Diagnostic Accuracy of Point-of-Care Tests for Detecting Albuminuria: A Systematic Review and Meta-analysis
The DOI matches its Crossref record, with a small difference in title or year; 53 of the 53 references it cites were found in a registry or an index.
Abstract
Background Experts recommend screening for albuminuria in patients at risk for kidney disease. Purpose To systematically review evidence about the diagnostic accuracy of point-of-care (POC) tests for detecting albuminuria in individuals for whom guidelines recommend such detection. Data sources Cochrane Library, EMBASE, Medion database, MEDLINE, and Science Citation Index from 1963 through 5 December 2013; hand searches of other relevant journals; and reference lists. Study selection Cross-sectional studies, published in any language, that compared the accuracy of machine-read POC tests of urinary albumin-creatinine ratio with that of laboratory measurement. Data extraction Two independent reviewers extracted study data and assessed study quality using the QUADAS-2 (Quality Assessment of Diagnostic Accuracy Studies 2) tool. Data synthesis Sixteen studies (n = 3356 patients) that evaluated semiquantitative or quantitative POC tests and used random urine samples collected in primary or secondary ambulatory care settings met inclusion criteria. Pooling results from a bivariate random-effects model gave sensitivity and specificity estimates of 76% (95% CI, 63% to 86%) and 93% (CI, 84% to 97%), respectively, for the semiquantitative test. Sensitivity and specificity estimates for the quantitative test were 96% (CI, 78% to 99%) and 98% (CI, 93% to 99%), respectively. The negative likelihood ratios for the semiquantitative and quantitative tests were 0.26 (CI, 0.16 to 0.40) and 0.04…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Clinical operators have a lower sensitivity but better specificity than laboratory technologists (500)”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [500]
In the citing paper
“Screening using two tests with either being positive interpreted as a positive (leading to subsequent quantitative testing) increases the sensitivity to 97% but reduces the specificity to 83% (500,501).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [500]
Citations
Cites
Preferred Reporting Items for Systematic Reviews and Meta-Analyses: the PRISMA statement
Moher, D., Liberati, A., Tetzlaff, J., Altman, D. G., & for the PRISMA Group (2009) · BMJ · doi:10.1136/bmj.b2535 · Cited by 83,823
Preferred reporting items for systematic reviews and meta-analyses: the PRISMA Statement.
(2009) · PubMed Central · Cited by 44,859
QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W. S., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M. M. G., Sterne, J. A. C., Bossuyt, P. M. M., & the QUADAS-2 Group* (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-8-201110180-00009 · Cited by 14,199
Standards of medical care in diabetes—2013
American Diabetes Association (2013) · Diabetes Care · doi:10.2337/dc13-s011 · Cited by 4,369
(2013) · Kidney International Supplements · doi:10.1038/kisup.2012.73 · Cited by 3,206
Deeks, J. J., Macaskill, P., & Irwig, L. (2005) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2005.01.016 · Cited by 2,910 · cited 1×
“We explored publication bias by using a regression of the log of the diagnostic odds ratio against the inverse of the square root of the effective sample size, weighted by effective sample size (29).”
MethodFox, C. S., Matsushita, K., Woodward, M., Bilo, H. J., Chalmers, J., Heerspink, H. J. L., Lee, B. J., Perkins, R. M., Rossing, P., Sairenchi, T., et al. (2012) · The Lancet · doi:10.1016/s0140-6736(12)61350-6 · Cited by 1,401
van der Velde, M., the Chronic Kidney Disease Prognosis Consortium, Matsushita, K., Coresh, J., Astor, B. C., Woodward, M., Levey, A. S., de Jong, P. E., & Gansevoort, R. T. (2011) · Kidney International · doi:10.1038/ki.2010.536 · Cited by 1,082
Gansevoort, R. T., Matsushita, K., van der Velde, M., Astor, B. C., Woodward, M., Levey, A. S., de Jong, P. E., & Coresh, J. (2011) · Kidney International · doi:10.1038/ki.2010.531 · Cited by 883
Research works printed but not linked: 5
- Knudsen ST, 2012, Dan Med J, 59, A4502
- Le Floch JP, 2001, Diabetes Metab, 27, 36
- Parsons M, 1999, Clin Nephrol, 51, 220
- 10.1002/(SICI)1098-2825(1998)12:5<280::AID-JCLA5>3.0.CO;2-A · doi:10.1002/(sici)1098-2825(1998)12:5<280::aid-jcla5>3.0.co;2-a
- Poulsen PL, 1992, Diabete Metab, 18, 395
Cited by
2018 ESC/ESH Guidelines for the management of arterial hypertension
Williams, B., Mancia, G., Spiering, W., Agabiti Rosei, E., Azizi, M., Burnier, M., Clement, D. L., Coca, A., de Simone, G., Dominiczak, A., et al. (2018) · European Heart Journal · doi:10.1093/eurheartj/ehy339 · Cited by 10,286
