Clinical workflow of sonographers performing fetal anomaly ultrasound scans: Deep-learning-based analysis
Open accessAbstract
Objective: Despite decades of obstetric scanning, the field of sonographer workflow remains largely unexplored. In the second trimester, sonographers use scan guidelines to guide their acquisition of standard planes and structures; however, the scan‐acquisition order is not prescribed. Using deep‐learning‐based video analysis, the aim of this study was to develop a deeper understanding of the clinical workflow undertaken by sonographers during second‐trimester anomaly scans. Methods: We collected prospectively full‐length video recordings of routine second‐trimester anomaly scans. Important scan events in the videos were identified by detecting automatically image freeze and image/clip save. The video immediately preceding and following the important event was extracted and labeled as one of 11 commonly acquired anatomical structures. We developed and used a purposely trained and tested deep‐learning annotation model to label automatically the large number of scan events. Thus, anomaly scans were partitioned as a sequence of anatomical planes or fetal structures obtained over time. Results: A total of 496 anomaly scans performed by 14 sonographers were available for analysis. UK guidelines specify that an image or videoclip of five different anatomical regions must be stored and these were detected in the majority of scans: head/brain was detected in 97.2% of scans, coronal face view (nose/lips) in 86.1%, abdomen in 93.1%, spine in 95.0% and femur in 92.3%. Analyzing the clinical workflow, we observed that sonographers were most likely to begin their scan by capturing the head/brain (in 24.4% of scans), spine (in 23.2%) or thorax/heart (in 22.8%). The most commonly identified two‐structure transitions were: placenta/amniotic fluid to maternal anatomy, occurring in 44.5% of scans; head/brain to coronal face (nose/lips) in 42.7%; abdomen to thorax/heart in 26.1%; and three‐dimensional/four‐dimensional face to sagittal face (profile) in 23.7%. Transitions between three or more consecutive structures in sequence were uncommon (up to 13% of scans). None of the captured anomaly scans shared an entirely identical sequence. Conclusions: We present a novel evaluation of the anomaly scan acquisition process using a deep‐learning‐based analysis of ultrasound video. We note wide variation in the number and sequence of structures obtained during routine second‐trimester anomaly scans. Overall, each anomaly scan was found to be unique in its scanning sequence, suggesting that sonographers take advantage of the fetal position and acquire the standard planes according to their visibility rather than following a strict acquisition order. © 2022 The Authors. Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of International Society of Ultrasound in Obstetrics and Gynecology.
AI snapshot
Written from the paper’s full text.
TL;DR
Using deep-learning-based video analysis of 496 routine second-trimester anomaly scans, this study found wide variation in sonographer workflow, with no two scans sharing an entirely identical acquisition sequence. The findings suggest anomaly scanning is opportunistic, guided by fetal position rather than a fixed order.
Key findings
- Among 496 scans performed by 14 sonographers, the five UK FASP mandatory structures were detected in most scans: head/brain 97.2%, coronal face (nose/lips) 86.1%, abdomen 93.1%, spine 95.0% and femur 92.3%. At least four of five mandatory structures were detected in 92.9% of scans, and all five in 72.4%.
- Sonographers most often began by capturing head/brain (24.4%), spine (23.2%) or thorax/heart (22.8%). The most common two-structure transitions were placenta/amniotic fluid to maternal anatomy (44.5%), head/brain to coronal face (42.7%), abdomen to thorax/heart (26.1%) and 3D/4D face to sagittal face (23.7%).
- Transitions between three or more consecutive structures were uncommon, occurring in up to 13% of scans. None of the captured anomaly scans shared an entirely identical sequence.
- Mean anomaly scan duration was 32.1 ± 11.4 min, representing approximately 265 h of scan video, with a mean of 39 ± 13 important scan events per scan.
Methods
- Prospective observational study of routine anomaly scans in singleton pregnancies at 19–21 weeks, conducted May 2018–February 2020 at a single UK maternity unit. Of 518 recruited women, 496 scans were included after 22 exclusions for technical recording problems.
- Full-length scan videos were recorded via HDMI. Image freeze, image save and clip save events were detected automatically; 5-s clips around each event were extracted and labeled as one of 11 anatomical structures.
- A deep-learning model was trained using manual labels from 62 full-length scans and then used to label all scan events automatically. Timelines of structure labels were analyzed with descriptive statistics for structure capture, order and transitions.
Limitations
- The paper states that it could not judge whether specific sequences perform better or result in shorter scanning times, because it collected prospective observational data on routine real-world practice with minimal interference.
- The study was conducted in a single maternity unit, which may not represent practice at other centers.
- Automatic detection missed some structures in some scans; this may reflect incomplete scans or misclassification, including structures labeled as mixed. The average maternal BMI was 25, so generalizability to women with higher BMI remains to be established.
- Evident from the text: the observational design cannot establish causal relationships between workflow patterns and scan quality or efficiency, and sonographers were aware that scans were being recorded, although they were not informed of the current analysis aim.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 6 cited sources checked · 0 supported · 0 flagged
- Findingown data
The two-structure transition head/brain to coronal face (nose/lips) occurred in 42.7% of scans.
