MR spectroscopy of the human brain with enhanced signal intensity at ultrashort echo times on a clinical platform at 3T and 7T
Open accessAbstract
Recently, the spin‐echo full‐intensity acquired localized (SPECIAL) spectroscopy technique was proposed to unite the advantages of short TEs on the order of milliseconds (ms) with full sensitivity and applied to in vivo rat brain. In the present study, SPECIAL was adapted and optimized for use on a clinical platform at 3T and 7T by combining interleaved water suppression (WS) and outer volume saturation (OVS), optimized sequence timing, and improved shimming using FASTMAP. High‐quality single voxel spectra of human brain were acquired at TEs below or equal to 6 ms on a clinical 3T and 7T system for six volunteers. Narrow linewidths (6.6 ± 0.6 Hz at 3T and 12.1 ± 1.0 Hz at 7T for water) and the high signal‐to‐noise ratio (SNR) of the artifact‐free spectra enabled the quantification of a neurochemical profile consisting of 18 metabolites with Cramér‐Rao lower bounds (CRLBs) below 20% at both field strengths. The enhanced sensitivity and increased spectral resolution at 7T compared to 3T allowed a two‐fold reduction in scan time, an increased precision of quantification for 12 metabolites, and the additional quantification of lactate with CRLB below 20%. Improved sensitivity at 7T was also demonstrated by a 1.7‐fold increase in average SNR (= peak height/root mean square [RMS]‐of‐noise) per unit‐time. Magn Reson Med, 2009. © 2009 Wiley‐Liss, Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Estimation of metabolite concentrations from localized in vivo proton NMR spectra
Provencher, S. W. (1993) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910300604 · Cited by 3,655
Proton NMR chemical shifts and coupling constants for brain metabolites
Govindaraju, V., Young, K., & Maudsley, A. A. (2000) · NMR in Biomedicine · doi:10.1002/1099-1492(200005)13:3<129::aid-nbm619>3.0.co;2-v · Cited by 1,689
Spatial Localization in NMR Spectroscopy in Vivo
BOTTOMLEY, P. A. (1987) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1987.tb32915.x · Cited by 1,378
In vivo1H NMR spectroscopy of rat brain at 1 ms echo time
Tk�?, I., Star?uk, Z., Choi, I.-Y., & Gruetter, R. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199904)41:4<649::aid-mrm2>3.0.co;2-g · Cited by 1,005
Self-citationAutomatic, localized in Vivo adjustment of all first‐and second‐order shim coils
Gruetter, R. (1993) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910290613 · Cited by 870
Self-citationLocalized proton spectroscopy using stimulated echoes
Frahm, J., Merboldt, K.-D., & Hänicke, W. (1987) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(87)90154-5 · Cited by 826
Image-selected in Vivo spectroscopy (ISIS). A new technique for spatially selective nmr spectroscopy
Ordidge, R. J., Connelly, A., & Lohman, J. A. B. (1986) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(86)90031-4 · Cited by 811
Field mapping without reference scan using asymmetric echo-planar techniques
Gruetter, R., & Tk�?, I. (2000) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(200002)43:2<319::aid-mrm22>3.0.co;2-1 · Cited by 569
Self-citationSchizophrenia: glutathione deficit in cerebrospinal fluid and prefrontal cortex in vivo
Do, K. Q., Trabesinger, A. H., Kirsten‐Krüger, M., Lauer, C. J., Dydak, U., Hell, D., Holsboer, F., Boesiger, P., & Cuénod, M. (2000) · European Journal of Neuroscience · doi:10.1046/j.1460-9568.2000.00229.x · Cited by 540
Analysis of macromolecule resonances in 1H NMR spectra of human brain
Behar, K. L., Rothman, D. L., Spencer, D. D., & Petroff, O. A. C. (1994) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910320304 · Cited by 467
In vivo 1 H NMR spectroscopy of the human brain at 7 T
Tkáč, I., Andersen, P., Adriany, G., Merkle, H., Uǧurbil, K., & Gruetter, R. (2001) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1213 · Cited by 394
