Photoplethysmograph signal reconstruction based on a novel hybrid motion artifact detection–reduction approach. Part I: Motion and noise artifact detection
Abstract
Motion and noise artifacts (MNA) are a serious obstacle in utilizing photoplethysmogram (PPG) signals for real-time monitoring of vital signs. We present a MNA detection method which can provide a clean vs. corrupted decision on each successive PPG segment. For motion artifact detection, we compute four time-domain parameters: (1) standard deviation of peak-to-peak intervals (2) standard deviation of peak-to-peak amplitudes (3) standard deviation of systolic and diastolic interval ratios, and (4) mean standard deviation of pulse shape. We have adopted a support vector machine (SVM) which takes these parameters from clean and corrupted PPG signals and builds a decision boundary to classify them. We apply several distinct features of the PPG data to enhance classification performance. The algorithm we developed was verified on PPG data segments recorded by simulation, laboratory-controlled and walking/stair-climbing experiments, respectively, and we compared several well-established MNA detection methods to our proposed algorithm. All compared detection algorithms were evaluated in terms of motion artifact detection accuracy, heart rate (HR) error, and oxygen saturation (SpO2) error. For laboratory controlled finger, forehead recorded PPG data and daily-activity movement data, our proposed algorithm gives 94.4, 93.4, and 93.7% accuracies, respectively. Significant reductions in HR and SpO2 errors (2.3 bpm and 2.7%) were noted when the artifacts that were identified by SVM-MNA were removed from the original signal than without (17.3 bpm and 5.4%). The accuracy and error values of our proposed method were significantly higher and lower, respectively, than all other detection methods. Another advantage of our method is its ability to provide highly accurate onset and offset detection times of MNAs. This capability is important for an automated approach to signal reconstruction of only those data points that need to be reconstructed, which is the subject of the companion paper to this article. Finally, our MNA detection algorithm is real-time realizable as the computational speed on the 7-s PPG data segment was found to be only 7 ms with a Matlab code.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Chong et al. achieved an accuracy of 93.7% during a daily motion experiment and saw a significant reduction in error for SpO2 and heart rate measurements.
“They were able to achieve an accuracy of 93.7% during a daily motion experiment, and they saw a significant reduction in error for SpO2 and heart rate measurements.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [170]
Cites
A Practical Guide to Support Vector Classication
Hsu, C., Chang, C.-C., & Lin, C. (2008) · Cited by 5,070 · see citing works
EEG analysis based on time domain properties
Hjorth, B. (1970) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(70)90143-4 · Cited by 1,504 · see citing works
Maximal perfusion of skeletal muscle in man.
Andersen, P., & Saltin, B. (1985) · The Journal of Physiology · doi:10.1113/jphysiol.1985.sp015794 · Cited by 1,250 · see citing works
Consensus Filters for Sensor Networks and Distributed Sensor Fusion
Olfati‐Saber, R., & Shamma, J. S. (2006) · IEEE Conference on Decision and Control (CDC) · doi:10.1109/cdc.2005.1583238 · Cited by 1,005 · see citing works
Li, Q., Mark, R. G., & Clifford, G. D. (2007) · Physiological Measurement · doi:10.1088/0967-3334/29/1/002 · Cited by 427 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
Artifact Removal in Physiological Signals—Practices and Possibilities
Sweeney, K. T., Ward, T. E., & McLoone, S. (2012) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2012.2188536 · Cited by 358 · see citing works
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345 · see citing works
Automatic Real Time Detection of Atrial Fibrillation
Dash, S. S., Chon, K. H., Lu, S., & Raeder, E. A. (2009) · Annals of Biomedical Engineering · doi:10.1007/s10439-009-9740-z · Cited by 314 · see citing works
Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data
Krishnan, R., Natarajan, B., & Warren, S. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2039568 · Cited by 263 · see citing works
Nakajima, K., TAMURA, T., & Miike, H. (1996) · Medical Engineering & Physics · doi:10.1016/1350-4533(95)00066-6 · Cited by 244 · see citing works
