Investigation of peripheral photoplethysmographic morphology changes induced during a hand-elevation study
Abstract
A hand-elevation study was carried out in the laboratory in order to alter peripheral blood flow with the aim of increasing understanding of factors affecting the morphology of peripheral photoplethysmographic signals. Photoplethysmographic (PPG) signals were recorded from twenty healthy volunteer subjects during a hand-elevation study in which the right hand was raised and lowered relative to heart level, while the left hand remained static. Red and infrared (IR) PPG signals were obtained from the right and left index fingers using a custom-made PPG processing system. PPG features were identified using a feature-detection algorithm based on the first derivative of the PPG signal. The systolic PPG amplitude, the reflection index, crest time, pulse width at half height, and delta T were calculated from 20 s IR PPG signals from three positions of the right hand with respect to heart level (-50, 0, +50 cm) in 19 volunteers. PPG features were found to change with hand elevation. On lowering the hand to 50 cm below heart level, ac systolic PPG amplitudes from the finger decreased by 68.32 %, while raising the arm increased the systolic amplitude by 69.99 %. These changes in amplitude were attributed to changes in hydrostatic pressure and the veno-arterial reflex. Other morphological variables, such as crest time, were found to be statistically significantly different across hand positions, indicating increased vascular resistance on arm elevation than on dependency. It was hypothesized that these morphological PPG changes were influenced by changes in downstream venous resistance, rather than arterial, or arteriolar, resistance. Changes in hand position relative to heart level can significantly affect the morphology of the peripheral ac PPG waveform. These alterations are due to a combination of physical effects and physiological responses to changes in hand position, which alter vascular resistance. Care should be taken when interpreting morphological data derived from PPG signals and methods should be standardized to take these effects into account.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The PPG can be affected by changes in the height of measurement relative to the heart.
“Finally, the PPG can be affected by changes in the height of measurement relative to the heart [53].”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [53]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Normal Vascular Aging: Differential Effects on Wave Reflection and Aortic Pulse Wave Velocity
McEniery, C. M., Yasmin, Y., Hall, I. R., Qasem, A., Wilkinson, I. B., & Cockcroft, J. R. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.07.037 · Cited by 1,361 · see citing works
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · see citing works
Photoplethysmography: Beyond the Calculation of Arterial Oxygen Saturation and Heart Rate
Shelley, K. H. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269512.82836.c9 · Cited by 450 · see citing works
Utility of the Photoplethysmogram in Circulatory Monitoring
Reisner, A. T., Shaltis, P., McCombie, D. B., Asada, H. H., Warner, D. S., & Warner, M. A. A. (2008) · Anesthesiology · doi:10.1097/aln.0b013e31816c89e1 · Cited by 365 · see citing works
Photoplethysmographic assessment of pulse wave reflection
Chowienczyk, P., Kelly, R. P., MacCallum, H., Millasseau, S., Andersson, T. L. G., Gosling, R. G., Ritter, J. M., & Änggård, E. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00441-6 · Cited by 363 · see citing works
Contour analysis of the photoplethysmographic pulse measured at the finger
Millasseau, S., Ritter, J. M., Takazawa, K., & Chowienczyk, P. (2006) · Journal of Hypertension · doi:10.1097/01.hjh.0000239277.05068.87 · Cited by 351 · see citing works
Noninvasive Assessment of the Digital Volume Pulse
Millasseau, S., Guigui, F. G., Kelly, R. P., Prasad, K., Cockcroft, J., Ritter, J. M., & Chowienczyk, P. (2000) · Hypertension · doi:10.1161/01.hyp.36.6.952 · Cited by 292 · see citing works
Photoplethysmographic derivation of respiratory rate: A review of relevant physiology
