu, C.L., and Zhang, Z.J. (2018). Assessing the elastic properties of skeletal muscle and tendon using shearwave ultrasound elastography and MyotonPRO. Sci. Rep., 8. · DOI · Paper page
- Rosicka, 2021, Comparison of different MyotonPRO probes for skin stiffness evaluation in young women, Skin Res. Technol., 27, 332, 10.1111/srt.12946 · DOI
- Junker, 2023, A quantitative comparison of devices for in vivo biomechanical characterization of human Skin, Mech. Soft Mater., 5, 5, 10.1007/s42558-023-00053-w · DOI · Paper page
- Vayron, 2014, Assessment of in vitro dental implant primary stability using an ultrasonic method, Ultrasound Med. Biol., 40, 2885, 10.1016/j.ultrasmedbio.2014.03.035 · doi:10.1016/j.ultrasmedbio.2014.03.035
- Vayron, 2018, Evaluation of dental implant stability in bone phantoms: Comparison between a quantitative ultrasound technique and resonance frequency analysis, Clin. Implant. Dent. Relat. Res., 20, 470, 10.1111/cid.12622 · doi:10.1111/cid.12622
- Bartsch, K., Brandl, A., Weber, P., Wilke, J., Bensamoun, S.F., Bauermeister, W., Klingler, W., and Schleip, R. (2023). Assessing reliability and validity of different stiffness measurement tools on a multi-layered phantom tissue model. Sci. Rep., 13. · DOI · Paper page
- Kissin, 2006, Durometry for the assessment of skin disease in systemic sclerosis, Arthritis Rheum., 55, 603, 10.1002/art.22093 · DOI
- Koch, V., and Wilke, J. (2022). Reliability of a New Indentometer Device for Measuring Myofascial Tissue Stiffness. J. Clin. Med., 11. · DOI
Cited by
Bouffandeau, A., Poisbleau, D., Ouidja, M. O. I., Karasinski, P., Papy, D., Flouzat-Lachaniette, C.-H., Bosc, R., Hersant, B., Meningaud, J.-P., & Haiat, G. (2026) · Journal of the Mechanical Behavior of Biomedical Materials · doi:10.1016/j.jmbbm.2026.107451 · Cited by 2
Self-citationBouffandeau, A., Poisbleau, D., Ouidja, M. O. I., Karasinski, P., Papy, D., Flouzat-Lachaniette, C.-H., Bosc, R., Hersant, B., Meningaud, J.-P., & Haiat, G. (2026) · arXiv · doi:10.48550/arxiv.2611.20949 · Cited by 2
Self-citation
Metadata
- Authors
- Bouffandeau, A., Poudrel, A.-S., Brossier, C., Rosi, G., Nguyen, V.-H., Flouzat-Lachaniette, C.-H., Meningaud, J.-P., & Haïat, G.
- Affiliations
- Laboratoire Modélisation et Simulation Multi Echelle, Centre National de la Recherche Scientifique, MSME UMR 8208 CNRS, 61 Avenue du Général de Gaulle, 94010 Créteil, France
- Laboratoire Modélisation et Simulation Multi Echelle, Université Paris Est Créteil, MSME UMR 8208 CNRS, 61 Avenue du Général de Gaulle, 94010 Créteil, France
- Service de Chirurgie Orthopédique et Traumatologique, Hôpital Henri Mondor AP-HP, Laboratoire IMRB INSERM U955—Université Paris—Est Créteil, 8 rue du Général Sarrail, 94010 Créteil, France
- Service de Chirurgie Plastique, Reconstructrice, Esthétique et Maxillo-Faciale, Hôpital Henri Mondor AP-HP, Laboratoire IMRB INSERM U955—Université Paris—Est Créteil, 8 rue du Général Sarrail, 94010 Créteil, France
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2025-02-22
- Volume
- 25
- Issue
- 5
- Pages
- 1344
- ISSN
- 1424-8220
- Cited by
- 8
- References
- 0
- Language
- en
- Funders
- European Research Council (101147066, ANR-21-CE19-0035-03, ANR-21-CE17-0004)
- Agence Nationale de la Recherche (101147066, ANR-21-CE19-0035-03, ANR-21-CE17-0004)
- CNRS Innovation for the Prematuration project CarImpact (101147066, ANR-21-CE19-0035-03, ANR-21-CE17-0004)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.