KDIGO 2024 clinical practice guideline for the evaluation and management of chronic kidney disease
Stevens, P. E., Ahmed, S. B., Carrero, J. J., Foster, B., Francis, A., Hall, R. K., Herrington, W. G., Hill, G., Inker, L. A., Kazancıoğlu, R., et al. (2024) · Kidney International · doi:10.1016/j.kint.2023.10.018 · Cited by 8,530
Self-citationWebster, A. C., Nagler, E. V., Morton, R. L., & Masson, P. (2017) · The Lancet · doi:10.1016/s0140-6736(16)32064-5 · Cited by 3,842
Mancia, G., Kreutz, R., Brunström, M., Burnier, M., Grassi, G., Januszewicz, A., Muiesan, M. L., Tsioufis, K., Agabiti-Rosei, E., Algharably, E. A. E., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003480 · Cited by 3,473
Williams, B., Mancia, G., Spiering, W., Agabiti Rosei, E., Azizi, M., Burnier, M., Clement, D. L., Coca, A., de Simone, G., Dominiczak, A., et al. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001940 · Cited by 2,885
Levey, A. S., Becker, C., & Inker, L. A. (2015) · JAMA · doi:10.1001/jama.2015.0602 · Cited by 591
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Florkowski, C., Don-Wauchope, A., Gimenez, N., Rodriguez-Capote, K., Wils, J., & Zemlin, A. (2017) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2017.1399336 · Cited by 189
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 205
Screening, identifying, and treating chronic kidney disease: why, who, when, how, and what?
Farrell, D. R., & Vassalotti, J. A. (2024) · BMC Nephrology · doi:10.1186/s12882-024-03466-5 · Cited by 113
“When studied, semi-quantitative tests have shown reasonable sensitivity, but do not match the quantitative uACR [53, 54].”
BackgroundMadero, M., Levin, A., Ahmed, S. B., Carrero, J. J., Foster, B. J., Francis, A., Hall, R. K., Herrington, W. G., Hill, G., Inker, L. A., et al. (2025) · Annals of Internal Medicine · doi:10.7326/annals-24-01926 · Cited by 90
“For albumin, another systematic review evaluated the diagnostic accuracy of quantitative and semiquantitative protein or albumin urine dipstick tests compared with laboratory tests (38).”
Self-citationScreening for chronic kidney disease in Canadian indigenous peoples is cost-effective
Ferguson, T. W., Tangri, N., Tan, Z., James, M. T., Lavallee, B., Chartrand, C. D., McLeod, L., Dart, A. B., Rigatto, C., & Komenda, P. V. J. (2017) · Kidney International · doi:10.1016/j.kint.2017.02.022 · Cited by 62
Iatridi, F., Carrero, J. J., Gall, É. C.-L., Kanbay, M., Luyckx, V. A., Shroff, R. C., & Ferro, C. J. (2024) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfae209 · Cited by 88
“POCT for urinary albumin-to-creatinine ratio (uACR) may be similarly convenient and would permit more complete CKD staging in appropriately selected individuals [29 ].”
Improving the quality of point-of-care testing
Price, C. P., Smith, I., & den Bruel, A. V. (2017) · Family Practice · doi:10.1093/fampra/cmx120 · Cited by 42
Self-citationMejia, J. R., Fernandez-Chinguel, J. E., Dolores-Maldonado, G., Becerra-Chauca, N., Goicochea-Lugo, S., Herrera-Añazco, P., Zafra-Tanaka, J. H., & Taype-Rondan, A. (2021) · Heliyon · doi:10.1016/j.heliyon.2021.e08253 · Cited by 35
“M ejia et al. H eliyon 7 (2021) e08253 8 Table 3 (continued ) Author (year) Population/ Setting Sample size Age Dipstick that assessed ACR Reference test (ACR) Funding received Risk of bias Brand (Manufacturer) Result Lecture Type of urine specimen Type of urine specimen Patient selection Index test Reference standard…”
BackgroundInfluentialSarafidis, P. A., Schmieder, R. E., Burnier, M., Persu, A., Januszewicz, A., Halimi, J., Arıcı, M., Ortíz, A., Wanner, C., Mancia, G., et al. (2024) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfae041 · Cited by 35
“…also suggested that a negative urinary dipstick test does not rule out A2 albuminuria, as many times it cannot detect ACR levels at the lower range [6 ], but it can offer information on other signs of kidney injury (i.e. microscopic hematuria, active urine sediment) and should be performed at…”
BackgroundFung, A. W. S. (2020) · Practical Laboratory Medicine · doi:10.1016/j.plabm.2020.e00187 · Cited by 40
Aguilar-Ramirez, D. J., Raña-Custodio, A., Villa, A. R., Rubilar, X., Olvera, N., Escobar, A., Johnson, R. J., Sánchez‐Lozada, L. G., Obrador, G. T., & Madero, M. (2020) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfaa041 · Cited by 28
The Risk Threshold for Hemoglobin A1c Associated With Albuminuria: A Population-Based Study in China
Lian, H., Wu, H., Ning, J., Lin, D., Huang, C., Li, F., Liang, Y. J., Qi, Y., Ren, M., Yan, L., et al. (2021) · Frontiers in Endocrinology · doi:10.3389/fendo.2021.673976 · Cited by 26
“The National Institute for Clinical Excellence (UK) and the Standards of Medical Care in Diabetes from ADA both recommend kidney care for all people with type 2 diabetes that includes measuring ACR or albumin levels annually (48, 49).”
