Monte Carlo Analysis of Optical Interactions in Reflectance and Transmittance Finger Photoplethysmography
Open accessAbstract
Photoplethysmography (PPG) is a non-invasive photometric technique that measures the volume changes in arterial blood. Recent studies have reported limitations in developing and optimising PPG-based sensing technologies due to unavailability of the fundamental information such as PPG-pathlength and penetration depth in a certain region of interest (ROI) in the human body. In this paper, a robust computational model of a dual wavelength PPG system was developed using Monte Carlo technique. A three-dimensional heterogeneous volume of a specific ROI (i.e., human finger) was exposed at the red (660 nm) and infrared (940 nm) wavelengths in the reflectance and transmittance modalities of PPG. The optical interactions with the individual pulsatile and non-pulsatile tissue-components were demonstrated and the optical parameters (e.g., pathlength, penetration depth, absorbance, reflectance and transmittance) were investigated. Results optimised the source-detector separation for a reflectance finger-PPG sensor. The analysis with the recorded absorbance, reflectance and transmittance confirmed the maximum and minimum impact of the dermis and bone tissue-layers, respectively, in the formation of a PPG signal. The results presented in the paper provide the necessary information to develop PPG-based transcutaneous sensors and to understand the origin of the ac and dc components of the PPG signal.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Prior Monte Carlo light propagation studies have not attempted to capture the PPG waveform in its entirety and have focused on an idealized skin model.
“In order to derive a better understanding of devices based on PPG, several groups have conducted light propagation models using Monte Carlo (MC) simulations29–31. Yet to our knowledge, these studies have not attempted to capture the PPG waveform in its entirety and have focused on an idealized skin model.”
From Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI) (2021) · cites this paper as [31]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Optical Constants of Water in the 200-nm to 200-μm Wavelength Region
Hale, G. M., & Querry, M. R. (1973) · Applied Optics · doi:10.1364/ao.12.000555 · Cited by 4,479 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
MCML—Monte Carlo modeling of light transport in multi-layered tissues
Wang, L. V., Jacques, S. L., & Zheng, L. (1995) · Computer Methods and Programs in Biomedicine · doi:10.1016/0169-2607(95)01640-f · Cited by 3,330 · see citing works
Diffuse radiation in the Galaxy
Henyey, L. C., & Greenstein, J. L. (1941) · The Astrophysical Journal · doi:10.1086/144246 · Cited by 2,971 · see citing works
Ash, C., Dubec, M., Donne, K., & Bashford, T. (2017) · Lasers in Medical Science · doi:10.1007/s10103-017-2317-4 · Cited by 1,040 · see citing works
Burkhoff, D., MIRSKY, I., & Suga, H. (2005) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00138.2005 · Cited by 767 · see citing works
Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
Chan, E. D., Chan, M. M., & Chan, M. M. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.02.004 · Cited by 712 · see citing works
Simpson, C. R., Kohl, M., Essenpreis, M., & Cope, M. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/9/003 · Cited by 651 · see citing works
A literature review and novel theoretical approach on the optical properties of whole blood
Bosschaart, N., Edelman, G. J., Aalders, M. C. G., van Leeuwen, T. G. J. M., & Faber, D. J. (2013) · Lasers in Medical Science · doi:10.1007/s10103-013-1446-7 · Cited by 463 · see citing works
(1997) · Institute of Physics eBooks · doi:10.1887/0750304677 · Cited by 455 · see citing works