“head/brain to coronal face (nose/lips) in 42.7%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
In 359 (72.4%) scans, all five mandatory structures were detected.
“in 359 (72.4%) scans, all five mandatory structures were detected”
Quote in the paperSubstantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Sonographers were most likely to begin their scan by capturing head/brain (24.4% of scans).
“sonographers were most likely to begin their scan by capturing the head/brain (in 24.4% of scans)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The three-structure transition placenta/amniotic fluid to maternal anatomy to placenta/amniotic fluid occurred in 53 (10.7%) scans.
“placenta/amniotic fluid to maternal anatomy to placenta/amniotic fluid in 53 (10.7%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingreasoning
A total of 496 second-trimester anomaly scans (one per woman) were included in the analysis.
“Therefore, 496 second‐trimester anomaly scans (one per woman) were included.”
Quote in the paperSubstantial claim
- Findingown data
Placenta/amniotic fluid was detected in 74.4% of scans.
“74.4% for placenta/amniotic fluid”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
All of the most common three- and four-structure transition sequences were performed by at least three sonographers.
“All of the most common three‐ and four‐structure transition sequences were performed by at least three sonographers.”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Scans in which a structure was detected automatically usually contained multiple capture episodes.
“Scans in which a structure was detected automatically usually contained multiple capture episodes”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The mean number of important scan events was 39 ± 13 per scan.
“The mean ± SD number of important scan events was 39 ± 13 per scan”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Findingown data
The structures most commonly evaluated last were maternal anatomy, sagittal face (profile) and spine, occurring in 111 (22.4%), 100 (20.2%) and 61 (12.3%) scans, respectively.
“The structures most commonly evaluated last were maternal anatomy, sagittal face (profile) and spine, occurring in 111 (22.4%), 100 (20.2%) and 61 (12.3%) scans, respectively.”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Sagittal face (profile) was detected in 78.8% of scans.
“78.8% for sagittal face (profile)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Each of the transition pairs was performed by multiple sonographers, making it impossible to attribute a particular transition pair to a specific sonographer.
“Each of these transition pairs was performed by multiple sonographers, making it impossible to attribute a particular transition pair to a specific sonographer.”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The two-structure transition abdomen to thorax/heart occurred in 26.1% of scans.
“abdomen to thorax/heart in 26.1%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
At least eight structures were detected automatically in 420 (84.7%) scans.
“at least eight structures were detected automatically in 420 (84.7%) scans”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Transitions between three or more consecutive structures in sequence were uncommon (up to 13% of scans).
“Transitions between three or more consecutive structures in sequence were uncommon (up to 13% of scans).”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Femur was detected in 92.3% of scans.
“femur in 92.3%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The average BMI of women included in the study was 25.
“the average BMI of women included in the study was 25”
Quote in the paperSubstantial claim - Findingown data
None of the captured anomaly scans shared an entirely identical sequence.
“None of the captured anomaly scans shared an entirely identical sequence.”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The four-structure transition head/brain to coronal face to abdomen to femur occurred in 19 (3.8%) scans.
“head/brain to coronal face to abdomen to femur in 19 (3.8%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
At least nine structures were detected in 313 (63.1%) scans.
“at least nine were detected in 313 (63.1%) scans”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Head/brain was detected in 97.2% of scans.
“head/brain was detected in 97.2% of scans”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Coronal face view (nose/lips) was detected in 86.1% of scans.
“coronal face view (nose/lips) in 86.1%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Abdomen was detected in 93.1% of scans.
“abdomen in 93.1%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The four-structure transition head/brain to coronal face to abdomen to thorax/heart occurred in 21 (4.2%) scans.
“head/brain to coronal face to abdomen to thorax/heart in 21 (4.2%) scans”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Spine was detected in 95.0% of scans.
“spine in 95.0%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The structures most commonly captured during anomaly scans were thorax/heart, spine and maternal anatomy, with 3254, 2430 and 1564 respective occurrences in the 496 scans.
“The structures most commonly captured during anomaly scans were thorax/heart, spine and maternal anatomy (including uterine artery), with 3254, 2430 and 1564 respective occurrences in the 496 scans.”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
3D/4D face was detected in 49.6% of scans.
“49.6% for 3D/4D face”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The most common three-structure transition sequence was thorax/heart to abdomen to thorax/heart, in 65 (13.1%) scans.
“thorax/heart to abdomen to thorax/heart in 65 (13.1%) scans”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The four-structure transition thorax/heart to head/brain to coronal face to abdomen occurred in 18 (3.6%) scans.
“thorax/heart to head/brain to coronal face to abdomen in 18 (3.6%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The three-structure transition head/brain to coronal face (nose/lips) to abdomen occurred in 61 (12.3%) scans.
“head/brain to coronal face (nose/lips) to abdomen in 61 (12.3%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The three-structure transition head/brain to coronal face to thorax/heart occurred in 53 (10.7%) scans.
