Sensitivity of MEG and EEG to Source Orientation
Open accessThe DOI matches its Crossref record; 37 of the 37 references it cites were found in a registry or an index.
Abstract
An important difference between magnetoencephalography (MEG) and electroencephalography (EEG) is that MEG is insensitive to radially oriented sources. We quantified computationally the dependency of MEG and EEG on the source orientation using a forward model with realistic tissue boundaries. Similar to the simpler case of a spherical head model, in which MEG cannot see radial sources at all, for most cortical locations there was a source orientation to which MEG was insensitive. The median value for the ratio of the signal magnitude for the source orientation of the lowest and the highest sensitivity was 0.06 for MEG and 0.63 for EEG. The difference in the sensitivity to the source orientation is expected to contribute to systematic differences in the signal-to-noise ratio between MEG and EEG.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures


Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source doesn’t support the claim as written.
In the citing paper
“Sources in the region of the SMA with a close to perpendicular cortical (dipole) orientation to the skull electrodes may be more readily detected by MEG than EEG (Ahlfors et al., 2010).”
From Oscillating central motor networks in pathological tremors and voluntary movements. What makes the difference? (2012) · cites this paper as Ahlfors et al., 2010
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This difference may be due to the use of MEG recordings in previous studies, and MEG may be more sensitive to cortical sources than EEG.”
From Oscillating central motor networks in pathological tremors and voluntary movements. What makes the difference? (2012) · cites this paper as Ahlfors et al., 2010
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“However, previous studies were all MEG-based, which may be slightly inferior to EEG in detecting deep radial sources (Ahlfors et al., 2010)”
From Oscillating central motor networks in pathological tremors and voluntary movements. What makes the difference? (2012) · cites this paper as Ahlfors et al., 2010
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Cortical Surface-Based Analysis
Dale, A. M., Fischl, B., & Sereno, M. I. (1999) · NeuroImage · doi:10.1006/nimg.1998.0395 · Cited by 10,910 · cited 1×
“A surface mesh of the gray-white matter border of the cerebral cortex with about 300 000 vertices was constructed from the MRI using the Free Surfer software (Dale et al. 1999; Fischl et al. 1999).”
MethodCortical Surface-Based Analysis
Fischl, B., Sereno, M. I., & Dale, A. M. (1999) · NeuroImage · doi:10.1006/nimg.1998.0396 · Cited by 6,147 · cited 1×
“A surface mesh of the gray-white matter border of the cerebral cortex with about 300 000 vertices was constructed from the MRI using the Free Surfer software (Dale et al. 1999; Fischl et al. 1999).”
MethodHämäläinen, M., Hari, R., Ilmoniemi, R. J., Knuutila, J., & Lounasmaa, O. V. (1993) · Reviews of Modern Physics · doi:10.1103/revmodphys.65.413 · Cited by 4,684 · cited 2×
“One of the main differences between magnetoencephalography (MEG) and electroencephalography (EEG) is thought to be that MEG is mainly sensitive to tangentially oriented sources, whereas EEG can detect sources of all orientations (Baillet et al. 2001; Cohen and Cuffin 1991; Hämäläinen et al. 1993).”
BackgroundSelf-citationBaillet, S., Mosher, J. C., & Leahy, R. M. (2001) · IEEE Signal Processing Magazine · doi:10.1109/79.962275 · Cited by 1,692 · cited 2×
“One of the main differences between magnetoencephalography (MEG) and electroencephalography (EEG) is thought to be that MEG is mainly sensitive to tangentially oriented sources, whereas EEG can detect sources of all orientations (Baillet et al. 2001; Cohen and Cuffin 1991; Hämäläinen et al. 1993).”
BackgroundGeddes, L. A., & Baker, L. E. (1967) · Medical & Biological Engineering · doi:10.1007/bf02474537 · Cited by 1,610 · cited 2×
“The conductivity ratios of 1:0.0125:1 were assumed for brain:skull:scalp ( Geddes and Baker 1967 ).”
BackgroundInfluentialRealistic conductivity geometry model of the human head for interpretation of neuromagnetic data
Hamalainen, M. S., & Sarvas, J. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.16463 · Cited by 989 · cited 4×
“…accordance with those found in previous studies comparing dipole localization accuracy in spherically symmetric volume conductor model vs. anatomically more realistic head models (Hämäläinen and Sarvas 1989; Haueisen et al. 2002; Schimpf et al. 2002; Tarkiainen et al. 2003; Wolters et al. 2006).”
