Regularized reconstruction of digital time-of-flight 68Ga-PSMA-11 PET/CT for the detection of recurrent disease in prostate cancer patients

Open access
Lindström, Elin, Velikyan, Irina, Regula, Naresh, Alhuseinalkhudhur, Ali, Sundin, Anders, Sörensen, Jens, Lubberink, Mark·2019·Theranostics

Abstract

Accurate localization of recurrent prostate cancer (PCa) is critical, especially if curative therapy is intended.With the aim to optimize target-to-background uptake ratio in 68 Ga-PSMA-11 PET, we investigated the image quality and quantitative measures of regularized reconstruction by block-sequential regularized expectation maximization (BSREM). Methods:The study encompassed retrospective reconstruction and analysis of 20 digital time-of-flight (TOF) PET/CT examinations acquired 60 min post injection of 2 MBq/kg of 68 Ga-PSMA-11 in PCa patients with biochemical relapse after primary treatment.Reconstruction by ordered-subsets expectation maximization (OSEM; 3 iterations, 16 subsets, 5 mm gaussian postprocessing filter) and BSREM (β-values of 100-1600) were used, both including TOF and point spread function (PSF) recovery.Background variability (BV) was measured by placing a spherical volume of interest in the right liver lobe and defined as the standard deviation divided by the mean standardized uptake value (SUV).The image quality was evaluated in terms of signal-to-noise ratio (SNR) and signal-to-background ratio (SBR), using SUVmax of the lesions.A visual assessment was performed by four observers.Results: OSEM reconstruction produced images with a BV of 15%, whereas BSREM with a β-value above 300 resulted in lower BVs than OSEM (36% with β 100, 8% with β 1300).Decreasing the acquisition duration from 2 to 1 and 0.5 min per bed position increased BV for both reconstruction methods, although BSREM with β-values equal to or higher than 800 and 1200, respectively, kept the BV below 15%.In comparison of BSREM with OSEM, the mean SNR improved by 25 to 66% with an increasing β-value in the range of 200-1300, whereas the mean SBR decreased with an increasing β-value, ranging from 0 to 125% with a β-value of 100 and 900, respectively.Decreased acquisition duration resulted in β-values of 800 to 1000 and 1200 to 1400 for 1 and 0.5 min per bed position, respectively, producing improved image quality measures compared with OSEM at a full acquisition duration of 2 min per bed position.The observer study showed a slight overall preference for BSREM β 900 although the interobserver variability was high.Conclusion: BSREM image reconstruction with β-values in the range of 400-900 resulted in lower BV and similar or improved SNR and SBR in comparison with OSEM.Key words: 68 Ga-PSMA-11, prostate cancer, PET/CT,

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.7150/thno.31970

Cites

17 of 17 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • In low-count conditions, the tendency to prefer smooth, low-noise images with beta values >600 over sharper images could become even more evident.

    “In low-count conditions, this tendency to prefer smooth, low-noise images with beta values >600 over “sharper” images could become even more evident [112,170].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [170]

    The Results (page unavailable) report a slight overall preference for β 900, with high interobserver variability, and show that higher β values reduced noise during shorter acquisitions. The paper does not report observer preferences under those shorter, lower-count conditions, so it does not establish that the preference would become more evident. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Several authors have reported improved SNR and image quality with PL reconstruction for different radiopharmaceuticals, including [18F]FDG, [18F]F-PSMA-1007, [68Ga]Ga-PSMA-11, [68Ga]Ga-DOTATATE, and 89Zr-labelled tracers.

    “several authors have reported improved SNR and image quality with different radiopharmaceuticals, such as [18F]FDG [107,112,113], [18F]F-PSMA-1007 [143], [68Ga]Ga-PSMA-11 [169,170], [68Ga]Ga-DOTATATE [171] or 89Zr-labelled tracers [172].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [170]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 1

  1. [6]Frenzel T Tienken M Abel M Berliner C Klutmann S Beyersdorff D The impact of [68Ga]PSMA I&T PET/CT on radiotherapy planning in patients with prostate cancer Strahlenther Onkol 2018 194 646 54 29572670

Cited by

Sorted by citations

Authors
Lindström, E., Velikyan, I., Regula, N., Alhuseinalkhudhur, A., Sundin, A., Sörensen, J., & Lubberink, M.
Affiliations
Uppsala University (SE)
Corresponding author
Lindström, Elin
Venue
Theranostics
Published
2019
Volume
9
Issue
12
Pages
3476-3484
ISSN
1838-7640 (ISSN-L)
OpenAlex
W2951519500
Cited by
32
References
17
Citations by year
2026: 2 · 2024: 4 · 2023: 6 · 2022: 8 · 2021: 5 · 2020: 6 · 2018: 1
FWCI
3.14
Type
journal article
Language
en
Open access
Licence
cc-by-nc
Locations
  • Theranostics (publishedVersion, cc-by-nc)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
  • PubMed Central (publishedVersion, cc-by-nc)
  • PubMed (publishedVersion)
  • Uppsala University Publications (Uppsala University) (submittedVersion, other-oa)
Funders
  • Cancerfonden (2016/835)
  • Akademiska Sjukhuset
  • Uppsala Universitet
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Prostate Cancer Diagnosis and Treatment
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiopharmaceutical Chemistry and Applications
Keywords
68Ga-PSMA-11, prostate cancer, recurrent disease detection, OSEM, image reconstruction, background variation, signal-to-background ratio, signal-to-noise ratio, acquisition time, point spread function, biochemical recurrence, PSMA PET imaging, SUVmax
MeSH terms
Positron Emission Tomography Computed Tomography, Aged, Gallium Isotopes, Gallium Radioisotopes, Humans, Image Processing, Computer-Assisted, Male, Membrane Glycoproteins, Middle Aged, Neoplasm Recurrence, Local, Organometallic Compounds, Prostatic Neoplasms, Retrospective Studies, Radiopharmaceuticals

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review