“head/brain to coronal face to thorax/heart in 53 (10.7%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The two-structure transition 3D/4D face to sagittal face (profile) occurred in 23.7% of scans.
“3D/4D face to sagittal face (profile) in 23.7%”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Kidneys were detected in 50.2% of scans.
“50.2% for kidneys”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The four-structure transition head/brain to coronal face to thorax/heart to abdomen occurred in 19 (3.8%) scans.
“head/brain to coronal face to thorax/heart to abdomen in 19 (3.8%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
Thorax/heart was detected in 95.4% of scans.
“95.4% for thorax/heart”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The most commonly identified two-structure transition was placenta/amniotic fluid to maternal anatomy, occurring in 44.5% of scans.
“placenta/amniotic fluid to maternal anatomy, occurring in 44.5% of scans”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
In 461 (92.9%) scans, at least four of five mandatory structures were detected.
“In 461 (92.9%) scans, at least four of five mandatory structures (head/brain, coronal face (nose/lips), abdomen, spine and femur) were detected”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Findingown data
The mean anomaly scan duration was 32.1 ± 11.4 min.
“The mean ± SD anomaly scan duration was 32.1 ± 11.4 min”
Quote in the paperNot a substantial claim - Interpretationown data
Sonographers take advantage of fetal position and acquire standard planes according to visibility rather than following a strict acquisition order.
“suggesting that sonographers take advantage of the fetal position and acquire the standard planes according to their visibility rather than following a strict acquisition order.”
Quote in the paperSubstantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Interpretationown data
The generalizability of the findings to women of higher BMI remains to be established.
“meaning the generalizability of our findings to women of higher BMI remains to be established”
Quote in the paperNot a substantial claim - Interpretationown data
Anomaly scans can be modeled as a non-ordered sequence of anatomical structure acquisitions.
“We showed that an anomaly scan can be modeled as a non‐ordered sequence of anatomical structure acquisitions.”
Quote in the paperSubstantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Interpretationown data
The most frequent imaging transitions occur between structures that are in proximity to each other.
“We also established that the most frequent imaging transitions occur between structures that are in proximity to each other”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- Interpretationown data
The lack of a universal scanning sequence supports the hypothesis that anomaly scanning is opportunistic by nature, continuously taking advantage of fetal position.
“The lack of a universal scanning sequence supports the hypothesis that anomaly scanning is opportunistic by nature, continuously taking advantage of fetal position.”
Quote in the paperSubstantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
- May combine two claims. 7
- Classified as probably not a substantial claim. 11
- The quote only partly states this claim. 2
May combine two claims.
Findingown dataThe structures most commonly captured multiple times were thorax/heart, spine and maternal anatomy, recorded on average seven, five and three times in each scan, respectively.
“The structures most commonly captured multiple times were thorax/heart, spine and maternal anatomy, recorded on average seven, five and three times in each scan, respectively”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown data22 (4.2%) women were excluded because of technical problems with recordings, such as incomplete videos.
“Of these, 22 (4.2%) were excluded because of technical problems with recordings, such as incomplete videos.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataThe scans were performed by 14 operators, of whom 11 were accredited sonographers and three fetal medicine doctors.
“Each scan was carried out by one of 14 operators, of whom 11 were accredited sonographers and three fetal medicine doctors”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Findingown dataSonographers were most likely to begin their scan by capturing thorax/heart (22.8% of scans).
“thorax/heart (in 22.8%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
Probably not a substantial claim: may not pass S1 (asserted), S3 (load-bearing). The quote only partly states this claim.
Findingown dataSonographers were most likely to begin their scan by capturing spine (23.2% of scans).
“spine (in 23.2%)”
Quote in the paperNot a substantial claim- Rests on an assumption: Image freeze and image/clip save events represent important scan events.
- Rests on an assumption: The 5-s clips extracted around each important event contain the anatomical structure being assessed.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe mean gestational age at scan was 20.3 ± 0.6 weeks.
“at a mean ± SD gestational age of 20.3 ± 0.6 weeks”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
PremisecitationArtificial neural networks are particularly good at image pattern recognition and can be utilized to improve the speed and accuracy of diagnosing patient conditions in medical fields that depend heavily on images and video, as in ultrasound.
“Artificial neural networks are particularly good at image pattern recognition and can be utilized to improve the speed and accuracy of diagnosing patient conditions in medical fields that depend heavily on images and video, as in the case of ultrasound 11 , 12 .”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [11] Drukker L, Noble JA, Papageorghiou AT. Introduction to artificial intelligence in ultrasound imaging in obstetrics and gynecology. Ultrasound Obstet Gynecol 2020; 56: 498– 505. 32530098 PMC7702141 source
Not checked in the free run.
- cites [12] Xie HN, Wang N, He M, Zhang LH, Cai HM, Xian JB, Lin MF, Zheng J, Yang YZ. Using deep‐learning algorithms to classify fetal brain ultrasound images as normal or abnormal. Ultrasound Obstet Gynecol 2020; 56: 579– 587. 31909548 source
Not checked in the free run.