High frequency volume coils for clinical NMR imaging and spectroscopy
Vaughan, J. T., Hetherington, H. P., Otu, J. O., Pan, J. W., & Pohost, G. M. (1994) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910320209 · Cited by 374
Localized short‐echo‐time proton MR spectroscopy with full signal‐intensity acquisition
Mlynárik, V., Gambarota, G., Frenkel, H., & Gruetter, R. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21043 · Cited by 274
Self-citationResolution Improvements inin Vivo1H NMR Spectra with Increased Magnetic Field Strength
Gruetter, R., Weisdorf, S. A., Rajanayagan, V., Terpstra, M., Merkle, H., Truwit, C. L., Garwood, M., Nyberg, S. L., & Ugurbil, K. (1998) · Journal of Magnetic Resonance · doi:10.1006/jmre.1998.1542 · Cited by 195
Self-citationMethodology of1H NMR spectroscopy of the human brain at very high magnetic fields
Tkáć, I., & Gruetter, R. (2005) · Applied Magnetic Resonance · doi:10.1007/bf03166960 · Cited by 184
Self-citationProFit: two‐dimensional prior‐knowledge fitting of J‐resolved spectra
Schulte, R. F., & Boesiger, P. (2006) · NMR in Biomedicine · doi:10.1002/nbm.1026 · Cited by 175
A Half-Volume Coil for Efficient Proton Decoupling in Humans at 4 Tesla
Adriany, G., & Gruetter, R. (1997) · Journal of Magnetic Resonance · doi:10.1006/jmre.1997.1113 · Cited by 169
Self-citationIn vivo proton MR spectroscopy of the human brain
Barker, P. B., & Lin, D. D. M. (2006) · Progress in Nuclear Magnetic Resonance Spectroscopy · doi:10.1016/j.pnmrs.2006.06.002 · Cited by 147
Mangia, S., Tkáč, I., Rolf Gruetter, A., de Moortele, P. V., Giove, F., Maraviglia, B., & Uǧurbil, K. (2006) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2005.12.023 · Cited by 128
Proton echo‐planar spectroscopic imaging of J ‐coupled resonances in human brain at 3 and 4 Tesla
Posse, S., Otazo, R., Caprihan, A., Bustillo, J., Chen, H., Henry, P., Marjańska, M., Gasparovic, C., Zuo, C. S., Magnotta, V. A., et al. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21287 · Cited by 123
Proton Spectroscopy of Human Brain with Very Short Echo Time Using High Gradient Amplitudes
Seeger, U., Klose, U., Seitz, D., Nägele, T., Lutz, O., & Grodd, W. (1998) · Magnetic Resonance Imaging · doi:10.1016/s0730-725x(97)00239-7 · Cited by 30
Cited by
Clinical Proton MR Spectroscopy in Central Nervous System Disorders
Öz, G., Alger, J. R., Barker, P. B., Bartha, R., Bizzi, A., Boesch, C., Bolan, P. J., Brindle, K. M., Cudalbu, C., Dinçer, A., et al. (2014) · Radiology · doi:10.1148/radiol.13130531 · Cited by 660
Self-citationCurrent practice in the use of MEGA-PRESS spectroscopy for the detection of GABA
Mullins, P. G., McGonigle, D. J., O'Gorman, R. L., Puts, N. A. J., Vidyasagar, R., Evans, C. J., & Edden, R. A. E. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2012.12.004 · Cited by 553
Stagg, C. J., Bestmann, S., Constantinescu, A. O., Moreno Moreno, L., Allman, C., Mekle, R., Woolrich, M., Near, J., Johansen‐Berg, H., & Rothwell, J. C. (2011) · The Journal of Physiology · doi:10.1113/jphysiol.2011.216978 · Cited by 401
Self-citationSimpson, R., Devenyi, G. A., Jezzard, P., Hennessy, T. J., & Near, J. (2017) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26091 · Cited by 368
Near, J., Harris, A. D., Juchem, C., Kreis, R., Marjańska, M., Öz, G., Slotboom, J., Wilson, M., & Gasparovic, C. (2021) · NMR in Biomedicine · doi:10.1002/nbm.4257 · Cited by 342
Metadata
- Authors
- Mekle, R., Mlynárik, V., Gambarota, G., Hergt, M., Krueger, G., & Gruetter, R.
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2009-03-24
- Volume
- 61
- Issue
- 6
- Pages
- 1279-1285
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W2127421809
- Cited by
- 324
- References
- 21
- FWCI
- 12.68
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.