Utility of the Photoplethysmogram in Circulatory Monitoring
Reisner, A. T., Shaltis, P., McCombie, D. B., Asada, H. H., Warner, D. S., & Warner, M. A. A. (2008) · Anesthesiology · doi:10.1097/aln.0b013e31816c89e1 · Cited by 365 · see citing works
Review of Monte Carlo modeling of light transport in tissues
Zhu, C., & Liu, Q. (2013) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.5.050902 · Cited by 352 · see citing works
Light scattering study of tissues
Tuchin, V. V. (1997) · Physics-Uspekhi · doi:10.1070/pu1997v040n05abeh000236 · Cited by 305 · see citing works
Jacques, S. L., & McAuliffe, D. J. (1991) · Photochemistry and Photobiology · doi:10.1111/j.1751-1097.1991.tb09891.x · Cited by 278 · see citing works
Rady, M. Y., Rivers, E. P., & Nowak, R. M. (1996) · The American Journal of Emergency Medicine · doi:10.1016/s0735-6757(96)90136-9 · Cited by 264 · see citing works
A new look at the essence of the imaging photoplethysmography
Kamshilin, A. A., Nippolainen, E., Sidorov, I. S., Васильев, П. В., Ерофеев, Н. П., Podolyan, N. P., & Владимирович, Р. Р. (2015) · Scientific Reports · doi:10.1038/srep10494 · Cited by 250 · see citing works
Origins of Tissue Optical Properties in the UVA, Visible, and NIR Regions
Jacques, S. L. (1996) · Advances in Optical Imaging and Photon Migration · doi:10.1364/aoipm.1996.opc364 · Cited by 247 · see citing works
The Cutaneous Microcirculation: Ultrastructure and Microanatomical Organization
Braverman, I. M. (1997) · Microcirculation · doi:10.3109/10739689709146797 · Cited by 239 · see citing works
The Light–Tissue Interaction of Pulse Oximetry
Mannheimer, P. D. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269522.84942.54 · Cited by 194 · see citing works
Computer simulation of the skin reflectance spectra
Meglinski, I. V., & Matcher, S. J. (2002) · Computer Methods and Programs in Biomedicine · doi:10.1016/s0169-2607(02)00099-8 · Cited by 179 · see citing works
New insights into the origin of remote PPG signals in visible light and infrared
Moço, A., Stuijk, S., & de Haan, G. (2018) · Scientific Reports · doi:10.1038/s41598-018-26068-2 · Cited by 170 · see citing works
Online object oriented Monte Carlo computational tool for the needs of biomedical optics
Doronin, A., & Meglinski, I. V. (2011) · Biomedical Optics Express · doi:10.1364/boe.2.002461 · Cited by 163 · see citing works
Khamdaeng, T., Luo, J., Vappou, J., Terdtoon, P., & Konofagou, E. E. (2012) · Ultrasonics · doi:10.1016/j.ultras.2011.09.006 · Cited by 156 · see citing works
Baker, P. D., Westenskow, D. R., & Kück, K. (1997) · Medical & Biological Engineering & Computing · doi:10.1007/bf02530049 · Cited by 141 · see citing works
Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry
Schmitt, J. M. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.137285 · Cited by 118 · see citing works
Effect of temperature on the optical properties ofex vivohuman dermis and subdermis
Laufer, J. G., Simpson, R., Kohl‐Bareis, M., Essenpreis, M., & Cope, M. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/9/004 · Cited by 99 · see citing works
Diffusion model of the optical absorbance of whole blood
Steinke, J. M., & Shepherd, A. P. (1988) · Journal of the Optical Society of America A · doi:10.1364/josaa.5.000813 · Cited by 98 · see citing works
Introduction to Monte-Carlo Methods for Transport and Diffusion Equations
Lapeyre, B., Pardoux, É., & Sentis, R. (2003) · doi:10.1093/oso/9780198525929.001.0001 · Cited by 97 · see citing works
Reflectance Photoplethysmography as Noninvasive Monitoring of Tissue Blood Perfusion
Abay, T. Y., & Kyriacou, P. A. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2417863 · Cited by 96 · see citing works
Self-citationSandberg, M., Zhang, Q., STYF, J. R., Gerdle, B. U. C., & Lindberg, L. (2005) · Acta Physiologica Scandinavica · doi:10.1111/j.1365-201x.2005.01412.x · Cited by 91 · see citing works