Classified as probably not a substantial claim.
PremisecitationThe five mandatory structures required by UK FASP guidelines are head/brain, coronal face view (nose/lips), abdomen, spine and femur.
“According to the UK FASP guidelines, five mandatory structures should be captured as images or clips: head/brain, coronal face view (nose/lips), abdomen, spine and femur.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] Public Health England. Fetal anomaly screening programme: standards. https://www.gov.uk/government/publications/fetal‐anomaly‐screening‐programme‐standards. 36680493
Not checked in the free run.
May combine two claims.
Study design and conductreasoningSonographers were aware that the scans were being recorded but had not been informed of the aim of the current analysis, making it unlikely that they acted differently from their usual practice.
“even though sonographers were aware that the scans were being recorded, they had not been informed of the aim of the current analysis, meaning that it is unlikely that they acted differently from their usual practice while participating in this study.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationSonographers should assess, but are not obliged to store an image or clip of, sagittal face, thorax/heart, kidneys, bladder, limbs and placenta/amniotic fluid.
“sonographers should assess, but are not obliged to store an image or a clip of, the following structures: sagittal face (fetal profile), thorax/heart, kidneys, bladder, limbs and placenta/amniotic fluid 14”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] Public Health England. Fetal anomaly screening programme: standards. https://www.gov.uk/government/publications/fetal‐anomaly‐screening‐programme‐standards. 36680493
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenConfusion matrices were used to compare interannotator and manual vs automatic labeling.
“Confusion matrices were used to compare interannotator and manual vs automatic labeling.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductreasoningTo reduce the risk of bias, the study included 14 sonographers, and the findings were consistent, making external validity more likely.
“To reduce the risk of bias, we included 14 sonographers, and we noted that the findings were consistent, making external validity more likely.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
PremisecitationScanning guidelines comprise a checklist of standard planes, structures and anatomies to be surveyed, but do not dictate the order in which the checklist should be populated.
“Such scanning guidelines comprise typically a checklist of standard planes, structures and anatomies to be surveyed, but do not dictate the order in which the checklist should be populated.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [9] Salomon LJ, Alfirevic Z, Berghella V, Bilardo CM, Chalouhi GE, Da Silva Costa F, Hernandez‐Andrade E, Malinger G, Munoz H, Paladini D, Prefumo F, Sotiriadis A, Toi A, Lee W. ISUOG Practice Guidelines (updated): performance of the routine mid‐trimester fetal ultrasound scan. Ultrasound Obstet Gynecol 2022; 59: 840– 856. 35592929 source
Not checked in the free run.
- cites [10] AIUM‐ACR‐ACOG‐SMFM‐SRU Practice Parameter for the Performance of Standard Diagnostic Obstetric Ultrasound Examinations. J Ultrasound Med 2018; 37: E13– 24. 30308091 source
Not checked in the free run.
Classified as probably not a substantial claim.
Study design and conductown dataThe deep-learning model was purposely trained and tested to label automatically the large number of scan events.
“We developed and used a purposely trained and tested deep‐learning annotation model to label automatically the large number of scan events.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenThe study was conducted in a single maternity unit which may not represent practice at other centers.
“the study was conducted in a single maternity unit which may not represent practice at other centers”
Quote in the paperNot a substantial claim
The paper relies on this without support.
Image freeze and image/clip save events represent important scan events.
“we analyzed the full‐length anomaly scans automatically and identified sonographer actions that represent important scan events by detecting the occurrences of image freeze, image save and clip save.”
The paper relies on this without support.
The 5-s clips extracted around each important event contain the anatomical structure being assessed.
“After identifying the important scan events, we extracted 5‐s clips that occurred around the event (i.e. immediately preceding and following the event).”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A limitation in workflow analysis applications was the impossibility of the generalization of workflow sequences due to maternal−fetal factors and different skill levels of sonographers for different anatomical regions during a full routine scan.
“Limitations in application were reported to be the impossibility of the generalization of workflow sequences due to maternal−fetal factors and different skill levels of sonographers for different anatomical regions during a full routine scan [155,158].”
From Evolving the Era of 5D Ultrasound? A Systematic Literature Review on the Applications for Artificial Intelligence Ultrasound Imaging in Obstetrics and Gynecology (2023) · cites this paper as [158]
The Results and Discussion (page unavailable) report that no two full scans followed an identical sequence and that fetal position and movement influence acquisition order. They do not establish that workflow sequences cannot be generalized because of maternal–fetal factors or differing sonographer skill by anatomical region; the Discussion instead calls for research to establish useful workflow patterns. (LLM, conf=0.87)
Experimental — accuracy not yet measuredSharma et al. and Drukker et al. analyzed full-length scanning videos to assess workflow by building timeline models illustrating the scanning sequence of anatomical regions over time.
“Sharma et al. and Drukker et al. analyzed full-length scanning videos to assess workflow by building timeline models illustrating the scanning sequence of anatomical regions over time [155,158].”