Meredith, D. J., Clifton, D., Charlton, P., Brooks, J., Pugh, C. W., & Tarassenko, L. (2012) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2011.638965 · Cited by 226 · see citing works
Gamble, G. D., Zorn, J., Sanders, G. P., MacMahon, S. W., & Sharpe, N. (1994) · Stroke · doi:10.1161/01.str.25.1.11 · Cited by 204 · see citing works
Predicting Arterial Stiffness From the Digital Volume Pulse Waveform
Alty, S. R., Angarita-Jaimes, N., Millasseau, S., & Chowienczyk, P. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.897805 · Cited by 192 · see citing works
Lymphedema Management: The Comprehensive Guide for Practitioners
Zuther, J. E., & Norton, S. (2009) · Cited by 164 · see citing works
The relationship between the photoplethysmographic waveform and systemic vascular resistance
Alian, A. A., Haddadin, A. S., Tantawy, H., Badr, T., Stout, R. G., Silverman, D. G., & Shelley, K. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9097-5 · Cited by 136 · see citing works
The analysis of PPG morphology: Investigating the effects of aging on arterial compliance
Yousef, Q., Reaz, M. B. I., & Ali, M. A. M. (2012) · Measurement Science Review · doi:10.2478/v10048-012-0036-3 · Cited by 126 · see citing works
Vessel diameter changes during the cardiac cycle
Chen, H. C., Patel, V., Wiek, J., Rassam, S., & KOHNER, E. M. (1994) · Eye · doi:10.1038/eye.1994.19 · Cited by 122 · see citing works
Budidha, K., & Kyriacou, P. A. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/2/111 · Cited by 67 · see citing works
Self-citationEffect of Changes in Limb Posture on Peripheral Blood Flow
Beaconsfield, P., & Ginsburg, J. (1955) · Circulation Research · doi:10.1161/01.res.3.5.478 · Cited by 62 · see citing works
Photoplethysmography Analysis of Artery Properties in Patients with Cardiovascular Diseases
Rubīns, U., Grabovskis, A., Grūbe, J., & Kukulis, I. (2008) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-540-69367-3_85 · Cited by 55 · see citing works
The effect of vascular changes on the photoplethysmographic signal at different hand elevations
Hickey, M., Phillips, J. P., & Kyriacou, P. A. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/3/425 · Cited by 43 · see citing works
Self-citationMonitoring of reactive hyperemia using photoplethysmographic pulse amplitude and transit time
Selvaraj, N., Jaryal, A. K., Santhosh, J., Anand, S., & Deepak, K. K. (2009) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-009-9199-3 · Cited by 38 · see citing works
The Effect of the Dependent Position upon Blood Flow in the Limbs
Wilkins, R. W., Halperin, M. H., & Litter, J. (1950) · Circulation · doi:10.1161/01.cir.2.3.373 · Cited by 36 · see citing works
Almond, N. E., Jones, D. P., & Cooke, E. D. (1988) · Angiology · doi:10.1177/000331978803900906 · Cited by 33 · see citing works
Design and development of a novel multi-channel photoplethysmographic research system
Rybynok, V. O., May, J. M., Budidha, K., & Kyriacou, P. A. (2013) · doi:10.1109/pht.2013.6461336 · Cited by 32 · see citing works
Self-citationGu, W. B., Poon, C. C. Y., & Zhang, Y. (2008) · doi:10.1109/issmdbs.2008.4575023 · Cited by 32 · see citing works
Darmanin, G., Jaggard, M. K. J., Hettiaratchy, S. P., Nanchahal, J., & Jain, A. (2013) · Journal of Plastic Reconstructive & Aesthetic Surgery · doi:10.1016/j.bjps.2013.02.024 · Cited by 30 · see citing works
Allen, J., Murray, A., Maria, C. D., & Newton, J. L. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/2/231 · Cited by 30 · see citing works
Direct Pulse Oximetry Within the Esophagus, on the Surface of Abdominal Viscera, and on Free Flaps
Kyriacou, P. A. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3182a1bef6 · Cited by 28 · see citing works
Self-citationMyocardial infarction and Antiphospholipid Syndrome: A first study on finger PPG waveforms effects
Angius, G., Barcellona, D., Cauli, E., Meloni, L., & Raffo, L. (2012) · UNICA IRIS Institutional Research Information System (University of Cagliari) · Cited by 27 · see citing works
Gavish, B., & Gavish, L. Y. (2011) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834ae315 · Cited by 26 · see citing works