The physical significance of time domain descriptors in EEG analysis
Hjorth, B. (1973) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(73)90260-5 · Cited by 210 · see citing works
Effects of hypoxic hypoxia on O<sub>2</sub>uptake and heart rate kinetics during heavy exercise
Engelen, M., Porszasz, J., Riley, M., Wasserman, K., Maehara, K., & Barstow, T. J. (1996) · Journal of Applied Physiology · doi:10.1152/jappl.1996.81.6.2500 · Cited by 207 · see citing works
The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
Barker, S. J., & Shah, N. K. (1997) · Anesthesiology · doi:10.1097/00000542-199701000-00014 · Cited by 197 · see citing works
Signal processing methods for pulse oximetry
Rusch, T. L., Sankar, R. T., & Scharf, J. E. (1996) · Computers in Biology and Medicine · doi:10.1016/0010-4825(95)00049-6 · Cited by 185 · see citing works
The Effect of Motion on Pulse Oximetry and Its Clinical Significance
Petterson, M. T., Begnoche, V. L., & Graybeal, J. M. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000278134.47777.a5 · Cited by 177 · see citing works
Signal quality measures for pulse oximetry through waveform morphology analysis
Sukor, J. A., Redmond, S. J., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/3/008 · Cited by 166 · see citing works
Reduction of motion artifact in pulse oximetry by smoothed pseudo Wigner-Ville distribution
Yan, Y. S., Poon, C. C. Y., & Zhang, Y. (2005) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-2-3 · Cited by 157 · see citing works
Lee, B., Han, J., Baek, H. J., Shin, J. H., Park, K. S., & Yi, W. J. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/12/003 · Cited by 155 · see citing works
Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
Karlen, W., Kobayashi, K., Ansermino, J. M., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1617 · Cited by 123 · see citing works
Statistical approach for the detection of motion/noise artifacts in Photoplethysmogram
Selvaraj, N., Mendelson, Y., Shelley, K. H., Silverman, D. G., & Chon, K. H. (2011) · 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6091232 · Cited by 119 · see citing works
Effects of motion, ambient light, and hypoperfusion on pulse oximeter function
Trivedi, N. S., Ghouri, A. F., Shah, N. K., Lai, E., & Barker, S. J. (1997) · Journal of Clinical Anesthesia · doi:10.1016/s0952-8180(97)00039-1 · Cited by 119 · see citing works
The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals
Lee, H.-W., Lee, J., Jung, W.-G., & Lee, G.-K. (2007) · International Journal of Control Automation and Systems · Cited by 114 · see citing works
A Method for Automatic Identification of Reliable Heart Rates Calculated from ECG and PPG Waveforms
Yu, C., Liu, Z., McKenna, T., Reisner, A. T., & Jaques Reifman, A. (2006) · Journal of the American Medical Informatics Association · doi:10.1197/jamia.m1925 · Cited by 100 · see citing works
Design of filter to reject motion artifact of pulse oximetry
Lee, J., Jung, W.-G., Kang, I., Kim, Y., & Lee, G.-K. (2003) · Computer Standards & Interfaces · doi:10.1016/s0920-5489(03)00077-1 · Cited by 91 · see citing works
Gil, E., Vergara, J. M., & Laguna, P. (2008) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2007.12.002 · Cited by 88 · see citing works
Foo, J. Y. A., & Wilson, S. J. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-005-0008-y · Cited by 74 · see citing works
Li, K., & Warren, S. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2167717 · Cited by 73 · see citing works
Motion resistant pulse oximetry in neonates
Sahni, R., Gupta, A., Ohira-Kist, K., & Rosen, T. S. (2003) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/fn.88.6.f505 · Cited by 61 · see citing works
Li, K., Warren, S., & Natarajan, B. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2157822 · Cited by 55 · see citing works
Ratiometric Artifact Reduction in Low Power Reflective Photoplethysmography
Patterson, J. A. C., & Yang, G. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2161304 · Cited by 46 · see citing works
Reducing motion artifacts in photoplethysmograms by using relative sensor motion: Phantom study
Wijshoff, R. W. C. G. R., Mischi, M., Veen, J., & Van Der Lee, A. M. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.11.117007 · Cited by 40 · see citing works
A characterization of motion affecting pulse oximetry in 350 patients.