Cites
Salomon, L. J., Alfirevic, Z., Berghella, V., Bilardo, C. M., Chalouhi, G. E., Da Silva Costa, F., Hernandez‐Andrade, E., Malinger, G., Munoz, H., Paladini, D., et al. (2022) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.24888 · Cited by 436 · see citing works
SonoNet: Real-Time Detection and Localisation of Fetal Standard Scan Planes in Freehand Ultrasound
Baumgartner, C. F., Kamnitsas, K., Matthew, J., Fletcher, T. P., Smith, S., Koch, L. M., Kainz, B., & Rueckert, D. (2017) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2712367 · Cited by 416 · see citing works
Karim, J. N., Roberts, N. W., Salomon, L. J., & Papageorghiou, A. T. (2017) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.17246 · Cited by 266 · see citing works
Self-citationIntroduction to artificial intelligence in ultrasound imaging in obstetrics and gynecology
Drukker, L., Noble, J. A., & Papageorghiou, A. T. (2020) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.22122 · Cited by 240 · see citing works
Self-citationBurgos-Artizzu, X. P., Coronado-Gutiérrez, D., Valenzuela-Alcaraz, B., Bonet-Carne, E., Eixarch, E., Crispi, F., & Gratacós, E. (2020) · Scientific Reports · doi:10.1038/s41598-020-67076-5 · Cited by 240 · see citing works
(2018) · Journal of Ultrasound in Medicine · doi:10.1002/jum.14831 · Cited by 216 · see citing works
Using deep-learning algorithms to classify fetal brain ultrasound images as normal or abnormal
Xie, H. N., Wang, N., He, M., Zhang, L. H., Cai, H. M., Xian, J. B., Lin, M. F., Zheng, J., & Yang, Y. Z. (2020) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.21967 · Cited by 176 · see citing works
Tele-Ultrasound in Resource-Limited Settings: A Systematic Review
Britton, N., Miller, M. A., Safadi, S., Siegel, A., Levine, A. R., & McCurdy, M. T. (2019) · Frontiers in Public Health · doi:10.3389/fpubh.2019.00244 · Cited by 166 · see citing works
Ultrasound Standard Plane Detection Using a Composite Neural Network Framework
Chen, H., Wu, L., Dou, Q., Qin, J., Li, S., Cheng, J.-Z., Ni, D., & Heng, P.-A. (2017) · IEEE Transactions on Cybernetics · doi:10.1109/tcyb.2017.2685080 · Cited by 164 · see citing works
AIUM Practice Parameter for the Performance of Fetal Echocardiography
(2019) · Journal of Ultrasound in Medicine · doi:10.1002/jum.15188 · Cited by 134 · see citing works
Singh, N., Knight, K. A., Wright, C. A., Baird, M. A., Akroyd, D., Adams, R., & Schneider, M. E. (2016) · Journal of Medical Imaging and Radiation Oncology · doi:10.1111/1754-9485.12547 · Cited by 90 · see citing works
Salomon, L. J., Winer, N., Bernard, J. P., & Ville, Y. (2008) · Prenatal Diagnosis · doi:10.1002/pd.2016 · Cited by 85 · see citing works
Thilaganathan, B., Wormald, B. W., Zanardini, C., Sheldon, J., Ralph, E., & Papageorghiou, A. T. (2010) · Obstetrics and Gynecology · doi:10.1097/aog.0b013e3181dd5137 · Cited by 80 · see citing works
Self-citationLin, M., He, X., Guo, H., He, M., Zhang, L., Xian, J., Lei, T., Xu, Q., Zheng, J., Feng, J., et al. (2022) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.24843 · Cited by 65 · see citing works
Drukker, L., Sharma, H., Droste, R., Alsharid, M., Chatelain, P., Noble, J. A., & Papageorghiou, A. T. (2021) · Scientific Reports · doi:10.1038/s41598-021-92829-1 · Cited by 59 · see citing works
Self-citationYaqub, M., Kelly, B., Papageorghiou, A. T., & Noble, J. A. (2015) · Lecture Notes in Computer Science · doi:10.1007/978-3-319-24574-4_82 · Cited by 56 · see citing works
Self-citationO’Gorman, N., Tampakoudis, G. P., Wright, A., Wright, D., & Nicolaides, K. H. (2015) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.15819 · Cited by 53 · see citing works
Does the Use of Automated Fetal Biometry Improve Clinical Work Flow Efficiency?
Espinoza, J., Good, S., Russell, E., & Lee, W. (2013) · Journal of Ultrasound in Medicine · doi:10.7863/jum.2013.32.5.847 · Cited by 53 · see citing works
Sharma, H., Drukker, L., Chatelain, P., Droste, R., Papageorghiou, A. T., & Noble, J. A. (2021) · Medical Image Analysis · doi:10.1016/j.media.2021.101973 · Cited by 48 · see citing works
Self-citationQuantification of scan-specific ergonomic risk-factors in medical sonography
Burnett, D. R., & Campbell-Kyureghyan, N. H. (2010) · International Journal of Industrial Ergonomics · doi:10.1016/j.ergon.2009.11.005 · Cited by 47 · see citing works
Benacerraf, B. R., Minton, K. K., Benson, C. B., Bromley, B., Coley, B. D., Doubilet, P. M., Lee, W., Maslak, S. H., Pellerito, J. S., Perez, J. J., et al. (2017) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2017.06.033 · Cited by 37 · see citing works
Does the Use of Automated Fetal Biometry Improve Clinical Work Flow Efficiency?
