A novel porous collagen scaffold around an implantable biosensor for improving biocompatibility. II. Long‐term in vitro/in vivo sensitivity characteristics of sensors with NDGA‐ or GA‐crosslinked collagen scaffolds
Abstract
We have developed a new 3D porous and biostable collagen scaffold for implantable glucose sensors. The scaffolds were fabricated around the sensors and crosslinked using nordihydroguaiaretic acid (NDGA) or glutaraldehyde (GA) to enhance physical and biological stability. The effect of the scaffolds on sensor function and biocompatibility was examined during long‐term (≥28 days) in vitro and in vivo experiments and compared with control bare sensors. To evaluate the effect of the sensor length on micromotion and sensor function, we also fabricated short and long sensors. 3D porous scaffold application around glucose sensors did not significantly affect the long‐term in vitro sensitivity of the sensors. The scaffolds, crosslinked by either NDGA or GA, remained stable around the sensors during the 4 week in vitro study. In the long‐term in vivo study, the sensitivity of the short sensors was higher than the sensitivity of long sensors presumably because of less micromotion in the subcutis of the rats. The sensors with NDGA‐crosslinked scaffolds had a higher sensitivity than the sensors with GA‐crosslinked scaffolds. Histological examination showed that NDGA‐crosslinked scaffolds retained their physical structure with reduced inflammation when compared with the GA‐crosslinked scaffolds. Therefore, the application of NDGA‐crosslinked collagen scaffolds might be a good method for enhancing the function and lifetime of implantable biosensors by minimizing the in vivo foreign body response. © 2009 Wiley Periodicals, Inc. J Biomed Mater Res, 2010
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Histological analysis showed improved biocompatibility of NDGA-crosslinked scaffolds compared with GA-crosslinked scaffolds, with less inflammatory cell localization and infiltration.
“Histological analysis after the experiment revealed improved biocompatibility of the NDGA-crosslinked compared to the GA-cross-linked scaffolds as evidenced by less inflammatory cell localization and infiltration into the porous membranes.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [148]
Despite the histological benefits, the analytical performance of control sensors lacking collagen was better than that of collagen-based sensors.
“Despite the observed histological benefits, the analytical performance of control sensors (i.e., lacking collagen) was better than that of the collagen-based sensors.148”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [148]
Cites
Home blood glucose biosensors: a commercial perspective
Newman, J. D., & Turner, A. P. F. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.11.012 · Cited by 901 · see citing works
Characterization of implantable biosensor membrane biofouling
Wisniewski, N. A., Moussy, F. G., & Reichert, W. M. (2000) · Fresenius Journal of Analytical Chemistry · doi:10.1007/s002160051556 · Cited by 351 · see citing works
Self-citationLong-term engraftment of hepatocytes transplanted on biodegradable polymer sponges
Mooney, D. J., Sano, K., Kaufmann, P. M., Majahod, K., Schloo, B. L., Vacanti, J. P., & Langer, R. (1997) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19971205)37:3 · Cited by 233 · see citing works
Application of Chronic Intravascular Blood Glucose Sensor in Dogs
Armour, J. C., Lucisano, J. Y., Bd, M., & Gough, D. A. (1990) · Diabetes · doi:10.2337/diab.39.12.1519 · Cited by 175 · see citing works
Quinn, C. P., Pathak, C. P., Heller, A., & Hubbell, J. A. (1995) · Biomaterials · doi:10.1016/0142-9612(95)98856-9 · Cited by 172 · see citing works
Wright, C. M. (1987) · Journal of Vascular Surgery · doi:10.1016/s0741-5214(87)70041-x · Cited by 164 · see citing works