Tobin, R. M., Pologe, J. A., & Batchelder, P. B. (2002) · PubMed · Cited by 38 · see citing works
Comtois, G., Mendelson, Y., & Ramuka, P. R. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4352592 · Cited by 36 · see citing works
Naraharisetti, K. V. P., & Bawa, M. (2011) · 2011 IEEE INTERNATIONAL CONFERENCE ON ELECTRO/INFORMATION TECHNOLOGY · doi:10.1109/eit.2011.5978571 · Cited by 35 · see citing works
Analysis of the photoplethysmographic signal by means of the decomposition in principal components
quez, R. H. E. I., Castellanos, M. S., guez, J. F. O. N. R. I., & Hernández-Cáceres, J. L. (2002) · Physiological Measurement · doi:10.1088/0967-3334/23/3/402 · Cited by 22 · see citing works
Real time reconstruction of quasiperiodic multi parameter physiological signals
Ganeshapillai, G., & Guttag, J. V. (2012) · EURASIP Journal on Advances in Signal Processing · doi:10.1186/1687-6180-2012-173 · Cited by 20 · see citing works
Pulse Rate Derivation and Its Correlation with Heart Rate
Chang, K.-M., & Chang, K.-M. (2009) · Journal of Medical and Biological Engineering · Cited by 14 · see citing works
Use of spectral estimation methods for computation of SpO2 from artifact reduced PPG signals
Ram, M. R., Madhav, K. V., Krishna, E. H., Reddy, K. N., & Reddy, K. A. (2011) · IEEE Recent Advances in Intelligent Computational Systems · doi:10.1109/raics.2011.6069349 · Cited by 6 · see citing works
Mannacio, V. A., Tommaso, L. D., Amicis, V. D., Stassano, P., & Vosa, C. (2011) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2011.06.011 · Cited by 5 · see citing works
Application of a fuzzy integral for weak classifiers boosting
Samorodov, A. V. (2011) · Pattern Recognition and Image Analysis · doi:10.1134/s1054661811020945 · Cited by 4 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Current progress of photoplethysmography and SPO2 for health monitoring
TAMURA, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 260 · see citing works
Min, S., An, J., Lee, J. H., Kim, J. H., Joe, D. J., Eom, S. H., Yoo, C. D., Ahn, H., Hwang, J.-Y., Xu, S. H., et al. (2025) · Nature Reviews Cardiology · doi:10.1038/s41569-025-01127-0 · Cited by 132 · see citing works
Dao, D. K., Salehizadeh, S. M. A., Noh, Y., Chong, J. W., Cho, C. H., McManus, D., Darling, C. E., Mendelson, Y., & Chon, K. H. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2612059 · Cited by 104 · see citing works
An Algorithm for Real-Time Pulse Waveform Segmentation and Artifact Detection in Photoplethysmograms
Fischer, C., Dömer, B., Wibmer, T., & Penzel, T. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2518202 · Cited by 91 · see citing works
A Supervised Approach to Robust Photoplethysmography Quality Assessment
Pereira, T., Gadhoumi, K., Ma, M., Liu, X., Xiao, R., Colorado, R. A., Keenan, K. J., Meisel, K. M., & Hu, X. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2909065 · Cited by 80 · see citing works
Robust PPG motion artifact detection using a 1-D convolution neural network
Goh, C., Tan, L. K., Lovell, N. H., Ng, S., Tan, M. P., & Lim, E. (2020) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2020.105596 · Cited by 80 · see citing works
Emerging Technologies for Identifying Atrial Fibrillation
Ding, E., Marcus, G. M., & McManus, D. D. (2020) · Circulation Research · doi:10.1161/circresaha.119.316342 · Cited by 77 · see citing works
Liu, S.-H., Li, R.-X., Wang, J.-J., Chen, W., & Su, C.-H. (2020) · Applied Sciences · doi:10.3390/app10134612 · Cited by 58 · see citing works
Ding, E., Han, D., Whitcomb, C., Bashar, S. K., Adaramola, O., Soni, A., Saczynski, J. S., Fitzgibbons, T. P., Moonis, M., Lubitz, S. A., et al. (2019) · JMIR Cardio · doi:10.2196/13850 · Cited by 52 · see citing works