BackgroundScreening Strategies for Unrecognized CKD
Komenda, P. V. J., Rigatto, C., & Tangri, N. (2016) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.04190416 · Cited by 20
“Although the seeKD Program did incorporate dipstick urinalysis into its screening algorithm, this semiquantitative test is not sufficient to predict risk or stratify those who will likely benefit from treatment (21).”
MethodAn ontological analysis of medical Bayesian indicators of performance
Barton, A., Éthier, J., Régis Duvauferrier, A., & Burgun, A. (2017) · Journal of Biomedical Semantics · doi:10.1186/s13326-016-0099-4 · Cited by 18
“However, other studies are pooled with this one by McTaggart and colleagues [8] to provide a better estimate of f2(g,IT0,RT0).”
BackgroundMethodInfluentialShore, J., Green, M., Hardy, A., & Livesey, D. (2019) · Expert Review of Pharmacoeconomics & Outcomes Research · doi:10.1080/14737167.2019.1650024 · Cited by 17
Flood, D., García, P., Douglas, K., Hawkins, J., & Rohloff, P. J. (2018) · BMJ Open · doi:10.1136/bmjopen-2017-019778 · Cited by 16
“Low-cost dipsticks to detect albuminuria are in development,(52) 53 but currently available models are insensitive and require use of an analyser.(54) Regarding POC serum creatinine assessment, our institution has had success using POC tests for HbA1c,(25) and reliable POC serum creatinine assays are available.”
MethodÖsterreichischer Blutdruckkonsens 2019
Hypertensiologie, Ö. G. F., Weber, T., Atherosklerosegesellschaft, Ö., Gesellschaft, Ö. D., Angiologie, Ö. G. F. I., Nephrologie, Ö. G. F., Gesellschaft, Ö. K., Neurologie, Ö. G. F., Schlaganfall-Gesellschaft, Ö., Allgemeinmedizin, Ö. G. F., et al. (2019) · Wiener klinische Wochenschrift · doi:10.1007/s00508-019-01565-0 · Cited by 18
Hayashi, Y. (2019) · Diagnostics · doi:10.3390/diagnostics9040133 · Cited by 17
“Moreover, POC tests for semiquantitative UACR can change the care process by enabling faster decision-making and reducing the need for further consultations [52].”
BackgroundYeo, M., Lee, S., Choi, E., Choi, J., Lee, K., Kwon, S., Ahn, H., Kim, B., Han, K., Oh, S., et al. (2025) · Cardiovascular Diabetology · doi:10.1186/s12933-025-02590-2 · Cited by 6
“Second, the urine dipstick test is semiquantitative, and its diagnostic sensitivity is reported to be approximately 76% [27].”
Chronic Kidney Disease in Disadvantaged Populations
Madero, M., & Aguilar-Ramirez, D. J. (2022) · Wiley eBooks · doi:10.1002/9781119105954.ch6 · Cited by 4
Nasuuna, E. M., Kalyesubula, R., Tomlinson, L. A., Castelnuovo, B., Okello, E., Chikwari, C. D., & Weiss, H. A. (2024) · BMJ Open · doi:10.1136/bmjopen-2023-083221 · Cited by 3
“We did not collect data on the smoking status of the participants, but this Open access would not impact on the validity of the POC test rather on who develops kidney disease.”
