Optimization of [18F]PSMA-1007 PET-CT using regularized reconstruction in patients with prostate cancer

Open access

Abstract

Background: Prostate-specific membrane antigen (PSMA) radiotracers such as [18 F]PSMA-1007 used with positron emission tomography-computed tomography (PET-CT) is promising for initial staging and detection of recurrent disease in prostate cancer patients. The block-sequential regularization expectation maximization algorithm (BSREM) is a new PET reconstruction algorithm, which provides higher image contrast while also reducing noise. The aim of the present study was to evaluate the influence of different acquisition times and different noise-suppressing factors in BSREM (β values) in [18 F]PSMA-1007 PET-CT regarding quantitative data as well as a visual image quality assessment. We included 35 patients referred for clinical [18 F]PSMA-1007 PET-CT. Four megabecquerels per kilogramme were administered and imaging was performed after 120 min. Eighty-four image series per patient were created with combinations of acquisition times of 1–4 min/bed position and β values of 300–1400. The noise level in normal tissue and the contrast-to-noise ratio (CNR) of pathological uptakes versus the local background were calculated. Image quality was assessed by experienced nuclear medicine physicians. Results: The noise level in the liver, spleen, and muscle was higher for low β values and low acquisition times (written as activity time products (ATs = administered activity × acquisition time)) and was minimized at maximum AT (16 MBq/kg min) and maximum β (1400). There was only a small decrease above AT 10. The median CNR increased slowly with AT from approximately 6 to 12 and was substantially lower at AT 4 and higher at AT 14–16. At AT 4–6, many images were regarded as being of unacceptable quality. For AT 8, β values of 700–900 were considered of acceptable quality. Conclusions: An AT of 8 (for example as in our study, 4 MB/kg with an acquisition time of 2 min) with a β value of 700 performs well regarding noise level, CNR, and visual image quality assessment.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/s40658-020-00298-8

Cites

20 of 20 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • A low product of injected activity per kilogram and acquisition time per bed position results in excessive image noise and possibly increased false-positive rates and decreased reader confidence.

    “A low product of the two factors results in excessive image noise and possibly increased rates of false-positive results and decreased reader confidence [76,143,144].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [143]

    The Results (page unavailable) show higher image noise and more images judged unacceptable at low activity–time products. The Methods define poor image quality to include noise-induced hotspots that mimic pathological uptake, supporting the possibility of false positives. The paper does not report decreased reader confidence. (LLM, conf=0.88)

    Experimental — accuracy not yet measured
  • Several authors have reported improved SNR and image quality with PL reconstruction for different radiopharmaceuticals, including [18F]FDG, [18F]F-PSMA-1007, [68Ga]Ga-PSMA-11, [68Ga]Ga-DOTATATE, and 89Zr-labelled tracers.

    “several authors have reported improved SNR and image quality with different radiopharmaceuticals, such as [18F]FDG [107,112,113], [18F]F-PSMA-1007 [143], [68Ga]Ga-PSMA-11 [169,170], [68Ga]Ga-DOTATATE [171] or 89Zr-labelled tracers [172].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [143]

    For [18F]PSMA-1007, the Results (Quantitative assessment) report lower image noise with some reconstruction settings, and the Conclusion reports good CNR and visual image quality at the recommended settings. The paper does not report SNR specifically, and its study compares BSREM settings rather than demonstrating an overall improvement over another reconstruction method. The other tracers in the citing sentence have separate citations and are not attributed to this paper. (LLM, conf=0.84)

    Experimental — accuracy not yet measured
  • Trägårdh et al. recommended 8 MBq/kg*min for [18F]F-PSMA-1007 based on the same scanner and PL reconstruction.

    “Moreover, the authors recommended 8 MBq/kg*min for [18F]F-PSMA-1007 based on the same scanner and PL reconstruction [143].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [143]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Metadata

Authors
Trägårdh, E., Minarik, D., Brolin, G., Bitzén, U., Olsson, B., & Oddstig, J.
Affiliations
Lund University (SE)
Corresponding author
Trägårdh, Elin
Venue
EJNMMI Physics
Published
2020-05-12
Volume
7
Issue
1
Pages
31-31
ISSN
2197-7364 (ISSN-L)
OpenAlex
W3027957804
Cited by
27
References
20
Citations by year
2026: 2 · 2025: 2 · 2024: 5 · 2023: 6 · 2022: 7 · 2021: 5
FWCI
2.19
Type
journal article
Language
en
Open access
Locations
  • EJNMMI Physics (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Lund University Publications (Lund University) (submittedVersion)
Funders
  • Knut och Alice Wallenbergs Stiftelse (-)
  • Prostatacancerförbundet (-)
  • Medicinska Fakulteten, Lunds Universitet (-)
  • Region Skåne (-)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiopharmaceutical Chemistry and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
Keywords
18F-PSMA-1007, prostate-specific membrane antigen, prostate cancer, PET/CT, BSREM, PET image reconstruction, acquisition time, noise reduction, contrast-to-noise ratio, image quality

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review