The intrinsic signal-to-noise ratio in NMR imaging
The DOI matches its Crossref record; 13 of the 13 references it cites were found in a registry or an index.
Abstract
The fundamental limit for NMR imaging is set by an intrinsic signal‐to‐noise ratio (SNR) for a particular combination of rf antenna and imaging subjects. The intrinsic SNR is the signal from a small volume of material in the sample competing with electrical noise from thermally generated, random noise currents in the sample. The intrinsic SNR has been measured for a number of antenna‐body section combinations at several different values of the static magnetic field and is proportional to B 0 . We have applied the intrinsic and system SNR to predict image SNR and have found satisfactory agreement with measurements on images. The relationship between SNR and pixel size is quite different in NMR than it is with imaging modalities using ionizing radiation, and indicates that the initial choice of pixel size is crucial in NMR. The analog of “contrast‐detail‐dose” plots for ionizing radiation imaging modalities is the “contrast‐detail‐time” plot in NMR, which should prove useful in choosing a suitable pixel array to visualize a particular anatomical detail for a given NMR receiving antenna. © 1986 Academic Press, Inc.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Imaging at a higher spatial resolution can decrease partial volume effects but can come at the cost of a reduced MRI signal-to-noise ratio (SNR) (Edelstein et al., 1986).”
From Phasic and sustained fear in humans elicits distinct patterns of brain activity (2011) · cites this paper as Edelstein et al., 1986
Citations
Cites
Bottomley, P. A., Foster, T. H., Argersinger, R. E., & Pfeifer, L. M. (1984) · Medical Physics · doi:10.1118/1.595535 · Cited by 1,192
The Sensitivity Performance of the Human Eye on an Absolute Scale*
Rose, A. (1948) · Journal of the Optical Society of America · doi:10.1364/josa.38.000196 · Cited by 931
Image Reconstruction from Projections
(2004) · Biomedical Engineering · doi:10.1201/9780203492543.ch9 · Cited by 915
Spin warp NMR imaging and applications to human whole-body imaging
Edelstein, W. A., Hutchison, J. M. S., Johnson, G., & Redpath, T. (1980) · Physics in Medicine & Biology · doi:10.1088/0031-9155/25/4/017 · Cited by 803
Self-citationThe sensitivity of the zeugmatographic experiment involving human samples
Hoult, D. I., & Lauterbur, P. C. (1979) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(79)90019-2 · Cited by 776
Kumar, A., Welti, D., & Ernst, R. R. (1975) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(75)90224-3 · Cited by 311
Comparison of linear and circular polarization for magnetic resonance imaging
Glover, G. H., Hayes, C. E., Pelc, N. J., Edelstein, W. A., Mueller, O. M., Hart, H. R., Hardy, C. J., O'Donnell, M., & Barber, W. D. (1985) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(85)90349-x · Cited by 265
Self-citationEdelstein, W. A., Bottomley, P. A., & Pfeifer, L. M. (1984) · Medical Physics · doi:10.1118/1.595484 · Cited by 230
Self-citationSignal, Noise, and Contrast in Nuclear Magnetic Resonance (NMR) Imaging
Edelstein, W. A., Bottomley, P. A., Hart, H. R., & Smith, L. S. (1983) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198306000-00001 · Cited by 201
Self-citationImage Reconstruction from Projections
Norton-Wayne, L. (1980) · Optica Acta: International Journal of Optics · doi:10.1080/713820221 · Cited by 184
Research works printed but not linked: 2
- W. A.Edelstein T. H.Foster andJ. F.Schenck Society for Magnetic Resonance in Medicine Foruth Annual Meeting Abstracts pp.964–965 1985.