Pereira, T., Ding, C., Gadhoumi, K., Tran, N., Colorado, R. A., Meisel, K. M., & Hu, X. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab5b84 · Cited by 53 · see citing works
Warren, K., Harvey, J. S., Chon, K. H., & Mendelson, Y. (2016) · Sensors · doi:10.3390/s16030342 · Cited by 47 · see citing works
Mahmoudzadeh, A., Azimi, I., Rahmani, A. M., & Liljeberg, P. (2021) · Procedia Computer Science · doi:10.1016/j.procs.2021.03.025 · Cited by 47 · see citing works
Hravnak, M., Chen, L., Dubrawski, A. W., Bose, E. L., Clermont, G., & Pinsky, M. R. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9788-2 · Cited by 43 · see citing works
Guo, Z., Ding, C., Hu, X., & Rudin, C. D. (2021) · Physiological Measurement · doi:10.1088/1361-6579/ac3b3d · Cited by 40 · see citing works
Adaptive template matching of photoplethysmogram pulses to detect motion artefact
Lim, P. K., Ng, S., Lovell, N. H., Yu, Y. P., Tan, M. P., McCombie, D. B., Lim, E., & Redmond, S. J. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aadf1e · Cited by 39 · see citing works
Bashar, S. K., Han, D., Soni, A., McManus, D. D., & Chon, K. H. (2018) · IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI ...) · doi:10.1109/bhi.2018.8333374 · Cited by 38 · see citing works
Tarvirdizadeh, B., Golgouneh, A., Tajdari, F., & Khodabakhshi, E. (2018) · Neural Computing and Applications · doi:10.1007/s00521-018-3767-8 · Cited by 37 · see citing works
Liu, S.-H., Wang, J.-J., Chen, W., Pan, K., & Su, C.-H. (2020) · Applied Sciences · doi:10.3390/app10041476 · Cited by 35 · see citing works
Hajeb-Mohammadalipour, S., Ahmadi, M., Shahghadami, R., & Chon, K. H. (2018) · Sensors · doi:10.3390/s18072090 · Cited by 34 · see citing works
Rehman, R. Z. U., Chatterjee, M., Manyakov, N. V., Daans, M., Jackson, A., O’Brisky, A., Telesky, T., Smets, S., Berghmans, P.-J., Yang, D., et al. (2024) · Sensors · doi:10.3390/s24216826 · Cited by 36 · see citing works
Liu, S.-H., Liu, H.-C., Chen, W., & Tan, T.-H. (2020) · IEEE Access · doi:10.1109/access.2020.3029842 · Cited by 33 · see citing works
PPGMotion: Model-based detection of motion artifacts in photoplethysmography signals
Maity, A. K., Veeraraghavan, A., & Sabharwal, A. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.103632 · Cited by 32 · see citing works
Tabei, F., Kumar, R., Phan, T. N., McManus, D. D., & Chong, J. W. (2018) · IEEE Access · doi:10.1109/access.2018.2875873 · Cited by 32 · see citing works
A Survey of Photoplethysmography and Imaging Photoplethysmography Quality Assessment Methods
Desquins, T., Bousefsaf, F., Pruski, A., & Maaoui, C. (2022) · Applied Sciences · doi:10.3390/app12199582 · Cited by 33 · see citing works
Shoushan, M. M., Reyes, B. A., Mejia-Rodriguez, A. R., & Chong, J. W. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.2998399 · Cited by 30 · see citing works
Detection and Characterization of Physical Activity and Psychological Stress from Wristband Data
Sevil, M., Rashid, M. M., Askari, M. R., Maloney, Z., Hajizadeh, I., & Çınar, A. (2020) · Signals · doi:10.3390/signals1020011 · Cited by 29 · see citing works
Feli, M., Azimi, I., Anzanpour, A., Rahmani, A. M., & Liljeberg, P. (2023) · Smart Health · doi:10.1016/j.smhl.2023.100390 · Cited by 29 · see citing works
Harvey, J. R., Salehizadeh, S. M. A., Mendelson, Y., & Chon, K. H. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2837499 · Cited by 29 · see citing works
Reljin, N., Zimmer, G., Malyuta, Y., Shelley, K. H., Mendelson, Y., Blehar, D., Darling, C. E., & Chon, K. H. (2018) · PLoS ONE · doi:10.1371/journal.pone.0195087 · Cited by 28 · see citing works