BackgroundMethodResultInfluentialEEG and MEG: forward solutions for inverse methods
Mosher, J. C., Leahy, R. M., & Lewis, P. S. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.748978 · Cited by 839 · cited 2×
“Silent source elements are a special case of silent source distributions defined by the null space of the forward matrix (Mosher et al. 1999; Riera et al. 2006).”
BackgroundMethodA sensor-weighted overlapping-sphere head model and exhaustive head model comparison for MEG
Huang, M. X., Mosher, J. C., & Leahy, R. M. (1999) · Physics in Medicine and Biology · doi:10.1088/0031-9155/44/2/010 · Cited by 612 · cited 2×
“…guarantee that the spherical model will provide an accurate MEG forward solution for those locations, previous studies have suggested that often the spherical model does indeed provide an accurate forward model for MEG, especially for most of the superficial cortical regions (Huang et al. 1999).”
BackgroundMethodWolters, C. H., Anwander, A., Tricoche, X., Weinstein, D., Koch, M. A., & MacLeod, R. S. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2005.10.014 · Cited by 452 · cited 4×
“…accordance with those found in previous studies comparing dipole localization accuracy in spherically symmetric volume conductor model vs. anatomically more realistic head models (Hämäläinen and Sarvas 1989; Haueisen et al. 2002; Schimpf et al. 2002; Tarkiainen et al. 2003; Wolters et al. 2006).”
MethodResultInfluentialA Quantitative Assessment of the Sensitivity of Whole-Head MEG to Activity in the Adult Human Cortex
Hillebrand, A., & Barnes, G. R. (2002) · NeuroImage · doi:10.1006/nimg.2002.1102 · Cited by 428 · cited 2×
“Few cortical sources, however, are expected to have exactly the orientation of the lowest sensitivity in MEG (Hillebrand and Barnes 2002).”
BackgroundMonte Carlo simulation studies of EEG and MEG localization accuracy
Liu, A. K., Dale, A. M., & Belliveau, J. W. (2002) · Human Brain Mapping · doi:10.1002/hbm.10024 · Cited by 349 · cited 2×
“The relative magnitude of the MEG signal generated by a radial and a tangential source at the same location was examined experimentally by Melcher and Cohen (1988), who found that the signals differed by a factor of about 6 for several locations in the rabbit brain.”
BackgroundMethodSelf-citationDemonstration of useful differences between magnetoencephalogram and electroencephalogram
Cohen, D., & Cuffin, B. N. (1983) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(83)90005-6 · Cited by 300 · cited 1×
“This is consistent with MEG being most sensitive to superficial sources, and the sensitivity being much reduced for deep sources (Cohen and Cuffin 1983; de Jongh et al. 2005; Goldenholz et al. 2009; Hillebrand and Barnes 2002).”
BackgroundGoldenholz, D. M., Ahlfors, S. P., Hämäläinen, M. S., Sharon, D., Ishitobi, M., Vaina, L. M., & Stufflebeam, S. M. (2009) · Human Brain Mapping · doi:10.1002/hbm.20571 · Cited by 266 · cited 2×
“This is consistent with MEG being most sensitive to superficial sources, and the sensitivity being much reduced for deep sources (Cohen and Cuffin 1983; de Jongh et al. 2005; Goldenholz et al. 2009; Hillebrand and Barnes 2002).”
BackgroundSelf-citationThe Influence of Brain Tissue Anisotropy on Human EEG and MEG
Haueisen, J., Tuch, D. S., Ramon, C., Schimpf, P. H., Wedeen, V. J., George, J. S., & Belliveau, J. W. (2002) · NeuroImage · doi:10.1006/nimg.2001.0962 · Cited by 228 · cited 3×
“…accordance with those found in previous studies comparing dipole localization accuracy in spherically symmetric volume conductor model vs. anatomically more realistic head models (Hämäläinen and Sarvas 1989; Haueisen et al. 2002; Schimpf et al. 2002; Tarkiainen et al. 2003; Wolters et al. 2006).”
BackgroundResultSelf-citationModel Studies of the Magnetocardiogram
Grynszpan, F., & Geselowitz, D. B. (1973) · Biophysical Journal · doi:10.1016/s0006-3495(73)86034-5 · Cited by 226 · cited 1×
“Theoretically, a radially oriented current dipole produces no magnetic field outside a spherically symmetric volume conductor (Baule and McFee 1965; Grynszpan and Geselowitz 1973).”