Multilayer Modeling of Reflectance Pulse Oximetry
Reuss, J. L. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.840188 · Cited by 72 · see citing works
Rady, M., Rivers, E., & Nowak, R. (1996) · Resuscitation · doi:10.1016/0300-9572(96)89045-9 · Cited by 71 · see citing works
Abay, T. Y., & Kyriacou, P. A. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0030-2 · Cited by 69 · see citing works
Self-citationThe measurement of oxygen saturation in arterial and venous blood
Nitzan, M., & Taitelbaum, H. (2008) · IEEE Instrumentation & Measurement Magazine · doi:10.1109/mim.2008.4534373 · Cited by 62 · see citing works
Transcutaneous Optical Measurement of Hyperbilirubinemia in Neonates
Saidi, I. S. (1992) · Rice Research Repository (Rice University) · Cited by 61 · see citing works
In vivo investigation of ear canal pulse oximetry during hypothermia
Budidha, K., & Kyriacou, P. A. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-9975-4 · Cited by 60 · see citing works
Self-citationHickey, M., Phillips, J. P., & Kyriacou, P. A. (2016) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9761-0 · Cited by 49 · see citing works
Self-citationThe pulse in reflectance pulse oximetry: Modeling and experimental studies
Reuss, J. L., & Siker, D. (2004) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-005-2909-6 · Cited by 47 · see citing works
Finger tissue model and blood perfused skin tissue phantom
Tuchin, V. V., Bashkatov, A. N., Genina, E. A., Kochubey, V. I., Валерьевич, Л. В., Portnov, S. A., Trunina, N. A., Miller, D. R., Cho, S.-M., Oh, H., et al. (2011) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.881604 · Cited by 39 · see citing works
Photoplethysmography for the Assessment of Haemorheology
Njoum, H., & Kyriacou, P. A. (2017) · Scientific Reports · doi:10.1038/s41598-017-01636-0 · Cited by 37 · see citing works
Self-citationKim, J., Kim, J.-H., & Ko, H. (2015) · Sensors · doi:10.3390/s16010046 · Cited by 36 · see citing works
Baltgaile, G. (2012) · Perspectives in Medicine · doi:10.1016/j.permed.2012.02.049 · Cited by 33 · see citing works
Darowish, M., Brenneman, R., & Bigger, J. (2014) · Hand · doi:10.1007/s11552-014-9679-x · Cited by 31 · see citing works
Chatterjee, S., Abay, T. Y., Phillips, J. P., & Kyriacou, P. A. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.7.075005 · Cited by 29 · see citing works
Self-citationAssessment of the calibration curve for transmittance pulse-oximetry
Doronin, A., Fine, I., & Meglinski, I. V. (2011) · Laser Physics · doi:10.1134/s1054660x11190078 · Cited by 27 · see citing works
Proof of concept non-invasive estimation of peripheral venous oxygen saturation
Khan, M., Pretty, C. G., Amies, A. C., Balmer, J., Banna, H. E., Shaw, G. M., & Chase, J. G. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-017-0351-x · Cited by 27 · see citing works
Influence of a skin status on the light interaction with dermis
Kamshilin, A. A., Mamontov, O. V., Koval, V. T., Zayats, G. A., & Владимирович, Р. Р. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.004326 · Cited by 25 · see citing works
Shafique, M., & Kyriacou, P. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/12/2065 · Cited by 15 · see citing works
Self-citationYang, Y., Rupani, A., Wimpenny, I., Weightman, A. P., & Bagnaninchi, P. O. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.8.081417 · Cited by 15 · see citing works
Budidha, K., & Kyriacou, P. A. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609855 · Cited by 11 · see citing works
Self-citationNjoum, H., & Kyriacou, P. A. (2016) · Physiological Measurement · doi:10.1088/1361-6579/38/2/87 · Cited by 8 · see citing works
Self-citationDoronin, A. (2014) · Otago University Research Archive (University of Otago) · doi:10.13140/2.1.4323.1365 · Cited by 7 · see citing works