Chan, G. S. H., Fazalbhoy, A., Birznieks, I., Macefield, V. G., Middleton, P. M., & Lovell, N. H. (2011) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00970.2011 · Cited by 26 · see citing works
Blood flow rate in arteriovenous anastomoses and capillaries in thumb, first toe, ear lobe, and nose
Midttun, M., & Sejrsen, P. (1996) · Clinical Physiology · doi:10.1111/j.1475-097x.1996.tb00574.x · Cited by 25 · see citing works
Multivariate classification of systemic vascular resistance using photoplethysmography
Lee, Q. Y., Chan, G. S. H., Redmond, S. J., Middleton, P. M., Steel, E., Malouf, P. J., Critoph, C., Flynn, G. M., O’Lone, E., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/8/008 · Cited by 24 · see citing works
Wang, A., Yang, L., Liu, C., Cui, J., Li, Y., Yang, X., Zhang, S., & Zheng, D. (2015) · BioMed Research International · doi:10.1155/2015/752570 · Cited by 19 · see citing works
Noninvasive Continuous Monitoring of Digital Pulse Waves during Hemodialysis
Burkert, A., Scholze, A., & Tepel, M. (2006) · ASAIO Journal · doi:10.1097/01.mat.0000199892.82612.8e · Cited by 17 · see citing works
Xin, S.-Z., Hu, S., Crabtree, V. P., Zheng, J., Azorin-Peris, V., Echiadis, A., & Smith, P. R. (2007) · Journal of Zhejiang University. Science A · doi:10.1631/jzus.2007.a0916 · Cited by 11 · see citing works
Kammila, S., Campbell, N. R. C., Brant, R., deJong, R., & Culleton, B. F. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200204000-00007 · Cited by 10 · see citing works
Influence of venous emptying on the reactive hyperemic blood flow response
Bahadir, Z., Tisdell, E., Arce‐Esquivel, A. A., Dobrosielski, D. A., & Welsch, M. A. (2007) · Dynamic Medicine · doi:10.1186/1476-5918-6-3 · Cited by 9 · see citing works
Hildebrandt, W., Herrmann, J., & Stegemann, J. (1994) · European Journal of Applied Physiology · doi:10.1007/bf00609404 · Cited by 7 · see citing works
Flagged first, then by citations
1 of the 39 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Self-citationAn optimal filter for short photoplethysmogram signals
Liang, Y., Elgendi, M., Chen, Z., & Ward, R. (2018) · Scientific Data · doi:10.1038/sdata.2018.76 · Cited by 242 · see citing works
The 2023 wearable photoplethysmography roadmap
Charlton, P. H., Allen, J., Bailón, R., Baker, S., Behar, J. A., Chen, F., Clifford, G. D., Clifton, D. A., Davies, H. J., Ding, C., et al. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acead2 · Cited by 210 · see citing works
Self-citationWearable Photoplethysmography for Cardiovascular Monitoring.
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · KTUePubl (Repository of Kaunas University of Technology) · doi:10.17863/cam.81148 · Cited by 113 · see citing works
Self-citationChatterjee, S., & Kyriacou, P. A. (2019) · Sensors · doi:10.3390/s19040789 · Cited by 86 · see citing works
Self-citationAbay, T. Y., & Kyriacou, P. A. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0030-2 · Cited by 69 · see citing works
Self-citationInvestigating the origin of photoplethysmography using a multiwavelength Monte Carlo model
Chatterjee, S., Budidha, K., & Kyriacou, P. A. (2020) · Physiological Measurement · doi:10.1088/1361-6579/aba008 · Cited by 67 · see citing works
Self-citationA Wristwatch-Based Wireless Sensor Platform for IoT Health Monitoring Applications
Kumar, S., Buckley, J. L., Barton, J. J., Pigeon, M., Newberry, R., Rodencal, M., Hajzeraj, A., Hannon, T., Rogers, K., Casey, D. P., et al. (2020) · Sensors · doi:10.3390/s20061675 · Cited by 66 · see citing works
Moscato, S., Giudice, S. L., Massaro, G., & Chiari, L. (2022) · Sensors · doi:10.3390/s22155831 · Cited by 51 · see citing works
Adaptive template matching of photoplethysmogram pulses to detect motion artefact
Lim, P. K., Ng, S., Lovell, N. H., Yu, Y. P., Tan, M. P., McCombie, D. B., Lim, E., & Redmond, S. J. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aadf1e · Cited by 39 · see citing works
Determinants of photoplethysmography signal quality at the wrist
Charlton, P. H., Marozas, V., Mejía‐Mejía, E., Kyriacou, P. A., & Mant, J. (2025) · PLOS Digital Health · doi:10.1371/journal.pdig.0000585 · Cited by 38 · see citing works
