Simulations of the Frequency Response of Implantable Glucose Sensors
Abstract
The response of enzyme electrode glucose sensors implanted in tissues to physiologic blood glucose oscillations is simulated. Models describe both oxygen-based and peroxide-based glucose sensors in spatially homogeneous medium simulating some mass transfer properties of tissue. Pass-through ratios and delays are reported as a function of frequency for the oxygen-based sensor, and the effects on continuous blood glucose monitoring are illustrated using data from the literature. Certain peroxide-based sensor designs may produce common signals for different glucose concentrations, a characteristic not found in oxygen-based sensors. The dynamic response depends on the frequency of glucose oscillation and is sensitive to sensor type, enzyme activity, and diffusional resistance.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Jablecki and Gough's simulations concluded that increased mass transfer would increase lag and could potentially reduce the magnitude and differences when glucose sensor signals fluctuate between high and low.
“Simulations of glucose concentration oscillations by Jablecki and Gough concluded that increases in mass transfer would increase lag and could potentially decrease the magnitude and differences when fluctuating between high and low glucose sensor signals.110”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [110]
The positive correlation between lag time and capsule thickness supported the models described above.
“The positive correlation between lag time and capsule thickness supported the models described above.107,110–111”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [110]
The increases in mass transfer in those simulations were not attributed to any specific part of the FBR and could arise from collagen capsule thickness, blood vessel density, or other unanticipated factors.
“the increases to mass transfer were not attributed to any specific part of the FBR and could originate from collagen capsule thickness, blood vessel density, or other unanticipated factors.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [110]
Cites
Two-dimensional enzyme electrode sensor for glucose
Gough, D., Lucisano, J. Y., & Tse, P. H. S. (1985) · Analytical Chemistry · doi:10.1021/ac00289a042 · Cited by 223 · see citing works
Self-citationApplication of Chronic Intravascular Blood Glucose Sensor in Dogs
Armour, J. C., Lucisano, J. Y., Bd, M., & Gough, D. A. (1990) · Diabetes · doi:10.2337/diab.39.12.1519 · Cited by 175 · see citing works
Self-citationCsoeregi, E., Schmidtke, D. W., & Heller, A. (1995) · Analytical Chemistry · doi:10.1021/ac00103a015 · Cited by 129 · see citing works
Model of a two-substrate enzyme electrode for glucose
Leypoldt, J. K., & Gough, D. (1984) · Analytical Chemistry · doi:10.1021/ac00278a063 · Cited by 125 · see citing works
Self-citationMoatti-Sirat, D., Capron, F., Poitout, V., Reach, G., Bindra, D. S., Zhang, Y., Wilson, G. S., & Th�venot, D. R. (1992) · Diabetologia · doi:10.1007/bf00400921 · Cited by 123 · see citing works
Shichiri, M., Asakawa, N., Yamasaki, Y., Kawamori, R., & Abe, H. (1986) · Diabetes Care · doi:10.2337/diacare.9.3.298 · Cited by 96 · see citing works
Time‐dependent inactivation of immobilized glucose oxidase and catalase
Tse, P. H. S., & Gough, D. A. (1987) · Biotechnology and Bioengineering · doi:10.1002/bit.260290607 · Cited by 95 · see citing works
Self-citationSubcutaneous implantation of a ferrocene-mediated glucose sensor in pigs
Claremont, D. J., Sambrook, E., Penton, C., et al. (1986) · Diabetologia · doi:10.1007/bf00873223 · Cited by 94 · see citing works
Does fall in tissue glucose precede fall in blood glucose?