Rady, M. Y., & Rivers, E. P. (1995) · Critical Care Medicine · doi:10.1097/00003246-199501001-00230 · Cited by 4 · see citing works
Characterizing optical properties in fibrous tissues
Shuaib, A. S. (2011) · doi:10.32469/10355/14451 · Cited by 4 · see citing works
Rushambwa, M. C., Gezimati, M., & Jeeva, J. B. (2017) · IOP Conference Series Materials Science and Engineering · doi:10.1088/1757-899x/263/5/052052 · Cited by 3 · see citing works
Investigating optical path in reflectance pulse oximetry using a multilayer Monte Carlo model
Chatterjee, S., Phillips, J. P., & Kyriacou, P. A. (2017) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2281612 · Cited by 2 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 4
- [23]Tuchin V.V. Advanced Biophotonics: Tissue Optical Sectioning Taylor and Francis, CRC Press Boca Raton, FL, USA 2013
- [27]Gosling J.A. Harris P.F. Humpherson J.R. Whitmore I. Willan P.L. Human Anatomy, Color Atlas and Textbook E-Book Elsevier Health Sciences Amsterdam, The Netherlands 2016
- [33]Marks J.G. Miller J.J. Lookingbill and Marks’ Principles of Dermatology E-Book Elsevier Health Sciences Amsterdam, The Netherlands 2017
- [60]Njoum H. Investigations of Photoplethysmography in the Assessment Of Haemodynamics, Vascular Mechanics and Haemorheology Ph.D. Thesis University of London London, UK 2017
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138 · see citing works
Self-citationScardulla, F., Cosoli, G., Spinsante, S., Poli, A., Iadarola, G., Pernice, R., Busacca, A., Pasta, S., Scalise, L., & D’Acquisto, L. (2023) · Measurement · doi:10.1016/j.measurement.2023.113150 · Cited by 120 · see citing works
A Review of Skin-Wearable Sensors for Non-Invasive Health Monitoring Applications
Mao, P., Li, H., & Yu, Z. (2023) · Sensors · doi:10.3390/s23073673 · Cited by 85 · see citing works
Ajmal, Boonya-Ananta, T., Rodriguez, A. J., Du Le, V. N., & Ramella-Roman, J. C. (2021) · Biomedical Optics Express · doi:10.1364/boe.439893 · Cited by 79 · see citing works
Boonya-ananta, T., Rodriguez, A. J., Ajmal, A., Du Le, V. N., Hansen, A. K., Hutcheson, J. D., & Ramella-Roman, J. C. (2021) · Scientific Reports · doi:10.1038/s41598-021-82124-4 · Cited by 70 · see citing works
Investigating the origin of photoplethysmography using a multiwavelength Monte Carlo model
Chatterjee, S., Budidha, K., & Kyriacou, P. A. (2020) · Physiological Measurement · doi:10.1088/1361-6579/aba008 · Cited by 67 · see citing works
Self-citationA Wristwatch-Based Wireless Sensor Platform for IoT Health Monitoring Applications
Kumar, S., Buckley, J. L., Barton, J. J., Pigeon, M., Newberry, R., Rodencal, M., Hajzeraj, A., Hannon, T., Rogers, K., Casey, D. P., et al. (2020) · Sensors · doi:10.3390/s20061675 · Cited by 66 · see citing works
The origin of photoplethysmography
Kyriacou, P. A., & Chatterjee, S. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00004-9 · Cited by 47 · see citing works
Self-citationLee, H. S., Noh, B., Kong, S. U., Hwang, Y. H., Cho, H., Jeon, Y., & Choi, K. C. (2023) · npj Flexible Electronics · doi:10.1038/s41528-023-00248-1 · Cited by 47 · see citing works
Design of Multi-Wavelength Optical Sensor Module for Depth-Dependent Photoplethysmography
Han, S., Roh, D., Park, J., & Shin, H. (2019) · Sensors · doi:10.3390/s19245441 · Cited by 41 · see citing works
Skin colour affects the accuracy of medical oxygen sensors
Keller, M. D., Harrison-Smith, B., Patil, C., & Arefin, M. S. (2022) · Nature · doi:10.1038/d41586-022-03161-1 · Cited by 38 · see citing works
The past, present, and future of sleep quality assessment and monitoring
Chen, Y., Zhou, E., Wang, Y., Xu, G., Xu, G., & Chen, L. (2023) · Brain Research · doi:10.1016/j.brainres.2023.148333 · Cited by 37 · see citing works