BackgroundBoundary element method volume conductor models for EEG source reconstruction
Fuchs, M., Wagner, M., & Kastner, J. (2001) · Clinical Neurophysiology · doi:10.1016/s1388-2457(01)00589-2 · Cited by 165 · cited 1×
“Some caution is necessary in interpreting our results, as numerical accuracy of the forward model for sources close to the skull may suffer from the sparseness of the BEM vertices (Fuchs et al. 2001).”
The value of multichannel MEG and EEG in the presurgical evaluation of 70 epilepsy patients
Knake, S., Halgren, E., Shiraishi, H., Hara, K., Hamer, H. M., Grant, P. E., Carr, V. A., Foxe, D., Camposano, S., Busa, E., et al. (2006) · Epilepsy Research · doi:10.1016/j.eplepsyres.2006.01.001 · Cited by 165 · cited 1×
“Several studies have reported that epileptic spikes are sometimes detected in either EEG or MEG but not necessarily in both, even when EEG and MEG are recorded simultaneously (e.g., Knake et al. 2006; Lin et al. 2003).”
BackgroundSelf-citationSources of auditory brainstem responses revisited: Contribution by magnetoencephalography
Parkkonen, L., Fujiki, N., & Mäkelä, J. P. (2009) · Human Brain Mapping · doi:10.1002/hbm.20788 · Cited by 161 · cited 1×
“This is particularly important for deep cortical and subcortical regions (Parkkonen et al. 2009; Tesche et al. 1996).”
BackgroundCancellation of EEG and MEG signals generated by extended and distributed sources
Ahlfors, S. P., Han, J., Lin, F., Witzel, T., Belliveau, J. W., Hämäläinen, M., & Halgren, E. (2009) · Human Brain Mapping · doi:10.1002/hbm.20851 · Cited by 161 · cited 2×
“More likely, however, the SNR for these spikes is higher in MEG than in EEG because of selective cancellation of background brain noise from tangentially oriented sources, such that in EEG the spikes are obscured by radially oriented background sources (Ahlfors et al. 2010).”
BackgroundSelf-citationDipole models for the EEG and MEG
Schimpf, P. H., Ramon, C., & Haueisen, J. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.995679 · Cited by 156 · cited 2×
“…accordance with those found in previous studies comparing dipole localization accuracy in spherically symmetric volume conductor model vs. anatomically more realistic head models (Hämäläinen and Sarvas 1989; Haueisen et al. 2002; Schimpf et al. 2002; Tarkiainen et al. 2003; Wolters et al. 2006).”
BackgroundResultEffects of head shape on EEGs and MEGs
Cuffin, B. N. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.43614 · Cited by 153 · cited 2×
“However, since the human head is not exactly spherically symmetric, the radial orientation is not well defined, and an approximately radial source in the brain is not necessarily silent in MEG ( Cuffin 1990 ).”
BackgroundTheory of Magnetic Detection of the Heart's Electrical Activity
Baule, G., & McFee, R. (1965) · Journal of Applied Physics · doi:10.1063/1.1714404 · Cited by 112 · cited 1×
“Theoretically, a radially oriented current dipole produces no magnetic field outside a spherically symmetric volume conductor (Baule and McFee 1965; Grynszpan and Geselowitz 1973).”
BackgroundA novel integrated MEG and EEG analysis method for dipolar sources
Huang, M., Song, T., Hagler, D. J., Podgorny, I. A., Jousmäki, V., Cui, L., Gaa, K., Harrington, D. L., Dale, A. M., Lee, R. R., et al. (2007) · NeuroImage · doi:10.1016/j.neuroimage.2007.06.002 · Cited by 112 · cited 2×
“…M. S. Hämäläinen MGH/HST Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital/Harvard Medical School, 149 13th Street, Rm 2301, Charlestown, MA 02129, USA e-mail: seppo@nmr.mgh.harvard.edu S. P. Ahlfors J. W. Belliveau M. S. Hämäläinen Harvard-MIT Division…”
BackgroundRole of Soft Bone, CSF and Gray Matter in EEG Simulations
Ramon, C., Schimpf, P. H., Haueisen, J., Holmes, M., & Ishimaru, A. (2004) · Brain Topography · doi:10.1023/b:brat.0000032859.68959.76 · Cited by 91 · cited 1×
“Further anatomical information about conductivity properties of the head could be taken into account by using finite element modeling (FEM) (Haueisen et al. 1995; Haueisen et al. 2002; Ramon et al. 2004; Wolters et al. 2006).”