- Glover G. H., 1986, J. Magn. Reson., 64, 255
Cited by
Items 701–800 of 858
Submillimeter T2 weighted BOLD fMRI of human visual cortex
Kemper, V. G. (2016) · Research Publications (Maastricht University) · doi:10.26481/dis.20160421vk · Cited by 1
Methodological aspects for improving clinical value of SPECT and MRI
Sipilä, O. (2000) · Aaltodoc (Aalto University) · Cited by 1
Image Analysis for the Life Sciences - Computer-assisted Tumor Diagnostics and Digital Embryomics
Kaster, F. O. (2011) · heiDOK (Heidelberg University) · doi:10.11588/heidok.00011820 · Cited by 1
Yu, Y. (2000) · OhioLink ETD Center (Ohio Library and Information Network) · Cited by 1
Design and implementation of internal MRI coils using ultimate intrinsic
Eryaman, Y. (2006) · Bilkent University Institutional Repository (Bilkent University) · Cited by 1
Instrumentation and Strategies for in Vivo Human Cardiac Phosphorus NMR Spectroscopy
Bottomley, P. A. (1993) · Springer eBooks · doi:10.1007/978-1-4615-3490-7_3 · Cited by 1
van den Boom, M., Miller, K. J., Ramsey, N. F., & Hermes, D. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.12.27.424363 · Cited by 1
Autonomous animal heating and cooling system for temperature-regulated MR experimentsPreprint
Verghese, G., Vöröslakos, M., Marković, S., Tal, A., Dehkharghani, S., Yaghmazadeh, O., & Alon, L. (2023) · arXiv (Cornell University) · doi:10.48550/arxiv.2305.01129 · Cited by 1
Imagerie par Résonance Magnétique à très bas champ pour l'imagerie du cerveau
Lafayette, S. L. (2021) · theses.fr (ABES) · doi:10.70675/54a107b8z2481z4d1dz9fbfz6064f5762bf0 · Cited by 1
Multi-study fMRI outlooks on subcortical BOLD responses in the stop-signal paradigmPreprint
Isherwood, S. J., Kemp, S. A., Miletić, S., Stevenson, N., Bazin, P., & Forstmann, B. U. (2024) · eLife · doi:10.7554/elife.88652.2 · Cited by 1
Multi-study fMRI outlooks on subcortical BOLD responses in the stop-signal paradigmPreprint
Isherwood, S. J., Kemp, S. A., Miletić, S., Stevenson, N., Bazin, P., & Forstmann, B. U. (2025) · eLife · doi:10.7554/elife.88652.3 · Cited by 1
Simulation of fMRI data: a statistical approach
Welvaert, M. (2013) · Ghent University Academic Bibliography (Ghent University) · Cited by 1
Magnetic resonance-guided endovascular catheter tracking
Draper, J. N. (2006) · Libraries and Cultural Resources (University of Calgary) · doi:10.11575/prism/706 · Cited by 1
Liffers, A. (2002) · Dokumentenrepositorium der RUB (Ruhr University Bochum) · Cited by 1
The Measurement of an NMR Signal with Halbach Arrays
Breimer, W. (2019) · Leiden Repository (Leiden University) · Cited by 1
An fmri study of brain functional connectivity in parkinson's disease patients
Evangelisti, S. (2013) · AMS Degree Thesis (University of Bologna) · Cited by 1
Current density impedance imaging
Ma, W. (2004) · TSpace (University of Toronto) · Cited by 1
Towards real-time diffusion imaging : noise correction and inference of the human brain connectivity
Brion, V. (2013) · theses.fr (ABES) · doi:10.70675/8a4ef0c5za6afz45a4zb9f8z0f28124c4141 · Cited by 1
Poujol, J. (2017) · theses.fr (ABES) · doi:10.70675/5214ebe6z9faaz4584zaf73z14a05f794938 · Cited by 1
Rowe, D. B., Jiang, J., & Haacke, E. M. (2011) · Wiley eBooks · doi:10.1002/9780470905203.ch31
The role of early visual areas in human working memory
Xing, Y. (2014) · Nottingham ePrints (University of Nottingham)
Magnetic Resonance with Conducting and High-Dielectric Samples
Hoult, D. I. (1996) · Wiley eBooks · doi:10.1002/9780470034590.emrstm1182
Clinical Multi-nuclear Magnetic Resonance Spectroscopy.