Al-Halawani, R., Qassem, M., & Kyriacou, P. A. (2024) · Scientific Reports · doi:10.1038/s41598-024-58435-7 · Cited by 36 · see citing works
Self-citationNinety years of pulse oximetry: history, current status, and outlook
Quaresima, V., Ferrari, M., & Scholkmann, F. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33307 · Cited by 28 · see citing works
A Noninvasive Blood Glucose Estimation System Using Dual-Channel PPGs and Pulse-Arrival Velocity
Lee, P., Wang, K., & Hsiao, C. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3306343 · Cited by 26 · see citing works
Nomoni, M., May, J. M., & Kyriacou, P. A. (2020) · Sensors · doi:10.3390/s20154246 · Cited by 24 · see citing works
Self-citationA 186μW Photoplethysmography-Based Noninvasive Glucose Sensing SoC
Hina, A., & Saadeh, W. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3180893 · Cited by 24 · see citing works
Perfusion Changes at the Forehead Measured by Photoplethysmography during a Head-Down Tilt Protocol
Ysehak Abay, T., Shafqat, K., & Kyriacou, P. A. (2019) · Biosensors · doi:10.3390/bios9020071 · Cited by 22 · see citing works
Self-citationNoninvasive In Vivo Estimation of Blood-Glucose Concentration by Monte Carlo Simulation
Haque, C. A., Hossain, S., Kwon, T.-H., & Kim, K. (2021) · Sensors · doi:10.3390/s21144918 · Cited by 19 · see citing works
Park, J., Park, J., Jeon, J., Chae, S., Kim, G. B., Han, G., Park, H., Jeong, Y., & Jeong, K. (2025) · Advanced Science · doi:10.1002/advs.202417625 · Cited by 19 · see citing works
Haque, C. A., Kwon, T.-H., & Kim, K. (2022) · Sensors · doi:10.3390/s22031175 · Cited by 18 · see citing works
Tissue mimicking materials and finger phantom design for pulse oximetry
Rodriguez, A., Vasudevan, S., Farahmand, M., Weininger, S., Vogt, W. C., Scully, C. G., Ramella‐Roman, J. C., & Pfefer, T. J. (2024) · Biomedical Optics Express · doi:10.1364/boe.518967 · Cited by 17 · see citing works
Hossain, S., & Kim, K. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3135305 · Cited by 17 · see citing works
Monte Carlo based simulations of racial bias in pulse oximetry
Arefin, M. S., Dumont, A. P., & Patil, C. A. (2022) · doi:10.1117/12.2610483 · Cited by 16 · see citing works
Photoplethysmography upon cold stress—impact of measurement site and acquisition mode
Fleischhauer, V., Bruhn, J., Rasche, S., & Zaunseder, S. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1127624 · Cited by 18 · see citing works
Al-Halawani, R., Qassem, M., & Kyriacou, P. A. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33305 · Cited by 17 · see citing works
Self-citationChen, S., Chuang, Y.-C., & Chang, C. (2020) · Sensors · doi:10.3390/s20226556 · Cited by 15 · see citing works
Photoplethysmography in oxygenation and blood volume measurements
Abay, T. Y., & Kyriacou, P. A. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00003-7 · Cited by 16 · see citing works
Self-citationQuantification of the Phenomena Affecting Reflective Arterial Photoplethysmography
Rovas, G., Bikia, V., & Stergiopulos, N. (2023) · Bioengineering · doi:10.3390/bioengineering10040460 · Cited by 14 · see citing works
A Novel Photoplethysmography Sensor for Vital Signs Monitoring from the Human Trachea
May, J. M., Phillips, J. P., Fitchat, T., Ramaswamy, S., Snidvongs, S., & Kyriacou, P. A. (2019) · Biosensors · doi:10.3390/bios9040119 · Cited by 12 · see citing works
Self-citationDual-ratio approach to pulse oximetry and the effect of skin tone
Blaney, G. P., Frias, J., Tavakoli, F., Sassaroli, A., & Fantini, S. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33311 · Cited by 12 · see citing works
Lapitan, D. G., Tarasov, A. P., & Рогаткин, Д. А. (2022) · Journal of Biomedical Photonics & Engineering · doi:10.18287/jbpe22.08.030306 · Cited by 11 · see citing works
Will Smart Ring be Next Wave of Wearables?