The 3D topography of MEG source localization accuracy: effects of conductor model and noise
Tarkiainen, A., Liljeström, M., Seppä, M., & Salmelin, R. H. (2003) · Clinical Neurophysiology · doi:10.1016/s1388-2457(03)00195-0 · Cited by 90 · cited 2×
“…accordance with those found in previous studies comparing dipole localization accuracy in spherically symmetric volume conductor model vs. anatomically more realistic head models (Hämäläinen and Sarvas 1989; Haueisen et al. 2002; Schimpf et al. 2002; Tarkiainen et al. 2003; Wolters et al. 2006).”
BackgroundResultCombining fMRI with EEG and MEG in order to relate patterns of brain activity to cognition
FREEMAN, W. (2009) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2008.12.019 · Cited by 90 · cited 1×
“For extended sources, selective cancellation of MEG and EEG signals generated by tangential sources may occur (Ahlfors et al. 2010; Eulitz et al. 1997; Freeman et al. 2009).”
BackgroundEEG versus MEG localization accuracy: Theory and experiment
Cohen, D., & Cuffin, B. N. (1991) · Brain Topography · doi:10.1007/bf01132766 · Cited by 88 · cited 2×
“One of the main differences between magnetoencephalography (MEG) and electroencephalography (EEG) is thought to be that MEG is mainly sensitive to tangentially oriented sources, whereas EEG can detect sources of all orientations (Baillet et al. 2001; Cohen and Cuffin 1991; Hämäläinen et al. 1993).”
Backgroundde Jongh, A., de Munck, J. C., alves, S. N. I. G., & Ossenblok, P. (2005) · Journal of Clinical Neurophysiology · doi:10.1097/01.wnp.0000158947.68733.51 · Cited by 88 · cited 2×
“This is consistent with MEG being most sensitive to superficial sources, and the sensitivity being much reduced for deep sources (Cohen and Cuffin 1983; de Jongh et al. 2005; Goldenholz et al. 2009; Hillebrand and Barnes 2002).”
BackgroundOn the influence of volume currents and extended sources on neuromagnetic fields: A simulation study
Haueisen, J., Ramon, C., Czapski, P., & Eiselt, M. (1995) · Annals of Biomedical Engineering · doi:10.1007/bf02584472 · Cited by 87 · cited 8×
“This measure of relative sensitivity to the ‘‘suppressed’’ source orientation is analogous to the suppression ratio for radial vs. tangential dipoles studied previously (Haueisen et al. 1995; Melcher and Cohen 1988).”
BackgroundMethodResultInfluentialMagnetoencephalographic yield of interictal spikes in temporal lobe epilepsy
Lin, Y., Shih, Y., Hsieh, J., Yu, H., Yiu, C. H., Wong, T., Yeh, T., Kwan, S.-Y., Ho, L. ‐., Yen, D., et al. (2003) · NeuroImage · doi:10.1016/s1053-8119(03)00181-2 · Cited by 76 · cited 1×
“Several studies have reported that epileptic spikes are sometimes detected in either EEG or MEG but not necessarily in both, even when EEG and MEG are recorded simultaneously (e.g., Knake et al. 2006; Lin et al. 2003).”
BackgroundRegional Differences in the Sensitivity of MEG for Interictal Spikes in Epilepsy
Huiskamp, G. J. M., Agirre‐Arrizubieta, Z., & Leijten, F. S. S. (2010) · Brain Topography · doi:10.1007/s10548-010-0134-1 · Cited by 60 · cited 2×
“ative signal-to-noise ratio (SNR) of MEG and EEG (de Jongh et al. 2005; Goldenholz et al. 2009; Huiskamp et al. 2010).”
BackgroundNon-invasive detection of neuronal population activity in human hippocampus
Tesche, C. D., Karhu, J., & Tissari, S. O. (1996) · Cognitive Brain Research · doi:10.1016/0926-6410(95)00044-5 · Cited by 56 · cited 1×
“This is particularly important for deep cortical and subcortical regions (Parkkonen et al. 2009; Tesche et al. 1996).”
BackgroundYvert, B., Bertrand, O., Echallier, J. F., & Pernier, J. (1996) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0921-884x(96)95691-x · Cited by 46 · cited 1×
“Previous studies have suggested that the resolution used here provides a reasonable accuracy for most cortical source locations (Yvert et al. 1996).”
BackgroundDependence of the MEG on dipole orientation in the rabbit head.
Melcher, J., & Cohen, D. (1988) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(88)90024-7 · Cited by 39 · cited 8×
“This measure of relative sensitivity to the ‘‘suppressed’’ source orientation is analogous to the suppression ratio for radial vs. tangential dipoles studied previously (Haueisen et al. 1995; Melcher and Cohen 1988).”
BackgroundMethodResultInfluentialHead position in the MEG helmet affects the sensitivity to anterior sources.