Self-citationCuffless Continuous Blood Pressure Estimation From Pulse Morphology of Photoplethysmograms
Yan, W., Peng, R., Zhang, Y., & Ho, D. L. (2019) · IEEE Access · doi:10.1109/access.2019.2942936 · Cited by 35 · see citing works
Postoperative Pain Assessment Indices Based on Photoplethysmography Waveform Analysis
Yang, Y. L., Seok, H. S., Noh, G., Choi, B., & Shin, H. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01199 · Cited by 33 · see citing works
Alterations of blood pulsations parameters in carotid basin due to body position change
Kamshilin, A. A., Krasnikova, T. V., Volynsky, M. A., Miridonov, S. V., & Mamontov, O. V. (2018) · Scientific Reports · doi:10.1038/s41598-018-32036-7 · Cited by 34 · see citing works
Postoperative Pain Assessment Model Based on Pulse Contour Characteristics Analysis
Seok, H. S., Choi, B., Noh, G., & Shin, H. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2018.2890482 · Cited by 28 · see citing works
Photoplethysmography upon cold stress—impact of measurement site and acquisition mode
Fleischhauer, V., Bruhn, J., Rasche, S., & Zaunseder, S. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1127624 · Cited by 18 · see citing works
Wearable Blood Pressure Monitoring Devices: Understanding Heterogeneity in Design and Evaluation
Cheung, M. Y., Sabharwal, A., Coté, G. L., & Veeraraghavan, A. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3434344 · Cited by 16 · see citing works
Saugel, B., Bendjelid, K., Critchley, L. A. H., Rex, S., & Scheeren, T. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-9976-3 · Cited by 13 · see citing works
Hamid, H. A., & Nayan, N. A. (2020) · International Journal of Online and Biomedical Engineering (iJOE) · doi:10.3991/ijoe.v16i09.13577 · Cited by 13 · see citing works
Impairing Factors in Remote-PPG Pulse Transit Time Measurements on the Face
Moço, A., Stuijk, S., van Gastel, M., & de Haan, G. (2018) · IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops (CVPRW) · doi:10.1109/cvprw.2018.00184 · Cited by 11 · see citing works
Burton, T., Saiko, G., Cao, M., & Douplik, A. (2022) · Journal of Biophotonics · doi:10.1002/jbio.202200187 · Cited by 10 · see citing works
Awelisah, Y. M., Li, G., Ijaz, M., & Lin, L. (2020) · Optics & Laser Technology · doi:10.1016/j.optlastec.2020.106515 · Cited by 10 · see citing works
Xie, Z., Wang, G., Cheng, Y., Zhang, Q., Le, M. N., & Wang, R. K. (2020) · Quantitative Imaging in Medicine and Surgery · doi:10.21037/qims-20-778 · Cited by 10 · see citing works
Karimpour, P., Ferizoli, R., May, J. M., & Kyriacou, P. A. (2024) · Sensors · doi:10.3390/s24051681 · Cited by 9 · see citing works
Self-citationLee, J., Yang, S.-M., Park, J., Kim, H. C., Kim, E., Jang, Y., Kim, J., & Kim, H. (2018) · Pediatric Anesthesia · doi:10.1111/pan.13415 · Cited by 9 · see citing works
Park, J., Yang, S.-M., Lee, J.-H., Kim, J., Kim, H., & Kim, H. C. (2018) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-018-0183-7 · Cited by 9 · see citing works
Bendjelid, K., Rex, S., Scheeren, T., & Saugel, B. (2016) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-016-9838-4 · Cited by 8 · see citing works
Intelligent Electrocardiogram Acquisition Via Ubiquitous Photoplethysmography Monitoring
Liu, Z., Zhu, T., Lü, L., Zhang, Y., & Clifton, D. A. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3344187 · Cited by 7 · see citing works
Low-frequency variability in photoplethysmography and autonomic function assessment
Allen, J., & Chen, F. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00008-6 · Cited by 7 · see citing works
Hersant, J., Ramondou, P., Thouveny, F., Daligault, M., Feuilloy, M., Saulnier, P., Abraham, P., & Henni, S. (2021) · Diagnostics · doi:10.3390/diagnostics11061128 · Cited by 6 · see citing works
Hersant, J., Ramondou, P., Josse, C., Lecoq, S., Henni, S., & Abraham, P. (2022) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2022.803919 · Cited by 5 · see citing works
Toward Objectification of Subjective Chronic Pain Based on Implicit Response in Biosignals
Seok, H. S., Kim, S. S., Kim, D.-W., & Shin, H. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3452708 · Cited by 3 · see citing works
Analysis of Pulse Wave Morphology Measured by Photoplethysmography During Arm Position Changes