Wang, M., Chen, C., Wu, H., Zhang, J., Zhou, S., & Wang, G. (2024) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas61083.2024.10798374 · Cited by 10 · see citing works
Bhusal, A., Farahmand, M., Hasan, M. S., Vasudevan, S., Vogt, W. C., Ibarra, B., Weininger, S., Scully, C. G., Zhang, X. H., Chen, Y., et al. (2025) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33314 · Cited by 11 · see citing works
Al-Halawani, R., Qassem, M., & Kyriacou, P. A. (2025) · Sensors · doi:10.3390/s25020372 · Cited by 11 · see citing works
Self-citationAl-Halawani, R., Qassem, M., & Kyriacou, P. A. (2025) · Sensors · doi:10.3390/s25020559 · Cited by 10 · see citing works
Self-citationUnderstanding and modeling finger vascular pattern imaging
Normakristagaluh, P., Laanstra, G. J., Spreeuwers, L., & Veldhuis, R. N. J. (2022) · IET Image Processing · doi:10.1049/ipr2.12408 · Cited by 10 · see citing works
Monte Carlo Simulation of Diffuse Optical Spectroscopy for 3D Modeling of Dental Tissues
Moradi, M., & Chen, Y. (2023) · Sensors · doi:10.3390/s23115118 · Cited by 10 · see citing works
Modeling light-tissue interactions in pulse oximetry: effect of device design and skin pigmentation
Moradi, M., Vasudevan, S., Bhusal, A., Weininger, S., Chen, Y., & Pfefer, J. (2024) · Design and Quality for Biomedical Technologies · doi:10.1117/12.3004189 · Cited by 8 · see citing works
IoT PULSE OXIMETRY STATUS MONITORING FOR HOME QUARANTINED COVID-19 PATIENTS
MIRON-ALEXE, V. (2020) · Journal of Science and Arts · doi:10.46939/j.sci.arts-20.4-c04 · Cited by 8 · see citing works
Hossain, S., & Kim, K. (2023) · Sensors · doi:10.3390/s23010540 · Cited by 8 · see citing works
Chatterjee, S., Patel, Z., Thaha, M. A., & Kyriacou, P. A. (2020) · Journal of Biomedical Optics · doi:10.1117/1.jbo.25.2.027001 · Cited by 8 · see citing works
Self-citationApplications of Optical Cardiovascular Monitoring
Lemay, M., Bertschi, M., Solà, J., Renevey, P., Genzoni, E., Proença, M., Ferrario, D., Braun, F., Parák, J., & Korhonen, I. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-819246-7.00018-8 · Cited by 8 · see citing works
Monte Carlo Simulation of the Effect of Human Skin Melanin in Light-Tissue Interactions
Al-Halawani, R., Chatterjee, S., & Kyriacou, P. A. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871350 · Cited by 7 · see citing works
Self-citationZhang, C., Yan, H., Liu, Q., Yang, K., Liu, F., & Lin, L. (2024) · IEEE Transactions on Molecular Biological and Multi-Scale Communications · doi:10.1109/tmbmc.2023.3349326 · Cited by 7 · see citing works
Fine, J. N., McShane, M. J., Coté, G. L., & Scully, C. G. (2022) · Biosensors · doi:10.3390/bios12080598 · Cited by 7 · see citing works
Photoplethysmography: New trends and future directions
Kyriacou, P. A., & May, J. M. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00012-8 · Cited by 6 · see citing works
Self-citationHu, H., Li, J., & Chen, X. (2022) · International Congress on Image and Signal Processing (CISP) · doi:10.1109/cisp-bmei56279.2022.9979897 · Cited by 6 · see citing works
Hersant, J., Ramondou, P., Durand, S., Feuilloy, M., Daligault, M., Abraham, P., & Henni, S. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2021.765174 · Cited by 6 · see citing works
On the Origin of the Photoplethysmography Signal: Modeling of Volumetric and Aggregation Effects
Lapitan, D. G., Tarasov, A. P., Shtyflyuk, M. E., & Рогаткин, Д. А. (2024) · Photonics · doi:10.3390/photonics11070637 · Cited by 5 · see citing works
Wrist-Worn Devices for Remote Monitoring of Cardiovascular Disease: A Survey
Sadiq, I., Ijaz, A., & Imran, A. (2025) · IEEE Access · doi:10.1109/access.2025.3602893 · Cited by 6 · see citing works
Bachir, W. (2023) · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · doi:10.1016/j.saa.2023.123181 · Cited by 5 · see citing works