Sternberg, F., Meyerhoff, C., Mennel, F. J., Mayer, H. G., Bischof, F., & Pfeiffer, E. F. (1996) · Diabetologia · doi:10.1007/bf00403309 · Cited by 90 · see citing works
Evaluation of a Subcutaneous Glucose Sensor out to 3 Months in a Dog Model
GILLIGAN, B. J., Shults, M. C., Rhodes, R. K., & Updike, S. J. (1994) · Diabetes Care · doi:10.2337/diacare.17.8.882 · Cited by 90 · see citing works
In vivo evaluation of an electroenzymatic glucose sensor implanted in subcutaneous tissue
Johnson, K. W., Mastrototaro, J. J., Howey, D. C., Brunelle, R. L., Burden-Brady, P. L., Bryan, N. A., Andrew, C. C., Rowe, H., Allen, D. J., Noffke, B. W., et al. (1992) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(92)85053-d · Cited by 73 · see citing works
A Miniaturized Nafion-Based Glucose Sensor: in vitro and in vivo evaluation in dogs
Moussy, F., Harrison, D. J., & Rajotte, R. V. (1994) · The International Journal of Artificial Organs · doi:10.1177/039139889401700206 · Cited by 67 · see citing works
Theoretical evaluation of the steady-state response of potentiometric enzyme electrodes
Brady, J. E., & Carr, P. W. (1980) · Analytical Chemistry · doi:10.1021/ac50056a046 · Cited by 66 · see citing works
Amperometric response enhancement of the immobilized glucose oxidase enzyme electrode
Mell, L. D., & Maloy, J. T. (1976) · Analytical Chemistry · doi:10.1021/ac50005a045 · Cited by 66 · see citing works
Lerner, H., Giner, J., Soeldner, J. S., et al. (1979) · Journal of The Electrochemical Society · doi:10.1149/1.2129011 · Cited by 59 · see citing works
Ertefai, S., & Gough, D. A. (1989) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(89)90097-6 · Cited by 52 · see citing works
Self-citationIn vitro stability of an oxygen sensor
Lucisano, J. Y., Armour, J. C., & Gough, D. (1987) · Analytical Chemistry · doi:10.1021/ac00132a012 · Cited by 48 · see citing works
Self-citationTransient Studies of Glucose, Oxygen, and Hydroquinone at a Membrane‐Covered Rotated Disk Electrode
Gough, D., & Leypoldt, J. K. (1980) · Journal of The Electrochemical Society · doi:10.1149/1.2129872 · Cited by 44 · see citing works
Self-citationDynamic Delay and Maximal Dynamic Error in Continuous Biosensors
Baker, D. A., & Gough, D. (1996) · Analytical Chemistry · doi:10.1021/ac960030d · Cited by 39 · see citing works
Self-citationBiosensor for continuos glucose monitoring
Atanasov, P., & Wilkins, E. S. (1994) · Biotechnology and Bioengineering · doi:10.1002/bit.260430310 · Cited by 32 · see citing works
Atkinson, B., & Lester, D. E. (1974) · Biotechnology and Bioengineering · doi:10.1002/bit.260161003 · Cited by 19 · see citing works
Rhodes, R. K., Shults, M. C., & Updike, S. J. (1994) · Analytical Chemistry · doi:10.1021/ac00081a026 · Cited by 15 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals
Gough, D. A., Kumosa, L. S., Routh, T. L., Lin, J. T., & Lucisano, J. Y. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3001148 · Cited by 179 · see citing works
Self-citationThe effect of nitric oxide surface flux on the foreign body response to subcutaneous implants
Nichols, S. P., Koh, A., Brown, N. L., Rose, M. B., Sun, B., Slomberg, D. L., Riccio, D. A., Klitzman, B., & Schoenfisch, M. H. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.05.053 · Cited by 67 · see citing works
Increased<i>In Vivo</i>Glucose Recovery via Nitric Oxide Release
Nichols, S. P., Le, N. N., Klitzman, B., & Schoenfisch, M. H. (2011) · Analytical Chemistry · doi:10.1021/ac103070t · Cited by 26 · see citing works
- Authors
- Jablecki, M., & Gough, D. A.
- Affiliations
- University of California San Diego (US)
- Corresponding author
- Gough, David A.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2000-03-17
- Volume
- 72
- Issue
- 8
- Pages
- 1853-1859
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W1971946532
- Cited by
- 24
- References
- 22
- Citations by year
- 2019: 1 · 2017: 2 · 2015: 1 · 2013: 2 · 2012: 1
- FWCI
- 0.86
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Keywords
- implantable glucose sensor, frequency response, continuous glucose monitoring, sensor delay, diffusion resistance, sensor modeling
- MeSH terms
- Algorithms, Blood Glucose, Computer Simulation, Models, Biological, Biosensing Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review