Espinoza, J., Good, S., Russell, E., et al. (2013) · Journal of Ultrasound in Medicine · doi:10.7863/ultra.32.5.847 · Cited by 31 · see citing works
Expected‐value bias in routine third‐trimester growth scans
Drukker, L., Droste, R., Chatelain, P., Noble, J. A., & Papageorghiou, A. T. (2020) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.21929 · Cited by 29 · see citing works
Self-citationDrukker, L., Droste, R., Chatelain, P., Noble, J. A., & Papageorghiou, A. T. (2020) · Ultraschall in der Medizin - European Journal of Ultrasound · doi:10.1055/a-1074-0722 · Cited by 27 · see citing works
Self-citationAn intelligent platform for ultrasound diagnosis of thyroid nodules
Ye, H., Hang, J., Chen, X., Xu, D., Chen, J., Ye, X., & Zhang, D. (2020) · Scientific Reports · doi:10.1038/s41598-020-70159-y · Cited by 23 · see citing works
Utility of follow-up standard sonography for fetal anomaly detection
Byrne, J. J., Morgan, J. L., Twickler, D. M., McIntire, D. D., & Dashe, J. S. (2020) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2020.01.001 · Cited by 15 · see citing works
The effect of maternal body mass index on fetal ultrasound image quality
Yaqub, M. K., Kelly, B., Noble, J. A., & Papageorghiou, A. T. (2021) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2021.04.248 · Cited by 15 · see citing works
Self-citationFisher vector encoding for detecting objects of interest in ultrasound videos
Maraci, M. A., Napolitano, R., Papageorghiou, A. T., & Noble, J. A. (2015) · IEEE International Symposium on Biomedical Imaging (ISBI) · doi:10.1109/isbi.2015.7163957 · Cited by 10 · see citing works
Self-citationSanyal, R., Kraft, B., Alexander, L. F., Ismail, A., Lockhart, M. E., & Robbin, M. L. (2016) · American Journal of Roentgenology · doi:10.2214/ajr.15.15030 · Cited by 10 · see citing works
Quality assessment of the detailed anomaly ultrasound scan
Souka, A. P., Pilalis, A., Papastefanou, I., Eleftheriadis, M., & Papadopoulos, G. K. (2017) · The Journal of Maternal-Fetal & Neonatal Medicine · doi:10.1080/14767058.2017.1388366 · Cited by 10 · see citing works
Sorrentino, K. (2019) · Journal of diagnostic medical sonography · doi:10.1177/8756479319838234 · Cited by 10 · see citing works
Jone, P., Gould, R., Barrett, C., Younoszai, A. K., & Fonseca, B. (2018) · Pediatric Cardiology · doi:10.1007/s00246-018-1812-0 · Cited by 7 · see citing works
Fetal imaging: past, present, and future. A journey of marvel
Abu‐Rustum, R. S., & Abuhamad, A. Z. (2018) · BJOG: An International Journal of Obstetrics & Gynaecology · doi:10.1111/1471-0528.15343 · Cited by 4 · cited 1× · see citing works
“The continuous improvement of obstetric ultrasound is attributed to many factors, including developments in education, accreditation, guidelines, quality assurance, anatomical and physiological knowledge and imaging quality 1–4 .”
BackgroundPublic Health England and Co-Production with the Fetal Anomaly Screening Programme
Lloyd, C., Corcoran, E., & Murray, L. (2023) · The New Bioethics · doi:10.1080/20502877.2022.2147444 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Progress in the Application of Artificial Intelligence in Ultrasound-Assisted Medical Diagnosis
Li, Y., Li, Q., Fu, K., Zhou, X., & Zhang, K. (2025) · Bioengineering · doi:10.3390/bioengineering12030288 · Cited by 71 · see citing works
Shahrivar, R. Y., Karami, F., & Karami, E. (2023) · Biomimetics · doi:10.3390/biomimetics8070519 · Cited by 57 · see citing works
Jost, E., Kosian, P., Jimenez Cruz, J., Albarqouni, S., Gembruch, U., Strizek, B., & Recker, F. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12216833 · Cited by 44 · see citing works
Płotka, S., Grzeszczyk, M. K., Szenejko, P. I., Żebrowska, K., Szymecka-Samaha, N., Łęgowik, T., Lipa, M., Kosińska–Kaczyńska, K., Brawura-Biskupski-Samaha, R., Išgum, I., et al. (2023) · American Journal of Obstetrics & Gynecology MFM · doi:10.1016/j.ajogmf.2023.101182 · Cited by 38 · see citing works
Lei, T., Zheng, Q., Feng, J., Zhang, L., Zhou, Q., He, M., Lin, M.-H., & Xie, H. (2024) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.29101 · Cited by 20 · see citing works
Fukumura, Y. E., Sommerich, C. M., Evans, K. D., & Roll, S. C. (2023) · Journal of diagnostic medical sonography · doi:10.1177/87564793231205612 · Cited by 9 · see citing works
Wang, L., Fatemi, M., & Alizad, A. (2025) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2025.110312 · Cited by 8 · see citing works
Sonographer interaction with artificial intelligence: collaboration or conflict?