Exploring the bias: how skin color influences oxygen saturation readings via Monte Carlo simulations
Swamy, S. K. N., Liu, C., Correia, R. G., Hayes‐Gill, B., & Morgan, S. P. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33308 · Cited by 5 · see citing works
Jones, Z. D., Reitzle, D., & Kienle, A. (2025) · Optics Letters · doi:10.1364/ol.562315 · Cited by 5 · see citing works
Brain Light-Tissue Interaction Modelling: Towards a non-invasive sensor for Traumatic Brain Injury
Roldán, M., Chatterjee, S., & Kyriacou, P. A. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630909 · Cited by 5 · see citing works
Self-citationBadolato, E., Little, A., & Le, V. N. D. (2024) · Optics Express · doi:10.1364/oe.510977 · Cited by 5 · see citing works
Design of Personal Health Monitoring Devices for Early Detection of Silent Hypoxia
Kuswoyo, H., Susana, E., & Tjahjadi, H. (2022) · Teknik · doi:10.14710/teknik.v43i1.42752 · Cited by 5 · see citing works
ESP32-Powered PPG Signal Acquisition: Open-Source Hardware and Software for Research and Education
Miranda-Vega, J. E., Nuñez-Patrón, E. Y., Prieto-Avalos, G., Flores‐Fuentes, W., Sergiyenko, O., García-González, W., Márquez-Ramirez, L. V., Contreras, R. C., & Ayala-Figueroa, R. I. (2025) · Hardware · doi:10.3390/hardware3040015 · Cited by 4 · see citing works
Reiser, M., Breidenassel, A., & Amft, O. (2025) · Scientific Reports · doi:10.1038/s41598-025-26560-6 · Cited by 4 · see citing works
Nair, B., Mysorekar, C., Srivastava, R., & Kale, S. N. (2024) · IEEE Applied Sensing Conference (APSCON) · doi:10.1109/apscon60364.2024.10466125 · Cited by 3 · see citing works
Sahli, I., Bachir, W., & El‐Daher, M. S. (2025) · Journal of Biophotonics · doi:10.1002/jbio.202400413 · Cited by 2 · see citing works
Jones, Z. D., Foschum, F., & Kienle, A. (2026) · Sensors · doi:10.3390/s26072070 · Cited by 2 · see citing works
Misran, M. A. B., Bilgaiyan, A., & Hattori, R. (2022) · Computational And Experimental Research In Materials And Renewable Energy · doi:10.19184/cerimre.v5i2.31668 · Cited by 2 · see citing works
Source-Detector Geometry Analysis of Reflective PPG by Measurements and Simulations
Reiser, M., Mueller, T., Breidenassel, A., & Amft, O. (2025) · IEEE Open Journal of Engineering in Medicine and Biology · doi:10.1109/ojemb.2025.3546771 · Cited by 2 · see citing works
Characterizing the influence of skin pigmentation on pulse oximetry
Putcha, A., Schichlein, K. D., Nguyen, T., Davis, B., Choffin, R., Malik, S., Sharma, A., Sosa, G., Saffari, S., Ribaudo, J. G., et al. (2025) · Biophotonics discovery. · doi:10.1117/1.bios.2.3.032506 · Cited by 2 · see citing works
Patel, Z., Chatterjee, S., Thaha, M. A., & Kyriacou, P. A. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8856763 · Cited by 2 · see citing works
Self-citationThe effect of contact pressure on photoplethysmography
Gong, X., Yu, C., & Xuan, Y. (2024) · Indonesian Journal of Electrical Engineering and Computer Science · doi:10.11591/ijeecs.v33.i3.pp1589-1603 · Cited by 1 · see citing works
Mechanical Testing of Artificial Vessels and Tissues for Photoplethysmography Phantoms
May, J. M., Nomoni, M., Budidha, K., Choi, C., & Kyriacou, P. A. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871830 · Cited by 1 · see citing works
Self-citationMonte Carlo simulation-based analysis of unobtrusive PPG monitoring through clothes
Badiola, I., Linschmann, O., Willms, L., Blažek, V., Leonhardt, S. K., & Lueken, M. (2023) · Proceedings - IEEE Symposium on Computer-Based Medical Systems · doi:10.1109/cbms58004.2023.00318 · Cited by 1 · see citing works
Khatoun, A., Petre, A.-R., Peeters, B., van Breemen, A. J. J. M., & Akkerman, H. B. (2023) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/sensors56945.2023.10324908 · Cited by 1 · see citing works