Gautam, Y., & Jebelli, H. (2024) · Automation in Construction · doi:10.1016/j.autcon.2024.105537 · Cited by 27 · see citing works
Quantifying Movement in Preterm Infants Using Photoplethysmography
Zuzarte, I., Indic, P., Sternad, D., & Paydarfar, D. (2018) · Annals of Biomedical Engineering · doi:10.1007/s10439-018-02135-7 · Cited by 25 · see citing works
Motion and Noise Artifact-Resilient Atrial Fibrillation Detection Using a Smartphone
Chong, J. W., Cho, C. H., Tabei, F., Le-Anh, D., Esa, N., McManus, D. D., & Chon, K. H. (2018) · IEEE Journal on Emerging and Selected Topics in Circuits and Systems · doi:10.1109/jetcas.2018.2818185 · Cited by 25 · see citing works
Characterization of Artifact Signals in Neck Photoplethysmography
García-López, I., & Rodriguez–Villegas, E. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.2972378 · Cited by 22 · see citing works
SmartFit: Smartphone Application for Garment Fit Detection
Foysal, K. H., Chang, H. J., Bruess, F., & Chong, J. W. (2021) · Electronics · doi:10.3390/electronics10010097 · Cited by 21 · see citing works
Low-Pass NGD Numerical Function and STM32 MCU Emulation Test
Ravelo, B., Guerin, M., Rahajandraibe, W., Giès, V., Rajaoarisoa, L., & Lalléchère, S. (2021) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2021.3109543 · Cited by 19 · see citing works
Lin, W.-J., & Ma, H. (2016) · International Conference on Communications · doi:10.1109/icc.2016.7511485 · Cited by 18 · see citing works
Prada, E. J. A., & García, J. F. C. (2024) · Sensors · doi:10.3390/s24227193 · Cited by 18 · see citing works
Motion Artifact Detection in PPG Signals Based on Gramian Angular Field and 2-D-CNN
Liu, X., Hu, Q., Yuan, H., & Yang, C. (2020) · International Congress on Image and Signal Processing (CISP) · doi:10.1109/cisp-bmei51763.2020.9263630 · Cited by 15 · see citing works
Tabei, F., Zaman, R., Foysal, K. H., Kumar, R., Kim, Y., & Chong, J. W. (2019) · PLoS ONE · doi:10.1371/journal.pone.0218248 · Cited by 13 · see citing works
Smartwatch Based Atrial Fibrillation Detection from Photoplethysmography Signals
Bashar, S. K., Han, D., Ding, E., Whitcomb, C., McManus, D. D., & Chon, K. H. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8856928 · Cited by 12 · see citing works
Detection of artifacts on photoplethysmography signals using random distortion testing
Cherif, S., Pastor, D., Nguyen, Q.-T., & L’Her, E. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7592148 · Cited by 11 · see citing works
Kasbekar, R. S., & Mendelson, Y. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aacfe5 · Cited by 10 · see citing works
McLean, M. K., Weaver, R. G., Lane‐Cordova, A. D., Smith, M. T., Parker, H., Stone, B., McAninch, J., Matolak, D. W., Burkart, S. A., Chandrashekhar, M. V. S., et al. (2023) · Sensors · doi:10.3390/s23073429 · Cited by 11 · see citing works
TAMURA, T., & Maeda, Y. (2017) · Springer eBooks · doi:10.1007/978-3-319-69362-0_6 · Cited by 7 · see citing works
Obi, A. I. (2022) · International Electronic Conference on Sensors and Applications (ECSA) · doi:10.3390/engproc2021010077 · Cited by 7 · see citing works
A Novel Signal Restoration Method of Noisy Photoplethysmograms for Uninterrupted Health Monitoring
Vraka, A., Zangróniz, R., Quesada, A., Hornero, F., Alcaraz, R., & Rieta, J. J. (2023) · Sensors · doi:10.3390/s24010141 · Cited by 7 · see citing works
Artifacts classification and apnea events detection in neck photoplethysmography signals
García-López, I., Pramono, R. X. A., & Rodriguez–Villegas, E. (2022) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02666-1 · Cited by 6 · see citing works