Day, T. G., Matthew, J., Budd, S., Hajnal, J. V., Simpson, J. M., Razavi, R., & Kainz, B. (2023) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.26238 · Cited by 10 · see citing works
Detecting anomalies in fetal electrocardiogram records using deep learning models
Santhanam, S., & Jose, D. (2023) · Journal of Intelligent & Fuzzy Systems · doi:10.3233/jifs-231681 · Cited by 8 · see citing works
Automatic Identification of Fetal Abdominal Planes from Ultrasound Images Based on Deep Learning
Ciobanu, Ș. G., Enache, I.-A., Iovoaica-Rămescu, C., Berbecaru, E. I. A., Vochin, A., Băluță, I. D., Istrate-Ofiţeru, A.-M., Comănescu, C. M., Nagy, R. D., Şerbănescu, M.-S., et al. (2025) · Journal of Imaging Informatics in Medicine · doi:10.1007/s10278-025-01409-6 · Cited by 6 · see citing works
Computational methods for the analysis of early-pregnancy brain ultrasonography: a systematic review
Bastiaansen, W., Klein, S., Koning, A. H. J., Niessen, W. J., Steegers‐Theunissen, R. P. M., & Rousian, M. (2023) · EBioMedicine · doi:10.1016/j.ebiom.2023.104466 · Cited by 5 · see citing works
Advancements in artificial intelligence for pelvic floor ultrasound analysis
Huang, X. (2024) · American Journal of Translational Research · doi:10.62347/jxqq5395 · Cited by 5 · see citing works
Cost-utility analysis of prenatal diagnosis of congenital cardiac diseases using deep learning
Ginsberg, G. M., Drukker, L., Pollak, U., & Brezis, M. (2024) · Cost Effectiveness and Resource Allocation · doi:10.1186/s12962-024-00550-3 · Cited by 4 · see citing works
Self-citationDeep Learning Model for Real-Time Nuchal Translucency Assessment at Prenatal US
Zhang, Y., Yang, X., Ji, C., Hu, X., Cao, Y., Chen, C., Sui, H., Li, B., Zhen, C., Huang, W., et al. (2025) · Radiology Artificial Intelligence · doi:10.1148/ryai.240498 · Cited by 4 · see citing works
Enhancement of phenotyping for fetal investigation using next‐generation sequencing
Mone, F., Homfray, T., Kagan, K. O., & Kilby, M. D. (2023) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.26301 · Cited by 4 · see citing works
Zhou, X., Wan, J., Sun, F., Wang, R., Yan, Y., Li, P., & Wang, C. (2026) · Scientific Reports · doi:10.1038/s41598-026-41635-8 · Cited by 3 · see citing works
Bai, J., Zhou, Z., Tang, Y., Gan, J., Liang, Z., Fan, J., Mcguire, L., Clarke, J., Cai, W., Spurway, J., et al. (2026) · Medical Image Analysis · doi:10.1016/j.media.2026.104043 · Cited by 4 · see citing works
Drukker, L. (2024) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.27703 · Cited by 3 · see citing works
Self-citationZoom Pattern Signatures for Fetal Ultrasound Structures
Alsharid, M., Yasrab, R., Drukker, L., Papageorghiou, A. T., & Noble, J. A. (2024) · Lecture notes in computer science · doi:10.1007/978-3-031-72083-3_73 · Cited by 1 · see citing works
Self-citationPrenatal Ultrasound Diagnosis of Sirenomelia Sequence in First Trimester: A Case Report
Shao, K., Feng, Z., Ding, S., Wang, Y., Liu, F., Li, T., & Ma, B. (2026) · Journal of Clinical Ultrasound · doi:10.1002/jcu.70223 · Cited by 1 · see citing works
Zoom is Meaningful: Discerning Ultrasound Images’ Zoom Levels
Alsharid, M., Yasrab, R., Sarker, M. M. K., Drukker, L., Papageorghiou, A. T., & Noble, J. A. (2024) · IEEE International Symposium on Biomedical Imaging (ISBI) · doi:10.1109/isbi56570.2024.10635854 · Cited by 1 · see citing works
Self-citationNovel Techniques in Antenatal Imaging of Spinal Dysraphisms
Royal, C., Chertin, L., Alfawzan, M. F., & Killian, M. E. (2025) · Current Urology Reports · doi:10.1007/s11934-025-01258-4 · Cited by 1 · see citing works
Cost-Utility Analysis of Prenatal Diagnosis of Congenital Cardiac Diseases using Deep LearningPreprint
Ginsberg, G. M., Drukker, L., Pollak, U., & Brezis, M. (2024) · doi:10.22541/au.171045405.59757841/v1 · Cited by 1 · see citing works
Self-citationA picture is worth a thousand words: textual analysis of the routine 20‐week scan
