Principles of Metered-Dose Inhaler Design
The DOI matches its OpenAlex record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
The pressurized metered-dose inhaler (pMDI) was introduced to deliver asthma medications in a convenient and reliable multi-dose presentation. The key components of the pMDI device (propellants, formulation, metering valve, and actuator) all play roles in the formation of the spray, and in determining drug delivery to the lungs. Hence the opportunity exists to design a pMDI product by adjusting the formulation, metering-valve size, and actuator nozzle diameter in order to obtain the required spray characteristics and fine-particle dose. Breath-actuated pMDIs, breath-coordinated pMDIs, spray-velocity modifiers, and spacer devices may be useful for patients who cannot use a conventional press-and-breathe pMDI correctly. Modern pMDI devices, which contain non-ozone-depleting propellants, should allow inhalation therapy via pMDI to extend well into the 21st century for a variety of treatment indications.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Stratospheric sink for chlorofluoromethanes: chlorine atom-catalysed destruction of ozone
Molina, M. J., & Rowland, F. S. (1974) · Nature · doi:10.1038/249810a0 · Cited by 4,332
Device Selection and Outcomes of Aerosol Therapy: Evidence-Based Guidelines
Dolovich, M. B., Ahrens, R. C., Hess, D. R., Anderson, P., Dhand, R., Rau, J. L., Smaldone, G. C., & Guyatt, G. (2005) · Chest · doi:10.1378/chest.127.1.335 · Cited by 794
Misuse of corticosteroid metered-dose inhaler is associated with decreased asthma stability
Giraud, V., & Roche, N. (2002) · European Respiratory Journal · doi:10.1183/09031936.02.00218402 · Cited by 538 · cited 1×
“39 Another problem with CFC pMDIs and some HFA pMDIs is that even with good inhaler technique they deposit only 10–20% of the dose in the lungs, with most of the dose being deposited in the oropharynx.”
Respiratory Drug Delivery (1989)
Byron, P. R. (2018) · CRC Press eBooks · doi:10.4324/9780203710456 · Cited by 449
Leach, C., Davidson, P., & Boudreau, R. (1998) · European Respiratory Journal · doi:10.1183/09031936.98.12061346 · Cited by 430 · cited 2×
“16,17 The drug-delivery characteristics of standard press-and-breathe pMDIs bear on the possible uses of pMDIs.”
Newman, S. P., Weisz, A. W., Talaee, N., & Clarke, S. W. (1991) · Thorax · doi:10.1136/thx.46.10.712 · Cited by 326 · cited 2×
“38 Misuse of corticosteroid pMDIs is associated with decreased asthma stability, especially when misuse involves poor coordination.”
Self-citationInappropriate inhaler use: assessment of use and patient preference of seven inhalation devices
LENNEY, J., INNES, J. A., & CROMPTON, G. K. (2000) · Respiratory Medicine · doi:10.1053/rmed.1999.0767 · Cited by 266 · cited 1×
“40 Some breath-actuated inhalers are described below, and several others are currently in development.”
Busse, W. W., Brazinsky, S., Jacobson, K., Stricker, W., Schmitt, K., Vanden Burgt, J., Donnell, D., Hannon, S., & Colice, G. L. (1999) · Journal of Allergy and Clinical Immunology · doi:10.1016/s0091-6749(99)70016-3 · Cited by 260 · cited 2×
“16,17 The drug-delivery characteristics of standard press-and-breathe pMDIs bear on the possible uses of pMDIs.”
Smyth, H. D. C. (2003) · Advanced Drug Delivery Reviews · doi:10.1016/s0169-409x(03)00079-6 · Cited by 159 · cited 3×
“The design of the actuator is important, particularly because the aerosol particle size is determined partly by the nozzle diameter, which ranges between 0.14 mm and 0.6 mm. 14 Aerosol particle size varies directly with nozzle diameter, 4,5 and particle size influences lung deposition.”
InfluentialPressurised aerosol deposition in the human lung with and without an "open" spacer device.
Newman, S. P., Clark, A. R., Talaee, N., & Clarke, S. W. (1989) · Thorax · doi:10.1136/thx.44.9.706 · Cited by 153 · cited 1×
“45,53 One tube spacer, Syncroner (Aventis, Bridgwater, New Jersey), had an open section in its upper surface, 54 which served as a coordination aid, because the patient could see the spray emerging from the open section if actuation and inhalation were not correctly coordinated.”
Self-citation
Cited by
Formulation strategy and use of excipients in pulmonary drug delivery
Pilcer, G., & Amighi, K. (2010) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2010.03.017 · Cited by 791
Aerosol drug delivery: developments in device design and clinical use
Dolovich, M. B., & Dhand, R. (2011) · The Lancet · doi:10.1016/s0140-6736(10)60926-9 · Cited by 505
Inhalation drug delivery devices: technology update
Garcia-Contreras, L., Ibrahim, M., & Verma, R. (2015) · Medical Devices: Evidence and Research · doi:10.2147/mder.s48888 · Cited by 306
A Comprehensive Map of FDA-Approved Pharmaceutical Products
Zhong, H., Chan, G., Hu, Y., Hu, H., & Ouyang, D. (2018) · Pharmaceutics · doi:10.3390/pharmaceutics10040263 · Cited by 288
Particle Transport and Deposition: Basic Physics of Particle Kinetics
Tsuda, A., Henry, F. S., & Butler, J. P. (2013) · Comprehensive Physiology · doi:10.1002/cphy.c100085 · Cited by 264
Advances in Metered Dose Inhaler Technology: Formulation Development
Myrdal, P. B., Sheth, P., & Stein, S. W. (2014) · AAPS PharmSciTech · doi:10.1208/s12249-013-0063-x · Cited by 184
“CFCs 11 and 114 mainly functioned to modify the vapor pressure of CFC 12 and to facilitate manufacturing when used in formulations with propellant blends (10).”
MethodLongest, P. W., Tian, G., Walenga, R. L., & Hindle, M. (2012) · Pharmaceutical Research · doi:10.1007/s11095-012-0691-y · Cited by 182
Lung Deposition of BDP/Formoterol HFA pMDI in Healthy Volunteers, Asthmatic, and COPD Patients
De Backer, W., Devolder, A., Poli, G., Acerbi, D., Monno, R., Herpich, C., Sommerer, K., Meyer, T., & Mariotti, F. (2010) · Journal of Aerosol Medicine and Pulmonary Drug Delivery · doi:10.1089/jamp.2009.0772 · Cited by 171
“(8) Inhalation of smaller drug particles leads to increased total lung deposition, farther distal airway penetration, and more peripheral lung deposition, (9) which would be beneficial for asthmatic and COPD patients since the peripheral airways are an important site of inflammation in both diseases.”
Gaur, P. K., Mishra, S., Gupta, V. B., Rathod, M. S., Purohit, S., & Savla, B. A. (2010) · Drug Development and Industrial Pharmacy · doi:10.3109/03639040903410300 · Cited by 24
Metadata
- Authors
- Newman, S. P.
- Venue
- Respiratory Care
- Publisher
- American Association for Respiratory Care
- Published
- 2005-09-01
- Volume
- 50
- Issue
- 9
- Pages
- 1177-1190
- ISSN
- 0020-1324 (ISSN-L)
- OpenAlex
- W2114489922
- Cited by
- 244
- References
- 18
- FWCI
- 3.70
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.