Norton, L. W., Koschwanez, H. E., Wisniewski, N. A., Klitzman, B., & Reichert, W. M. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31088 · Cited by 128 · see citing works
Biocompatible, glucose-permeable hydrogel for in situ coating of implantable biosensors
Quinn, C. A. P., Connor, R. E., & Heller, A. (1997) · Biomaterials · doi:10.1016/s0142-9612(97)00125-7 · Cited by 126 · see citing works
Moussy, F. G., Jakeway, S., Harrison, D. J., & Rajotte, R. V. (1994) · Analytical Chemistry · doi:10.1021/ac00094a007 · Cited by 117 · see citing works
Self-citationCahn, R. W., Haasen, P., Kramer, E. J., et al. (1991) · Medical Entomology and Zoology · Cited by 102 · see citing works
Mediation of <i>in vivo</i> glucose sensor inflammatory response via nitric oxide release
Gifford, R., Batchelor, M. M., Lee, Y., Gokulrangan, G., Meyerhoff, M. E., & Wilson, G. S. (2005) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.30359 · Cited by 96 · see citing works
A Hydrogel-Based Implantable Micromachined Transponder for Wireless Glucose Measurement
Lei, M., Baldi, A., Nuxoll, E. E., et al. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.112 · Cited by 95 · see citing works
Evaluation of a Subcutaneous Glucose Sensor out to 3 Months in a Dog Model
GILLIGAN, B. J., Shults, M. C., Rhodes, R. K., & Updike, S. J. (1994) · Diabetes Care · doi:10.2337/diacare.17.8.882 · Cited by 90 · see citing works
Fernández, E., López, D., López‐Cabarcos, E., et al. (2005) · Polymer · doi:10.1016/j.polymer.2004.12.039 · Cited by 85 · see citing works
In vitro and in vivo mineralization of Nafion membrane used for implantable glucose sensors
Mercado, R. C., & Moussy, F. (1998) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(97)00112-7 · Cited by 73 · see citing works
Self-citationJu, Y. M., Yu, B., Koob, T. J., Moussy, Y., & Moussy, F. G. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31756 · Cited by 70 · see citing works
Self-citationBiomimetic approaches to tendon repair
Koob, T. J. (2002) · Comparative Biochemistry and Physiology Part A Molecular & Integrative Physiology · doi:10.1016/s1095-6433(02)00247-7 · Cited by 69 · see citing works
A Miniaturized Nafion-Based Glucose Sensor: in vitro and in vivo evaluation in dogs
Moussy, F., Harrison, D. J., & Rajotte, R. V. (1994) · The International Journal of Artificial Organs · doi:10.1177/039139889401700206 · Cited by 67 · see citing works
Self-citationEnzymatic glucose sensors. Improved long-term performance in vitro and in vivo.
Updike, S. J., Shults, M. C., Rhodes, R. K., GILLIGAN, B. J., LUEBOW, J. O., & von Heimburg, D. (1994) · PubMed · Cited by 57 · see citing works
Ertefai, S., & Gough, D. A. (1989) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(89)90097-6 · Cited by 52 · see citing works
Updike, S. J., Shults, M. C., Rhodes, R. K., GILLIGAN, B. J., LUEBOW, J. O., & von Heimburg, D. (1994) · ASAIO Journal · doi:10.1097/00002480-199404000-00007 · Cited by 52 · see citing works
Pickup, J. C., Claremont, D. J., & Shaw, G. W. (1993) · Acta Diabetologica · doi:10.1007/bf00572858 · Cited by 47 · see citing works
Study of the effects of tissue reactions on the function of implanted glucose sensors
Dungel, P., Long, N., Yu, B., Moussy, Y., & Moussy, F. (2008) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31593 · Cited by 47 · see citing works
Self-citationPfeiffer, E. F. (1987) · Diabetologia · doi:10.1007/bf00274572 · Cited by 45 · see citing works
Winet, H., & Bao, J. Y. (1998) · Journal of Biomedical Materials Research · doi:10.1002/(sici)1097-4636(19980615)40:4 · Cited by 44 · see citing works
Glucose sensors: toward closed loop insulin delivery
Chia, C. W., & Saudek, C. D. (2004) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2003.12.001 · Cited by 42 · see citing works
Nishida, K., Sakakida, M., Ichinose, K., et al. (1995) · PubMed · Cited by 41 · see citing works