Motion and Noise Artifact-Resilient Atrial Fibrillation Detection Using a Smartphone
Zaman, R., Chong, J. W., Cho, C. H., Esa, N., McManus, D. D., & Chon, K. H. (2016) · IEEE/ACM Conference on Connected Health: Applications, Systems and Engineering Technologies · doi:10.1109/chase.2016.75 · Cited by 6 · see citing works
A Novel Abnormal Segments Detection Method for Photopletysmography Signal
Teng, Y., Ge, L., & Chou, Y. (2018) · Chinese Control Conference (CCC) · doi:10.23919/chicc.2018.8483815 · Cited by 6 · see citing works
Chou, Y., Xu, B., Zhang, R., Feng, Y., & Jin, Y. (2016) · International Conference on Control, Automation and Information Sciences (ICCAIS) · doi:10.1109/iccais.2016.7822460 · Cited by 5 · see citing works
Pre-processing of Photoplethysmographic Waveform for Amplitude Regularization
Kim, J., Lee, J., & Shin, H. (2019) · Journal of Electrical Engineering and Technology · doi:10.1007/s42835-019-00185-y · Cited by 5 · see citing works
A Novel Re-Targetable Application Development Platform for Healthcare Mobile ApplicationsPreprint
Cho, C. H., Tabei, F., Phan, T. N., Kim, Y., & Chong, J. W. (2019) · arXiv (Cornell University) · doi:10.48550/arxiv.1903.05783 · Cited by 5 · see citing works
Holmer, M., Sandberg, F., Solem, K., Olde, B., & Sörnmo, L. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/9/1499 · Cited by 4 · see citing works
Gu, J., Tomioka, Y., Kaneko, A., Enomoto, S., Saito, I., Okazaki, M., Someya, T., & Sekino, M. (2019) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2019.100257 · Cited by 4 · see citing works
Mao, J., Li, H., Yu, J., Wu, H., López, M. B., & Zhao, Y. (2024) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2024.107191 · Cited by 3 · see citing works
Huang, T., Yuan, Z., Ran, W., & Shen, G. (2026) · Science China Materials · doi:10.1007/s40843-026-4069-0 · Cited by 2 · see citing works
A novel heart rate monitoring method using a smartphone
Zaman, R., Cho, C. H., Kim, Y., & Chong, J. W. (2016) · IEEE Healthcare Innovations and Point of Care Technologies Conference (HI-POCT) · doi:10.1109/hic.2016.7797683 · Cited by 2 · see citing works
Real-Time Pulse Waveform Profiling Algorithm for Wearable Applications
Liu, H., Fujita, D., Zhang, L., & Suzuki, A. (2018) · IEEE Access · doi:10.1109/access.2018.2875548 · Cited by 2 · see citing works
Tabei, F., Askarian, B., & Chong, J. W. (2019) · IEEE Healthcare Innovations and Point of Care Technologies Conference (HI-POCT) · doi:10.1109/hi-poct45284.2019.8962833 · Cited by 2 · see citing works
Photoplethysmography Signal Quality Assessment using Attentive-CNN Models
Silva, L., Lima, R., Lucafo, G. D., Sandoval, I., Freitas, P. G., & Penatti, O. A. B. (2024) · doi:10.5753/sbcas.2024.2206 · Cited by 1 · see citing works
Robotic surgery and cardiac biosignals: Bridging human–artificial intelligence collaboration
Chandra, P., Ali, Z., Fatma, N., & Sachan, N. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-45242-0.00008-6 · Cited by 1 · see citing works
Pulse‐based photoplethysmography quality assessment improves wearable seizure detection performance
Shahbakhti, M., Bosch, A. T., Long, X., van Dijk, J. P., & Thijs, R. D. (2026) · Epilepsia Open · doi:10.1002/epi4.70242 · Cited by 1 · see citing works
A Reflectance–based multimodal wearable photoplethysmography (PPG) sensor
Durbha, R., & Koomson, V. J. (2026) · Review of Scientific Instruments · doi:10.1063/5.0304023 · Cited by 1 · see citing works
Autonomous Heart Rate Tracking Methodology Using Kalman Filter and the EM Algorithm