Klomp, D. (2007) · Radboud Repository (Radboud University)
Statistical wavelet analysis of functional images of the brain
Feilner, M. (2002) · Infoscience (Ecole Polytechnique Fédérale de Lausanne) · doi:10.5075/epfl-thesis-2671
Functional neuroimaging of the somatosensory system with Ultra-high-field fMRI and MEG
Wang, F. (2012) · Nottingham ePrints (University of Nottingham)
Besbes, H., Hmani, W., Essafi, F., & Salah, R. B. (2006) · doi:10.1109/ictta.2006.1684537
Development and initial characterization of a nuclear magnetic resonance dosimetry system
Thomasson, D. (1990) · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · doi:10.2172/6240477
Signal - To - Noise Ratio In Fourier Transform NMR Imaging
Yoda, K., Itagaki, H., & Fujimura, S. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1990.690944
Hao, J. (2010) · University of Birmingham Institutional Research Archive (University of Birmingham)
Advanced parallel magnetic resonance imaging methods with applications to MR spectroscopic imaging
Otazo, R. (2008) · UNM’s Digital Repository (University of New Mexico)
In memoriam: William A. Edelstein, 1944–2014
Bottomley, P. A. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25296
MRI at 3 T of brain functions and fibre tracts adjacent to intracranial tumors
van Westen, D. (2006) · Lund University Publications (Lund University)
Indirect detection of myelin water by T2-relaxation during the RF pulse
Al-Abasse, Y. (2015) · Lund University Publications Student Papers (Lund University)
A comparison study of different RF shields for an 8-element transceive small animal array at 9.4T
Jin, J., Li, Y., Liu, F., Ewald, W., & Crozier, . (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6090100
Jiang, S., & Zhang, P. (2014) · International Journal of Imaging Systems and Technology · doi:10.1002/ima.22082
Optimum Voxel Volume for Maximizing Contrast-to-Noise Ratio in BOLD Functional MRI
Fujita, N., Okada, K., Tanaka, H., & Murase, K. (2008) · The Open Magnetic Resonance Journal · doi:10.2174/1874769800801010054
Enhancing the image quality of high‐resolution MR angiography
Sano, K., Koizumi, H., & Takane, A. (1995) · Systems and Computers in Japan · doi:10.1002/scj.4690260207
Observations on the choice of reconstruction matrix in magnetic resonance imaging
YOUNG, I. R., & Hall, A. S. (1991) · Computerized Medical Imaging and Graphics · doi:10.1016/0895-6111(91)90080-f
Statistical properties of spin noise in MRI
Lei, T., Song, H. K., Wehrli, F. W., & Roberts, T. P. L. (2008) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.772145
Multi-functional phantom for MR imaging using surface coils
Ishii, M. (1996) · Oral Radiology · doi:10.1007/bf02351576
A re-evaluation of magnetic resonance imaging damping loops
McRobbie, D. W. (1993) · British Journal of Radiology · doi:10.1259/0007-1285-66-790-934
Sources of signal fluctuations in functional magnetic resonance imaging at 7 Tesla
Jorge, J., Figueiredo, P., van der Zwaag, W., & Marques, J. P. (2011) · IEEE Portuguese Meeting on Bioengineering · doi:10.1109/enbeng.2011.6026092
Bradley, W. G. (2012) · Neuroimaging Clinics of North America · doi:10.1016/j.nic.2012.04.001
FIELD INHOMOGENEITY COMPENSATION IN HIGH FIELD MAGNETIC RESONANCE IMAGING (MRI)
Zhang, Z. (2007) · D-Scholarship@Pitt (University of Pittsburgh)
A bandwidth indicator for magnetic resonance imagers
Date, T., Redpath, T. W., & Bussell, D. M. (1997) · Medical Physics · doi:10.1118/1.597994
Observations of the choice of reconstruction matrix in magnetic resonance imaging
YOUNG, I. R., & Hall, A. S. (2003) · Proceedings - International Conference on Pattern Recognition/Proceedings/International Conference on Pattern Recognition · doi:10.1109/icpr.1988.28468
Shea, S. M., Schirf, B. E., Bi, X., Tang, R. L. Q., Lu, B., Omary, R. A., & Li, D. (2005) · Computers in cardiology · doi:10.1109/cic.2004.1442892
EM Modeling and MR: overview and safety topics
Collins, C. M. (2007) · International Zurich Symposium on Electromagnetic Compatibility (EMC Zurich) · doi:10.1109/emczur.2007.4388285