An Investigation of Long-Term Performance of Minimally Invasive Glucose Biosensors
Yu, B., Ju, Y., West, L. A., Moussy, Y., & Moussy, F. G. (2007) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.0020 · Cited by 35 · see citing works
Self-citationThe role of continuous glucose sensors in diabetes care
Garg, S. K., Hoff, H. K., & Chase, H. P. (2004) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2004.01.001 · Cited by 32 · see citing works
[6] Long-term implantation of voltammetric oxidase/peroxide glucose sensors in the rat peritoneum
Clark, L. C., Noyes, L. K., Spokane, R. B., Sudan, R., & Miller, M. L. (1988) · Methods in enzymology on CD-ROM/Methods in enzymology · doi:10.1016/0076-6879(88)37008-4 · Cited by 31 · see citing works
Klueh, U., Dorsky, D. I., & Kreutzer, D. L. (2003) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.20041 · Cited by 26 · see citing works
Shichiri, M., Yamasaki, Y., Nao, K., et al. (1988) · PubMed · Cited by 14 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500317 · Cited by 125 · see citing works
Collagen and its modifications—crucial aspects with concern to its processing and analysis
Rýglová, Š., Braun, M., & Suchý, T. (2017) · Macromolecular Materials and Engineering · doi:10.1002/mame.201600460 · Cited by 106 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500318 · Cited by 102 · see citing works
The effect of nitric oxide surface flux on the foreign body response to subcutaneous implants
Nichols, S. P., Koh, A., Brown, N. L., Rose, M. B., Sun, B., Slomberg, D. L., Riccio, D. A., Klitzman, B., & Schoenfisch, M. H. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.05.053 · Cited by 67 · see citing works
A dexamethasone‐loaded PLGA microspheres/collagen scaffold composite for implantable glucose sensors
Ju, Y. M., Yu, B., West, L., Moussy, Y., & Moussy, F. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32512 · Cited by 44 · see citing works
Self-citationLong-term calibration considerations during subcutaneous microdialysis sampling in mobile rats
Mou, X., Lennartz, M. R., Loegering, D. J., & Stenken, J. A. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.02.016 · Cited by 33 · see citing works
Ward, W. K., Hansen, J. C., Massoud, R. G., Engle, J. M., Takeno, M. M., & Hauch, K. D. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32651 · Cited by 26 · see citing works
- Authors
- Ju, Y. M., Yu, B., West, L., Moussy, Y., & Moussy, F.
- Affiliations
- University of South Florida (US)
- Corresponding author
- Moussy, Francis
- Venue
- Journal of Biomedical Materials Research Part A
- Publisher
- Wiley
- Published
- 2009-02-20
- Volume
- 92A
- Issue
- 2
- Pages
- 650-658
- ISSN
- 1549-3296 (ISSN-L)
- 1552-4965
- OpenAlex
- W2054261405
- Cited by
- 52
- References
- 32
- Citations by year
- 2025: 1 · 2022: 2 · 2021: 4 · 2020: 2 · 2019: 2 · 2018: 2 · 2017: 5 · 2016: 4 · 2015: 5 · 2014: 6 · 2013: 4 · 2012: 3
- FWCI
- 2.20
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Biomedical Materials Research Part A (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institute of Biomedical Imaging and Bioengineering (R01EB001640)
- Topics
- Physical Sciences › Materials Science › Biomaterials › Collagen: Extraction and Characterization
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Keywords
- collagen scaffolds, implantable glucose sensor, nordihydroguaiaretic acid, glutaraldehyde, crosslinking, biocompatibility, foreign body reaction, sensor sensitivity, porous scaffolds, inflammation, subcutaneous implantation, sensor lifetime
- MeSH terms
- Animals, Biocompatible Materials, Collagen, Cross-Linking Reagents, Glutaral, Male, Materials Testing, Masoprocol, Biosensing Techniques, Porosity, Rats, Sprague-Dawley, Prostheses and Implants, Rats, Tissue Scaffolds
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review