Souza, T. T., Balasingam, B., & Maev, R. G. (2019) · International Conference on Information Fusion (FUSION) · doi:10.23919/fusion43075.2019.9011407 · Cited by 1 · see citing works
Optical difference in the frequency domain to suppress disturbance for wearable electronics
Li, H., Wang, Z., Cao, Y. K., Ma, Y., & Feng, X. (2020) · Biomedical Optics Express · doi:10.1364/boe.403033 · Cited by 1 · see citing works
Monitoring of Breathing Activity using Smartphone-acquired Signals
Reyes, B. A. (2015) · OpenCommons at University of Connecticut (University of Connecticut) · Cited by 1 · see citing works
Movement as a vital sign in neonates
Zuzarte, I. (2020) · doi:10.17760/d20385584 · Cited by 1 · see citing works
Wearable Wireless Sensor for Multi-Scale Physiological Monitoring
Chon, K. H., & Mendelson, Y. (2013) · doi:10.21236/ada590832
Salehizadeh, S. M. A., & Chon, K. H. (2018) · International Conference on Information and Communication Technology Convergence (ICTC) · doi:10.1109/ictc.2018.8539413
Kaduk, S. I., Roberts, A. P. J., & Stanton, N. A. (2019) · Advances in intelligent systems and computing · doi:10.1007/978-3-030-20148-7_19
Disease Classification Using Photoplethysmography
Govind, I., & Surling, N. (2018) · Journal of Emerging Technologies and Innovative Research
Telemedicine systems to manage chronic disease
Biswas, U., Goh, C., Ooi, S., Lim, E., Redmond, S. J., & Lovell, N. H. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-818914-6.00020-x
Estimation of Non-invasive Cuff-less Blood Pressure Using the Photoplethysmogram SignalPreprint
Rajagopal, S., Manimegalai, P., & Ramachandaran, D. (2020) · doi:10.21203/rs.3.rs-22297/v1
An in Situ Embedded System for Electrocardiography and Photoplethysmography Acquisition
Huang, Z., Luo, Z., Lyu, L., & Wu, X. Y. (2022) · International Conference on Electronics Technology (ICET) · doi:10.1109/icet55676.2022.9824681
Hardware and Algorithmic Approaches to Combat Motion Artifacts in Photoplethysmographic Data
Han, D., Mohagheghian, F., & Chon, K. H. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-822548-6.00135-7
Motion noise cancellation algorithm in PPG signals based on temporal variability analysisPreprint
Cordero-Fernández, G., Cebada-Fuentes, R., & Sanchez-Perez, C. A. (2025) · SSRN Electronic Journal · doi:10.2139/ssrn.5477679
Benkert, A., Bludau, J., Boborzi, L., Prueckner, S., & Schniepp, R. (2026) · Sensors · doi:10.3390/s26010324 · Cited by 1 · see citing works
- Authors
- Chong, J. W., Dao, D. K., Salehizadeh, S. M. A., McManus, D. D., Darling, C. E., Chon, K. H., & Mendelson, Y.
- Affiliations
- Worcester Polytechnic Institute (US)
- University of Massachusetts Chan Medical School (US)
- Corresponding author
- Chon, Ki H.
- Venue
- Annals of Biomedical Engineering
- Published
- 2014-08-05
- Volume
- 42
- Issue
- 11
- Pages
- 2238-2250
- ISSN
- 0090-6964 (ISSN-L)
- 1573-9686
- OpenAlex
- W2042388577
- Cited by
- 79
- References
- 41
- Citations by year
- 2026: 5 · 2025: 2 · 2024: 5 · 2023: 3 · 2022: 6 · 2021: 6 · 2020: 12 · 2019: 12 · 2018: 15 · 2017: 1 · 2016: 9 · 2015: 2 · 2013: 1
- FWCI
- 2.07
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- The Journal of the American Medical Association (JAMA) Network (American Medical Association) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- photoplethysmogram, motion artifact detection, support vector machine, time-domain features, heart rate estimation, oxygen saturation, vital signs monitoring, real-time processing
- MeSH terms
- Algorithms, Heart Rate, Humans, Monitoring, Physiologic, Motion, Oximetry, Artifacts, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review