A multi spin echo pulse sequence with optimized excitation pulses and a 3D cone readout for hyperpolarized 13C imaging
Abstract
Purpose: Imaging tumor metabolism in vivo using hyperpolarized [1‐13 C]pyruvate is a promising technique for detecting disease, monitoring disease progression, and assessing treatment response. However, the transient nature of the hyperpolarization and its depletion following excitation limits the available time for imaging. We describe here a single‐shot multi spin echo sequence, which improves on previously reported sequences, with a shorter readout time, isotropic point spread function (PSF), and better signal‐to‐noise ratio. Methods: The sequence uses numerically optimized spectrally selective excitation pulses set to the resonant frequencies of pyruvate and lactate and a hyperbolic secant adiabatic refocusing pulse, all applied in the absence of slice selection gradients. The excitation pulses were designed to be resistant to the effects of B 0 and B 1 field inhomogeneity. The gradient readout uses a 3D cone trajectory composed of 13 cones, all fully refocused and distributed among 7 spin echoes. The maximal gradient amplitude and slew rate were set to 4 G/cm and 20 G/cm/ms, respectively, to demonstrate the feasibility of clinical translation. Results: The pulse sequence gave an isotropic PSF of 2.8 mm. The excitation profiles of the optimized pulses closely matched simulations and a 46.10 ± 0.04% gain in image SNR was observed compared to a conventional Shinnar–Le Roux excitation pulse. The sequence was demonstrated with dynamic imaging of hyperpolarized [1‐13 C]pyruvate and [1‐13 C]lactate in vivo. Conclusion: The pulse sequence was capable of dynamic imaging of hyperpolarized 13 C labeled metabolites in vivo with relatively high spatial and temporal resolution and immunity to system imperfections.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Although advanced computational techniques have been instrumental for designing compensated RF pulses such as composite, adiabatic, and numerically optimized pulses (6, 12–18, 19, 20–29), high- and ultra-high–field NMR and MRI spectroscopy require RF pulses with larger bandwidths, higher fidelity, and compensation for instrumental inhomogeneities.”
From High-fidelity control of spin ensemble dynamics via artificial intelligence: from quantum computing to NMR spectroscopy and imaging (2022) · cites this paper as [20]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
No references could be resolved yet.
Research works printed but not linked: 35
- 10.1073/pnas.1733835100 · DOI · Paper page
- 10.1021/jacs.5b03300 · DOI · Paper page
- 10.1259/bjr/81120511 · DOI
- 10.1038/nm1650 · DOI · Paper page
- 10.1002/mrm.22999 · DOI · Paper page
- 10.1158/0008-5472.CAN-08-0749 · DOI · Paper page
- 10.1158/1078-0432.CCR-16-0159 · DOI
- 10.1038/nature12627 · DOI · Paper page
- 10.1002/mrm.20164 · doi:10.1002/mrm.20164
- 10.1148/radiol.2019182391 · DOI · Paper page
- 10.1158/0008-5472.CAN-06-2564 · DOI · Paper page
- 10.1002/mrm.21025 · DOI
- 10.1016/j.jmr.2008.06.010 · DOI · Paper page
- 10.1002/mrm.22525 · doi:10.1002/mrm.22525
- 10.1002/mrm.27136 · DOI
- 10.1002/mrm.25730 · doi:10.1002/mrm.25730
- 10.1002/nbm.4004 · doi:10.1002/nbm.4004
- 10.1002/mrm.26168 · doi:10.1002/mrm.26168
- 10.1002/mrm.27179 · DOI
- 10.1007/s10334-009-0169-z · doi:10.1007/s10334-009-0169-z
- 10.1109/TMI.2013.2285120 · DOI
- 10.1002/mrm.26004 · doi:10.1002/mrm.26004
- 10.1016/j.mri.2012.06.037 · doi:10.1016/j.mri.2012.06.037
- 10.1002/mrm.26260 · doi:10.1002/mrm.26260
- 10.1002/mrm.21973 · DOI
- 10.1002/mrm.22473 · doi:10.1002/mrm.22473
- 10.1002/mrm.20796 · DOI · Paper page
- 10.1002/mrm.1910340323 · doi:10.1002/mrm.1910340323
- 10.1002/nbm.3301 · doi:10.1002/nbm.3301
- 10.1002/mrm.1910330414 · doi:10.1002/mrm.1910330414
- 10.1152/ajpheart.00224.2002 · DOI
- 10.1002/mrm.22650 · DOI · Paper page
- 10.1002/mrm.24353 · doi:10.1002/mrm.24353
- 10.1016/j.jmr.2011.11.010 · DOI
- 10.1016/j.jmr.2011.10.006 · DOI
Cited by
Subrahmanian, M. V., Pavuluri, K., Olivieri, C., & Veglia, G. (2022) · PNAS Nexus · doi:10.1093/pnasnexus/pgac133 · Cited by 24
Metadata
- Authors
- Somai, V., Wright, A. J., Fala, M., Hesse, F., & Brindle, K. M.
- Affiliations
- Cancer Research UK Cambridge Institute University of Cambridge Cambridge United Kingdom
- ORCID
- Brindle, Kevin M.
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2020-03-15
- Volume
- 84
- Issue
- 4
- Pages
- 1895-1908
- ISSN
- 0740-3194
- 1522-2594
- Cited by
- 4
- References
- 0
- Language
- en
- Funders
- Cancer Research UK (A17197, A25177, A28397, A29580)
- Rosetrees Trust (A1698)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A multi spin echo pulse sequence with optimized excitation pulses and a 3D cone readout for hyperpolarized 13C imaging. https://app.cochat.ai/papers/10.1002/mrm.28248
- The paper
- Somai, V., Wright, A. J., Fala, M., Hesse, F., & Brindle, K. M. (2020). A multi spin echo pulse sequence with optimized excitation pulses and a 3D cone readout for hyperpolarized 13C imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.28248
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.