A Comparative Evaluation of Point-of-Care and Laboratory HbA1c Testing in Diabetes Care: An Indian Perspective

Open access
Rathod, Lokendra, Khan, Sameera, Shubham, Swasti, Bisne, Niranjan, Singh, Sammradhi, Kumar, Manoj, Tiwari, Rajnarayan, Sarma, Devojit K·2024·Cureus

Abstract

AIM: This study aimed to assess the diagnostic performance of a point-of-care (POC) glycated hemoglobin (HbA1c) device in the Indian population against the standard laboratory (high-performance liquid chromatography {HPLC}) method for the effective management of diabetes in India. METHODS: This study on the diagnostic accuracy of a POC HbA1c device involved 121 participants. These participants were categorized into two groups according to their HbA1c levels - one group with HbA1c values below 6.5% and another group with HbA1c values equal to or greater than 6.5%. The HbA1c levels in enrolled participants were estimated using both the POC device (HemoCue HbA1c 501; Ängelholm, Sweden: HemoCue AB) and the standard HPLC-based method. The level of agreement and concordance between the two test results were assessed by the Bland-Altman plot and Lin's concordance correlation coefficient. Sensitivity, specificity, diagnostic accuracy, positive likelihood, and negative likelihood ratios of POC-HbA1c device were assessed with a 6.5% HbA1c cut-off. RESULTS: The mean HbA1c values obtained by the two methods showed no statistically significant difference with a minimum effect size (Cohen's d = 0.035), indicating there was a negligible difference between these methods. The Bland-Altman plot revealed that most values were within acceptable limits (95% CI: -0.5 to 0.7) and Lin's concordance correlation coefficient showed strong agreement (p < 0.0001). The POC-HbA1c device demonstrated an area under the curve (AUC) of 0.991 (95% CI: 0.953-1.000) with sensitivity, specificity, diagnostic accuracy, positive likelihood, and negative likelihood ratios of 93.62%, 97.30%, 95.87%, 34.64 and 0.07, respectively, compared to the standard diagnostic assay. CONCLUSIONS: The diagnostic accuracy, sensitivity, and specificity demonstrated by the POC-HbA1c device to the standard HPLC method offers a viable and practical solution for diabetes management in India. Its ability to provide rapid and reliable results at the point of care can improve patient outcomes, reduce healthcare costs, and enhance access to diabetes care, especially in primary care, remote areas, and resource-limited settings of developing countries like India.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.7759/cureus.69956

Cites

21 of 21 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • The systematic review and meta-analysis included 11 studies involving 1,593 participants and 3,285 paired HbA1c measurements comparing POC devices with reference HPLC-based laboratory methods.

    “A total of 11 studies involving 1,593 participants and 3,285 paired measurements were included.”

    From Comparison of point‑of‑care HbA1c devices with laboratory HPLC methods: A systematic review and meta‑analysis (2026) · cites this paper as [33]

    The Materials and methods section describes a single diagnostic-accuracy study in Bhopal, not a systematic review or meta-analysis. The Results section reports 121 participants whose HbA1c was measured by both POC and HPLC methods, contrary to the attributed totals of 11 studies, 1,593 participants, and 3,285 paired measurements. The indexed passages provide no page numbers. (LLM, conf=0.90)

    Checked by a reasoning model — may be wrong.

    Experimental — accuracy not yet measured
  • Some included studies evaluated POC devices using venous blood instead of capillary fingerstick samples, despite the devices being designed for the latter.

    “From a practical standpoint, some studies evaluated POC devices using venous blood instead of capillary fingerstick samples, despite the devices being designed for the latter (30,33).”

    From Comparison of point‑of‑care HbA1c devices with laboratory HPLC methods: A systematic review and meta‑analysis (2026) · cites this paper as [33]

    The Discussion says this study used venous blood for both the POC and laboratory HbA1c tests, whereas most previous studies used capillary blood for POC testing. It does not establish that the device was designed specifically for capillary fingerstick samples. (LLM, conf=0.84)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Rathod, L., Khan, S., Shubham, S., Bisne, N., Singh, S., Kumar, M., Tiwari, R., & Sarma, D. K.
Affiliations
  • Rajiv Gandhi Technical University (IN)
  • Indian Council of Medical Research (IN)
Corresponding author
Sarma, Devojit K
Venue
Cureus
Published
2024-09-22
Volume
16
Issue
9
Pages
e69956-e69956
ISSN
2168-8184 (ISSN-L)
OpenAlex
W4402723009
Cited by
1
References
21
Citations by year
2025: 1
FWCI
0.33
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • Cureus (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
Keywords
point-of-care testing, HbA1c, diabetes management, HPLC, diagnostic accuracy, measurement agreement, Indian population, diabetes screening, glycemic control monitoring, sensitivity and specificity, likelihood ratios, cost-effectiveness

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review