Ssali, T. (2015) · Scholarship@Western (Western University)
Hyperpolarized noble gas magnetic resonance imaging of human lungs in 150G magnetic field
Zhang, A. X. (2000) · DSpace@MIT (Massachusetts Institute of Technology)
Brain connectivity analysis with ICA
Yang, K. (2005) · DR-NTU (Nanyang Technological University) · doi:10.32657/10356/2546
Artifact Control in Magnetic Resonance Spectroscopic Imaging
Kirchner, T. (2015) · Repository for Publications and Research Data (ETH Zurich) · doi:10.3929/ethz-a-010594058
Methods and tools for the analysis of cerebral activation in fMRI
Aguerre, C., & Braquelaire, A. J. P. (2006) · OpenGrey (Institut de l'Information Scientifique et Technique)
Chung, Y. (2015) · Springer eBooks · doi:10.1007/978-3-319-22141-0_6
Kuperman, V. Y. (2000) · Elsevier eBooks · doi:10.1016/b978-012429150-8/50007-3
Dashner, R. A. (2003) · OhioLink ETD Center (Ohio Library and Information Network)
Calculation Method of NMR Image Signal to Noise Ratio Using Probe Sensitivity
Yoda, K., Itagaki, H., & Fujimura, S. (1988) · IEEJ Transactions on Electronics Information and Systems · doi:10.1541/ieejeiss1987.108.12_989
Design and Construction of a Highly Sensitive Coil for MRI of the Spinal Cord
Foias, A. (2015) · PolyPublie (École Polytechnique de Montréal)
Resting-state Connectivity Dynamics in the Human Brain using High-speed fMRI
Vakamudi, K. (2016) · UNM’s Digital Repository (University of New Mexico)
Investigating reward systems in the human midbrain with fMRI
Limbrick‐Oldfield, E. H. (2012) · Spiral (Imperial College London) · doi:10.25560/14511
Jivan, A. (1998) · Figshare
Ultrahigh-Field Whole-Body MRI for Cartilage Imaging: Technical Challenges
Johnson, C. P., de Moortele, P. V., Wang, L., Uǧurbil, K., & Ellermann, J. (2016) · Royal Society of Chemistry eBooks · doi:10.1039/9781782623663-00671
Holtrop, J. L. (2012) · Illinois Digital Environment for Access to Learning and Scholarship (University of Illinois at Urbana-Champaign)
Molecular MRI using exogenous enzymatic sensors and endogenous chemical exchange contrast
Taylor, A. J. (2016) · Nottingham ePrints (University of Nottingham)
Water and Fat Image Reconstruction in Magnetic Resonance Imaging
Huang, F. (2011) · OhioLink ETD Center (Ohio Library and Information Network)
Kopanoğlu, E. (2012) · Bilkent University Institutional Repository (Bilkent University)
Paar, S. (2016) · Dokumentenrepositorium der RUB (Ruhr University Bochum)
Predictive Gaussian Classification of Functional MRI Data
Yourganov, G. (2014) · TSpace (University of Toronto)
Design and Characterization of a Compact Hybrid FMT-MRI System
Elmer, A. (2017) · Repository for Publications and Research Data (ETH Zurich) · doi:10.3929/ethz-b-000255280
Menezes, L. P. (2018) · Digital Library of Theses and Dissertations (Universidade de São Paulo) · doi:10.11606/d.76.2018.tde-07032018-095408
Neuroimaging functional pain networks in health and disease
Brawn, J. (2017) · Oxford University Research Archive (ORA) (University of Oxford) · doi:10.5287/ora-kejexkeod
TUNABLE NMR RECEIVER FRONT-END USING VARIABLE MEMS CAPACITOR
Ghasemi, M. M., Bahrami, M., & Danishvar, S. (2018) · Biomedical Engineering Applications Basis and Communications · doi:10.4015/s1016237218500369
Ngo, G. (2018) · Illinois Digital Environment for Access to Learning and Scholarship (University of Illinois at Urbana-Champaign)
Linear Input as a Fundamental Motif of Brain Connectivity OrganizationPreprint
O’Rawe, J. F., & Leung, H. (2019) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/549345
Advanced Encoding Concepts for Magnetic Resonance Elastography
Guenthner, C. (2019) · Repository for Publications and Research Data (ETH Zurich) · doi:10.3929/ethz-b-000337639
Imaging Diagnosis of SLAP Tears and Microinstability
Krepkin, K., Tuite, M. J., & Bencardino, J. T. (2019) · Springer eBooks · doi:10.1007/978-3-030-06240-8_8
Optimal Sampling & Reconstruction: Theory and ApplicationsPreprint
Haldar, J. P. (2019) · arXiv (Cornell University) · doi:10.48550/arxiv.1911.09595
