Abstract
Venous pulsations are thought, but to our knowledge have never been demonstrated, to result in underestimation of oxygen saturation measured by pulse oximetry (SpO2); however, to our knowledge, this has never been demonstrated. A patient with ischemic cardiomyopathy requiring intra-aortic balloon counterpulsation (IABC) had to withstand a temporary cessation of IABC (period I), during which she had marked central venous pulsations and a pulse oximeter bias of -18%. After IABC was reinstituted (period II), the venous pulsations disappeared and the bias was reduced to 0%. Cardiac output, systemic vascular resistance, and temperature during both periods were comparable. Pulmonary vascular resistance, however, was much lower during period II. We postulate that the bias was caused by venous pulsations significantly contributing to the AC, or variable component, of light absorption.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Pulse oximetry values were both lower in cases where CVP contributed to the signal.
“The pulse oximetry values were both lower in cases where CVP contributed to the signal.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [127]
Shelley et al. and Sami et al. both demonstrated the influence of the venous signal on SpO2, showing that forehead and earlobe pulse oximeters produced a significantly biased variable component of the PPG determined to originate from central venous pressure (CVP).
“Shelley et al. and Sami et al. both demonstrated the influence of the venous signal on SpO2, showing that forehead and earlobe pulse oximeters (respectively) produced a significantly biased variable component of the PPG determined to originate from central venous pressure (CVP) [125,127].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [127]
Cites
Tremper, K. K., & Barker, S. J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00019 · Cited by 438 · see citing works
Effect of anemia on pulse oximeter accuracy at low saturation
Severinghaus, J. W., & Koh, S. O. (1990) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf02828282 · Cited by 129 · see citing works
Kim, J. M., Arakawa, K., Benson, K. T., & Fox, D. K. (1986) · Anesthesia & Analgesia · doi:10.1213/00000539-198612000-00015 · Cited by 86 · see citing works
Pulse oximetry versus measurement of transcutaneous oxygen
New, W. (1985) · The Journal of Clinical Monitoring · doi:10.1007/bf02832200 · Cited by 54 · see citing works
Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery
PALVE, H., & Vuori, A. (1989) · Critical Care Medicine · doi:10.1097/00003246-198901000-00014 · Cited by 47 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Current progress of photoplethysmography and SPO2 for health monitoring
TAMURA, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 260 · see citing works
The Light–Tissue Interaction of Pulse Oximetry
Mannheimer, P. D. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269522.84942.54 · Cited by 194 · see citing works
Hunt, C. E., Corwin, M. J., Lister, G., Weese‐Mayer, D. E., Neuman, M. R., Tinsley, L. R., Baird, T. M., Keens, T. G., & Cabral, H. (1999) · The Journal of Pediatrics · doi:10.1016/s0022-3476(99)70056-9 · Cited by 136 · see citing works
Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia
YAMAYA, Y., Bogaard, H. J., Wagner, P. D., Niizeki, K., & Hopkins, S. R. (2002) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00409.2001 · Cited by 115 · see citing works
Pulmonary Gas Exchange and Acid‐Base Balance During Exercise
Stickland, M. K., Lindinger, M. I., Olfert, I. M., Heigenhauser, G. J. F., & Hopkins, S. R. (2013) · Comprehensive physiology · doi:10.1002/cphy.c110048 · Cited by 106 · see citing works
Shelley, K. H., Tamai, D., Jablonka, D., Gesquiere, M., Stout, R. G., & Silverman, D. G. (2005) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000145063.01043.4b · Cited by 91 · see citing works
Origin of Photoplethysmographic Waveform at Green Light
Kamshilin, A. A., & Борисович, М. Н. (2017) · Physics Procedia · doi:10.1016/j.phpro.2017.01.024 · Cited by 90 · see citing works
Effect of Peripheral Vasoconstriction on Pulse Oximetry
Talke, P. O., & Stapelfeldt, C. (2006) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-006-9022-3 · Cited by 59 · see citing works
Effect of local limb temperature on pulse oximetry and the Plethysmographic pulse wave
Schramm, W. M., Bartunek, A. E., & Gilly, H. (1997) · International journal of clinical monitoring and computing · doi:10.1007/bf03356574 · Cited by 53 · see citing works
The detection of peripheral venous pulsation using the pulse oximeter as a plethysmograph
Shelley, K. H., Dickstein, M., & Shulman, S. M. (1993) · Journal of Clinical Monitoring · doi:10.1007/bf02886699 · Cited by 48 · see citing works
Pulse oximetry ‐ clinical implications and recent technical developments
Lindberg, L., Lennmarken, C., & Vegfors, M. (1995) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1995.tb04062.x · Cited by 48 · see citing works
The pulse in reflectance pulse oximetry: Modeling and experimental studies
Reuss, J. L., & Siker, D. (2004) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-005-2909-6 · Cited by 47 · see citing works
Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing
Ascha, M., Bhattacharyya, A., RAMOS, J. A., & Tonelli, A. R. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001658 · Cited by 44 · see citing works
Limitations of forehead pulse oximetry
Jørgensen, J. S., Schmid, E. R., König, V., Faisst, K., Huch, A., & Huch, R. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617520 · Cited by 25 · see citing works
Pulse oximetry and photoplethysmographic waveform analysis of the esophagus and bowel
Phillips, J. P., Kyriacou, P. A., Jones, D. P., Shelley, K. H., & Langford, R. M. (2008) · Current Opinion in Anaesthesiology · doi:10.1097/aco.0b013e328317794d · Cited by 23 · see citing works
Pulmonary Gas Exchange and Acid‐Base Balance During Exercise
Stickland, M. K., Lindinger, M. I., Olfert, I. M., Heigenhauser, G. J. F., & Hopkins, S. R. (2013) · Comprehensive physiology · doi:10.1002/j.2040-4603.2013.tb00500.x · Cited by 22 · see citing works
Effect of location of the sensor on reflectance pulse oximetry
Dassel, A. C. M., Graaff, R., Aardema, M. W., Zijlstra, W. G., & Aarnoudse, J. G. (1997) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1997.tb14350.x · Cited by 22 · see citing works
Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbits
Hummler, H., Pohlandt, F., & Franz, A. R. (2002) · Critical Care Medicine · doi:10.1097/00003246-200211000-00016 · Cited by 15 · see citing works
Khushhal, A. A., Abdelaal, A. A. M., & Elsayed, M. (2024) · Journal of Multidisciplinary Healthcare · doi:10.2147/jmdh.s449071 · Cited by 14 · see citing works
Extracting Venous and Respiration Signals from Photoplethysmographs
Yousefi, R., & Nourani, M. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2334697 · Cited by 12 · see citing works
Detection and monitoring of Hypoxemia and Oxygen Therapy
HESS, D. R. (2000) · Respiratory Care · doi:10.1177/002013240004500106 · Cited by 11 · see citing works
TAMURA, T., & Maeda, Y. (2017) · Springer eBooks · doi:10.1007/978-3-319-69362-0_6 · Cited by 7 · see citing works
Marley, R. A. (1998) · Journal of PeriAnesthesia Nursing · doi:10.1016/s1089-9472(98)80010-5 · Cited by 6 · see citing works
Techniques and Devices for Monitoring Oxygenation
Hess, D., & Kacmarek, R. M. (1993) · Respiratory Care · doi:10.1177/194336549303800607 · Cited by 5 · see citing works
Hayashi, K. (2021) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2021.110198 · Cited by 1 · see citing works
Kurki, T. S. O., & Eisenkraft, J. B. (2011) · Cambridge University Press eBooks · doi:10.1017/cbo9780511974083.018 · Cited by 1 · see citing works
Monitoring during mechanical ventilation
HESS, D. R. (2006) · Paediatric Respiratory Reviews · doi:10.1016/j.prrv.2006.04.177
Non-Invasive Method and Experimental Study for Measurement of Oxygen Saturation Rate in Wide Range
Ye, S., Zhou, F., Chen, H., Yan, H. Z., & Li, Y. (2008) · International Conference on Bioinformatics and Biomedical Engineering · doi:10.1109/icbbe.2008.179
Falsely low pulse oximetry values caused by aortocaval fistula
Martin, D. C., & Chui, C. (2012) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2011.06.024
Finger pulse oximetry detects an intense congestion
Niwa, H., Kimura, F., Kudo, T., Shito, A., & Hirota, K. (2013) · European Journal of Anaesthesiology · doi:10.1097/eja.0b013e328363196a
Cardiac gating technique for arterial oxygen saturation measurement
Yousefi, R., & Nourani, M. (2014) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2014.6981638
Kosaka, M., Yoshida, A., & Ohe, K. (2016) · Journal of the Japanese Society of Intensive Care Medicine · doi:10.3918/jsicm.23.625
PECULIARITIES OF USING THE EMD ALGORITHM TO ESTIMATE THE LEVEL OF OXYGEN IN THE BLOOD
Boloban, O., & Petrenko, A. I. (2023) · Scientific notes of Taurida National V I Vernadsky University Series Technical Sciences · doi:10.32782/2663-5941/2023.2.1/20
Pal, R., Rudas, Á., Chiang, J. N., Barney, A. M., & Cannesson, M. P. (2025) · medRxiv · doi:10.64898/2025.12.30.25343231
- Authors
- Sami, H. M., Kleinman, B. S., & Lonchyna, V. A.
- Affiliations
- Loyola University Chicago (US)
- Venue
- Journal of Clinical Monitoring
- Published
- 1991-10
- Volume
- 7
- Issue
- 4
- Pages
- 309-312
- ISSN
- 1387-1307 (ISSN-L)
- 0748-1977
- 1573-2614
- OpenAlex
- W2058795083
- Cited by
- 36
- References
- 5
- Citations by year
- 2025: 1 · 2024: 1 · 2023: 2 · 2021: 2 · 2019: 1 · 2018: 1 · 2017: 2 · 2016: 1 · 2014: 2 · 2013: 3 · 2012: 1
- FWCI
- 0.86
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Journal of Clinical Monitoring and Computing (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Physical Sciences › Engineering › Biomedical Engineering › Mechanical Circulatory Support Devices
- Keywords
- pulse oximetry, intra-aortic balloon pump, ischemic cardiomyopathy, pulmonary vascular resistance, cardiac output
- MeSH terms
- Aged, Carbon Dioxide, Catheterization, Peripheral, Catheterization, Swan-Ganz, Central Venous Pressure, Coronary Disease, Femoral Artery, Heart Rate, Humans, Intra-Aortic Balloon Pumping, Male, Oximetry, Oxygen, Pulmonary Artery, Pulse, Vascular Resistance, Counterpulsation, Monitoring, Intraoperative
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review