BackgroundTanintheerakul, C., Peerakam, R., Nanthawong, P., Kittikunnathum, S., & Bunpo, P. (2026) · Diagnostics · doi:10.3390/diagnostics16111678 · Cited by 2
“These findings support the conclusion that quantitative measurements (UACR or UPCR) are superior to dipstick testing for albuminuria screening, particularly for detecting microalbuminuria in its early stages when intervention may be most effective [31,32].”
Critical appraisal in the practice of laboratory medicine
Price, C. P., & Christenson, R. H. (2015) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/0004563215599008 · Cited by 2
“clinical and technical, in relation to the impact on diagnostic accuracy.(25) It is important therefore that the technical performance of the analytical methods is described in detail.”
BackgroundMethodSelf-citationCorrelation between Urinary Angiotensinogen (AGT) and Albuminuria in Chronic Kidney Disease (CKD)
Ali, S. M., Liu, Z., Hassan, D. S., Rexidan, & Aslam, N. (2024) · International Journal of Innovative Science and Research Technology (IJISRT) · doi:10.38124/ijisrt/ijisrt24mar1868 · Cited by 2
Phongphithakchai, A., Wongna, K., Netphakdee, R., Klangbud, W. K., Thanasai, J., Kawakami, F., & Chatatikun, M. (2026) · Medical Sciences · doi:10.3390/medsci14050521 · Cited by 1
Urine Microchip Sensing System
Luo, C., Syu, M., Shiesh, S., Lai, S.-C., Ma, W.-J., Juan, Y.-H., & Juang, W.-H. (2016) · Springer eBooks · doi:10.1007/978-3-319-33201-7_13 · Cited by 1
Cabo, J., & Favresse, J. (2023) · Clinica Chimica Acta · doi:10.1016/j.cca.2023.117534 · Cited by 1
Miloslavsky, D. К., Koval, S. M., & Schenyavska, O. M. (2024) · HYPERTENSION · doi:10.22141/2224-1485.1.2.2024.361
Health Technology Assessment of Center for Innovation and Bio-Design (CIBioD)
Prinja, S. (2025) · Springer eBooks · doi:10.1007/978-981-97-0244-2_4
Mosepele, M., Ponatshego, P., Molebatsi, K., Kebotsamang, K., Williams, C. G., Mokgatlhe, L., Lockman, S., Youssouf, N. F., Gross, R., Jarvis, J. N., et al. (2025) · AIDS Research and Therapy · doi:10.1186/s12981-025-00809-6
“…risk for developing cardiovascular disease (CVD) in people living with HIV (PLWH) is almost 1.5-to 2-fold higher than that in their counterparts without HIV, this increased prevalence of albuminuria is associated with increased morbidity and mortality from cardiovascular and renal disease [12–15].”
InfluentialEller, K., Koller, T., Buchta, C., Exner, M., Greiner, G., Griesmacher, A., Hellbert, K., Karimi, V., Cj, M., Säemann, M., et al. (2026) · Wiener klinische Wochenschrift · doi:10.1007/s00508-026-02749-1
Plengsuriyakarn, T., Chai่jaroenkul, W., Youngvises, N., Sirisabhabhorn, K., Bubpha, W., & Na‐Bangchang, K. (2026) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.70322
Metadata
- Authors
- McTaggart, M. P., Newall, R. G., Hirst, J. A., Bankhead, C. R., Lamb, E. J., Roberts, N. W., & Price, C. P.
- Affiliations
- East Kent Hospitals University NHS Foundation Trust (GB)
- Venue
- Annals of Internal Medicine
- Publisher
- American College of Physicians
- Published
- 2014-04-15
- Volume
- 160
- Issue
- 8
- Pages
- 550-557
- ISSN
- 0003-4819 (ISSN-L)
- 1539-3704
- OpenAlex
- W2016924560
- Semantic Scholar
- ec5af8a2770f8e7ade8c1406879482d328b1b53b
- Cited by
- 39
- References
- 52
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 7 · 2023: 4 · 2022: 1 · 2021: 2 · 2020: 2 · 2019: 3 · 2018: 3 · 2017: 4 · 2016: 3 · 2015: 2
- FWCI
- 0.44
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.acpjournals.org/journal/aim/text-and-data-mining (tdm)
- Locations
- Annals of Internal Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Nephrology › Chronic Kidney Disease and Diabetes
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Nephrology › Acute Kidney Injury Research
- Keywords
- albuminuria, point-of-care testing, urinary albumin-creatinine ratio, diagnostic accuracy, sensitivity and specificity, kidney disease screening, chronic kidney disease, negative likelihood ratio, outpatient care
- MeSH terms
- Albuminuria, Humans, Quality Assurance, Health Care, Sensitivity and Specificity, Likelihood Functions, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.