Marinković, K., Cox, B., Reid, K., & Halgren, E. (2004) · PubMed · Cited by 32
Differential outcomes from magneto- and electroencephalography for the analysis of human cognition
Eulitz, C., Eulitz, H., & Elbert, T. R. (1997) · Neuroscience Letters · doi:10.1016/s0304-3940(97)00333-9 · Cited by 31 · cited 2×
“For extended sources, selective cancellation of MEG and EEG signals generated by tangential sources may occur (Ahlfors et al. 2010; Eulitz et al. 1997; Freeman”
BackgroundA theoretical formulation of the electrophysiological inverse problem on the sphere
Riera, J. J., Valdes-Sosa, P. A., Tanabe, K., & Kawashima, R. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/7/007 · Cited by 10 · cited 1×
“Silent source elements are a special case of silent source distributions defined by the null space of the forward matrix (Mosher et al. 1999; Riera et al. 2006).”
Background
Cited by
EEG and MEG: Relevance to Neuroscience
Lopes da Silva, F. (2013) · Neuron · doi:10.1016/j.neuron.2013.10.017 · Cited by 900
“Theoretically, a radially oriented dipolar source does not give rise to a magnetic field outside a spherical volume conductor; consequently the MEG is not sensitive to radial components of dipolar sources but to the tangential components (Ahlfors et al., 2010).”
Electroencephalographic source imaging: a prospective study of 152 operated epileptic patients
Brodbeck, V., Spinelli, L., Lascano, A. M., Wissmeier, M., Vargas, M.-I., Vulliemoz, S., Pollo, C., Schaller, K., Michel, C. M., & Seeck, M. (2011) · Brain · doi:10.1093/brain/awr243 · Cited by 427
“There are concerns that MEG may miss deep sources and that it is insensitive to sources with radial orientation (Ahlfors et al., 2010), which appears to be less of an issue in EEG (Lejten et al.”
BackgroundRoux, F., Wibral, M., Singer, W., Aru, J., & Uhlhaas, P. J. (2013) · The Journal of Neuroscience · doi:10.1523/jneurosci.5778-12.2013 · Cited by 169
Muthuraman, M., Heute, U., Arning, K., Anwar, A. R., Elble, R., Deuschl, G., & Raethjen, J. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2012.01.088 · Cited by 134
“However, previous studies were all MEG-based, which may be slightly inferior to EEG in detecting deep radial sources (Ahlfors et al., 2010), and these studies did not establish the tremor network with respect to the rhythmic EMG signal but to the cortical region with the strongest coherence with EMG (typically M1).”
Babo-Rebelo, M., Wolpert, N., Adam, C., Hasboun, D., & Tallon-Baudry, C. (2016) · Philosophical Transactions of the Royal Society B: Biological Sciences · doi:10.1098/rstb.2016.0004 · Cited by 106
Metadata
- Authors
- Ahlfors, S. P., Han, J., Belliveau, J. W., & Hämäläinen, M. S.
- Affiliations
- Harvard University (US)
- Corresponding author
- Ahlfors, Seppo P.
- Venue
- Brain Topography
- Published
- 2010-07-18
- Volume
- 23
- Issue
- 3
- Pages
- 227-232
- ISSN
- 0896-0267 (ISSN-L)
- 1573-6792
- PMC
- PMC2914866
- OpenAlex
- W1565410510
- Semantic Scholar
- 061ed32cfd3870a9eedaa09036dc1811843b7af3
- Cited by
- 1
- References
- 37
- Citations by year
- 2023: 1
- FWCI
- 0.00
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc
- Locations
- Brain Topography (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- DSpace@MIT (Massachusetts Institute of Technology) (submittedVersion, cc-by-nc)
- Springer US
- Funders
- National Institutes of Health (NS037462)
- National Institute of Neurological Disorders and Stroke (R01NS057500, R56NS057500, -NS037462, R01NS037462, NS057500, R56NS037462)
- National Center for Research Resources (P41RR014075)
- Eunice Kennedy Shriver National Institute of Child Health and Human Development (HD 040712, R01HD040712)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Life Sciences › Neuroscience › Cognitive Neuroscience › Neural dynamics and brain function
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Keywords
- magnetoencephalography, EEG, source orientation, realistic head model, forward model, signal-to-noise ratio, cortical sources
- MeSH terms
- Adult, Brain Mapping, Cerebral Cortex, Electroencephalography, Humans, Male, Models, Neurological, Orientation, Sensitivity and Specificity, Signal Processing, Computer-Assisted, Magnetoencephalography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.