Alsharid, M., Drukker, L., Sharma, H., Noble, J. A., & Papageorghiou, A. T. (2022) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.24972
Self-citationCost-Utility Analysis of Prenatal Diagnosis of Congenital Cardiac Diseases using Deep LearningPreprint
Ginsberg, G. M., Drukker, L., Pollak, U., & Brezis, M. (2024) · Research Square · doi:10.21203/rs.3.rs-3981770/v1
Self-citationLous, M. L., Vasconcelos, F., Vece, C. D., Dromey, B., Napolitano, R., Yoo, S., Edwards, E., Huaulmé, A., Peebles, D., Stoyanov, D. V., et al. (2024) · European Journal of Obstetrics & Gynecology and Reproductive Biology · doi:10.1016/j.ejogrb.2024.04.042
Aadhavan, K., Balasubramanian, M., Sowmya, S., Jose, D. V., B., L., & Al-Zaidi, R. (2024) · International Conference on Integrated Intelligence and Communication Systems (ICIICS) · doi:10.1109/iciics63763.2024.10860052
AP-Net: Semi-Supervised Ultrasound Cardiac Segmentation Using Enhanced Anatomical Prior
Lu, Y., Li, J., Lin, J., Yuan, Y., Rajapakse, J. C., Zhu, N., Wang, C., & Li, K. (2025) · IEEE International Conference on Acoustics Speech and Signal Processing · doi:10.1109/icassp49660.2025.10888667
Goff, D. A. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-24098-0.00006-2
Ginsberg, G. M., Drukker, L., Pollak, U., & Brezis, M. (2024) · Figshare · doi:10.6084/m9.figshare.26723981
Self-citationAtia, H., Sadek, S. M., Ahmad, R. A., Elgammal, H., & Aziz, T. M. A. (2025) · International Journal of Gynecology & Obstetrics · doi:10.1002/ijgo.70779
Kashani, M., Naghavi, F., & Rostami, M. (2026) · Research Square · doi:10.21203/rs.3.rs-8521430/v1
Mahmoud, R., soliman, S. H., Tahir, F., & Elmageed, N. E. A. (2026) · The Egyptian Journal of Radiology and Nuclear Medicine · doi:10.1186/s43055-026-01728-w
Varrier, A. S., Alférez, S., Riera-Marín, M., Velayudhan, J., Sabrià, J., & K, S. O. (2026) · IEEE International Symposium on Biomedical Imaging (ISBI) · doi:10.1109/isbi61048.2026.11515723
FUSEP: A Multi-Center Benchmark for Diverse Tasks in Fetal Ultrasound Screening Early Pregnancy
Pu, B., Yang, J., Wang, L., Tan, Y., He, G., Dong, X., Lin, Q., Guo, J., Yang, L., Liu, Z., et al. (2026) · ACM SIGKDD Conference on Knowledge Discovery and Data Mining (KDD) · doi:10.1145/3770855.3817453
Wang, J., Jia, Z., Wu, K., Liu, P., Du, J., & Li, L. (2026) · Journal of Assisted Reproduction and Genetics · doi:10.1007/s10815-026-04006-w
- Authors
- Drukker, L., Sharma, H., Karim, J. N., Droste, R., Noble, J. A., & Papageorghiou, A. T.
- Affiliations
- Tel Aviv University (IL)
- University of Oxford (GB)
- John Radcliffe Hospital (GB)
- Corresponding author
- Papageorghiou, A. T.
- Venue
- Ultrasound in Obstetrics & Gynecology
- Publisher
- Wiley
- Published
- 2022-12
- Volume
- 60
- Issue
- 6
- Pages
- 759-765
- ISSN
- 0960-7692 (ISSN-L)
- 1469-0705
- PMC
- PMC10107110
- OpenAlex
- W4283217943
- Cited by
- 36
- References
- 33
- Citations by year
- 2026: 8 · 2025: 8 · 2024: 11 · 2023: 8 · 2022: 1
- FWCI
- 7.68
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Ultrasound in Obstetrics and Gynecology (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- St George's Online Research Archive (St George's University of London) (acceptedVersion)
- Funders
- H2020 European Research Council (ERC‐ADG‐2015 694581)
- Topics
- Health Sciences › Medicine › Pediatrics, Perinatology and Child Health › Prenatal Screening and Diagnostics
- Health Sciences › Medicine › Health Informatics › Artificial Intelligence in Healthcare and Education
- Health Sciences › Medicine › Surgery › Congenital Anomalies and Fetal Surgery
- Keywords
- clinical workflow, ultrasound video analysis, abdominal imaging
- MeSH terms
- Deep Learning, Female, Fetus, Humans, Pregnancy, Pregnancy Trimester, Second, Ultrasonography, Prenatal, Workflow
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-10.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-10.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review