Skripal’, A. V., Геннадиевич, В. Д., Dobdin, S. Y., Сагайдачный, А. А., Usanov, A. D., Antonov, A. V., & Оленко, Е. С. (2025) · Journal of Biomedical Photonics & Engineering · doi:10.18287/jbpe25.11.020310 · Cited by 3 · see citing works
Vascular stiffness in cold pressor test hyper-reactors
Adnan, S. K., Neda, F. S., & Al‐Gailani, B. T. (2023) · Polish Journal of Medical Physics And Engineering · doi:10.2478/pjmpe-2023-0016 · Cited by 3 · see citing works
A Model for Waveform Dissimilarities in Dual-Depth Reflectance-PPG
Moço, A., Stuijk, S., de Haan, G., Wang, R. K., & Verkruysse, W. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8513484 · Cited by 2 · see citing works
In-Vitro Investigation of Flow Profiles in Arteries Using the Photoplethysmograph
Pilt, K., May, J. M., & Kyriacou, P. A. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9629713 · Cited by 1 · see citing works
Self-citationRemote photoplesymography unmasks glabrous skin temporal lead over non-glabrous
Burton, T., Saiko, G., Cao, M., & Douplik, A. (2022) · IEEE Photonics Conference · doi:10.1109/ipc53466.2022.9975569 · Cited by 1 · see citing works
Zhang, X. M., Xu, Y., Li, Y., Shan, J., Hu, T., Gao, L., Sun, Y., Chen, M., Zhao, X., & Ma, Z. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105088 · Cited by 1 · see citing works
Lecoq, S., Hersant, J., & Abraham, P. (2024) · Physiological Measurement · doi:10.1088/1361-6579/ad65b1 · Cited by 1 · see citing works
Hersant, J., Lecoq, S., Ramondou, P., Papon, X., Feuilloy, M., Abraham, P., & Henni, S. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.726315 · Cited by 1 · see citing works
Abdullah, S. A., Essa, S. I., & Talib, B. (2022) · Pakistan Journal of Medical and Health Sciences · doi:10.53350/pjmhs22166552 · Cited by 1 · see citing works
Medidas ópticas não invasivas de tempo de resposta da perfusão sanguínea
de Souza, R. P. (2020) · doi:10.11606/d.59.2020.tde-22072020-165223 · Cited by 1 · see citing works
Investigation of Morphological Variations of Photoplethysmography Signal in Human Epilepsy
Safavi, S. M., Valisharifabad, N., Sabino, R. C., Tran, D., Lin, J. J., Lopour, B. A., & Chou, P. H. (2020) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc44109.2020.9175400
Custom Silicone Vessels for In Vitro Investigations on Vascular Ageing using Photoplethysmography
Karimpour, P., Ferizoli, R., May, J. M., & Kyriacou, P. A. (2024) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc53108.2024.10781560
Self-citationHeart rate increase results in case of positional venous entrapment
Petit, Q., Lecoq, S., Congnard, F., Cronier, N., de Müllenheim, P., Abraham, P., & Noury-Desvaux, B. (2026) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.70041
Sato, S., & Nakayama, S. (2026) · Biomedical Engineering Advances · doi:10.1016/j.bea.2026.100220
Individuality in Visual Representation of Pulse: A Comprehensive Analysis
Sun, L., Wang, L., Liu, C., Bai, F., & Ma, Z. (2026) · Sensors · doi:10.3390/s26196224
- Authors
- Hickey, M., Phillips, J. P., & Kyriacou, P. A.
- Affiliations
- City, University of London (GB)
- Corresponding author
- Hickey, Michelle
- Venue
- Journal of Clinical Monitoring and Computing
- Published
- 2015-08-29
- Volume
- 30
- Issue
- 5
- Pages
- 727-736
- ISSN
- 1387-1307 (ISSN-L)
- 1573-2614
- OpenAlex
- W1862208204
- Cited by
- 49
- References
- 39
- Citations by year
- 2026: 3 · 2025: 2 · 2024: 7 · 2023: 5 · 2022: 8 · 2021: 3 · 2020: 7 · 2019: 3 · 2018: 8 · 2017: 2 · 2016: 1
- FWCI
- 1.82
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Journal of Clinical Monitoring and Computing (publishedVersion)
- PubMed (publishedVersion)
- City Research Online (City University London) (acceptedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- photoplethysmography, peripheral blood flow, hydrostatic pressure, vascular resistance, venous resistance, finger PPG
- MeSH terms
- Adult, Algorithms, Arteries, Blood Pressure, Female, Hand, Heart Rate, Humans, Hydrostatic Pressure, Male, Systole, Vascular Resistance, Veins, Photoplethysmography, Young Adult, Healthy Volunteers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review