Imaging of Epidural Spinal Cord Compression
Shah, L. M., Gordon, J. P., Phillips, C. D., & Schiff, D. (2008) · Elsevier eBooks · doi:10.1016/b978-012370863-2.50054-3
SNR gains with superconducting coils and hyperpolarized inert gas
Black, R. D. (1995) · Proceedings annual meeting Electron Microscopy Society of America · doi:10.1017/s0424820100136805
Novel techniques and innovative designs for the RF chain of magnetic resonance imaging scanners
Sadeghi‐Tarakameh, A. (2020) · Bilkent University Institutional Repository (Bilkent University)
A Multinuclear Low Field Magnetic Resonance Minitomograph
Frolov, V. V., Tyutyukin, K. V., Shubin, S. A., Lavrov, S. A., & Bogachev, Y. V. (2020) · Instruments and Experimental Techniques · doi:10.1134/s0020441220050127
Sinha, D. (2020) · Scientific Reports · doi:10.1038/s41598-020-77206-8
Bratch, A. (2020) · University of Minnesota Digital Conservancy (University of Minnesota)
Forced Current Excitation in Selectable Field of View Coils for 7T MRI and MRS
Cui, J. (2017) · OakTrust (Texas A&M University Libraries)
Passive high-temperature superconducting microwave devices
(2007) · Springer eBooks · doi:10.1007/bfb0111190
Technology for Ultrahigh Field Imaging
Uǧurbil, K. (2022) · Wiley eBooks · doi:10.1002/9783527827244.ch4
Prost, R. W. (2002) · Wiley eBooks · doi:10.1002/0471443395.img114
New Developments in Breast Magnetic Resonance Imaging
(2007) · Springer eBooks · doi:10.1007/978-0-387-73507-8_13
Lei, T. (2011) · Chapman & Hall/CRC machine learning & pattern recognition/Chapman & Hall/CRC machine learning & pattern recognition series · doi:10.1201/b11413-8
Hutton, C., Balteau, E., Lutti, A., Josephs, O., & Weiskopf, N. (2013) · PLoS ONE · doi:10.1371/annotation/ad7981aa-87ee-458a-ae72-00972b3af933
Diseño de un sistema de bobinas de RF para aplicaciones de imagen de resonancia magnética en flúor
Valverde, P. V. (2023) · UPM Digital Archive (Technical University of Madrid) · doi:10.20868/upm.thesis.72581
Autonomous animal heating and cooling system for temperature-regulated MR experiments.Preprint
Verghese, G., Vöröslakos, M., Marković, S., Tal, A., Dehkharghani, S., Yaghmazadeh, O., & Alon, L. (2023) · PubMed Central
Polzehl, J., & Tabelow, K. (2023) · Use R! · doi:10.1007/978-3-031-38949-8_5
Ultrahigh field FMRI - principles and applications
Uǧurbil, K. (2024) · Elsevier eBooks · doi:10.1016/b978-0-12-820480-1.00183-2
Author response: Multi-study fMRI outlooks on subcortical BOLD responses in the stop-signal paradigm
Isherwood, S. J., Kemp, S. A., Miletić, S., Stevenson, N., Bazin, P., & Forstmann, B. U. (2025) · eLife · doi:10.7554/elife.88652.4.sa2
Automatic SNR measurement of brain MR images using a deep learning-based approach
Kojima, S., Higuchi, S., Hayashi, T., Kariyasu, T., Nishikawa, M., Yamaguchi, H., & Machida, H. (2025) · Acta Radiologica Open · doi:10.1177/20584601251387564
Metadata
- Authors
- Edelstein, W. A., Glover, G. H., Hardy, C. J., & Redington, R. W.
- Affiliations
- General Electric (United States) (US)
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 1986-08
- Volume
- 3
- Issue
- 4
- Pages
- 604-618
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W2032038095
- Cited by
- 758
- References
- 13
- Citations by year
- 2026: 16 · 2025: 19 · 2024: 21 · 2023: 25 · 2022: 22 · 2021: 18 · 2020: 35 · 2019: 25 · 2018: 22 · 2017: 25 · 2016: 35 · 2015: 25 · 2014: 31 · 2013: 20 · 2012: 27
- FWCI
- 6.33
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Magnetic Resonance in Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Chemistry › Spectroscopy › Advanced NMR Techniques and Applications
- Physical Sciences › Chemistry › Spectroscopy › NMR spectroscopy and applications
- Keywords
- magnetic resonance imaging, thermal noise, receiving antenna, pixel size, static magnetic field dependence
- MeSH terms
- Abdomen, Biometry, Biophysics, Head, Humans, Magnetic Resonance Spectroscopy, Pelvis, Biophysical Phenomena
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.