Swamy, S. K. N., He, C., Liu, C., Correia, R. G., Hayes‐Gill, B., Clark, D. J., Green, S., & Morgan, S. P. (2024) · Design and Quality for Biomedical Technologies · doi:10.1117/12.3000705 · Cited by 1 · see citing works
Monte Carlo Characterization of Short-Wave Infrared Optical Wavelengths for Biosensing Applications
Budidha, K., Chatterjee, S., Qassem, M., & Kyriacou, P. A. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630061 · Cited by 1 · see citing works
Self-citationHarrison-Smith, B., Arefin, M. S., Dumont, A., & Patil, C. A. (2022) · doi:10.1117/12.2610519
Sensor technologies for mobile and wearable applications in mobile respiratory management
Wacker, J., Bonnal, B., Braun, F., Chételat, O., Ferrario, D., Lemay, M., Rapin, M., Renevey, P., & Yılmaz, G. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-823447-1.00006-3
Scaling of Algorithmic Bias in Pulse Oximetry with Signal-to-Noise Ratio
Cao, J., Mehta, N. A., Wu, J., Wood, S., Kainerstorfer, J. M., & Grover, P. (2023) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc40787.2023.10341034
Valladares-Pérez, J. H., Cortés-Ortegón, F., Garcı́a-Valenzuela, A., & Sanchez-Perez, C. A. (2023) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-49401-7_15
Goenka, C., & Yu, L. (2024) · Elsevier eBooks · doi:10.1016/b978-0-44-318840-4.00008-5
Modeling Light Propagation Through Finger Tissue for Sensor Alignment in Transmission Mode
Vo, D. K., Nguyen, A. T. P., Nguyen, D. M., Dang, V. T., Nguyen, T. T., Tran, H. D. T., & Tran, T. N. (2026) · Journal of Physics Conference Series · doi:10.1088/1742-6596/3180/1/012012
PPG-Based Respiration Monitoring Using CycleGAN
Loganathan, D., Nihal, B., Kamalpur, R. S., & Gowtham, R. (2026) · International Conference on Advances in Electrical, Computing, Communication and Sustainable Technologies (ICAECT) · doi:10.1109/icaect68478.2026.11426020
Smart textile wearables for human-environment interaction
Chen, Y., Ren, J., Zhang, M., Rajapakse, D., Liu, F., Lee, B. W., Won, S. M., & Zhu, X. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.177140
Park, T. W., Zhang, H., Park, C., Goldstein, S. D., Wade, L., Rytkin, E., McLuckie, A. J., Külahçı, Y., Ramu, J., Kim, T. Y., et al. (2026) · Science Advances · doi:10.1126/sciadv.aeg8187
The precision ceiling: information-theoretic accuracy bounds for responsible AI health screening
Ding, K., Li, Z., Zou, X., Jia, J., Li, C., & Jiang, Z. (2026) · Frontiers in Digital Health · doi:10.3389/fdgth.2026.1956262
Du, K., Zhao, M., Wang, T., & Hongxing, 蔡. C. (2026) · Journal of Russian Laser Research · doi:10.1007/s10946-026-10329-x
Wolling, F., Filipp, A., & Michahelles, F. (2026) · ACM International Symposium on Wearable Computers (ISWC) · doi:10.1145/3830727.3834825
Methods for Optimizing Photoplethysmography Signal Quality
Ode, O., Anderson, D. V., & Inan, O. T. (2026) · Sensors · doi:10.3390/s26206669
- Authors
- Chatterjee, S., & Kyriacou, P. A.
- Affiliations
- City, University of London (GB)
- Corresponding author
- Chatterjee, Subhasri
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2019-02-15
- Volume
- 19
- Issue
- 4
- Pages
- 789
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC6412556
- OpenAlex
- W2914695435
- Cited by
- 86
- References
- 56
- Citations by year
- 2026: 9 · 2025: 11 · 2024: 15 · 2023: 14 · 2022: 14 · 2021: 12 · 2020: 7 · 2019: 4
- FWCI
- 3.26
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- City Research Online (City University London) (acceptedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Keywords
- photoplethysmography, finger photoplethysmography, Monte Carlo light transport simulation, optical path length, penetration depth, source-detector separation, dermis, bone, arterial blood volume
- MeSH terms
- Blood Volume Determination, Computer Simulation, Fingers, Humans, Monte Carlo Method, Oximetry, Oxygen, Signal Processing, Computer-Assisted, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review